• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall DATEXOHMEDA NETWORK AND ICENTRAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall DATEXOHMEDA NETWORK AND ICENTRAL see related information
Date Initiated by Firm June 25, 2010
Date Posted August 12, 2010
Recall Status1 Terminated 3 on August 14, 2012
Recall Number Z-2211-2010
Recall Event ID 55502
510(K)Number K052972  
Product Classification System, network and communication, physiological monitors - Product Code MSX
Product DATEX-OHMEDA NETWORK AND ICENTRAL '05, SALES REVISION 4.3 ( Datek-Ohmeda Network and iCentral (including iCentral Client)) The Datex-Ohmeda Network and iCentral (including iCentral Client) transfers information between networked Datex-Ohmeda devices in the Monitor Network as well as between Monitor Network and networked devices in GE Unity NetworkTM. It also allows information transfer between several iCentrals. Within one Monitor Network it allows a networked device to display, store, print and otherwise process information received from other networked devices. The iCentral maintains the network connections between the Datex-Ohmeda bedside monitors and other networked devices in Monitor Network. Network connections consist of hardwired network cables and/or Wireless LAN (WLAN) connections. Furthermore, it coordinates the transfer of information between devices in the Datex-Ohmeda Network, between the Datex-Ohmeda Network and devices in GE Unity Network as well as between the Datex-Ohmeda Network and Hospital Information Systems (HIS). The iCentral can be used for remote monitor management, storing, printing, viewing, reviewing or otherwise processing of information from several bedside monitors or other networked devices. The iCentral Client can be used for remote monitor management, printing, viewing, reviewing or otherwise processing of information from several bedside monitors or other networked devices. The Datex-Ohmeda Network will be used for patients in the hospital and it is meant for continuous use. The device is for use by qualified personnel only.
Code Information Parent SN CZC53003BQ CZC53003BR CZC53003BX CZC7414GJD CZC7414GLN CZC5510Y64 CZC721500D CZC7272RM2 CZC7414GFS CZC7520S3Y CZC53003B3 CZC53003B8 CZC53003B9 CZC7520S16 CZC7520S18 CZC6322Q7L CZC64511L8 CZC84095RN FRB4210XQ9 FRB4210XQZ FRB4260139 CZC5240Q1L CZC5472F8J CZC6482WKL CZC73312RN CZC6130L3B CZC6261MW4 CZC6261MWZ CZC5280BS7 CZC94890SW CZC7520RZ0 CZC7520S2X CZC5240Q1G GB8627CX31 CZC642559Z CZC5391VGL 5100648 5100649 5100650 5100651 5100653 5100654 5100655 5115314 5123573 5123574 5123575 5123578 5132670 5132671 5132672 5132673 5132674 5132675 5132676 5132677 5132678 5132679 5132680 5132681 5140814 5148398 5148399 5148404 5148405 5148406 5148407 5148408 5156076 5156077 5156078 5156079 5156080 5156081 5156284 5165853 5165868 5171824 5173135 5178892 5178893 5178894 5178895 5178896 5178898 5178899 5178900 5182706 5182709 5182710 5182711 5182718 5199055 5199063 5199080 5213727 5213728 5213729 5213730 5213731 5213732 5213733 5213734 5213741 6161335 6186167 6299938 CZC7364RTL CZC7414GK1 CZC7520RZH CZC7520S49 CZC931344X CZC931344Z CZC7272RM6 FRB4250SZY CZC6181MT8 CZC64006K3 CZC6322Q85 CZC642559H CZC0044VYY CZC0044VZ5 CZC0044VZ7 CZC0044VZ8 CZC92520VB CZC92520VD CZC92520VL CZC62803TK CZC62803TR CZC7244NXG CZC7244NXJ CZC721500R CZC721500T 5205304 5213704 5213705 5225449 CZC6130L3F CZC7520S3S CZC61030G8 CZC92520TS CZC92520TZ CZC92520V0 CZC92520W0 CZC64255BK 5182691 CZC7520RYL CZC7520RYM CZC7520RYN CZC7520RYP CZC7520RYW CZC7520RYY CZC7520RZ2 CZC7520RZ3 CZC7520RZS CZC7520S0G CZC7520S0Z CZC7520S13 CZC7520S14 CZC7520S2M CZC7520S44 CZC7132V6M CZC7132V83 CZC7215009 CZC721500M CZC0044VYG CZC6261MWT CZC63128WT CZC8304C9W CZC7250WNJ CZC6181MT1 CZC5240Q16 CZC5240Q18 CZC5240Q1J CZC8304C9Z CZC8304CB1 CZC8304CBW CZC8304CC0 CZC63128WC CZC84912DT CZC7215010 CZC61030GC FRB3470CPS CZC7132V79 CZC6394DFG CZC7520RZF CZC8304CBC CZC84095RB CZC84095RH CZC84095S0 CZC84095S2 FRB4210XSV FRB4250SM8 CZC5492LKS CZC63128X5 CZC5391VFJ CZC7520RZM CZC93311FP CZC93311G3 CZC7364RV6 CZC6181MSZ CZC64511LK CZC94890RS CZC94890S3 CZC5510Y6G CZC5510Y6J CZC61030GH CZC61030H0 FRB4250QGW CZC528244Y FRB4250T11 CZC64006JT CZC64006JY CZC94438TD CZC5280BSD CZC5391VG5 FRB4250SM6 CZC7272RLH CZC6130L30 CZC6482WKT CZC6394DGV CZC7272RLS C2C528245B CZC528245B CZC5492LLM CZC6261MW5 CZC84912DR