• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall HeartStart FR2 automated external defibrillator (AED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall HeartStart FR2 automated external defibrillator (AED)see related information
Date Initiated by FirmMay 28, 2010
Date PostedNovember 26, 2010
Recall Status1 Terminated 3 on September 13, 2013
Recall NumberZ-0470-2011
Recall Event ID 55682
510(K)NumberK003565 K003819 K013425 K014157 K051632 
Product Classification Automated external defibrillators (non-wearable) - Product Code MKJ
ProductHeartStart FR2+ automated external defibrillator (AED) Model M3860: Philips brand; configurable ECG on screen; configurable manual charge in advance mode.
Code Information MODEL 3860A SERIAL NUMBERS:

DeviceSN, 0108267946, 0108269457, 0208270821, 0607239273, 0707243380, 0807248084, 0807248830, 0807248894, 0807248896, 0807248900, 0807249110, 0807249118, 0807249133, 0807249150, 0807249226, 0807249237, 0807249692, 0807250834, 0807250838, 0807250892, 0807250894, 0807250900, 0807251119, 0907254462, 0907254500, 0907254544, 0907254554, 0907254586, 0907254592, 0907254596, 0907254909, 0907254913, 0907254927, 0909305230, 1007259202, 1007259206, 1007259210, 1007259220, 1007259228, 1007259232, 1007259299, 1007259327, 1007259516, 1007259524, 1007259528, 1007259538, 1007259542, 1007259546, 1007259552, 1007259556, 1007259558, 1007259560, 1207265816, 1207265818, 1207265822, 1207265824, 1207265826, 1207265830, 1207265834, 1207265838, 1207265842, 1207265846, 1207265848, 0108267573, 0108267647, 0108267651, 0108267681, 0108267794, 0108267952, 0108267962, 0108268101, 0108268103, 0108268105, 0108268127, 0108268129, 0108268131, 0108268133, 0108268135, 0108268137, 0108268212, 0108268228, 0108268230, 0108268234, 0108268238, 0108268240, 0108268242, 0108268318, 0108268324, 0108268554, 0108268556, 0108268558, 0108268560, 0108268568, 0108269459, 0208271423, 0208271435, 0208271447, 0208271449, 0208271451, 0208271453, 0308272802, 0308272810, 0308272812, 0308272814, 0308272816, 0308272820, 0308272824, 0308272826, 0308272850, 0308272852, 0308272854, 0308272856, 0308272893, 0308272897, 0308273007, 0308273013, 0308273019, 0308273021, 0308273023, 0308273025, 0308273027, 0308273029, 0308273031, 0308273033, 0308273039, 0308273041, 0308273043, 0308273047, 0308273049, 0308273051, 0308273053, 0308273057, 0308273059, 0308273061, 0308273195, 0308273363, 0308273367, 0308273379, 0408274276, 0408274278, 0408274311, 0408274312, 0408274313, 0408274315, 0408274316, 0408274323, 0408274325, 0408274329, 0408274333, 0408274335, 0408274355, 0408274357, 0408274378, 0408274392, 0408274588, 0408274613, 0408274679, 0409300816, 0507235246, 0507235248, 0507235252, 0507235258, 0507235741, 0507235785, 0507235787, 0507235791, 0507235795, 0507235829, 0507235837, 0507235845, 0507235847, 0507235851, 0507235909, 0507235911, 0507235913, 0507235971, 0507235989, 0507236025, 0507236027, 0507236029, 0507236031, 0507236041, 0507236047, 0507236049, 0507236448, 0507236450, 0507236452, 0507236454, 0507236474, 0507236484, 0507236486, 0507236494, 0507236500, 0507236501, 0507236505, 0507236513, 0507236547, 0507236549, 0507236551, 0507236553, 0507236555, 0507236569, 0507236573, 0507236575, 0507236581, 0507236587, 0507236593, 0507236595, 0507237307, 0507237386, 0508275204, 0508275210, 0508275224, 0508275230, 0508275240, 0508275244, 0508275306, 0508275345, 0508275359, 0508275361, 0508275385, 0508275387, 0508275395, 0508275548, 0508275554, 0508275560, 0508275562, 0508275566, 0508275572, 0508275574, 0508275604, 0508275608, 0508275673, 0508275685, 0508275689, 0508275691, 0508275697, 0508275724, 0508275728, 0508275746, 0508275748, 0508276002, 0508276006, 0508276008, 0508276012, 0508276014, 0508276018, 0508276097, 0508276099, 0508276100, 0508276107, 0508276109, 0508276117, 0508276119, 0508276136, 0508276138, 0508276140, 0508276158, 0508276779, 0607239255, 0607239532, 0607239864, 0607239918, 0607239934, 0607239936, 0607239938, 0607239941, 0607239984, 0607239988, 0607239992, 0607240040, 0607240079, 0607240081, 0607240085, 0607240098, 0607240100, 0607240104, 0607240144, 0607240152, 0607240180, 0607240186, 0607240216, 0607240254, 0607240325, 0607241391, 0607241395, 0607241405, 0607241413, 0607241417, 0607241419, 0607241423, 0607241425, 0607241433, 0607241477, 0607241479, 0607241609, 0607241613, 0607241629, 0607241631, 0607241641, 0607241643, 0607241667, 0607241670, 0607241671, 0607241675, 0607241683, 0607241704, 0607241749, 0607241852, 0607241856, 0607241858, 0607241915, 0607241932, 0607241936, 0607241938, 0607241952, 0607242277, 0607242281, 0607242286, 0607242290, 0607242292, 0607242294, 0607242296, 0607242297, 0607242357, 0607242367, 0607242369, 0607242397, 0607242399, 0607242400, 0607242403, 0607242513, 0607242527, 0607242529, 0607242545, 0607242547, 0607242551, 0607242561, 0607242562, 0607242579, 0608279398, 0608279708, 0608279714, 0608279716, 0608279720, 0608279722, 0608279724, 0608279726, 0608279728, 0608279736, 0608279744, 0608279746, 0608279752, 0608279756, 0608279758, 0608279760, 0608279766, 0608279772, 0608279774, 0608279784, 0608279824, 0608279853, 0608279861, 0608279881, 0608279899, 0608280096, 0608280107, 0608280119, 0608280120, 0608280125, 0608280166, 0608280294, 0608280298, 0608280300, 0608280305, 0608280313, 0608280315, 0608280317, 0608280337, 0608280351, 0608280357, 0608280367, 0608280393, 0707243030, 0707244071, 0707244086, 0707244121, 0707244238, 0707244265, 0707244267, 0707244363, 0707244411, 0707244417, 0707244476, 0707245333, 0707245611, 0707245613, 0707245615, 0707245617, 0707245619, 0707245621, 0707245623, 0707245625, 0707245627, 0707245629, 0707245631, 0707245633, 0707245635, 0707245643, 0707245649, 0707245651, 0707245653, 0707245659, 0707245665, 0707245667, 0707245669, 0707245671, 0707245678, 0707245681, 0707245682, 0707245683, 0707245684, 0707245689, 0707245691, 0707245692, 0707245701, 0707245705, 0707245718, 0707245720, 0707245728, 0707245730, 0707245731, 0707245732, 0707245734, 0707245748, 0707245959, 0707245963, 0707245967, 0707245969, 0707245971, 0707245973, 0707245981, 0707245983, 