• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Class 2 Device Recall ADVANTAGE Turbo 2Button Shaver

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
 Class 2 Device Recall ADVANTAGE Turbo 2Button Shaversee related information
Date Initiated by FirmOctober 29, 2009
Date PostedJuly 17, 2010
Recall Status1 Terminated 3 on August 19, 2010
Recall NumberZ-2026-2010
Recall Event ID 54521
Product Classification Motor, surgical instrument, ac-powered - Product Code GEY
ProductADVANTAGE Turbo 2-Button Shaver. Rx Only, Made In USA, NON STERILE. CONMED LINVATEC, 11311 CONCEPT BLVD, LARGO, FL 33773-4908 USA.
Code Information Lot Number:  00001000 00001023 00001024 00001025 00001026 00001027 00001028 00001029 00001030 00001022 00001031 00001032 00001033 00001034 00001035 00001036 00001037 00001038 00001039 00001040 00001041 00001043 00001042 00001044 00001045 00001046 00001047 00001048 00001050 00001051 00001052 00001049 00001053 00001054 00001055 00001056 00001057 00001058 00001059 00001060 00001062 00001063 00001064 00001065 00001061 00001066 00001067 00001068 00001069 00001070 00001071 00001072 00001073 00001074 00001075 00001076 00001077 00001078 00001079 00001080 00001081 00001082 00001083 00001084 00001085 00001086 00001087 00001088 00001089 00001090 00001091 00001096 00001098 00001099 00001100 00001101 00001102 00001103 00001104 00001105 00001106 00001107 00001108 00001097 00001109 00001110 00001111 00001112 00001114 00001115 00001116 00001117 00001118 00001119 00001120 00001121 00001122 00001113 00001123 00001124 00001125 00001126 00001127 00001129 00001130 00001128 00001131 00001132 00001133 00001134 00001135 00001136 00001137 00001138 00001139 00001140 00001141 00001142 00001143 00001144 00001145 00001146 00001147 00001148 00001149 00001150 00001151 00001152 00001153 00001154 00001155 00001156 00001157 00001158 00001160 00001161 00001162 00001163 00001164 00001165 00001166 00001167 00001168 00001169 00001170 00001171 00001172 00001173 00001174 00001175 00001176 00001177 00001159 00001178 00001179 00001180 00001181 00001182 00001183 00001185 00001186 00001187 00001188 00001189 00001190 00001191 00001192 00001193 00001194 00001195 00001196 00001197 00001198 00001200 00001201 00001202 00001203 00001204 00001205 00001206 00001207 00001208 00001209 00001210 00001211 00001212 00001213 00001214 00001216 00001215 00001217 00001218 00001219 00001220 00001221 00001222 00001223 00001224 00001225 00001226 00001227 00001228 00001229 00001230 00001231 00001232 00001233 00001235 00001234 00001236 00001237 00001238 00001239 00001240 00001241 00001242 00001243 00001244 00001245 00001246 00001247 00001248 00001249 00001250 00001252 0001251 00001184 00001253 00001254 00001255 00001256 00001257 00001258 00001260 00001261 00001263 00001264 00001265 00001266 00001267 00001199 00001259 00001262 00001268 00001269 00001270 00001271 00001272 00001273 00001274 00001276 00001275 00001277 00001278 00001279 00001280 00001281 00001282 00001283 00001284 00001285 00001286 00001287 00001288 00001289 00001290 00001291 00001292 00001293 00001294 00001295 00001296 00001297 00001298 00001299 00001300 00001301 00001302 00001303 00001304 00001305 00001306 00001307 00001309 00001310 00001311 00001312 00001313 00001314 00001315 00001316 00001308 00001317 00001318 00001319 00001320 00001321 00001322 00001323 00001324 00001325 00001326 00001327 00001328 00001329 00001330 00001331 00001332 00001333 00001334 00001335 00001336 00001337 00001338 00001339 00001340 00001341 00001342 00001343 00001344 00001345 00001346 00001347 