CZC9151RLG CZC92520TW CZC9313455 CZC9387DM7 CZC94438SH CZC7363P2Z CZC7363P3M CZC7414GH8 CZC5391VF4 CZC7414GGD CZC7414GJH GB8717D5DA CZC5391VGC CZC6291WTP CZC6291WTR CZC6103GG CZC64006KB CZC721500C CZC5492LLK CZC603076D CZC0044VZB CZC0016T08 CZC0016T09 CZC0044VY8 CZC0044VYL CZC0044VZ1 CZC642559M CZC7414GMB CZC7290DZ1 CZC7132V7D CZC84857GK CZC7364RV2 CZC8304C9Q CZC7290DYF CZC7520S3X CZC94890S6 CZC6181MT0 CZC6261MWM CZC7520S4F CZC84095QX CZC5510Y6M CZC5492LK5 CZC5492LLD CZC5492LLH CZC5510Y60 CZC5510Y63 CZC5510Y6B CZC5510Y6N CZC7212G4K CZC7212G52 CZC83797HC CZC84857GN CZC7520S02 CZC7520S0C CZC64255BX CZC7212G51 FRB4250T0Q CZC6482WKM CZC72414B9 CZC73312RZ CZC6130L2N CZC6130L2P CZC6130L2S CZC84912F2 CZC83053JZ CZC7244NXM CZC7132V76 CZC6322Q88 CZC0044VZ6 CZC7132V6P CZC7132V7Z CZC7520S0N CZC94438SN CZC94438TB 5140820 5140821 CZC84912GC CZC9071RL3 CZC6394DH7 CZC83053JK CZC62803TL CZC62803TS 5065154 CZC7364RSK CZC7364RSL CZC7414GF1 CZC7414GK7 5065155 5156275 5156276 5165861 5213703 5225427 5225445 CZC0016T0H CZC0016T0J CZC0016T0T CZC0044VYM CZC0044VYP CZC528245K CZC5391VFC CZC64006KM CZC92520TV CZC92520V2 CZC92520V3 CZC92520V8 CZC92520VP CZC9263WVH CZC9313449 CZC931344L CZC93311GC CZC9341W7W CZC9341W7X CZC9341W8D CZC9387DLV CZC94438SS CZC94438SY CZC9451GN6 CZC94890SD CZC64006JK CZC64006JR CZC83053HW CZC84912F3 CZC84912F7 CZC9071RKR CZC9071RKX CZC7244NXF CZC7272RLP CZC7520S2K 6197653 CZC5391VFR CZC6425599 CZC5391VGN CZC5391VFW CZC64511LS 5077174 5140806 CZC7414GLW CZC74146J3 CZC5472F8L 5065153 5126775 CZC8304CC1 5126776 5182701 5199054 CZC7414GJM 5140819 CZC7244NX5 CZC7414GF5 CZC7520S1X 5182695 CZC7105QDP CZC7105QDZ 5156052 5156053 5156282 5156285 CZC7414GH7 CZC62803T5 CZC7520S4S 5126779 5225429 CZC83053JS CZC84912DY CZC84912FS CZC84912G4 CZC84912G7 CZC84912G9 CZC84912GD CZC642558J CZC7105QDT CZC7105QFR CZC5492LJY CZC7212G4Q CZC7132V7X CZC7520RZ7 CZC63128WW CZC64255CH CZC7212G4P CZC7364RSV CZC64511LN CZC64511M9 CZC7105QFG CZC7105QFS CZC84912FC CZC9071RK9 CZC9071RKT CZC9071RLB CZC5391VFG CZC5510Y5T CZC7520S22 CZC5460C7K CZC7105QFF 5065152 5065156 5065157 5065158 5065159 5077167 5077168 5156278 5156279 5156281 5175445 5175446 5175447 5175449 5199076 5199077 5199078 5199079 CZC5280BS5 CZC5280BSK CZC5460C7Y CZC5462HVY CZC5462HW0 CZC5462HW8 CZC5472F8T CZC5492LJS CZC5521M44 CZC6130L2J CZC6130L2L CZC6130L2T CZC63128WF CZC63128WZ CZC63128X7 CZC63128X8 CZC63128XQ CZC64006JH CZC64255BC CZC64255BJ CZC64255C2 CZC64255CD CZC6482WKG CZC7105QDY CZC7105QF8 CZC7105QFB CZC7105QFQ CZC7132V72 CZC7212G50 CZC721500K CZC7244NX4 CZC7244NX8 CZC7244NXC CZC7250WNP CZC7414GFT CZC7414GG6 CZC7414GHD CZC7414GHT CZC7080QHY CZC7364RTD CZC7520RZR CZC7520S11 CZC7520S29 CZC7520S3H CZC7520S43 CZC7520S50 CZC8304CBP CZC83053HL CZC83053HQ CZC83053HY CZC83053JD CZC83053JH CZC83053JJ CZC8379760 CZC83797G5 CZC83797GN CZC83797GT CZC83797H8 CZC84095RD CZC9071RLC CZC91217MC CZC91217MF CZC91217MM CZC91217MP CZC92520VG CZC92520VT CZC92520VW CZC931344M CZC93311F9 CZC93311FB CZC93311FF CZC93311FH CZC93311FX CZC93311GB CZC9387DM1 CZC94890SC CZC94890SK CZC7364RSH CZC7364RSS CZC7414GGL CZC7244NXD CZC7244NXP CZC7364RRY CZC7520S4Y CZC83053K8 CZC0016T06 CZC9151RLJ CZC7414GFV CZC721500B CZC7364RV9 CZC84095QQ CZC84857GB CZC0016T0G CZC6383VVK CZC6391G9W CZC6391G9X CZC64006KN CZC64006KT CZC64006KV CZC64006L3 CZC72414B4 CZC7414GFZ CZC7414GG0 CZC7414GG3 CZC7414GG4 CZC7414GGG CZC7414GHV CZC7414GKC CZC7414GLK CZC94890SP CZC94890SS 5100633 5100643 5140822 5148392 5156049 5170744 5213709 5213746 5213747 5213749 CZC0016T0Q CZC5521M4K CZC61030GX CZC6130L3G CZC6181MT5 CZC6181MT9 CZC6181MTM CZC62803TC CZC62803TQ CZC62803TT CZC62803V0 CZC63128WD CZC63128WJ CZC63128WL CZC63182WL CZC642558L CZC6425595 CZC642559V CZC64255C1 CZC64255CJ CZC64511LQ CZC7272RM4 CZC7272RM5 CZC7272RMS CZC73312RY CZC7364RRR CZC7364RTQ CZC7364RVD CZC7520RZ4 CZC7520S0V CZC84095RT CZC84912FL CZC9071RKG CZC9151RLC CZC9471VGJ CZC9471VGX CZC9471VH2 FRB4250SMF FRB4250SMH FRB4250SWS FRB4250SZF FRB425SZB FRB3470L5J 5156055 5156057 5165854 5170738 5170748 5170749 5171820 5171821 5171823 5175464 5175465 5182683 5205319 5205772 5205773 5205774 5205775 CZC0016SZV CZC0016SZZ CZC0016T0C CZC5240Q13 CZC5240Q14 CZC5460C85 CZC5462HW5 CZC5492LL3 CZC61030G4 CZC61030GD CZC61030H2 CZC61544VD CZC61544VM CZC61544VZ CZC6261MWH CZC6322Q7Q CZC64255BT CZC64255BY CZC7105QF2 CZC7105QF9 CZC7105QFD CZC7105QFJ CZC72126A6 