0707245985, 0707245987, 0707245989, 0707245993, 0707245997, 0707245999, 0707246005, 0707246110, 0707246112, 0707246114, 0707246116, 0707246122, 0707246140, 0707246144, 0707246146, 0707246152, 0707246156, 0707246176, 0707246178, 0707246180, 0707246182, 0707246227, 0707246406, 0807246570, 0807246773, 0807246856, 0807246922, 0807246924, 0807247104, 0807247184, 0807247185, 0807247202, 0807247205, 0807247206, 0807247216, 0807247316, 0807247322, 0807247349, 0807247359, 0807247373, 0807247375, 0807247377, 0807247399, 0807247434, 0807247443, 0807247465, 0807247479, 0807247489, 0807247502, 0807247506, 0807247510, 0807247511, 0807247514, 0807247516, 0807247518, 0807247519, 0807247520, 0807247524, 0807247525, 0807247527, 0807247532, 0807247538, 0807247540, 0807247544, 0807247546, 0807247552, 0807247554, 0807247556, 0807247557, 0807247560, 0807247561, 0807247566, 0807247567, 0807247614, 0807247620, 0807247626, 0807247638, 0807247640, 0807247642, 0807247646, 0807247648, 0807247652, 0807247656, 0807247660, 0807247662, 0807247666, 0807247670, 0807247672, 0807247674, 0807247678, 0807247680, 0807247681, 0807247682, 0807247684, 0807247686, 0807247690, 0807247696, 0807247700, 0807247705, 0807247739, 0807247741, 0807247747, 0807247753, 0807247765, 0807247774, 0807247775, 0807247779, 0807247781, 0807247794, 0807247796, 0807247798, 0807247801, 0807247805, 0807247818, 0807247819, 0807247838, 0807247882, 0807247884, 0807247886, 0807247894, 0807247903, 0807247907, 0807247928, 0807247943, 0807247944, 0807247948, 0807247969, 0807247975, 0807247979, 0807247981, 0807247985, 0807247987, 0807247993, 0807248037, 0807248063, 0807248065, 0807248102, 0807248155, 0807248177, 0807248183, 0807248196, 0807248200, 0807248203, 0807248221, 0807248229, 0807248252, 0807248262, 0807248288, 0807248292, 0807248300, 0807248303, 0807248309, 0807248315, 0807248319, 0807248322, 0807248327, 0807248346, 0807248360, 0807248388, 0807248391, 0807248396, 0807248398, 0807248400, 0807248407, 0807248409, 0807248411, 0807248415, 0807248417, 0807248437, 0807248449, 0807248456, 0807248459, 0807248472, 0807248477, 0807248479, 0807248485, 0807248489, 0807248493, 0807248497, 0807248550, 0807248589, 0807248622, 0807248626, 0807248628, 0807248630, 0807248632, 0807248636, 0807248638, 0807248646, 0807248658, 0807248676, 0807248691, 0807248695, 0807248736, 0807248766, 0807248768, 0807248804, 0807248806, 0807248821, 0807248831, 0807248843, 0807248857, 0807248862, 0807248897, 0807248901, 0807248912, 0807248917, 0807248922, 0807248940, 0807248964, 0807248967, 0807248978, 0807248984, 0807249017, 0807249025, 0807249029, 0807249043, 0807249101, 0807249107, 0807249109, 0807249126, 0807249128, 0807249141, 0807249144, 0807249160, 0807249192, 0807249200, 0807249255, 0807249270, 0807249284, 0807249305, 0807249307, 0807249315, 0807249317, 0807249323, 0807250059, 0807250231, 0807250392, 0907251353, 0907251510, 0907251935, 0907252020, 0907252045, 0907252047, 0907252049, 0907252075, 0907252080, 0907252082, 0907252084, 0907252086, 0907252094, 0907252105, 0907252109, 0907252121, 0907252146, 0907252148, 0907252152, 0907252154, 0907252156, 0907252158, 0907252159, 0907252160, 0907252161, 0907252162, 0907252164, 0907252166, 0907252168, 0907252173, 0907252176, 0907252178, 0907252194, 0907252195, 0907252197, 0907252198, 0907252204, 0907252206, 0907252208, 0907252210, 0907252213, 0907252214, 0907252216, 0907252225, 0907252229, 0907252231, 0907252233, 0907252235, 0907252238, 0907252249, 0907252256, 0907252258, 0907252261, 0907252265, 0907252267, 0907252268, 0907252282, 0907252284, 0907252286, 0907252288, 0907252300, 0907252305, 0907252308, 0907252309, 0907252312, 0907252314, 0907252315, 0907252316, 0907252325, 0907252329, 0907252332, 0907252334, 0907252346, 0907252350, 0907252358, 0907252360, 0907252362, 0907252365, 0907252378, 0907252397, 0907252403, 0907252405, 0907252412, 0907252414, 0907252433, 0907252451, 0907253614, 0907253616, 0907253620, 0907253622, 0907253624, 0907253626, 0907253628, 0907253630, 0907253632, 0907253634, 0907253636, 0907253638, 0907253640, 0907253646, 0907253648, 0907253652, 0907253656, 0907253658, 0907253662, 0907254203, 0907254217, 0907254463, 0907254465, 0907254467, 0907254469, 0907254473, 0907254483, 0907254485, 0907254489, 0907254491, 0907254497, 0907254501, 0907254522, 0907254565, 0907254576, 0907254589, 0907254975, 0907254989, 0908287910, 0908287980, 0908287982, 0908288126, 0908288148, 0908288158, 0908288160, 0908288166, 1007257802, 1007257804, 1007257806, 1007257810, 1007257830, 1007257858, 1007257901, 1007257941, 1007257985, 1007258028, 1007258070, 1007258072, 1007258088, 1007258090, 1007258098, 1007258100, 1007258202, 1007258212, 1007258220, 1007258227, 1007258289, 1007258291, 1007258372, 1007258380, 1007258386, 1007258388, 1007258392, 1007258402, 1007258408, 1007258412, 1007258413, 1007258416, 1007258417, 1007258420, 1007258424, 1007258430, 1007258431, 1007258432, 1007258433, 1007258434, 1007258435, 1007258436, 1007258444, 1007258454, 1007258455, 1007258479, 1007258487, 1007258492, 1007258536, 1007258540, 1007258544, 1007258578, 1007258588, 1007258592, 1007258596, 1007258613, 1007258615, 1007258635, 1007258646, 1007258647, 1007258652, 1007258654, 1007258656, 1007258658, 1007258660, 1007258664, 1007258666, 1007258672, 1007258674, 1007258680, 1007258684, 1007258688, 1007258698, 1007258700, 1007258704, 1007258706, 1007258707, 1007258712, 1007258714, 1007258718, 1007258722, 1007258733, 1007258755, 1007258777, 1007258779, 1007258783, 1007258787, 1007258789, 1007258808, 1007258816, 1007258837, 1007258844, 1007258850, 1007258852, 1007258861, 1007258873, 1007258901, 1007258902, 1007258904, 1007258906, 1007258907, 1007258908, 1007258909, 1007258910, 1007258912, 1007258913, 1007258914, 1007258915, 1007258916, 1007258917, 1007258918, 1007258919, 1007258922, 1007258924, 1007258926, 1007258928, 1007258933, 1007258935, 1007258937, 1007258943, 1007258955, 1007258957, 1007258959, 1007258963, 1007258973, 1007258979, 1007258981, 1007258983, 1007258985, 1007258987, 1007258989, 1007258991, 1007258993, 1007258995, 1007258997, 1007258999, 