00001348 00001349 00001350 00001351 00001352 00001353 00001355 00001354 00001356 00001357 00001358 00001359 00001360 00001361 00001362 00001363 00001364 00001365 00001366 00001367 00001368 00001369 00001370 00001371 00001372 00001373 00001374 00001375 00001376 00001377 00001378 00001379 00001380 00001381 00001382 00001383 00001384 00001385 00001386 00001387 00001388 00001389 00001390 00001391 00001392 00001393 00001394 00001395 00001396 00001397 00001398 00001399 00001400 00001401 00001402 00001403 00001404 00001405 00001406 00001407 00001408 00001409 00001410 00001411 00001412 00001413 00001414 00001415 00001416 00001417 00001418 00001419 00001420 00001421 00001422 00001423 00001424 00001426 00001427 00001428 00001429 00001430 00001431 00001432 0001425 00001433 00001434 00001435 00001436 00001437 00001438 00001439 00001440 00001441 00001442 00001443 00001444 00001445 00001446 00001447 00001448 00001449 00001450 00001451 00001452 00001453 00001454 00001455 00001456 00001457 00001458 00001459 00001460 00001461 00001462 00001463 00001464 00001465 00001466 00001467 00001468 00001469 00001470 00001471 00001472 00001473 00001474 00001475 00001476 00001477 00001478 00001479 00001482 00001483 0001481 00001480 00001484 00001485 00001486 00001487 00001488 00001489 00001490 00001491 00001492 00001493 00001494 00001496 00001497 00001495 00001498 00001499 00001500 00001501 00001502 00001503 00001504 00001505 00001506 00001507 00001508 00001509 00001511 00001512 00001513 00001514 00001515 00001516 00001517 00001518 00001519 00001520 00001521 00001522 00001523 00001524 00001525 00001526 00001527 00001528 00001529 00001530 00001531 00001532 00001533 00001534 00001535 00001536 00001537 00001538 00001539 00001540 00001541 00001542 00001543 00001544 00001545 00001546 00001547 00001548 00001549 00001550 00001551 00001552 00001553 00001554 00001555 00001556 00001557 00001558 00001559 00001560 00001561 00001562 00001563 00001564 00001565 00001566 00001567 00001568 00001569 00001570 00001571 00001572 00001573 00001574 00001575 00001576 00001577 00001578 00001579 00001580 00001581 00001582 00001583 00001584 00001585 00001586 00001587 00001588 00001589 00001590 00001591 00001592 00001593 00001595 00001594 00001596 00001597 00001598 00001599 00001600 00001601 00001602 00001603 00001604 00001605 00001606 00001607 00001608 00001609 00001610 00001611 00001612 00001613 00001614 00001615 00001616 00001617 00001618 00001619 00001620 00001621 00001622 00001623 00001624 00001625 00001626 00001627 00001628 00001629 00001630 00001631 00001632 00001633 00001634 00001635 00001636 00001637 00001638 00001639 00001640 00001641 00001642 00001643 00001644 00001645 00001646 00001647 00001648 00001649 00001650 00001651 00001652 00001653 00001654 00001655 00001656 00001657 00001658 00001659 00001660 00001661 00001662 00001663 00001664 00001665 00001666 00001667 00001668 00001669 00001670 00001671 00001672 00001673 00001674 00001675 00001676 00001677 00001678 00001679 00001680 00001681 00001682 00001683 00001684 00001685 00001686 00001687 00001688 00001689 00001690 00001691 00001692 00001693 00001694 00001695 00001696 00001697 00001698 00001699 00001701 00001702 00001703 00001704 00001705 00001707 00001708 00001709 00001700 00001710 00001711 00001712 00001713 00001714 00001715 00001716 00001717 00001718 00001720 00001721 00001722 00001723 00001724 00001725 00001726 00001727 00001728 00001729 00001730 00001731 00001732 00001733 00001734 00001735 00001736 00001737 00001738 00001739 00001742 00001743 00001719 00001744 00001746 