CZC7212G4J CZC7364RTS CZC7414GFJ CZC7414GFN CZC7414GJN CZC7414GKN CZC7414GMC CZC7520RZZ CZC8304CB4 CZC8304CBL CZC8304CBT CZC83053J1 CZC83053K0 CZC84857GM CZC9260DNP CZC9260DNR CZC9260DNS CZC94438SJ 5089740 5123579 5123580 5123581 5156273 5156274 5156280 5182677 5199043 5199044 5199047 5199069 5199071 5199072 5225442 CZC5280BSS CZC53003B1 CZC5460C7B CZC5462BTZ CZC6130L3C CZC64006L6 CZC642559Q CZC642559T CZC642559W CZC64255B6 CZC64255B9 CZC64255BD CZC64255BL CZC64255BV CZC64255C7 CZC7080QJM CZC7105QDR CZC7132V7P CZC7414GFX CZC7414GG9 CZC7414GGC CZC7414GHY CZC7414GKS CZC7520537 CZC7520S00 CZC83797GG CZC83797GJ CZC83797H3 CZC84912DZ CZC9471VH1 CZC9471VH5 FRB4260120 FRB4260124 FRB4260125 FRB426013X 5140807 5156083 5213707 5225426 5225436 CZC53003B2 CZC53003B5 CZC5492LK4 CZC63128X1 CZC642559F CZC64255B1 CZC7105QF1 CZC724149T CZC7364RRP CZC83797GH FRB4210XSD FRB4250SWZ FRB426011G FRB426013M CZC94890S4 CZC83053JC CZC83053K3 5175454 5213735 5225441 CZC53003BK CZC5472F88* CZC5472F8S* CZC6130L2H CZC6291WTN CZC63128WM CZC64511M0 CZC7364RS2 CZC7414GHZ CZC83797HG CZC83797HL CZC9071RKN CZC9071RKP CZC92520V9 CZC9341W7Z CZC9451GNC CZC72414B3 CZC72414B6 CZC7250WNB CZC7364RRG CZC7364RT8 CZC7364RTK 5156272 5170746 5170747 5213726 5225430 5225435 CZC83053K6 CZC92520VF CZC94890SR CZC5240Q1F CZC5462BV3 CZC6181MTN CZC6181MT4 CZC6181MTQ 5165858 FRB426012P FRB426012T CZC603076N CZC61030G6 CZC5280BRW CZC5460C87 CZC5492LKL CZC6394DGG CZC642558P CZC6425594 CZC6425597 CZC642559B CZC7080QJ8 CZC7080QK7 CZC7414GH3 CZC7414GHC CZC7414GHR CZC7414GK3 CZC7414GL1 CZC7520S1J CZC6322Q7K CZC6322Q7S CZC7520S3N CZC8304C9X CZC8304CB3 5225425 CZC6322Q7F CZC9313445 CZC931344D CZC93311FR CZC93311FV CZC93311G4 CZC93311G7 CZC9451GMR CZC9451GN3 CZC9451GNJ CZC9451GNK CZC7364RRS CZC7132V6V 5182694 CZC84095QP CZC84095R1 CZC84095R9 CZC72414B5 CZC7244NXS FRB4210XS1 FRB4250SZN CZC551046P CZC7105QDV CZC7364RTW CZC7520S0L CZC8304CB6 5170760 CZC7520S1P CZC642558M CZC6425592 CZC84095RC 5132682 CZC5492LKB CZC5510Y5X CZC5521M4W CZC6181MTH CZC62803TN CZC6394D67 CZC7520S2D FRB4250SMJ 5148395 5148396 5148397 5165864 5165865 5165866 5165867 5171809 5182715 5182716 5182717 5199051 5205310 5205311 5225447 5170762 5171812 5175448 5182684 5182693 5182699 5182705 5182707 5182708 5182722 5199041 5199042 5205301 5222765 6004796 6095050 CZC0016SZS CZC0016T03 CZC0016T0S CZC5240Q1D CZC5240Q1H CZC5391VFQ CZC5391VFV CZC5391VG2 CZC5391VGD CZC5391VGJ CZC5460C7F CZC5460C7N CZC5460C7Q CZC5460C86 CZC5492LJX CZC5492LKC CZC5521M4M CZC5521M4Z CZC6130L2Y CZC6261MWC CZC63128WY CZC6322Q7R CZC64006KS CZC64006KY CZC642558Q CZC642559D CZC64255BW CZC7080QJ7 CZC7364RS9 CZC7414GGV CZC7414GJ8 CZC7414GM0 CZC7520RZ9 CZC7520RZG CZC8304C9V CZC8304CBY CZC8304CC4 CZC83053HN CZC83053K7 CZC83797G1 CZC83797GK CZC83797GL CZC83797H0 CZC92520TY CZC9313446 CZC931344C CZC9313456 CZC9341W81 CZC9451GN4 CZC9451GN7 CZC9451GNP CZC7520S0T CZC83053HH CZC83053K5 CZC93311G0 CZC9260DNT CZC9260DNV CZC9263WV5 CZC9263WV8 CZC7364RS5 5089745 FRB426014P 5089746 5089743 CZC91217MT CZC5492LL6 CZC9313442 CZC931344P CZC84095RX 5182721 5100640 5115311 CZC7364RTZ CZC8304C9Y CZC0016T04 CZC7414GJ9 CZC7414GKQ CZC7414GLX CZC84095RF CZC84095RW CZC84095S8 CZC84912FG CZC84912FK CZC84912FZ CZC84912G0 CZC84912G3 6095066 6161298 CZC5240Q1M CZC5280BS4 CZC5280BSF CZC5280BSM CZC5280BST CZC5282454 CZC5391VDV CZC5441LL0 CZC5462BV6 CZC5492LJV CZC5492LJZ CZC5510Y5W CZC5521M42 CZC5521M46 CZC5521M4P CZC5521M4Q CZC6130L2V CZC61544V9 CZC61544VH CZC61544VL CZC61544VQ CZC61544VV CZC61544VY CZC61544W0 CZC6181MTV CZC6394DFY CZC6394DG4 CZC6394DG8 CZC6394DGH CZC6394DGK CZC6394DH0 CZC6394DH9 CZC64006K9 CZC64006KL CZC64006KR CZC64006KW CZC642558N CZC64511L6 CZC6482WK7 CZC6482WKX CZC6482WL3 CZC6482WLB CZC7105QF0 CZC7132V6T CZC7132V70 CZC7132V73 CZC7132V7K CZC7212G42 CZC73312RQ CZC73312RR CZC73312RW CZC7364RTR CZC7414GFB CZC7414GFD CZC7414GFP CZC7414GFQ CZC7414GGT CZC7414GGZ CZC7414GH4 CZC7414GHQ CZC7414GJR CZC7414GLC CZC7414GML CZC7520S1N CZC7520S1W CZC7520S2L CZC7520S4T CZC8304CB5 CZC8304CBJ CZC83053JT CZC83053JX CZC83797GV CZC83797HH CZC84857GL CZC84912FT CZC84912GB CZC9151RL3 CZC9151RL4 CZC9151RL5 CZC9151RL6 CZC9151RLD CZC93311GD FRB3440KVB FRB4250QG5 FRB4250QGR FRB4250QGY FRB4250QH3 FRB4250SWV FRB4250SWW FRB4250SZ9 FRB4250SZW FRB4250SZZ FRB4250T02 FRB4250T04 FRB4250T0M FRB426011Q FRB4260129 FRB426013N FRB426013Z FRB4260140 CZC6391G9T