1007259001, 1007259019, 1007259021, 1007259023, 1007259025, 1007259027, 1007259029, 1007259033, 1007259035, 1007259045, 1007259047, 1007259049, 1007259053, 1007259055, 1007259059, 1007259061, 1007259063, 1007259065, 1007259067, 1007259069, 1007259071, 1007259073, 1007259075, 1007259077, 1007259079, 1007259081, 1007259083, 1007259085, 1007259091, 1007259093, 1007259094, 1007259095, 1007259096, 1007259098, 1007259099, 1007259100, 1007259102, 1007259103, 1007259104, 1007259106, 1007259107, 1007259108, 1007259109, 1007259110, 1007259111, 1007259112, 1007259113, 1007259114, 1007259115, 1007259116, 1007259117, 1007259118, 1007259119, 1007259120, 1007259121, 1007259124, 1007259125, 1007259126, 1007259128, 1007259129, 1007259130, 1007259131, 1007259132, 1007259133, 1007259134, 1007259136, 1007259137, 1007259138, 1007259139, 1007259140, 1007259141, 1007259142, 1007259143, 1007259144, 1007259145, 1007259147, 1007259149, 1007259150, 1007259151, 1007259152, 1007259153, 1007259157, 1007259158, 1007259159, 1007259162, 1007259164, 1007259165, 1007259166, 1007259168, 1007259169, 1007259170, 1007259171, 1007259172, 1007259173, 1007259174, 1007259175, 1007259176, 1007259177, 1007259179, 1007259180, 1007259181, 1007259182, 1007259183, 1007259184, 1007259185, 1007259187, 1007259188, 1007259189, 1007259191, 1007259192, 1007259193, 1007259194, 1007259195, 1007259198, 1007259199, 1007259200, 1007259501, 1007259507, 1007259511, 1007259513, 1007259515, 1007259517, 1007259521, 1007259525, 1007259527, 1007259529, 1007259531, 1007259533, 1007259535, 1007259537, 1007259539, 1107259882, 1107260202, 1107260204, 1107260206, 1107260289, 1107260291, 1107260299, 1107261402, 1107261412, 1107261420, 1107261432, 1107261448, 1107261456, 1107261470, 1107261484, 1107261489, 1107261495, 1107261498, 1107261503, 1107261515, 1107261527, 1107261549, 1107261606, 1107261608, 1107261614, 1107261624, 1107261643, 1107261644, 1107261647, 1107261649, 1107261651, 1107261653, 1107261655, 1107261657, 1107261659, 1107261668, 1107261681, 1107261691, 1107262643, 1107263150, 1107263183, 1107263196, 1107263310, 1107263411, 1107263597, 1107263829, 1107263901, 1107263905, 1107263975, 1107264087, 1207266015, 1207266019, 1207266042, 1207266044, 1207266046, 1207266048, 1207266063, 1207266065, 1207266066, 1207266067, 1207266068, 1207266069, 1207266071, 1207266073, 1207266074, 1207266076, 1207266092, 1207266099, 1207266202, 1207266206, 1207266218, 1207266224, 1207266226, 1207266236, 1207266238, 1207266240, 1207266246, 1207266250, 1207266254, 1207266256, 1207266258, 1207266260, 1207266262, 1207266266, 1207266267, 1207266268, 1207266269, 1207266270, 1207266272, 1207266273, 1207266274, 1207266276, 1207266279, 1207266280, 1207266282, 1207266285, 1207266290, 1207266292, 1207266293, 1207266294, 1207266352, 1207266354, 1207266356, 1207266358, 1207266360, 1207266362, 1207266364, 1207266366, 1207266368, 1207266370, 1207266372, 1207266374, 1207266376, 1207266378, 1207266380, 1207266382, 1207266384, 1207266386, 1207266388, 1207266390, 1207266392, 1207266394, 1207266398, 1207266400, 1207266401, 1207266408, 1207266410, 1207266413, 1207266417, 1207266421, 1207266425, 1207266430, 1207266431, 1207266432, 1207266434, 1207266438, 1207266439, 1207266440, 1207266441, 1207266442, 1207266443, 1207266446, 1207266448, 1207266456, 1207266459, 1207266465, 1207266468, 1207266470, 1207266479, 1207266485, 1207266489, 1207266493, 1207266496, 1207266498, 1207266511, 1207266605, 1207266627, 1207266641, 1207266643, 1207266647, 1207266651, 1207266659, 1207266666, 1207266668, 1207266686, 1207266835, 1207267299, 0508275620, 0707243299, 0807249745, 0907251461, 0110990023, 0110990035, 0110990079, 0110990083, 0110990087, 0110990432, 0110990436, 0110990438, 0110990450, 0210990234, 0210990297, 0210990314, 0210990315, 0210990320, 0210990347, 0210990351, 0210990358, 0210990359, 0210990362, 0210990366, 0210990368, 0210990371, 0210990391, 0210990393, 0210990428, 0210990429, 0210990430, 0210990509, 0210990519, 0210990529, 0210990540, 0310990044, 0310990054, 0310990322, 0310990332, 0310990363, 0310990392, 0310990395, 0310990406, 0310990477, 0310990505, 0408274657, 0608279094, 0608279754, 0608280301, 0807246834, 0907252483, 0907253104, 0907254578, 1007255292, 1007255656, 1007255773, 1007255994, 1007256845, 1007257175, 1107261060, 1107263386, 1209990007, 1209990008, 1209990013, 1209990019, 1209990029, 1209990031, 1209990036, 1209990037, 1209990038, 1209990042, 0108268236, 0308273055, 0408274274, 0408274346, 0408274372, 0408274374, 0408274812, 0408274816, 0507235254, 0507235256, 0507235284, 0507235789, 0507235841, 0507236496, 0507236503, 0507236507, 0507236511, 0507236561, 0507236583, 0507236589, 0507236591, 0507236881, 0508275214, 0508275228, 0508275393, 0508275602, 0508275704, 0508275949, 0508276010, 0508276016, 0508276733, 0508276773, 0607240077, 0607240102, 0607240847, 0607240901, 0607242263, 0607242279, 0607242410, 0607242584, 0707243600, 0707243687, 0707244019, 0707244672, 0707245121, 0707245326, 0707245415, 0707245416, 0707245904, 0707246041, 0707246120, 0707246145, 0707246154, 0807246520, 0807246542, 0807247119, 0807247203, 0807247298, 0807247300, 0807247320, 0807247363, 0807247379, 0807247383, 0807247437, 0807247504, 0807247508, 0807247515, 0807247521, 0807247522, 0807247526, 0807247528, 0807247534, 0807247539, 0807247548, 0807247550, 0807247555, 0807247558, 0807247565, 0807247570, 0807247628, 0807247630, 0807247658, 0807247664, 0807247668, 0807247676, 0807247685, 0807247687, 0807247691, 0807247694, 0807247698, 0807247713, 0807247714, 0807247716, 0807247717, 0807247726, 0807247734, 0807247736, 0807247738, 0807247740, 0807247742, 0807247746, 0807247756, 0807247757, 0807247764, 0807247767, 0807247773, 0807247776, 0807247795, 0807247797, 0807247803, 0807247804, 0807247807, 0807247808, 0807247813, 0807247815, 0807247816, 0807247817, 0807247821, 0807247822, 0807247824, 0807247828, 0807247832, 0807247842, 0807247848, 0807247852, 0807247855, 0807247863, 0807247864, 0807247866, 0807247872, 0807247874, 0807247875, 0807247899, 0807247901, 0807247905, 0807247917, 0807247926, 0807248100, 