00001748 00001749 00001750 00001751 00001752 00001753 00001754 00001755 00001756 00001757 00001758 00001759 00001760 00001761 00001762 00001764 00001706 00001745 00001763 00001765 00001766 00001767 00001768 00001769 00001770 00001771 00001772 00001773 00001774 00001775 00001776 00001777 00001778 00001779 00001741 00001780 00001781 00001782 00001783 00001784 00001785 00001786 00001787 00001789 00001790 00001793 00001794 00001795 00001796 00001797 00001798 00001740 00001788 00001791 00001792 00001799 00001800 00001801 00001802 00001803 00001804 00001805 00001806 00001807 00001808 00001809 00001810 00001811 00001812 00001813 00001814 00001815 00001816 00001817 00001818 00001819 00001820 00001821 00001822 00001823 00001824 00001825 00001826 00001827 00001828 00001829 00001830 00001831 00001832 00001833 00001834 00001835 00001836 00001837 00001838 00001839 00001840 00001841 00001842 00001843 00001844 00001845 00001846 00001847 00001848 00001849 00001850 00001851 00001852 00001853 00001854 00001855 00001856 00001857 00001859 00001860 00001861 00001862 00001858 00001863 00001864 00001865 00001866 00001867 00001868 00001869 00001870 00001871 00001872 00001873 00001874 00001875 00001876 00001877 00001878 00001879 00001880 00001881 00001882 00001883 00001884 00001885 00001886 00001887 00001888 00001889 00001890 00001891 00001893 00001894 00001895 00001897 00001898 00001899 00001900 00001901 00001902 00001903 00001904 00001905 00001906 00001907 00001910 00001911 00001912 00001908 00001909 00001913 00001914 00001915 00001916 00001917 00001918 00001919 00001920 00001921 00001922 00001923 00001924 00001925 00001926 00001927 00001928 00001929 00001930 00001931 00001932 00001933 00001934 00001935 00001936 00001937 00001892 00001896 00001938 00001940 00001941 00001942 00001943 00001944 00001945 00001946 00001947 00001948 00001949 00001950 00001951 00001952 00001953 00001954 00001955 00001956 00001957 00001958 00001959 00001960 00001939 00001961 00001962 00001963 00001964 00001965 00001966 00001967 00001968 00001969 00001970 00001971 00001972 00001973 00001974 00001975 00001976 00001977 00001978 00001979 00001980 00001981 00001982 00001983 00001984 00001985 00001986 00001987 00001988 00001989 00001990 00001991 00001992 00001993 00001994 00001995 00001996 00001997 00001998 00001999 00002000 00002001 00002002 00002003 00002004 00002005 00002006 00002007 00002008 00002009 00002010 00002011 00002012 00002013 00002014 00002015 00002016 00002017 00002018 00002019 00002020 00002021 00002022 00002023 00002024 00002025 00002026 00002027 00002028 00002029 00002030 00002031 00002032 00002033 00002034 00002035 00002036 00002037 00002038 00002039 00002040 00002041 00002042 00002043 00002044 00002045 00002046 00002047 00002048 00002049 00002050 00002051 00002052 00002053 00002054 00002055 00002056 00002057 00002058 00002059 00002060 00002061 00002062 00002063 00002064 00002065 00002066 00002067 00002068 00002069 00002070 00002071 00002072 00002073 00002074 00002075 00002076 00002077 00002078 00002079 00002080 00002081 00002082 00002083 00002084 00002085 00002086 00002087 00002088 00002089 00002090 00002091 00002092 00002093 00002094 00002095 00002096 00002097 00002098 00002099 00002100 00002101 00002102 00002103 00002104 00002105 00002106 00002107 00002108 00002109 00002110 00002111 00002112 00002113 00002114 00002115 00002116 00002117 00002118 00002119 00002120 00002121 00002122 00002123 00002124 00002125 00002126 00002127 00002128 00002129 00002130 00002131 00002132 00002133 00002134 00002135 00002136 00002137 00002138 