Date Initiated by Firm | October 21, 2009 |
Date Posted | March 10, 2010 |
Recall Status1 |
Terminated 3 on April 22, 2010 |
Recall Number | Z-1066-2010 |
Recall Event ID |
53675 |
Product Classification |
CMV IgG - Product Code LFX
|
Product | IMMULITE 2000 and IMMULITE 2500
IgG/IgM Sample Diluent, Cat. No. L2IGZ2, Lot Number 150 |
Code Information |
Lot Number 150 |
Recalling Firm/ Manufacturer |
Siemens Healthcare Diagnostics Inc. 5700 W 96th St Los Angeles CA 90045-5544
|
For Additional Information Contact | Mary Lou Mattes-Pound 914-524-2373 |
Manufacturer Reason for Recall | The recall was initiated because Siemens Healthcare Diagnostics has identified there is the potential for patient results to be biased low (-0.3 Sample to Cutoff [S/CO] ratio units) due to incompatibility of the Toxoplasma IgM kits with this particular lot of IgG/IgM Sample Diluent. All available lots of the IMMULITE 2000/2500 Toxoplasma IgM kits (L2KTM/L5KTM) are incompatible with Diluent lot 150 |
FDA Determined Cause 2 | Nonconforming Material/Component |
Action | The customer notification was initiated on 10/22/2009 with the firm forwarding an Urgent Field Safety notice with attached Field Correction Effectiveness Check form to the affected consignees informing them of the issue and the products affected. The consignees were informed that Siemens is initiating the Field Correction (FC) to prevent IMMULITE 2000/2500 Toxoplasma IgM from being run with Diluent lot 150.
The consignees were instructed in the interim, until replacement kit lots of RUB IgG and CMV IgG are received, your laboratory can continue using Rubella IgG, CMV IgG, or Toxoplasma IgM as long as Diluent Lot 150 is not loaded on the analyzer when running Toxoplasma IgM. Diluent Lot
150 can be used with RUB IgG and CMV IgG.
Once the replacements lots are received, any unopened kit lots listed in the above table MUST BE RETURNED to Siemens per instructions that will accompany a separate Fed Ex Call Tag that will explain the proper return of the affected product.
The consignees were also instructed to complete the attached Field Correction Effectiveness Check form and fax to the firm's technical solutions center.
Additional questions should be directed to the firm at 973-927-2828. |
Quantity in Commerce | 10 |
Distribution | Nationwide |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|