0807248188, 0807248192, 0807248198, 0807248199, 0807248239, 0807248290, 0807248382, 0807248384, 0807248390, 0807248401, 0807248412, 0807248457, 0807248461, 0807248476, 0807248496, 0807248644, 0807248798, 0807248845, 0807248856, 0807248930, 0807248948, 0807248990, 0807249000, 0807249039, 0807249134, 0807249137, 0807249185, 0807249239, 0807249262, 0807249272, 0807249311, 0907252090, 0907252169, 0907252171, 0907252199, 0907252202, 0907252237, 0907252264, 0907252310, 0907252348, 0907252380, 0907252749, 0907253642, 0907254201, 0907254211, 0907254487, 0907254499, 0907254503, 0907254561, 0907254572, 0907254945, 0908287996, 1007258096, 1007258216, 1007258218, 1007258396, 1007258398, 1007258426, 1007258438, 1007258670, 1007258678, 1007259009, 1007259057, 1007259101, 1007259135, 1107261422, 1107261468, 1107261501, 1107261625, 1107261639, 1207266025, 1207266288, 1207266472, 0110990424, 0110990430, 0110990434, 0210990276, 0508275236, 0607241089, 0707245695, 0807246550, 0807246552, 0807246558, 0807246613, 0807246617, 0807246627, 0807246657, 0807246659, 0807246685, 0807246729, 0807246737, 0807246761, 0807246763, 0807246765, 0807246767, 0807247328, 0807247336, 0807247338, 0807247340, 0807247342, 0807247348, 0807247381, 0807247395, 0807247397, 0807247449, 0807247455, 0807247457, 0807247459, 0807247461, 0807247467, 0807247470, 0807247472, 0807247475, 0807247483, 0807247491, 0807247494, 0807247497, 0807247601, 0807247605, 0807247609, 0807247897, 0807247964, 1107260944, 0507235226, 0507235228, 0507235230, 0507235234, 0507235236, 0507235242, 0507235244, 0507235264, 0507235266, 0507235268, 0507235270, 0507235272, 0507235274, 0507235278, 0507235286, 0507235288, 0507235290, 0507235292, 0507235294, 0507235296, 0507235298, 0507235300, 0507235711, 0507235797, 0507235799, 0507235801, 0507235803, 0507235812, 0507235819, 0507235820, 0507235821, 0507235859, 0507235861, 0507235863, 0507235865, 0507235867, 0507235869, 0507235871, 0507235873, 0507235875, 0507235877, 0507235879, 0507235881, 0507235883, 0507235885, 0507235887, 0507235889, 0507235891, 0507235892, 0507235903, 0507235905, 0507235921, 0507235923, 0507235925, 0507235927, 0507235929, 0507235931, 0507235933, 0507235935, 0507235937, 0507235939, 0507235941, 0507235943, 0507235945, 0507235947, 0507235949, 0507235951, 0507235953, 0507235955, 0507235957, 0507235959, 0507235961, 0507235963, 0507235967, 0507235969, 0507235973, 0507235977, 0507235979, 0507235981, 0507235993, 0507235999, 0507236007, 0507236013, 1007257812, 1007257814, 1007257816, 1007257818, 1007257886, 1007257909, 1007257919, 1007257927, 1007258086, 1007258262, 1007258276, 1007258280, 1007258469, 1007258475, 1007258542, 1007258546, 1007258582, 1007258602, 1007258616, 1007258618, 1007258621, 1007258628, 1007258629, 1007258630, 1007258632, 1007258634, 1007258636, 1007258638, 1007258640, 1007258642, 1007258644, 1007258697, 1007258747, 1007258889, 1007258896, 1007259010, 1007259012, 1007259016, 1007259041, 1007259043, 1007259066, 1007259072, 1007259074, 1007259076, 1007259080, 1007259082, 1007259084, 1007259090, 1007259092, 0108270066, 0108270078, 0108270080, 0508275107, 0508275168, 0508275630, 0508275678, 0508275732, 0508276059, 0508276063, 0508276073, 0508276080, 0508276084, 0508276086, 0508276092, 0508276094, 0508276096, 0508276098, 0508276104, 0508276106, 0508276146, 0508276148, 0508276154, 0508276160, 0508276162, 0508276166, 0508276168, 0508276172, 0508276354, 0508276366, 0508276735, 0508276741, 0508276755, 0508276757, 0508276759, 1207264636, 0408274398, 0507236585, 0508275202, 0508275208, 0508275722, 0508276004, 0608280296, 0608280331, 0707243892, 0708282243, 0807247189, 0807249276, 0807249303, 0907252189, 0907252218, 0907252242, 0907252270, 0907252319, 0907252321, 0907252327, 0907252331, 0907252374, 0907254971, 1007257813, 1007258394, 1007258406, 1007259148, 1007259519, 1107263934, 1107263979, 1207266027, 0408274303, 0408274304, 0408274306, 0408274334, 0408274356, 0408274390, 0408274537, 0408274539, 0408274571, 0408274575, 0408274635, 0408274818, 0408274854, 0409300792, 0507235985, 0507235991, 0507236565, 0507236571, 0507236597, 0508275636, 0508275726, 0508275789, 0508275914, 0508276055, 0508276057, 0508276065, 0508276085, 0508276368, 0508276761, 0607240316, 0607240809, 0607240811, 0607240864, 0607240878, 0607240880, 0607240904, 0607241005, 0607241398, 0607241518, 0607241528, 0607241532, 0607241536, 0607241546, 0607241548, 0607241552, 0607241556, 0607242476, 0607242585, 0607242599, 0607242601, 0608279330, 0608279396, 0608279401, 0608279405, 0608279522, 0707243293, 0707243368, 0707243390, 0707243775, 0707246273, 0807246776, 0807246835, 0807246869, 0807246893, 0807246897, 0807246899, 0807246903, 0807246913, 0807246927, 0807246934, 0807246936, 0807246942, 0807246957, 0807246968, 0807246995, 0807247018, 0807247020, 0807247026, 0807247030, 0807247068, 0807247074, 0807247096, 0807247101, 0807247105, 0807247109, 0807247113, 0807247117, 0807247127, 0807247129, 0807247180, 0807247183, 0807247197, 0807247207, 0807247208, 0807247209, 0807247211, 0807247213, 0807247217, 0807247233, 0807247235, 0807247237, 0807247257, 0807247261, 0807247267, 0807247269, 0807247271, 0807247273, 0807247513, 0807247517, 0807247536, 0807247542, 0807247564, 0807247688, 0807249688, 0807249904, 0807249981, 0907251302, 0907251391, 0907251393, 0907251449, 0907254782, 0907254788, 0907254966, 0907254968, 0907255013, 1007256804, 1007256806, 1007256812, 1007256839, 1007256847, 1007256883, 1007256997, 1007257073, 1007257102, 1007257104, 1007257126, 1007257128, 1007257131, 1007257150, 1007257160, 1007257161, 1007257163, 1007257164, 1007257168, 1007257174, 1007257189, 1007257193, 1007257197, 1007257506, 1007257512, 1107260528, 1107261759, 1107262430, 1207264508, 1207264512, 1207264518, 1207264524, 1207264526, 1207264532, 1207264550, 1207264562, 1207264565, 1207264612, 1207264623, 1207264624, 1207264648, 1207264649, 1207264650, 1207264653, 1207264657, 1207264675, 1207264683, 1207264692, 1207264693, 1207264717, 1207264729, 1207264747, 1207264749, 1207264760, 1207264766, 1207264832, 1207264893, 1207264909, 1207264917, 