00002139 00002140 00002141 00002142 00002143 00002144 00002145 00002146 00002147 00002148 00002149 00002150 00002151 00002152 00002153 00002154 00002155 00002156 00002157 00002158 00002159 00002160 00002161 00002162 00002163 00002164 00002165 00002166 00002167 00002168 00002169 00002170 00002171 00002172 00002173 00002174 00002176 00002177 00002178 00002179 00002180 00002181 00002182 00002183 00002184 00002185 00002186 00002187 00002188 00002189 00002190 00002191 00002192 00002193 00002194 00002195 00002196 00002197 00002198 00002199 00002200 00002201 00002202 00002203 00002204 00002205 00002206 00002207 00002208 00002209 00002210 00002211 00002212 00002213 00002214 00002215 00002216 00002217 00002218 00002219 00002220 00002221 00002222 00002223 00002224 00002225 00002226 00002227 00002228 00002229 00002230 00002231 00002232 00002233 00002234 00002235 00002236 00002237 00002238 00002239 00002240 00002241 00002242 00002243 00002244 00002245 00002246 00002247 00002248 00002249 00002250 00002251 00002252 00002253 00002254 00002255 00002256 00002257 00002259 00002260 00002261 00002262 00002258 00002263 00002264 00002265 00002266 00002267 00002268 00002269 00002270 00002271 00002272 00002273 00002274 00002275 00002276 00002277 00002278 00002279 00002280 00002281 00002282 00002283 00002284 00002285 00002286 00002287 00002288 00002289 00002290 00002291 00002292 00002293 00002294 00002295 00002296 00002297 00002298 00002299 00002300 00002301 00002302 00002303 00002304 00002305 00002306 00002307 00002308 00002309 00002310 00002311 00002312 00002313 00002314 00002315 00002316 00002317 00002318 00002319 00002320 00002321 00002323 00002324 00002325 00002326 00002327 00002328 00002329 00002330 00002331 00002332 00002333 00002334 00002335 00002336 00002337 00002338 00002339 00002340 00002342 00002343 00002341 00002344 00002345 00002346 00002347 00002348 00002349 00002350 00002351 00002352 00002353 00002354 00002355 00002356 00002357 00002358 00002359 00002360 00002361 00002362 00002363 00002364 00002365 00002366 00002367 00002368 00002369 00002370 00002371 00002372 00002373 00002374 00002375 00002376 00002377 00002378 00002379 00002380 00002381 00002382 00002383 00002384 00002385 00002386 00002387 00002388 00002389 00002390 00002391 00002392 00002393 00002394 00002395 00002396 00002397 00002398 00002399 00002400 00002401 00002402 00002403 00002404 00002405 00002406 00002407 00002408 00002409 00002410 00002411 00002412 00002413 00002414 00002415 00002416 00002417 00002418 00002419 00002420 00002421 00002422 00002423 00002424 00002425 00002426 00002427 00002428 00002429 00002430 00002431 00002432 00002433 00002434 00002436 00002437 00002438 00002439 00002440 00002441 00002442 00002443 00002445 00002446 00002447 00002435 00002444 00002448 00002449 00002450 00002451 00002452 00002453 00002454 00002455 00002456 00002457 00002458 00002459 00002460 00002461 00002462 00002463 00002464 00002465 00002466 00002467 00002468 00002469 00002470 00002471 00002472 00002473 00002474 00002475 00002476 00002477 00002478 00002479 00002480 00002481 00002482 00002483 00002484 00002485 00002486 00002487 00002488 00002489 00002490 00002491 00002492 00002493 00002494 00002497 00002495 00002496 00002498 00002499 00002500 00002501 00002502 00002503 00002504 00002505 00002506 00002507 00002508 00002509 00002510 00002511 00002512 00002513 00002514 00002515 00002516 00002517 00002518 00002519 00002520 00002521 00002522 00002523 00002524 00002525 00002526 00002527 00002528 00002529 00002530 00002531 00002532 00002533 00002534 