CZC84912FP CZC7080QJC CZC0016T05 CZC0016T07 CZC93311FT CZC93311G1 CZC9387DLQ CZC94438SC CZC94438SM CZC94438SQ CZC94438SR CZC94438T2 CZC94438T8 CZC94438TC CZC94438TG 5170753 5213740 CZC52303S6 CZC52303S8 CZC5391VF3 CZC5462BV7 CZC5492LJP CZC5492LKR CZC5492LKX CZC5492LL2 CZC5510Y5Y CZC5521M45 CZC5521M49 CZC5521M4B CZC5521M4C CZC5521M4X CZC5521M4Y CZC61030GM CZC6130L2X CZC6181MT7 CZC6261MWS CZC62803TD CZC62803TJ CZC62803TW CZC63128WB CZC63128WN CZC63128WR CZC63128X2 CZC63128X6 CZC6322Q82 CZC6322Q8B CZC6391G9V CZC6394DG0 CZC6394DG1 CZC64006JW CZC64006JZ CZC64006KD CZC64006KX CZC64006KZ CZC64006L2 CZC642558V CZC642558W CZC642559C CZC642559K CZC64255B0 CZC64255B3 CZC64255B5 CZC64255B7 CZC64255BF CZC64255BH CZC64255BM CZC73312RG CZC73312RH CZC73312RL CZC73312S1 CZC7363P2W CZC7363P34 CZC7363P38 CZC7363P3C CZC7363P3G CZC7363P3J CZC7363P3K CZC7364RT6 CZC74146L5 CZC7414GF4 CZC7520S3J CZC7520S42 VIRTUAL1 VIRTUAL2 CZC83797G9 CZC84912F5 CZC528244V CZC53003BC CZC53003BH CZC53003BJ CZC5521M40 CZC5521M4R CZC5521M4V CZC6425598 CZC64511LZ CZC7364RTV CZC7414GHL CZC7520RYQ CZC7520RZL CZC9313443 CZC9387DLT CZC7364RS1 5205305 CZC7105QFM CZC7363P3F CZC7414GH2 CZC9341W87 CZC9451GMY FRB4250T0C CZC7272RLX CZC7520S21 CZC7520S3Q CZC93311G8 CZC93311GF CZC63128X4 CZC7105QDX CZC7105QFK CZC7364RV8 CZC7520S33 CZC7520S3D CZC8304CBQ CZC84912F1 CZC84912FO FRB4250SMB FRB4250SMD 5089747 5126774 CZC5280BSR CZC53003B4 CZC5460C7H CZC5460C7M CZC5460C7R CZC5460C7S CZC5460C7T CZC5460C7V CZC5460C80 CZC5460C83 CZC5460C84 CZC5460C88 CZC5460C89 CZC5472F82 CZC5472F8B CZC5472F8D CZC5472F8P CZC5492LL0 CZC5492LLB CZC63128W9 CZC63128WP CZC6394DGF CZC64006K8 CZC64006KH CZC6482WLD 5182679 CZC7302PJX CZC73312RD CZC7414GF9 CZC7414GFL CZC7414GJP CZC7520S40 CZC84095QS CZC84095QW CZC84095RG CZC84095RQ CZC9071RKV CZC9071RLO CZC93311FQ CZC61030G0 CZC61030G5 CZC61030GF 6387660 CZC642559Y CZC642889N CZC7364RRH CZC7364RVC CZC7414GGQ CZC7414GLH CZC5460C7J CZC931344F CZC5472F80 CZC603076R CZC6030777 5115300 5115301 5115302 5115303 5115304 5115305 5115307 CZC9387DM5 CZC0016T0F CZC0016T0P CZC64006L9 CZC9313454 CZC9451GMT CZC9451GNL CZC9451GNM 5100634 5115309 5129989 5140813 5156291 5165863 5171826 5171827 5175461 5199056 5199057 5205309 6203592 6299949 CZC5391VGK CZC5472F87 CZC603076G CZC603076J CZC603076L CZC603076W CZC6030774 CZC61544V6 CZC61544VF CZC6181MTB CZC6181MTX CZC62803T4 CZC63128WH CZC6394DGZ CZC64006JN CZC64006K6 CZC64006KP CZC64006KQ CZC64255CK CZC6482WJR CZC6482WJT CZC6482WK1 CZC6482WKD CZC7080QJK CZC7080QJP CZC7212G4H CZC7520RZJ CZC7520S31 CZC83053J5 CZC83053J6 CZC83053JF CZC83053JL CZC83053KB FRB4210XQX FRB4210XR1 FRB4210XS7 FRB4250T0J CZC7520545 CZC5492LK1 CZC64006KK CZC7212G54 CZC8304CBZ CZC7290DYM CZC5472F84 CZC84095RL CZC7105QDS 5126772 5148402 5148403 5156082 5170750 5170758 5170759 5170761 5182700 5182719 5199053 5199066 5199070 5205303 5205771 5205318 5213708 5213736 5213744 5213750 5225434 CZC7520S0Q CZC61030GT CZC63128WK CZC84857GJ 6310167 6330140 6465595 CZC0016SZQ CZC5240Q1R CZC5391VF6 CZC6322Q87 CZC729ODZ4 CZC7520S2Y CZC83053JW CZC84095R0 CZC84095SB CZC9071RKW CZC9387DL7 CZC9387DM4 FRB4210XQM CZC84095RK CZC7364RVG CZC6261MW7 CZC7520S1M CZC84095QN CZC84857GH FRB4210XPX FRB3470L62 CZC84095R2 CZC84095RZ CZC84095SC CZC8304C9M CZC6130L36 FRB3470L9D FRB3470LBJ FRB4370L9D CZC84095S7 FRB4210X06 FRB4210XQ6 FRB42509X7 FRB4250SX7 CZC7212G55 CZC5462HVZ CZC6394DFT CZC7414GJQ CZC6261MWP CZC7414GL7 CZC5282456 CZC528245F CZC7212G4F CZC7244NXH CZC752062V CZC7520S0R CZC7520S2R CZC83053HS CZC9260DNW FR83470L8D FRB3470CPG FRB3470CRK FRB3470L8D FRB4210XSB FRB4250SZ6 FRB426013G FRB4260153 CZC5240Q1S CZC7363P32 CZC73312RS CZC5521M43 CZC6130L2M CZC7364RVB CZC6181MTR CZC5282455 CZC5460C7W CZC5472F85 CZC93311FD CZC5240Q1K CZC6322Q8F CZC83797GM CZC83797GP FRB4210XSH FRB4250SXJ FRB426014M CZC61544V8 CZC5521M4H CZC9341W7S CZC7520S2W CZC721500P CZC64255BQ CZC64006JS CZC53003BB CZC642558X CZC7414GG8 FRB3470L5Z CZC5492LKY CZC5492LLC CZC64006KG CZC64511M3 CZC7520S0H CZC7520S1B CZC64511L3 CZC64511LF CZC64511M2 CZC64511M4 FRB4250SWR CZC8304CBK CZC7244NXK CZC9387DM6 CZC6261MWX CZC83797GY CZC7290DYD FRB3470CQ4 FRB4260121 CZC7272RLW CZC6030771 CZC7520S4G CZC6261MWD CZC7520S2V CZC53003B6 CZC83053JR CZC91217ML CZC84095QZ CZC84095RM CZC84095RO CZC84095RS CZC83053HT CZC7520S2G CZC7520S3G CZC7290DYH CZC7290DYQ CZC53003BW CZC6130L34 CZC5492LL4 CZC6322Q7N 5140816 5148401 