1207264919, 1207264927, 1207264937, 1207264943, 1207264945, 1207264965, 0108267518, 0108267520, 0108267546, 0108267558, 0108267578, 0108267585, 0108267592, 0108267597, 0108267598, 0108267636, 0108267638, 0108267642, 0108267659, 0108267661, 0108267689, 0108267691, 0108267703, 0108267733, 0108267770, 0108267911, 0208270672, 0308272942, 0308272964, 0408274319, 0408274337, 0408274341, 0408274343, 0408274633, 0507235793, 0507235811, 0507235987, 0507235995, 0507236043, 0507236045, 0507236051, 0508275114, 0508275120, 0508275356, 0508275381, 0508275383, 0508275399, 0508275632, 0508275638, 0508275640, 0508275642, 0508275714, 0508275758, 0508275788, 0508275818, 0508275828, 0508275830, 0508275832, 0508275836, 0508275838, 0508275840, 0508275842, 0508275844, 0508275846, 0508275850, 0508275854, 0508275858, 0508275860, 0508276061, 0508276071, 0508276087, 0508276088, 0508276089, 0508276091, 0508276142, 0508276152, 0607239261, 0607239353, 0607240047, 0607240049, 0607240051, 0607240087, 0607240093, 0607240218, 0607240230, 0607240232, 0607240234, 0607240260, 0607240322, 0607240356, 0607240374, 0607240376, 0607240378, 0607241611, 0607241617, 0607241665, 0607241840, 0607241850, 0607241876, 0707243289, 0707243301, 0707243771, 0707243800, 0707244508, 0707245713, 0707245796, 0707245804, 0707246201, 0707246209, 0707246238, 0707246249, 0707246251, 0707246271, 0707246296, 0707246298, 0807248154, 0907251387, 1007255910, 1007255991, 1007256016, 1007256040, 1007256041, 1007256062, 1007256101, 1007256103, 1007256105, 1007256109, 1007256170, 1007256178, 1007256182, 1007256196, 1007257109, 1007257900, 1007257923, 1007258548, 1007258550, 1007258552, 1007258554, 1007258586, 1007258611, 1007258614, 1007258620, 1007258651, 1007258659, 1007258661, 1007258663, 1007258665, 1007258667, 1007258668, 1007258691, 1007258693, 1007258812, 1007258832, 1007258836, 1007258865, 1007258886, 1007258894, 1007259014, 1007259018, 1007259020, 1007259022, 1007259024, 1007259042, 1007259044, 1007259046, 1007259048, 1007259052, 1007259054, 1107261401, 1107261863, 1107262065, 1107262307, 1107262617, 1107262628, 1107262691, 1107263035, 1107263053, 1107263063, 1107263164, 1107263192, 1107263290, 1107263291, 1107263397, 1107263419, 1107263538, 1107263855, 1107263956, 1207264507, 1207264582, 1207264600, 1207264628, 1207264645, 1207264667, 1207264750, 1207264776, 1207264895, 1207264921, 1207264925, 1207264935, 1207264939, 1207265383, 1207265402, 1207265410, 1207265430, 1207265493, 1207265810, 0108267534, 0108267715, 0308272866, 0308272868, 0308272890, 0308272894, 0308272911, 0308272915, 0308272919, 0308272921, 0308272925, 0308272930, 0308272932, 0308272934, 0308272938, 0308272940, 0507235733, 0507235737, 0507236512, 0507236579, 0507236983, 0607238353, 0607240013, 0607240105, 0607240257, 0607240272, 0607240346, 0607242273, 0607242275, 0607242287, 0607242293, 0607242389, 0607242398, 0607242588, 0607242594, 0608279326, 0608279706, 0707243261, 0707243323, 0707243340, 0707243342, 0707243348, 0707243350, 0707243358, 0707243382, 0707243396, 0707243399, 0707243401, 0707243444, 0707243457, 0707243465, 0707243808, 0707243838, 0707243964, 0707243966, 0707243974, 0707243976, 0707243998, 0707244002, 0807246514, 0807246549, 0807246551, 0807246553, 0807246555, 0807246569, 0807246642, 0807246644, 0807246648, 0807246650, 0807246658, 0807248053, 0807248150, 0807248668, 0807249204, 0807250077, 0807250168, 0807250170, 0807250218, 0907251316, 0907251318, 0907251336, 0907251356, 0907251372, 0907251432, 0907251446, 0907251452, 0907251475, 0907251477, 0907251485, 0907251491, 1007256857, 1007257108, 1007257116, 1007257118, 1007257121, 1007257127, 1007257898, 1007257911, 1007258564, 1007258568, 1007258570, 1007258576, 1007258580, 1007258584, 1007258600, 1007258608, 1007258612, 1007258617, 1007258623, 1007258641, 1007258645, 1007258653, 1007258655, 1007258657, 1007258669, 1007258682, 1007258685, 1007258749, 1007258810, 1007258814, 1007258838, 1007258840, 1007258848, 1007258854, 1007258866, 1007258867, 1007258882, 1007258888, 1007258890, 1007258891, 1007258893, 1007258900, 1007259005, 1007259032, 1007259036, 1007259038, 1107261439, 1107261735, 1107261799, 1107261819, 1107261857, 1107261859, 1207264809, 1207265469, 1207265477, 1207267443, 0408274265, 0607240046, 0607240062, 1207264690, 0108267747, 0508275038, 0508275044, 0508275046, 0508275048, 0508275054, 0508275056, 0508275060, 0508275062, 0508275064, 0508275068, 0508275076, 0508275080, 0508275082, 0508275084, 0508275086, 0508275088, 0508275090, 0508275092, 0508275094, 0508275096, 0508275098, 0508275100, 0508275102, 0508275104, 0508275106, 0508275108, 0508275110, 0508275112, 0508275118, 0508275122, 0508275124, 0508275126, 0508275128, 0508275130, 0508275132, 0508275134, 0508275136, 0508275138, 0508275140, 0508275142, 0508275144, 0508275146, 0508275148, 0508275150, 0508275152, 0508275154, 0508275156, 0508275158, 0508275160, 0508275162, 0508275164, 0508275166, 0508275170, 0508275172, 0508275174, 0508275176, 0508275178, 0508275180, 0508275182, 0508275184, 0508275186, 0508275188, 0508275190, 0508275192, 0508275194, 0508275196, 0508275401, 0508275403, 0508275407, 0508275802, 0508275806, 0508275848, 0508275852, 0508275866, 0508275873, 0508275877, 0508275879, 0508275881, 0508275883, 0508275885, 0508275887, 0508275889, 0508275891, 0508275893, 0508275895, 0508277099, 0607239438, 0607240172, 0607240839, 0607241894, 0607241911, 0607242542, 0608277602, 0608277604, 0608277612, 0608277662, 0608277666, 0608277680, 0608277693, 0608277699, 0608278067, 0707243676, 0707243678, 0707244011, 0707244012, 0707244014, 0707244015, 0707244016, 0707244017, 0707244018, 0707244020, 0707244021, 0707244022, 0707244023, 0707244024, 0707244025, 0707244026, 0707244027, 0707244028, 0707244030, 0707244031, 0707244032, 0707244033, 0707244035, 0707244037, 0707244040, 0707244041, 0707244042, 0707244043, 0707244045, 0707244046, 0707244048, 0707244050, 0707244052, 0707244080, 0707244082, 0707244098, 0707244101, 0707244401, 0707244405, 0707244410, 0707244423, 0707244427, 0707244429, 0707244431, 0707244435, 0707244438, 0707244439, 0707244441, 0707244443, 