00002535 00002536 00002537 00002538 00002539 00002540 00002541 00002542 00002543 00002544 00002545 00002546 00002547 00002548 00002549 00002550 00002175 00002551 00002552 00002553 00002554 00002555 00002556 00002557 00002558 00002559 00002560 00002561 00002563 00002562 00002564 00002565 00002566 00002567 00002568 00002569 00002570 00002571 00002572 00002573 00002322 00002574 00002575 00002576 00002577 00002578 00002579 00002580 00002581 00002582 00002583 00002584 00002585 00002586 00002587 00002588 00002589 00002590 00002591 00002592 00002593 00002594 00002595 00002596 00002598 00002597 00002599 00002600 00002601 00002602 00002603 00002604 00002605 00002606 00002607 00002608 00002609 00002610 00002611 00002612 00002613 00002614 00002615 00002616 00002617 00002618 00002619 00002620 00002621 00002622 00002623 00002624 00002625 00002626 00002627 00002628 00002629 00002631 00002632 00002630 00002633 00002634 00002635 00002636 00002637 00002638 00002639 00002640 00002641 00002642 00002643 00002644 00002645 00002646 00002647 00002648 00002649 00002650 00002651 00002652 00002653 00002654 00002655 00002656 00002657 00002658 00002659 00002661 00002662 00002663 00002664 00002665 00002666 00002660 00002667 00002668 00002669 00002670 00002671 00002672 00002673 00002674 00002675 00002676 00002677 00002678 00002680 00002682 00002679 00002681 00002683 00002684 00002685 00002686 00002687 00002688 00002689 00002690 00002691 00002692 00002693 00002694 00002695 00002696 00002697 00002698 00002699 00002700 00002701 00002702 00002703 00002704 00002705 00002706 00002707 00002708 00002709 00002710 00002711 00002712 00002713 00002714 00002715 00002716 00002717 00002718 00002719 00002720 00002721 00002722 00002723 00002724 00002725 00002726 00002727 00002728 00002729 00002730 00002731 00002732 00002733 00002734 00002735 00002736 00002737 00002738 00002739 00002740 00002741 00002742 00002743 00002744 00002745 00002746 00002747 00002748 00002749 00002750 00002751 00002752 00002753 00002754 00002755 00002756 00001510 00002757 00002758 00002759 00002760 00002761 00002762 00002763 00002764 00002765 00002766 00002767 00002768 00002769 00002770 00002771 00002772 00002773 00002774 00002775 00002776 00002777 00002778 00002779 00002780 00002781 00002782 00002783 00002784 00002785 00002786 00002787 00002788 00002789 00002790 00002791 00002792 00002793 00002794 00002795 00002798 00002796 00002797 00002799 00002801 00002802 00002803 00002800 00002804 00002805 00002806 00002807 00002808 00002809 00002810 00002811 00002812 00002813 00002814 00002815 00002816 00002817 00002818 00002819 00002820 00002821 00002822 00002823 00002824 00002825 00002826 00002827 00002828 00002829 00002830 00002831 00002832 00002833 00002834 00002835 00002836 00002837 00002838 00002839 00002840 00002841 00002842 00002843 00002844 00002845 00002846 00002847 00002848 00002849 00002850 00002851 00002852 00002853 00002854 00002855 00002856 00002857 00002858 00002859 00002860 00002861 00002862 00002863 00002864 00002865 00002866 00002867 00002868 00002869 00002870 00002871 00002872 00002873 00002874 00002875 00002876 00002877 00002878 00002880 00002881 00002882 00002883 00002884 00002885 00002886 00002879 00002887 00002888 00002889 00002890 00002891 00002892 00002893 00002894 00002895 00002896 00002897 00002898 00002899 00002900 00002901 00002902 00002903 00002904 00002905 00002906 00002907 00002908 00002909 00002910 00002911 00002912 00002913 00002914 00002915 00002916 00002917 00002918 00002919 00002920 00002921 00002922 00002923 00002924 00002925 