5199058 5225428 CZC0016T01 CZC0016T0M CZC5280BS9 CZC5391VDX CZC5492LJM CZC5510Y6H CZC5510Y6L CZC5521M4S CZC6425591 CZC642559R CZC642559S CZC7132V6N CZC7132V7N CZC7364RTB CZC7364RTO CZC7414G7M CZC7414GH5 CZC7520509 CZC752Q54C CZC8304CB7 CZC83797GF CZC83797HB CZC84857GD CZC84857GG CZC9263WV6 CZC9341W85 FRB4210XRY FRB426012M CZC7520S4N CZC841912FN CZC53003BV CZC7520S4B CZC5472F7Z 6228600 CZC64511L2 CZC64511LR CZC64511LT CZC7364RVH CZC92520V6 CZC94438T6 CZC7364RSY CZC83797GR 5132683 CZC64006JQ 5205308 5225456 5089742 CZC61544VW CZC91217M9 CZC6482WLN CZC64255CC CZC6322Q84 CZC5521M4T CZC7520RYV CZC7520S17 CZC84857GF CZC63128WV CZC64006JD CZC7080QJW 5148390 5148391 5175439 5175440 5175441 5175443 5175444 CZC5510Y65 CZC528244Z FRB4250QGD FRB4250SM9 CZC931344B CZC94438SK CZC5462BV8 CZC5462BVC CZC6181MT2 5148394 5170764 5205300 5205302 CZC5521M4G CZC5280BS0 CZC5280BSL CZC5280BSN FRB4250SZ0 CZC7520S0M CZC5472F8Q CZC603077C CZC7414GHP 5123583 5123584 5123585 5140823 5175453 5213751 5225431 CZC5510Y61 CZC9451GMS CZC7520S07 FRB4250SZV CZC7363P36 CZC0016T0N CZC9451GN1 CZC9471VGR CZC94438SW CZC94438SX CZC94438SZ CZC9471VH4 CZC7520S4K CZC9387DL8 CZC94438SD CZC94438SG CZC94438SL CZC94438TH CZC94438T9 CZC92520W1 CZC7520S0X CZC7080QJB 5089748 5089749 5100629 5100657 CZC5472F7X CZC5472F7Y CZC5472F83 CZC5472F86 CZC5472F8C CZC5472F8H CZC7105QFT CZC7520S48 CZC7520S4J CZC7520S4L CZC7520S4M CZC7520S4V CZC7520S4W CZC83797G3 CZC83797G6 CZC83797H2 CZC84857GC CZC84857GS FRB4250SWX CZC61544V5 CZC61030GP CZC6130L35 6186143 CZC5510Y66 CZC61030GV CZC6130L2W CZC6181MTK CZC6322Q8D CZC6482WLH CZC6482WLK CZC7132V7M CZC724149S CZC724149Z CZC7414GGP 5077176 5077177 CZC92520W3 CZC9263WV1 5077178 5077179 5115312 5115313 5123576 5123577 5171816 5171817 5171818 5175458 5175460 5175462 5182685 5182686 5182688 5199049 5199059 5199060 5205778 5205779 5213702 5213706 5213737 5213745 6120246 6226668 6307432 6309945 6315267 6323792 6372879 6387646 C2C5462BV5 CZC132V6W CZC5240Q1N CZC5240Q1V CZC5280BRV CZC5280BRY CZC5280BS2 CZC5280BS6 CZC5280BSB CZC5280BSC CZC528244T CZC5282450 CZC5282453 CZC528245G CZC528245N CZC530039Y CZC5391VDW CZC5391VF7 CZC5391VFM CZC5391VFP CZC5391VFT CZC5460C7P CZC5460C81 CZC5462BTS CZC5462BTT CZC5462BTW CZC5462BTY CZC5462BVB CZC5462HVX CZC5462HW4 CZC5462HW7 CZC5472F7V CZC5472F81 CZC5472F89 CZC5472F8F CZC5472F8N CZC5492LJN CZC5492LJW CZC5492LK0 CZC5492LKP CZC5492LKZ CZC5492LLG CZC5510Y5V CZC5510Y5Z CZC5510Y6Q CZC603076Q CZC603076V CZC603076X CZC6030775 CZC6030778 CZC61030G2 CZC61030G3 CZC6130L2K CZC6130L31 CZC6130L38 CZC6130L3H CZC61544V3 CZC61544VG CZC6181MSY CZC6181MTC CZC6261MWV CZC62803T0 CZC62803T6 CZC62803T9 CZC6291WTH CZC6291WTQ CZC63128W8 CZC6322Q7X CZC6394DFR CZC6394DFV CZC6394DG2 CZC6394DGB CZC6394DGJ CZC6394DGM CZC6394DGW CZC6394DGX CZC6394DH6 CZC6394DHB CZC64006K0* CZC64006L5 CZC642558K CZC642558T CZC642558Y CZC642559X CZC64255BN CZC64255BR CZC64255C0 CZC64255C3 CZC64255CG CZC64511L4 CZC64511L9 CZC64511LD CZC64511LY CZC64511M6 CZC6482WJX CZC6482WKZ CZC7080QHX CZC7080QJ0 CZC7080QJ3 CZC7080QJ5 CZC7080QJ6 CZC7080QJG CZC7080QJJ CZC7080QK6 CZC7080QK8 CZC7105QF4 CZC7105QF6 CZC7105QFL CZC7132V6Q CZC7132V6R CZC7132V6Y CZC7132V75 CZC7132V7G CZC7132V7H CZC721500S CZC721500V CZC721500W CZC724149W CZC72414B8 CZC72414BB CZC72414BC CZC72414BD CZC72414BF CZC72414BQ CZC7244NXR CZC7250WMZ CZC7250WN0 CZC7250WN3 CZC7250WN4 CZC7250WN6 CZC7250WND CZC7250WNG CZC7250WNH CZC7250WNK CZC7250WNL CZC7272RLR CZC7272RLV CZC7272RM3 CZC7290DYN CZC7290DYR CZC7290DYZ CZC7363P35 CZC7363P39 CZC7363P3H CZC7364RS3 CZC7364RSB CZC7364RSF CZC7364RSZ CZC7364RT3 CZC7364RTG CZC7364RV3 CZC7364RV4 CZC7364RVF CZC741466M CZC74146DV CZC7414GF8 CZC7414GGF CZC7414GGN CZC7414GGS CZC7414GGY CZC7414GHB CZC7414GHM CZC7414GHN CZC7414GHW CZC7414GJ2 CZC7414GJL CZC7414GJS CZC7414GJT CZC7414GJW CZC7414GJX CZC7414GK0 CZC7414GMD CZC7414GMM CZC7414GMR CZC7414GMT CZC7520534 CZC752053Z CZC7520RYT CZC7520RZ8 CZC7520RZB CZC7520RZD CZC7520RZQ CZC7520RZT CZC7520S04 CZC7520S0P CZC7520S12 CZC7520S23 CZC7520S28 CZC7520S2H CZC7520S2J CZC7520S2N CZC7520S2T CZC7520S32 CZC7520S34 CZC7520S3M CZC7520S3P CZC7520S3R CZC7520S3V CZC7520S3W CZC7520S41 CZC7520S46 CZC7520S4D CZC7520S4R CZC7520S4X CZC8304C9N CZC8304CBD CZC8304CBX CZC8304CC5* CZC83053J2 CZC83053JM CZC83797FX CZC83797FY CZC83797G4 CZC83797GQ CZC83797GS CZC83797GZ CZC83797H4 CZC83797H9 CZC84095RJ CZC84095RV CZC84095S1 