0707244445, 0707244447, 0707244449, 0707244451, 0707244453, 0707244457, 0707244459, 0707244461, 0707244463, 0707244465, 0707244469, 0707244473, 0707244475, 0707244477, 0707244481, 0707244487, 0707244489, 0707244491, 0707244495, 0707244503, 0707244892, 0707245639, 0707245641, 0707245645, 0707245647, 0707245657, 0707245663, 0707245703, 0707245707, 0707245708, 0707245711, 0707245712, 0707245714, 0707245715, 0707245716, 0707245717, 0707245722, 0707245726, 0707245738, 0707245740, 0707245744, 0707245746, 0707245750, 0707245752, 0707245756, 0707245758, 0707245762, 0707245764, 0707245768, 0707245770, 0707245772, 0707245774, 0707245776, 0707245778, 0707245780, 0707245782, 0707246003, 0707246106, 0707246107, 0707246109, 0707246111, 0707246113, 0707246115, 0707246117, 0707246118, 0707246119, 0707246121, 0707246123, 0707246124, 0707246125, 0707246126, 0707246127, 0707246129, 0707246130, 0707246131, 0707246132, 0707246133, 0707246134, 0707246135, 0707246136, 0707246137, 0707246138, 0707246141, 0707246142, 0707246143, 0707246147, 0707246150, 0707246151, 0707246153, 0707246155, 0707246157, 0707246158, 0707246159, 0707246160, 0707246161, 0707246162, 0707246163, 0707246164, 0707246165, 0707246166, 0707246167, 0707246168, 0707246169, 0707246170, 0707246171, 0707246172, 0707246173, 0707246174, 0707246175, 0707246177, 0707246179, 0707246181, 0707246183, 0707246184, 0707246185, 0707246186, 0707246187, 0707246188, 0707246189, 0707246190, 0707246191, 0707246192, 0707246193, 0707246194, 0707246195, 0707246196, 0707246197, 0707246198, 0707246199, 0707246202, 0707246203, 0707246204, 0707246205, 0707246208, 0707246218, 0707246220, 0707246222, 0707246224, 0707246226, 0707246228, 0707246230, 0707246232, 0707246236, 0707246242, 0707246244, 0707246246, 0707246248, 0707246250, 0707246256, 0707246258, 0707246260, 0707246264, 0707246268, 0707246269, 0707246272, 0707246277, 0707246279, 0707246280, 0707246282, 0707246284, 0707246286, 0707246292, 0807247444, 0807248001, 0807248204, 0807248224, 0807248232, 0807248243, 0807248245, 0807248247, 0807248255, 0807248263, 0807248271, 0807248273, 0807248275, 0807248277, 0807248283, 0807248285, 0807248287, 0807248289, 0807248293, 0807248367, 0807248402, 0807248404, 0807248406, 0807248408, 0807248410, 0807248422, 0807248424, 0807248428, 0807248430, 0807248434, 0807248436, 0807248438, 0807248440, 0807248446, 0807248452, 0807248478, 0807248488, 0807248491, 0807249049, 0807249051, 0807249071, 0807249601, 0807249605, 0807249609, 0807249611, 0807249614, 0807249615, 0807249616, 0807249617, 0807249619, 0807249620, 0807249621, 0807249623, 0807249624, 0807249626, 0807249627, 0807249628, 0807249630, 0807249632, 0807249633, 0807249634, 0807249635, 0807249636, 0807249637, 0807249638, 0807249639, 0807249640, 0807249641, 0807249642, 0807249643, 0807249645, 0807249646, 0807249647, 0807249648, 0807249649, 0807249652, 0807249655, 0807249659, 0807249672, 0807249674, 0807249676, 0807249681, 0807249685, 0807249689, 0807249699, 0807249802, 0807249804, 0807249816, 0807249818, 0807249822, 0807249828, 0807249830, 0807249834, 0807249836, 0807249844, 0807249866, 0807249868, 0807249870, 0807249872, 0807249876, 0807249882, 0807249886, 0807249888, 0807249890, 0807249892, 0807249894, 0807249896, 0807249901, 0807249903, 0807249905, 0807249907, 0807249911, 0807249912, 0807249913, 0807249915, 0807249917, 0807249919, 0807249921, 0807249925, 0807249933, 0807249937, 0807249957, 0807249977, 0807249979, 0807249999, 0807250187, 0907254766, 0907255014, 0907255016, 0907255024, 0907255028, 0907255030, 0907255032, 0907255034, 0907255036, 0907255040, 0907255042, 0907255044, 0907255052, 0907255057, 0907255060, 0907255061, 0907255066, 0907255067, 0907255069, 0907255072, 0907255074, 0907255076, 0907255077, 0907255084, 0907255086, 0907255089, 0907255090, 0907255094, 0907255100, 0907255102, 0907255110, 0907255113, 0907255122, 0907255135, 0907255137, 0907255143, 0907255145, 0907255147, 0907255149, 0907255153, 0907255155, 0907255156, 0907255157, 0907255158, 0907255159, 0907255162, 0907255163, 0907255164, 0907255165, 0907255168, 0907255171, 0907255173, 0907255175, 0907255177, 0907255181, 0907255183, 0907255185, 0907255189, 1007255203, 1007255207, 1007255210, 1007255227, 1007255229, 1007255241, 1007255249, 1007255307, 1007255317, 1007255371, 1007256446, 1007256450, 1007256452, 1007256454, 1007256472, 1007256498, 1007256500, 1007256589, 1007256602, 1007256605, 1007256625, 1007256627, 1007256631, 1007256641, 1007256644, 1007256648, 1007256668, 1007256670, 1007256672, 1007256681, 1007256684, 1007256688, 1007256692, 1007256694, 1007256700, 1007256710, 1007256722, 1007256724, 1007256739, 1007256741, 1007256746, 1007256752, 1007256767, 1007256769, 1007256773, 1007256777, 1007256781, 1007256799, 1007256933, 1007256940, 1007256950, 1007257006, 1007257016, 1007257028, 1007257030, 1007257032, 1007257515, 1007257579, 1007258969, 1007259013, 1007259559, 1107260975, 1107261261, 1107261263, 1107261271, 1107261273, 1107261382, 1107261450, 1107261461, 1107261467, 1107261476, 1107261483, 1107261524, 1107261530, 1107261532, 1107261533, 1107261535, 1107261538, 1107261543, 1107261545, 1107261546, 1107261547, 1107261557, 1107261575, 1107261577, 1107261590, 1107261592, 1107261594, 1107261595, 1107261607, 1107261616, 1107261629, 1107261636, 1107261665, 1107261674, 1107261675, 1107261676, 1107261678, 1107261708, 1107261714, 1107261716, 1107261730, 1107261739, 1107261745, 1107261749, 1107261757, 1107261761, 1107261763, 1107261768, 1107261797, 1107261801, 1107261802, 1107261804, 1107261805, 1107261806, 1107261810, 1107261811, 1107261813, 1107261821, 1107261825, 1107261826, 1107261827, 1107261828, 1107261829, 1107261830, 1107261831, 1107261832, 1107261833, 1107261834, 1107261838, 1107261840, 1107261842, 1107261844, 1107261846, 1107261847, 1107261849, 1107261850, 1107261852, 1107261853, 1107261854, 1107261855, 1107261856, 1107261858, 1107261861, 1107261862, 1107261864, 1107261865, 1107261870, 1107261871, 1107261872, 1107261873, 1107261879, 1107261883, 1107261893, 1107261895, 1107261903, 1107261907, 1107261909, 1107262255, 1107262429, 