00002926 00002927 00002928 00002929 00002930 00002931 00002932 00002933 00002934 00002935 00002936 00002937 00002938 00002939 00002940 00002941 00002942 00002943 00002944 00002945 00002946 00002947 00002948 00002949 00002950 00002951 00002952 00002953 00002954 00002955 00002956 00002957 00002958 00002959 00002960 00002961 00002962 00002963 00002964 00002965 00002966 00002967 00002968 00002969 00002970 00002971 00002972 00002973 00002974 00002975 00002976 00002977 00002978 00002979 00002980 00002981 00002982 00002983 00002984 00002985 00002986 00002987 00002988 00002989 00002990 00002992 00002993 00002994 00002995 00002996 00002997 00002991 00002998 00002999 00003000 00003001 00003002 00003003 00003004 00003005 00003006 00003007 00003008 00003009 00003010 00003011 00003012 00003013 00003014 00003015 00003016 00003017 00003018 00003019 00003020 00003021 00003022 00003023 00003024 00003025 00003026 00003027 00003028 00003029 00003030 00003031 00003032 00003033 00003034 00003035 00003036 00003037 00003038 00003039 00003040 00003041 00003042 00003043 00003044 00003045 00003046 00003047 00003048 00003049 00003050 00003051 00003052 00003053 00003054 00003055 00003056 00003057 00003058 00003059 00003060 00003061 00003062 00003063 00003064 00003065 00003066 00003067 00003068 00003069 00003070 00003071 00003072 00003073 00003074 00003075 00003076 00003077 00003078 00003079 00003080 00003081 00003082 00003083 00003084 00003085 00003086 00003087 00003088 00003089 00003090 00003091 00003092 00003093 00003094 00003095 00003096 00003097 00003098 00003099 00003100 00003101 00003102 00003103 00003104 00003105 00003106 00003107 00003108 00003109 00003110 00003111 00003112 00003113 00003114 00003115 00003116 00003117 00003118 00003119 00003120 00003121 00003122 00003123 00003124 00003125 00003126 00003127 00003128 00003129 00003130 00003131 00003132 00003133 00003134 00003135 00003136 00003137 00003138 00003139 00003140 00003141 00003142 00003143 00003144 00003145 00003146 00003147 00003148 00003149 00003150 00003151 00003152 00003153 00003154 00003155 00003156 00003157 00003158 00003159 00003160 00003161 00003162 00003163 00003164 00003165 00003166 00003167 00003168 00003169 00003170 00003171 00003172 00003173 00003174 00003175 00003176 00003177 00003178 00003179 00003180 00003181 00003182 00003183 00003184 00003185 00003186 00003187 00003188 00003189 00003190 00003191 00003192 00003193 00003194 00003195 00003196 00003197 00003198 00003199 00003200 00003201 00003202 00003203 00003204 00003205 00003206 00003207 00003208 00003209 00003210 00003211 00003212 00003213 00003214 00003215 00003216 00003218 00003219 00003220 00003217 00003221 00003222 00003223 00003224 00003225 00003226 00003227 00003228 00003229 00003230 00003231 00003232 00003233 00003234 00003235 00003236 00003237 00003238 00003239 00003240 00003241 00003242 00003243 00003244 00003245 00003246 00003247 00003248 00003249 00003250 00003251 00003252 00003253 00003254 00003255 00003256 00003257 00003258 00003259 00003260 00003261 00003262 00003263 00003264 00003265 00003266 00003267 00003268 00003269 00003270 00003271 00003272 00003273 00003274 00003275 00003276 00003277 00003278 00003279 00003280 00003281 00003282 00003283 00003284 00003285 00003286 00003287 00003288 00003289 00003290 00003291 00003292 00003293 00003294 00003295 00003296 00003297 00003298 00003299 00003300 00003301 00003302 00003303 00003304 00003305 00003306 00003307 00003308 00003309 00003310 00003311 00003312 00003313 00003314 00003315 00003316 00003317 00003318 00003319 00003320 