CZC84095S3 CZC84095S6 CZC84095S9 CZC84857GP CZC9071RKL CZC9071RL2 CZC9151RL1 CZC9151RL7 CZC92520TX CZC92520V7 CZC92520VC CZC92520VY CZC9260DNM CZC9260DNX CZC9263WTZ CZC9263WV4 CZC9263WVC CZC9263WVD CZC9263WVF CZC9313444 CZC931344Y CZC93311FJ CZC93311FL CZC9341W7K CZC9341W7L CZC9341W84 CZC9341W8G CZC9387DL9 CZC94438SP CZC94438T0 CZC94438T5 CZC9451GMW CZC9451GN0 CZC9451GNN CZC9471VGM CZC9471VGN CZC9471VGT CZC9471VGZ CZC9471VHC FRB4210XQ5 FRB4210XRT FRB4250QG4 FRB4250QGP FRB4250QGX FRB4250SWY FRB4250T07 FRB4250T0W FRB4250TOL FRB426011K FRB4260138 CZC7520S1F 6372879B CZC64006JF CZC64255BZ CZC64255C4 5170742 5170765 5171819 5171828 5175455 5175456 5175463 5182674 5182687 5182702 5182703 5182704 5199073 5199074 5199075 5205312 5205313 5205314 5205315 5205325 5213710 5213712 5213713 5225443 5225452 5225453 5225454 5225455 6186155 6255644 6441590 CZC5240Q1P CZC5240Q1X CZC5240Q1Z CZC5280BSJ CZC53003BF CZC53003BS CZC5391VDP CZC5391VFH CZC5391VFZ CZC5391VG9 CZC5462BTR CZC5492LK7 CZC5492LL7 CZC603076S CZC61030GR CZC61030GW CZC6130L2Z CZC61544V1 CZC61544VP CZC6261MW2 CZC6261MW3 CZC6261MW8 CZC6261MWW CZC6261MX0 CZC62803T1 CZC62803T7 CZC62803TM CZC62803TV CZC6322Q7P CZC6322Q7T CZC6322Q7V CZC6322Q7Y CZC6322Q89 CZC6394DFS CZC6394DGN CZC6394DGT CZC64006JM CZC64006JP CZC64006K5 CZC64006K7 CZC642558R CZC642558S CZC6425593 CZC642559J CZC64255C6 CZC64511L1 CZC64511L5 CZC64511LB CZC64511LC CZC64511LM CZC64511LP CZC64511LW CZC6482WJY CZC6482WK0 CZC6482WKC CZC6482WKH CZC6482WKQ CZC6482WKR CZC6482WL1 CZC6482WL2 CZC6482WL6 CZC6482WL8 CZC6482WL9 CZC6482WLL CZC7080QJ4 CZC7080QJD CZC7080QJY CZC7080QK1 CZC7080QK2 CZC7080QK4 CZC7080QKC CZC7105QF3 CZC7132V7B CZC7212G4L CZC724149V CZC7241EB2 CZC7244NXN CZC7244NXV CZC7250RZV CZC7250WN2 CZC7250WN7 CZC7250WNC CZC7250WNM CZC7272RM1 CZC7290DYJ CZC7290DYV CZC7290DYW CZC7290DZ3 CZC73312R9 CZC73312RK CZC73312RT CZC73312RV CZC7363P2X CZC7363P2Y CZC7363P3B CZC7363P3D CZC7364RRK CZC7364RRT CZC7364RT4 CZC7364RT9 CZC7364RV1 CZC7364RVJ CZC7414GDX CZC7414GFF* CZC7414GG1 CZC7414GGH CZC7414GGK CZC7414GGX CZC7414GH6 CZC7414GHG CZC7414GHK CZC7414GHO CZC7414GJ4 CZC7414GJB CZC7414GJZ CZC7414GL4 CZC7414GLF CZC7414GLR CZC7414GMH CZC7520RYX CZC7520RZ1 CZC7520RZN CZC7520RZW CZC7520S03 CZC7520S05 CZC7520S06 CZC7520S0J CZC7520S1G CZC7520S1K CZC7520S1L CZC7520S1Y CZC7520S25 CZC7520S2C CZC7520S30 CZC7520S3T CZC7520S4H CZC7520S4Z CZC8304C9P CZC8304C9R CZC8304CB0 CZC8304CBF CZC8304CBM CZC83053HF CZC83053HK CZC83053HR CZC83053HV CZC83053J8 CZC83053JG CZC83053JN CZC83053JQ CZC83053JV CZC83053K9 CZC830S3JB CZC83797G7 CZC83797G8 CZC83797H7 CZC83797HF CZC83797HJ CZC83797HK CZC84095QT CZC84095QV CZC84095QY CZC84095R5 CZC84095R6 CZC84095R8 CZC84095RP CZC84857GQ CZC84857GT CZC84912F4 CZC84912F6 CZC84912F8 CZC84912F9 CZC84912FB CZC84912FD CZC84912FM CZC84912FQ CZC84912FW CZC84912FY CZC84912G1 CZC84912G5 CZC9071RKK CZC91217MJ CZC9121FMB CZC92520TT CZC92520VN CZC92520VQ CZC92520VR CZC92520VX CZC92520W4 CZC9263WV0 CZC9263WV2 CZC9263WV3 CZC9263WV9 CZC9263WVB CZC9263WVJ CZC9263WVN CZC931344G CZC931344T CZC9313450 CZC93311FG CZC9341W82 CZC9341W8H CZC9387DL3 CZC9387DL5 CZC9387DLB CZC9387DLJ CZC9387DLN CZC9387DLX CZC9387DLZ CZC9387DM0 CZC9387DM2 CZC9451GMQ CZC9451GNF CZC94890S9 FRB3470CPW FRB4210XS6 FRB4250QH0 FRB4250SMC FRB4250SZX FRB4250T0S FRB4250T0T FRB4250T12 FRB426014N CZC9341W7J CZC9341W8K CZC9341W8L CZC7364RT2 CZC6482WL5 CZC5492LJR CZC5492LL1 CZC5492LLL CZC603076Z CZC6030770 CZC6030773 CZC6030776 CZC64511L7 CZC64511M1 CZC64511M7 CZC6394DHF CZC7080QJS CZC7132V7R CZC713ZV7V CZC7364RST CZC7244NXQ*** CZC5472F8R CZC61030GY CZC63128WS CZC7105QF7 CZC721500X CZC63128WQ CZC64255CL CZC6261MWR CZC64255BB CZC9387DLW CZC94438ST CZC94438T1 5100636 5100638 5100639 5126771 5140805 5170754 5170755 5171813 5171814 5171815 5175457 5205324 CZC64006JG CZC7364RRQ CZC7364RRZ CZC0044VYC CZC931344Q CZC93311G2 CZC5462BTX CZC0016SZP CZC5521M41 CZC6181MTF CZC5462BV4 CZC6394DGC CZC7414GLJ CZC7414GLQ CZC83053HG CZC83053HM CZC83053K2 CZC7414GF2 CZC7414GFC CZC91217MD CZC92520VM CZC92520W2 CZC741446L8 CZC7520S38 CZC0044VYJ CZC642559G CZC9260DNL CZC5391VF2 CZC64006J CZC64006JC CZC64006K1 CZC7132V7T CZC7132V7W CZC7414GHS 5077173 5100631 5100632 5199045 CZC9341W7N 6451759 CZC7132V7F CZC7520S47 CZC7520S4P 5100646 5140809 5140810 5140811 5140812 5170737 5199061 6161315 CZ6394DFW CZC5240Q11 CZC61544V4 CZC61544VK CZC6181MTJ CZC6394DFZ CZC6394DGL