1107262648, 1107262811, 1107263186, 1107263328, 1107263362, 1107263364, 1107263762, 1107263770, 1107263772, 1107263790, 1107263822, 1107263845, 1107263846, 1107263877, 1107263883, 1107263885, 1107263887, 1107263889, 1107263891, 1107263893, 1107263897, 1107263942, 1107263973, 1107264001, 1107264038, 1107264043, 1107264049, 1107264051, 1107264053, 1107264056, 1107264059, 1107264061, 1107264063, 1107264065, 1107264067, 1107264069, 1107264070, 1107264071, 1107264073, 1107264075, 1107264077, 1107264079, 1107264088, 1107264144, 1107264148, 1107264160, 1107264182, 1107264186, 1107264188, 1107264190, 1107264192, 1107264194, 1107264196, 1107264198, 1107264200, 1107264202, 1107264203, 1107264206, 1107264207, 1107264208, 1107264209, 1107264210, 1107264211, 1107264217, 1107264219, 1107264221, 1107264222, 1107264223, 1107264224, 1107264225, 1107264227, 1107264228, 1107264229, 1107264230, 1107264231, 1107264233, 1107264235, 1107264236, 1107264237, 1107264238, 1107264239, 1107264240, 1107264242, 1107264243, 1107264247, 1107264251, 1107264253, 1107264261, 1107264263, 1107264265, 1107264271, 1107264273, 1107264275, 1107264277, 1107264279, 1107264281, 1107264283, 1107264287, 1107264289, 1107264293, 1107264297, 1107264299, 1107264302, 1107264304, 1107264306, 1107264310, 1107264314, 1107264316, 1107264320, 1107264322, 1107264324, 1107264326, 1107264328, 1107264330, 1107264332, 1107264334, 1107264336, 1107264338, 1107264340, 1107264342, 1107264344, 1107264346, 1107264348, 1107264352, 1107264354, 1107264356, 1107264358, 1107264360, 1107264366, 1107264370, 1107264374, 1107264376, 1107264378, 1207264517, 1207264521, 1207264523, 1207264527, 1207264529, 1207264531, 1207264535, 1207264537, 1207264539, 1207264540, 1207264541, 1207264543, 1207264546, 1207264547, 1207264551, 1207264553, 1207264554, 1207264556, 1207264557, 1207264566, 1207264568, 1207264569, 1207264570, 1207264573, 1207264574, 1207264577, 1207264578, 1207264581, 1207264584, 1207264586, 1207264599, 1207264602, 1207264626, 1207264629, 1207264631, 1207264651, 1207264691, 1207264702, 1207264712, 1207264724, 1207264738, 1207264741, 1207264742, 1207264745, 1207264752, 1207264758, 1207264762, 1207264764, 1207264778, 1207264781, 1207264782, 1207264783, 1207264792, 1207264797, 1207264807, 1207264811, 1207264813, 1207264845, 1207264848, 1207264854, 1207264877, 0508275333, 0508276093, 0108267530, 0108267532, 0108267588, 0108267617, 0108267701, 0108267707, 0108267708, 0108267711, 0108267717, 0108267723, 0108267725, 0108267731, 0108267746, 0208270692, 0208270698, 0208270805, 0308272936, 0308273710, 0308273712, 0308273714, 0308273718, 0308273732, 0308273738, 0308273740, 0308273742, 0308273744, 0308273746, 0308273748, 0308273752, 0308273754, 0308273756, 0308273758, 0308273760, 0308273762, 0308273764, 0308273787, 0308273789, 0308273791, 0308273797, 0408274200, 0408274251, 0408274271, 0408274307, 0408274342, 0408274348, 0408274350, 0408274352, 0408274363, 0408274402, 0408274406, 0408274549, 0408274553, 0408274555, 0408274557, 0408274561, 0408274567, 0408274569, 0408274573, 0408274810, 0408274884, 0408274886, 0408274898, 0507235262, 0507235725, 0507235805, 0507235807, 0507235809, 0507235855, 0507235907, 0507236333, 0507236508, 0507236981, 0508275137, 0508275145, 0508275147, 0508275339, 0508275366, 0508275397, 0508275740, 0508275744, 0508275797, 0508275799, 0508275856, 0508276067, 0508276144, 0508276350, 0508276356, 0607238233, 0607238249, 0607238352, 0607238359, 0607238393, 0607238406, 0607238410, 0607238412, 0607238610, 0607238615, 0607238621, 0607238625, 0607238646, 0607238650, 0607238654, 0607238676, 0607238721, 0607238752, 0607238756, 0607238762, 0607238774, 0607238776, 0607238821, 0607238858, 0607238930, 0607238942, 0607238944, 0607238952, 0607239063, 0607239281, 0607239331, 0607239929, 0607239935, 0607239940, 0607239952, 0607239956, 0607240008, 0607240088, 0607240113, 0607240151, 0607240160, 0607240174, 0607240188, 0607241190, 0607241194, 0607241281, 0607241344, 0607241348, 0607241350, 0607241352, 0607242207, 0607242280, 0607242307, 0607242362, 0607242384, 0607242392, 0607242613, 0608278933, 0608279370, 0608279524, 0707243259, 0707243291, 0707243765, 0707243769, 0707244559, 0707244776, 0707245261, 0707245652, 0707245656, 0707245658, 0707245660, 0707245662, 0707245664, 0707245666, 0707245668, 0707245670, 0707245672, 0707245674, 0707245676, 0707245680, 0707245688, 0707245696, 0707245700, 0707245702, 0707245704, 0707245871, 0707245943, 0707245945, 0707245947, 0707245949, 0707245951, 0707245957, 0707245961, 0707245965, 0707245975, 0707245977, 0707245979, 0707246128, 0707246293, 0807246904, 0807246914, 0807246920, 0807246926, 0807246930, 0807246939, 0807246944, 0807246946, 0807246948, 0807246950, 0807246952, 0807246956, 0807246960, 0807246962, 0807246964, 0807246966, 0807246967, 0807246972, 0807246974, 0807246975, 0807246976, 0807246980, 0807246982, 0807246984, 0807246986, 0807246988, 0807246990, 0807246992, 0807246993, 0807246998, 0807247000, 0807247102, 0807247106, 0807247108, 0807247110, 0807247112, 0807247114, 0807247116, 0807247120, 0807247123, 0807247124, 0807247125, 0807247126, 0807247128, 0807247130, 0807247135, 0807247139, 0807247140, 0807247142, 0807247144, 0807247145, 0807247146, 0807247147, 0807247149, 0807247150, 0807247151, 0807247152, 0807247153, 0807247154, 0807247155, 0807247156, 0807247158, 0807247160, 0807247161, 0807247162, 0807247163, 0807247164, 0807247168, 0807247177, 0807247178, 0807247201, 0807247210, 0807247215, 0807247227, 0807247239, 0807247259, 0807247263, 0807247277, 0807247279, 0807247281, 0807247289, 0807247295, 0807247435, 0807247488, 0807247530, 0807247572, 0807247578, 0807247600, 0807247701, 0807247702, 0807247704, 0807247706, 0807247708, 0807247710, 0807247720, 0807247728, 0807247743, 0807247749, 0807247752, 0807247754, 0807247759, 0807247761, 0807247766, 0807247768, 0807247771, 0807247778, 0807247784, 0807247786, 0807247788, 0807247799, 0807247800, 0807247812, 0807247823, 0807247825, 0807247827, 0807247833, 0807247834, 0807247835, 0807247837, 0807247839, 0807247840, 0807247841, 0807247844, 0807247845, 0807247849, 