00003321 00003322 00003323 00003324 00003325 00003326 00003327 00003328 00003329 00003330 00003331 00003332 00003333 00003334 00003335 00003336 00003337 00003338 00003339 00003340 00003341 00003342 00003343 00003344 00003345 00003346 00003347 00003348 00003349 00003350 00003351 00003352 00003353 00003354 00003355 00003356 00003357 00003360 00003361 00003558 00003559 00003358 00003359 00003362 00003363 00003364 00003365 00003366 00003367 00003368 00003369 00003370 00003371 00003372 00003373 00003374 00003375 00003376 00003377 00003378 00003379 00003380 00003381 00003382 00003383 00003384 00003385 00003386 00003387 00003388 00003389 00003390 00003391 00003392 00003393 00003394 00003395 00003396 00003397 00003398 00003399 00003400 00003401 00003402 00003403 00003404 00003405 00003406 00003407 00003408 00003409 00003410 00003411 00003412 00003413 00003414 00003415 00003416 00003417 00003418 00003419 00003420 00003421 00003422 00003423 00003424 00003425 00003426 00003427 00003428 00003429 00003430 00003431 00003432 00003433 00003434 00003435 00003436 00003438 00003439 00003440 00003441 00003437 00003442 00003443 00003444 00003445 00003446 00003447 00003448 00003449 00003450 00003451 00003452 00003453 00003454 00003455 00003456 00003457 00003458 00003459 00003460 00003461 00003462 00003463 00003464 00003465 00003466 00003467 00003468 00003469 00003470 00003471 00003472 00003473 00003475 00003476 00003477 00003478 00003479 00003480 00003481 00003482 00003483 00003484 00003485 00003486 00003487 00003488 00003489 00003490 00003491 00003492 00003493 00003494 00003495 00003496 00003497 00003498 00003499 00003500 00003501 00003502 00003503 00003504 00003505 00003506 00003507 00003508 00003509 00003510 00003511 00003512 00003513 00003514 00003515 00003516 00003517 00003518 00003519 00003520 00003521 00003522 00003523 00003524 00003525 00003526 00003528 00003529 00003530 00003531 00003532 00003533 00003534 00003535 00003536 00003537 00003538 00003539 00003540 00003541 00003542 00003543 00003544 00003545 00003546 00003547 00003548 00003549 00003550 00003551 00003552 00003553 00003554 00003555 00003556 00003557 00003560 00003561 00003562 00003563 00003564 00003565 00003566 00003567 00003568 00003569 00003570 00003571 00003572 00003573 00003574 00003575 00003576 00003577 00003578 00003579 00003581 00003582 00003583 00003584 00003585 00003586 00003587 00003580 00003588 00003589 00003590 00003591 00003592 00003593 00003596 00003594 00003595 00003597 00003598 00003599 00003600 00003601 00003602 00003603 00003604 00003605 00003606 00003474 00003607 00003608 00003609 00003610 00003611 00003612 00003613 00003614 00003615 00003616 00003617 00003618 00003619 00003620 00003621 00003622 00003623 00003624 00003625 00003626 00003627 00003628 00003629 00003630 00003631 00003632 00003633 00003634 00003635 00003636 00003638 00003527 00003637 00003639 00003640 00003641 00003642 00003643 00003644 00003645 00003646 00003647 00003648 00003649 00003650 00003651 00003652 00003653 00003654 00003655 00003656 00003657 00003658 00003659 00003660 00003661 00003662 00003663 00003664 00003665 00003666 00003667 00003668 00003669 00003670 00003671 00003672 00003673 00003674 00003675 00003676 00003677 00003678 00003679 00003680 00003681 00003682 00003683 00003684 00003685 00003686 00003687 00003688 00003689 00003690 00003691 00003692 00003693 00003694 00003695 00003696 00003697 00003698 00003699 00003700 00003701 00003702 00003703 00003704 00003705 00003706 00003707 00003708 00003709 00003710 00003711 00003712 00003713 00003714 