CZC7212G4R CZC7364R5C CZC7364RT5 CZC7414GKV CZC7520S2F CZC84912DX CZC9151RLF FRB4210XQN 5077170 5077172 5100630 5100647 5126777 5126778 5126780 5182680 5199046 6350190 CZC5462HW1 CZC5521M4F CZC5521M4J CZC61030GB CZC61030GJ CZC61030GN CZC61544VB CZC7080QJV CZC7080QK0 CZC7290DZ0 CZC7290DZZ CZC73312RJ CZC7364RTF CZC7364RTJ CZC7414GMW* CZC7414GMX* CZC83797G2 CZC83797H1 CZC83797H5 CZC83797H6 FRB4250QGV CZC61217MV CZC84857G9 CZC9313447 CZC9313448 CZC931344J CZC93311FW CZC93311G6 CZC93311G9 CZC9451GNG CZC9451GNT CZC9471VGB CZC9471VH8 CZC94890S0 CZC94890SF 5199064 5199065 CZC8304CB8 CZC8304CBV 6303248 CZC6322Q81 CZC7132V80 CZC721500Q CZC9071RKD CZC92520W5 CZC9451GN8 CZC6394DGP CZC6394DGY CZC7364RTC CZC8304C9S 5129986 5129987 5129988 5225424 CZC93311FZ CZC6261MW9 5126787 5170740 5170741 5170743 5175466 5205326 5205327 5205328 5222754 CZC0044VYB CZC5391VFK CZC5521M48 CZC6228603 CZC64255CB CZC6482WJV CZC7080QJL CZC7080QJZ CZC7364RSR CZC7364RTP CZC7414GJC CZC9151RL9 CZC9263WTX FRB4250QG3 CZC8304CB2 CZC8304CBH CZC5492LL9 CZC601030GK CZC931344H CZC931344K CZC931344R CZC9387DLD 6203667 6279463 CZC5280BRX CZC528244P CZC528244X CZC53003BL CZC5391VF0 CZC5391VG3 CZC5391VGR CZC5472F7W CZC5492LKH CZC5492LKK CZC5510Y67 CZC603076F CZC603076M CZC603076T CZC603076Y CZC61544VJ CZC6181MTL CZC6322Q7G CZC6322Q7W CZC6322Q80 CZC6322Q8C CZC6394DFQ CZC6394DH3 CZC6394DH4 CZC6482WKK CZC6482WKS CZC7272RLQ CZC7272RLY CZC74146HX CZC74146MD CZC7414GHH CZC7414GKX CZC7414GL3 CZC7414GLG CZC7414GLP CZC7414GM8 CZC7414GMV CZC7520S01 CZC7520S0K CZC7520S19 CZC7520S1S CZC7520S2Z CZC7520S3L FRB4210XPW FRB4210XPZ FRB4210XQD FRB4210XQF FRB4210XQG FRB4210XQV FRB4210XQY FRB4210XRD FRB4210XRP FRB4210XSS FRB4210XSX FRB4250QG9 FRB4250QGG FRB4250QGL FRB4250SM7 FRB4250SZR FRB4250T0N FRB4250T0V FRB426012N FRB426013F FRB426014Q CZC84095R3 CZC83797GB CZC9071RKH CZC9071RKF CZC9451GN9 CZC0044VYN CZC0044VZ0 CZC7520S0F CZC7520S0W 5182681 5182682 5225439 5225440 5182713 CZC7414GLL CZC724149X CZC7520S0Y CZC7080QKD CZC7363P37 6203582 CZC5521M47 CZC5521M50 CZC72414B7 CZC72414BI CZC7250WN8 CZC61030GL CZC7414GKW CZC6482WKP CZC721500G CZC721500H CZC721500J CZC7290DYK CZC7290DYL CZC7290DYT CZC7290DYY CZC7414GL2 CZC7520S1V CZC64511LX CZC73312RF CZC73312RP CZC9341W7V CZC9341W83 CZC7272RM7 CZC94438TF CZC7414GL6 5225446 CZC6130L3D CZC6103069 CZC6261MWF CZC7414GLT CZC5492LK6 5175442 6111290 6255417 6327416 6366309 96603193 CZC5240Q1T CZC528244S CZC528245H CZC528245J CZC5391VFB CZC5391VG4 CZC5492LJL CZC5492LK2 CZC5492LKF CZC5492LKV CZC5492LL8 CZC5521M4D CZC62803T8 CZC62803TF CZC62803TP CZC6383VV6 CZC6383VVF CZC6383VVH CZC6383VVJ CZC6394DGR CZC6394DGS CZC6425596 CZC64255B4 CZC64511LH CZC6482WLF CZC7132V6Z CZC7272RLJ CZC7272RLL CZC7272RLT CZC7272RLZ CZC7272RM0 CZC73312S0 CZC7364RRN CZC7414GF3 CZC7414GFG* CZC7414GFH CZC7414GGW CZC7414GH9 CZC7414GHF CZC7414GJ6 CZC7414GKR CZC7414GMF CZC7520S1Q CZC7520S20 CZC7520S27 CZC7520S3B CZC8304CG CZC83053J7 CZC83797FZ CZC83797GW CZC83797GX CZC83797HD CZC83797HM CZC9341W8C CZC9341W8F CZC94890RT FRB4210XS8 CZC6130L2R 5100628 5100642 5100644 5115308 5115310 5123586 5123587 5126768 5126773 5156048 5156074 5156283 5165855 5165860 5171810 5199048 5205306 5213742 5213748 5222752 CZC6261MWB CZC6261MWY 5165859 CZC7414665 CZC7414GFR CZC83053K4 5156288 5165839 5165845 5165847 5205307 5205329 5225448 6111282 6111283 6161280 6161281 6161302 6161303 6161318 6161319 6161320 6161334 6161354 6161355 6161356 6161357 6161358 6161361 6161362 6161363 6161364 6161365 6161366 6161367 6161368 6161369 6161370 6161371 6161372 6161373 6161374 6161375 6161376 6161377 6161392 6161393 6161395 6161396 6161397 6161398 6161404 6161405 6161408 6161409 6161410 6161411 6161412 6161413 6161414 6161437 6161446 6186166 6186171 6197698 6203596 6203824 6269008 6315249 6380555 6437393 CZC0016SZX CZC5240Q10 CZC5240Q17 CZC5240Q1Y CZC5240Q20 CZC5280BRZ CZC5280BS1 CZC5280BS8 CZC5280BSG CZC5280BSH CZC5280BSP CZC5280BSV CZC5282452 CZC5282458 CZC528245D CZC530039Z CZC53003B0 CZC53003BD CZC53003BT CZC53003BY CZC5391VDY CZC5391VF5 CZC5391VF8 CZC5391VG0 CZC5391VG1 CZC5391VG7 CZC5391VG8 CZC5391VGB CZC5391VGH CZC5391VGM CZC5391VGS CZC5460C7X CZC5460C7Z CZC5462BV2 CZC5462BV9 CZC5472F7T CZC5472F8G CZC5492LK8 CZC5492LKD CZC5492LKG CZC5492LKQ CZC5492LKW CZC5492LL5 CZC5510Y5Q CZC5510Y5R CZC5510Y5S CZC5510Y69 CZC5510Y6D CZC603077D CZC61544W1 CZC6181MT6 CZC6261MW1 CZC6261MW6 CZC6261MWG