0807247850, 0807247851, 0807247854, 0807247856, 0807247858, 0807247873, 0807247877, 0807247879, 0807247880, 0807247883, 0807247889, 0807247890, 0807247891, 0807247893, 0807247906, 0807247910, 0807247911, 0807247912, 0807247914, 0807247916, 0807247918, 0807247921, 0807247923, 0807247929, 0807247936, 0807247937, 0807247941, 0807247942, 0807247945, 0807247946, 0807247950, 0807247965, 0807247967, 0807247971, 0807247973, 0807247978, 0807247991, 0807248004, 0807248012, 0807248013, 0807248015, 0807248019, 0807248023, 0807248061, 0807248072, 0807248078, 0807248126, 0807248179, 0807248185, 0807248307, 0807248312, 0807248314, 0807248316, 0807248318, 0807248348, 0807250560, 0807250562, 0807250913, 0907251389, 0907251443, 0907251587, 1007256801, 1007256802, 1007256803, 1007256808, 1007256810, 1007256811, 1007256814, 1007256816, 1007256817, 1007256818, 1007256819, 1007256820, 1007256824, 1007256826, 1007256830, 1007256835, 1007256836, 1007256837, 1007256840, 1007256841, 1007256844, 1007256846, 1007256849, 1007256850, 1007256851, 1007256853, 1007256859, 1007256870, 1007256872, 1007256874, 1007256877, 1007256881, 1007256885, 1007256887, 1007256890, 1007257063, 1007257066, 1007257067, 1007257068, 1007257079, 1007257081, 1007257085, 1007257087, 1007257089, 1007257110, 1007257122, 1007257130, 1007257132, 1007257134, 1007257136, 1007257148, 1007257151, 1007257152, 1007257154, 1007257158, 1007257162, 1007257166, 1007257170, 1007257172, 1007257173, 1007257176, 1007257178, 1007257184, 1007257191, 1007257195, 1007257529, 1007258942, 1007258960, 1007258970, 1007258974, 1007259229, 1007259233, 1007259235, 1007259561, 1207264559, 1207264611, 1207265276, 1207265278, 1207265280, 1207265282, 1207265284, 1207265286, 1207265294, 1207265794, 1207265796, 1207265803, 1207265805, 1207265807, 1207265813, 1207265817, 1207265819, 1207265823, 1207265825, 1207265827, 1207265831, 1207265847, 1207265849, 1207265855, 1207265859, 1207265865, 1207265873, 1207265877, 1207266181, 1207267451, 1207267455, 1207267461, 0108267508, 0108267574, 0108267576, 0108267587, 0108267590, 0108267599, 0408274259, 0508275580, 0807249874, 0907251338, 0907251342, 0907251344, 0907251352, 0907251379, 0907251383, 0907251451, 0907251453, 0907251497, 1007257913, 1007257915, 1007257921, 1007258532, 1007258534, 1007258560, 1007258566, 1007258573, 1007258574, 1007258598, 1007258601, 1007258625, 1007258649, 1007258675, 1007258806, 1007258817, 1007258819, 1007258834, 1007258856, 1007258858, 1007258862, 1007258864, 1007258874, 1007258876, 1007258877, 1007258898, 1007258932, 1007258936, 1007258940, 1007258946, 1007258950, 1007258986, 1007258988, 1007258992, 1007258996, 1007259002, 1007259006, 1007259008, 1007259203, 1007259207, 1007259223, 1007259249, 1007259253, 1007259261, 1007259263, 1007259265, 1007259277, 1007259279, 1007259285, 1007259508, 1007259510, 1007259512, 1007259577, 1007259579, 1007259583, 1007259589, 1007259593, 1007259597, 1007259599, 1107262515, 1107262521, 1107262529, 1107262557, 1107262563, 1107262565, 1107262599, 1107262906, 1107262964, 1207265485, 1207265499, 0507236559, 0607242278, 0707246408, 0807246515, 1007255516, 0210990360, 0210990523, 0210990531, 0707244941, 0310990176, 0310990203, 0310990213, 0310990214, 0310990237, 0310990269, 0310990270, 0310990272, 0310990295, 0310990297, 0507236506, 1107260654, 1107263057, 0110990148, 0310990323, 1107261823, 0210990405, 0110990098, 0110990101, 0110990103, 0110990115, 0110990117, 0110990121, 0110990129, 0110990133, 0110990151, 0110990155, 0110990157, 0110990161, 0110990163, 0110990171, 0110990173, 0110990175, 0110990177, 0110990179, 0110990181, 0110990185, 0110990187, 0110990189, 0110990191, 0110990201, 0110990203, 0210990070, 0210990079, 0210990083, 0210990087, 0210990097, 0210990099, 0508275474, 0508277095, 0607239707, 0907252670, 1007255724, 1007255788, 1007257372, 1007258301, 1107262288, 1209990134, 1209990162, 1209990165, 1209990171, 0310990313, 0310990326, 0310990383, 0310990330, 0310990331, 0310990336, 0310990359, 0310990360, 0310990385, 0310990386, 0310990394, 0310990397, 0110990172, 0210990528, 0210990532, 0210990536, 0310990046, 0310990301, 0310990303, 0310990307, 0310990316, 0310990319, 0310990367, 0310990370, 0310990373, 0310990380, 0310990381, 0310990382, 0310990409, 0310990410, 0310990413, 0310990414, 0310990419, 0310990421, 0310990424, 0907252260, 1007258082, 0310990369, 0310990180, 0310990302, and 0310990308.
FEI Number 1000524572
Recalling Firm/
Manufacturer
Philips Medical Systems
2301 5th Ave Ste 200
Seattle WA 98121
For Additional Information Contact
206-664-5290
Manufacturer Reason
for Recall
Failure of the voltage detector, U8 may cause the AED battery to drain more rapidly than normal or make the AED unusable.
FDA Determined
Cause 2
Nonconforming Material/Component
ActionPhilips Medical Systems issued an Urgent Medical Device Recall letter dated May 28, 2010 identifying the affected device and actions to be taken. Customers were instructed to review their inventory for affected devices. If affected devices are found, the customer should contact Philips to arrange for a replacement device. The customer should also run a battery insertion test. If the device passes the test, the AED may remain in service until a replacement is received. For further information, contact Philips at 1 800 263-3342, option 6.
Quantity in Commerce72419 (20419 are subject to this recall) INCLUDES ALL MODELS/BRANDS
DistributionWorldwide Distribution: Throughout the USA, and to the following countries: ALBANIA, ARGENTINA, ARUBA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BERMUDA, BOSNIA AND HERZEGOVINA, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, FINLAND, FRANCE, GERMANY, HONG KONG, ICELAND, INDIA, INDONESIA, IRELAND, ISRAEL, ITALY, JAPAN, JORDAN, KENYA, REPUBLIC OF KOREA, KUWAIT, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PERU, PORTUGAL, PUERTO RICO, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, and VIETNAM.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
510(K) Database510(K)s with Product Code = MKJ
510(K)s with Product Code = MKJ
-
-