00003715 00003716 00003717 00003718 00003719 00003720 00003721 00003722 00003723 00003724 00003725 00003726 00003727 00003728 00003729 00003730 00003731 00003732 00003733 00003734 00003735 00003736 00003737 00003738 00003739 00003740 00003741 00003742 00003743 00003744 00003745 00003746 00003747 00003748 00003749 00003750 00003751 00003752 00003753 00003754 00003755 00003756 00003757 00003758 00003759 00003760 00003761 00003762 00003763 00003764 00003765 00003766 00003767 00003768 00003769 00003770 00003771 00003772 00003773 00003774 00003775 00003776 00003777 00003778 00003779 00003780 00003781 00003782 00003783 00003784 00003785 00003786 00003787 00003788 00003789 00003790 00003791 00003792 00003793 00003794 00003795 00003796 00003797 00003798 00003799 00003800 00003801 00003802 00003803 00003804 00003805 00003806 00003807 00003808 00003809 00003810 00003811 00003812 00003813 00003814 00003815 00003816 00003817 00003818 00003819 00003820 00003821 00003822 00003823 00003824 00003825 00003826 00003827 00003828 00003829 00003830 00003831 00003832 00003833 00003834 00003835 00003836 00003837 00003838 00003839 00003840 00003841 00003842 00003843 00003845 00003846 00003844 00003847 00003848 00003849 00003850 00003851 00003852 00003853 00003854 00003855 00003856 00003857 00003858 00003859 00003860 00003861 00003862 00003863 00003864 00003865 00003866 00003867 00003868 00003869 00003870 00003871 00003872 00003873 00003874 00003875 00003876 00003877 00003878 00003879 00003880 00003881 00003882 00003883 00003884 00003885 00003886 00003887 00003888 00003889 00003890 00003891 00003892 00003893 00003894 00003895 00003896 00003897 00003898 00003899 00003900 00003901 00003902 00003903 00003904 00003905 00003906 00003907 00003908 00003909 00003910 00003911 00003912 00003913 00003914 00003915 00003916 00003917 00003918 00003919 00003920 00003921 00003922 00003923 00003924 00003925 00003926 00003927 00003928 00003929 00003930 00003931 00003932 00003933 00003934 00003935 00003936 00003937 00003938 00003939 00003940 00003941 00003944 00003943 00003945 00003946 00003947 00003948 00003949 00003950 00003951 00003952 00003953 00003954 00003955 00003956 00003957 00003958 00003959 00003960 00003942 
FEI Number 1017294
Recalling Firm/
Manufacturer
Linvatec Corp. dba ConMed Linvatec
11311 Concept Blvd
Largo FL 33773-4908
For Additional Information Contact
727-392-6464
Manufacturer Reason
for Recall
ConMed Linvatec is recalling Advantage, 2-Button Turbo Handpiece Catalog Number D9924. These products may self-activate due to moisture intrusion into the housing body. This issue has been isolated to products manufactured between June 7, 2004 and June 1, 2006.
FDA Determined
Cause 2
Other
ActionConMed Linvatech notified affected customers by letter sent by FedEx next day delivery. Customers were asked to complete a response form and indicated if they have the product or have transferred the product. A Service Request number will be assigned for the returned product. Additional questions are directed to the firm's Customer Service department at 800-535-8536.
Quantity in Commerce2,373
DistributionWorldwide distribution. Products were distributed domestically throughout the U.S. Also United Arab Emirates, Argentina, Australia, Belgium, Switzerland, Chile, China, Germany, Denmark, Spain, France, United Kingdom, Hong Kong, India, Italy, Japan, Lithuania, Morocco, Netherlands, New Zealand, Singapore, Turkey, South Africa.
Total Product Life CycleTPLC Device Report

1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
-
-