| Class 2 Device Recall SCORPIOFLEX PS TIBIAL TRIAL, |  |
Date Initiated by Firm | July 13, 2010 |
Date Posted | March 16, 2011 |
Recall Status1 |
Terminated 3 on September 29, 2017 |
Recall Number | Z-1683-2011 |
Recall Event ID |
56332 |
Product Classification |
Prosthesis, knee, patellofemorotibial, semi-constrained, cemented, polymer/metal/polymer - Product Code LXH
|
Product | SCORPIO-FLEX PS TIBIAL TRIAL, non sterile, made in the USA
Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430.
The Tibial trial are utilized during surgery to assess the overall component fit, ligament stability, and joint range of motion prior to final implantation of the insert to the tray. |
Code Information |
T72-5-0308A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 3, 8 MM T72-5-0310A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 3, 10 MM T72-5-0312A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 3, 12 MM T72-5-0315A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 3, 15 MM T72-5-0318A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 3, 18 MM T72-5-0321A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 3, 21 MM T72-5-0324A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 3, 24 MM T72-5-0508A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 5, 8 MM T72-5-0510A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 5, 10 MM T72-5-0512A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 5, 12 MM T72-5-0515A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 5, 15 MM T72-5-0518A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 5, 18 MM T72-5-0521A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 5, 21 MM T72-5-0524A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 5, 24 MM T72-5-0708A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 7, 8 MM T72-5-0710A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 7, 10 MM T72-5-0712A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 7, 12 MM T72-5-0715A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 7, 15 MM T72-5-0718A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 7, 18 MM T72-5-0721A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 7, 21 MM T72-5-0724A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 7, 24 MM T72-5-0908A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 9, 8 MM T72-5-0910A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 9, 10 MM T72-5-0912A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 9, 12 MM T72-5-0915A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 9, 15 MM T72-5-0918A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 9, 18 MM T72-5-0921A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 9, 21 MM T72-5-0924A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 9, 24 MM T72-5-1108A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 11, 8 MM T72-5-1110A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 11, 10 MM T72-5-1112A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 11, 12 MM T72-5-1115 SCORPIO SUPERFLEX PS TIB.TRIAL, SIZE 11, 15 MM T72-5-1115A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 11, 15 MM T72-5-1118A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 11, 18 MM T72-5-1121A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 11, 21 MM T72-5-1124A SCORPIO-FLEX PS TIBIAL TRIAL, SIZE 11, 24 MM T72-3-1108M INSERTOS DE PRUEBA, SIZE 11, 8 MM T72-3-1110M INSERTOS DE PRUEBA, SIZE 11, 10 MM All lots |
Recalling Firm/ Manufacturer |
Stryker Howmedica Osteonics Corp. 325 Corporate Drive Mahwah NJ 07430
|
For Additional Information Contact | Richard Wolyn 201-972-2100 |
Manufacturer Reason for Recall | Stryker Orthopaedics became aware that there is the potential for the Scorpio T72 tibial Insert Trials to be oversized by a maximum of .024" (0.6mm). |
FDA Determined Cause 2 | Other |
Action | Stryker Howmedica Osteonics Corp.sent Urgent Product Correction letters Correction letters dated July 23, 2010, via Federal Express to all affected customers. were sent to all Stryker Branches on July 23, 2010 via Federal Express. Asia Pacific, Canada, Europe, Middle East, Africa, Japan and Latin America sites were sent letters on July 26, 2010 by Federal Express. The letters identified the product, the problem, and the action to be taken by the customer.
Customers were instructed to complete and return the attached Product Correction Acknowledgement Form by fax as soon as possible to 201-831-6069.
For any questions customers were to call 201-972-2100 or 201-831-5028. |
Quantity in Commerce | 159,095 all types |
Distribution | Worldwide Distribution - USA including OH, TX, MS, and CO and the countries of Australia, Hong Kong, Taiwan, India, Korea, Japan, Chile, Brazil, Columbia, Venezuela, , Panama, Santo Domingo, Mexico, Argentina, Costa Rica, Guatemala, Peru,Sweden, Germany, Netherlands, France, Spain, Italy, Romania, South Africa, United Kingdom. and Canada. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
|
|
|