CZC6261MWJ CZC6261MWL CZC6261MWQ CZC6261MX1 CZC62803TZ CZC63128WG CZC63128WX CZC6322Q7H CZC6394DGQ CZC6394DHC CZC64006JV CZC64006JX CZC64006K4 CZC6425587 CZC642558Z CZC6425590 CZC64255B2 CZC64255B8 CZC64255BG CZC64255BP CZC64255BS CZC64255C8 CZC64255C9 CZC64255CF CZC64511LL CZC6482WJZ CZC6482WK2 CZC6482WK9 CZC6482WKJ CZC7080QJ2 CZC7080QJ9 CZC7080QJF CZC7080QJN CZC7080QJT CZC7080QK9 CZC7105QFC CZC7132V7C CZC7132V81 CZC7132V82 CZC7212G4S CZC7212G4X CZC721500Z CZC7244NXT CZC7250WN9 CZC7250WNF CZC7290DYG CZC7290DYS CZC73312RB CZC7364RRL CZC7364RRM CZC7364RRV CZC7364RS8 CZC7364RSD CZC7364RSG CZC7364RSN CZC7364RSX* CZC7364RTT CZC7364RTY CZC74146M7 CZC7414GFW CZC7414GG2 CZC7414GGB CZC7414GGR CZC7414GH1 CZC7414GJJ CZC7414GKK CZC7414GKM CZC7414GKT CZC7414GKY CZC7414GLB CZC7414GM4 CZC7520RYR CZC7520RZ5 CZC7520RZC CZC7520RZP CZC7520RZX CZC7520RZY CZC7520S08 CZC7520S0B CZC7520S0D CZC7520S10 CZC7520S15 CZC7520S1C CZC7520S1H CZC7520S1R CZC7520S1Z CZC7520S2B CZC7520S35 CZC7520S3C CZC7520S3K CZC8304CBB CZC8304CC2 CZC8304CC3 CZC83053HJ CZC83053HP CZC83053HZ CZC83053JP CZC83053JY CZC83797GC CZC83797GD CZC84095RY CZC84095S4 CZC84912DS CZC84912FV CZC84912FX CZC84912G6 CZC84912GF CZC9071RK8 CZC9071RKC CZC9151RL0 CZC9151RLB CZC92520TQ CZC92520TR CZC92520V5 CZC92520VH CZC9263WVG CZC9263WVK CZC9263WVM CZC931344N CZC931344S CZC931344W CZC93311FC CZC93311FY CZC9341W7G CZC9341W7M CZC9341W7Y CZC9451GMV CZC9451GMX CZC9451GMZ CZC9451GNH CZC9451GNS CZC9471VGG CZC9471VH9 CZC94890SG CZC94890SM FRB3440KT3 FRB3470L9X FRB3470LDP FRB4210XR9 FRB4210XRC FRB4210XRR FRB4210XSL FRB4210XSP FRB4210XSY FRB4210XSZ FRB4250SM5 FRB4250SMG FRB4250SWT FRB4250T0P FRB426011S FRB426012C FRB4260133 FRB426013Q 5100635 5100637 5126769 5126770 5140808 5148400 5170751 5170752 5199050 5205317 5213739 5225432 5225433 6120270 6434923 CZC5240Q12 CZC5240Q15 CZC5240Q19 CZC5240Q1B CZC5460C82 CZC5462HW6 CZC6030779 CZC6130L37 CZC6181MTD CZC6181MTG CZC6181MTS CZC6181MTT CZC6181MTY CZC62803T3 CZC62803TG CZC62803TX CZC6322Q86 CZC6394DG6 CZC64255C5 CZC7080QKF CZC7132V7S CZC7132V7Y CZC7212G4Y CZC7212G53 CZC721500Y CZC7364RTM CZC7414GG7 CZC7414GHJ CZC7414GKP CZC7414GMS CZC7520RZ6 CZC7520S2Q CZC7520S4Q CZC91217MN CZC9151RKZ CZC9151RL8 CZC9263WTY CZC9263WV7 CZC9387DLY CZC94438T4 CZC9471VH6 CZC9471VH7 FRB4250T0F FRB426012Z CZC94438SF CZC5492LLJ CZC84095R7 5126781 5126782 5126783 5126784 5140824 CZC6482WJS CZC6482WKF CZC6482WLG CZC6482WLJ CZC7250WNN CZC6130L33 CZC752QS2P CZC8304CB9 CZC9341W89 CZC9341W8J CZC7080QK5 CZC7244NX7 CZC6394DG3 CZC6482WK8 CZC7520S1T 
Recalling Firm/
Manufacturer
Ge Healthcare, Llc
3000 N Grandview Blvd
Waukesha WI 53188-1615
For Additional Information Contact
262-544-3894
Manufacturer Reason
for Recall
The iCentral or iCentral client v5.0.2 with HP xw 4400 and 4600 may spontaneously shutdown, causing monitoring loss at the central monitoring station. Monitoring loss may result in prolonged delay of treatment of potentially lethal arrhythmia.
FDA Determined
Cause 2
Software design
Action GE Healthcare sent a "Urgent Medical Device Correction letter" dated 6/18/10 on 6/25/10 to consignees. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. The letter instructs consignees to follow specific detailed instructions for safe use of the device pending provision of a software upgrade by the firm. Consignees will be contacted by a GE Healthcare representative to schedule the software upgrade.
Quantity in Commerce 2962
Distribution AZ, AR, CA, CO, CT, FL, GA, HI, IL, IN, IA, KS, KY, LA, ME, MI, MN, MS, MO, NE, NH, NH, NY, NC, ND, OH, OK PA, RI, SC, TN, TX VT, VA, WA, WI. ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BAHRAIN, BANGLADESH, BARBADOS, BELGIUM, BERMUDA, BOLIVIA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, GREECE, GUATEMALA, GUYANA, HONG KONG, HUNGARY, ICELAND, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KOREA , KUWAIT, LITHUANIA, MACEDONIA, MALAYSIA, MEXICO, NAMIBIA, NETHERLANDS, NEW ZEALAND,NICARAGUA, NIGERIA, NORWAY, PHILIPPINES, POLAND, PORTUGAL QATAR, RUSSIAN FEDERATION SAUDI ARABIA, SINGAPORE, SLOVAKIA, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN THAILAND TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY, VENEZUELA, YEMEN.
Total Product Life Cycle TPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database 510(K)s with Product Code = MSX and Original Applicant = GE HEALTHCARE
-
-