|
Class 2 Device Recall Solar 8000i with Patient Data Module / Transport Pro Monitor |
 |
Date Initiated by Firm |
October 05, 2010 |
Date Posted |
November 15, 2010 |
Recall Status1 |
Terminated 3 on May 17, 2012 |
Recall Number |
Z-0363-2011 |
Recall Event ID |
56461 |
510(K)Number |
K071073
|
Product Classification |
Monitor, physiological, patient(with arrhythmia detection or alarms) - Product Code MHX
|
Product |
GE Healthcare Transport Pro¿ Monitor with the CARESCAPE" Patient Data Module. |
Code Information |
Parent SN AAD05390701GA AAD05390702GA AAD05390703GA AAD05390704GA AAD05390705GA AAD05390706GA AAD06031132GA AAD06031137GA AAD06131461GA AAD06131462GA AAD06131466GA AAD06131468GA AAD06131472GA AAD06171539GA AAD06221773GA AAD06221776GA AAD06221777GA AAD06251813GA AAD06251830GA AAD06251851GA AAD06251853GA AAD06251856GA AAD06251857GA AAD06261923GA AAD06261927GA AAD06281937GA AAD06362122GA AAD06372145GA SCS06500630GA SCS06500645GA SCS06500649GA SCS06500661GA SCS06500663GA SCS07010713GA SCS07080892GA SCS07080901GA SCS07201199GP SCS07201200GP SCS07201201GP SCS07201202GP SCS07201203GP SCS07201204GP SCS07201205GP SCS07201206GP SCS07201207GP SCS07281360GP SCS07281361GP SCS07281362GP SCS07281363GP SCS07281364GP SCS07281365GP SCS07281366GP SCS07281367GP SCS07281368GP SCS07281369GP SCS07281370GP SCS07281371GP SCS07281372GP SCS07281373GP SCS07281374GP SCS07391691GA SCS07441939GA SCS08383087GA SCS07482026GA SCS08493368GA SCS08493387GA SCS07221315GA SCS07221316GA SCS07221317GA SCS07221322GA SCS07361513GA SCS07361519GA SCS09464289GA SCS08352985GA SCS08352989GA SCS08362994GA SCS08362997GA SCS08362998GA SCS08363002GA SCS08363009GA SCS08363010GA SCS08363011GA SCS08363012GA SCS08363013GA SCS08363015GA SCS08363017GA SCS08363018GA SCS08363024GA SCS08423226GA SCS08122513GA SCS08142588GA SCS08142590GA SCS08142595GA SCS08142597GA SCS08162602GA SCS08292852GA SCS08292856GA SCS06440374GA AAD06261865GA AAD06261879GA AAD06362136GA AAD06362142GA AAD06372146GA AAD06372160GA AAD06372162GA AAD06372163GA AAD06372165GA SCS07371545GA SCS07371547GA SCS07461979GA SCS08132536GA SCS09173726GA AAD06031155GA AAD05430764GA AAD05430776GA AAD05430777GA SCS08393161GA SCS08413170GA SCS08413173GA SCS07080875GA SCS07080884GA SCS07080907GA SCS07391631GA SCS07391683GA SCS09474313GA SCS09474319GA SCS09474327GA SCS09303900GA SCS08172648GA SCS08192660GA SCS08192662GA SCS08202677GA SCS08202678GA SCS08212695GA SCS08212698GA SCS08212699GA SCS08473302GA SCS08473306GA SCS08473307GA SCS08473308GA SCS08473310GA SCS08473313GA SCS08473321GA SCS08473325GA SCS08473327GA SCS08473331GA SCS09454267GA SCS09474338GA SCS09153713GA SCS09173719GA SCS09333987GA SCS09333988GA SCS09333989GA SCS09333990GA SCS09333991GA SCS09333992GA SCS09333993GA SCS09333995GA SCS09333996GA SCS09333997GA SCS09424201GA SCS09424202GA SCS09263820GA SCS09263821GA SCS09263822GA SCS09263823GA SCS09263824GA SCS09263825GA SCS09263828GA SCS07431852GA AAD05240213GA AAD05310398GA AAD05310399GA AAD05310400GA AAD05310401GA AAD05310402GA AAD05310403GA AAD05310404GA AAD05310405GA AAD05430743GA AAD05430744GA AAD05430797GA AAD05430798GA AAD05521016GA AAD05521017GA AAD06171503GA AAD06171511GA AAD06171512GA AAD06171513GA AAD06171521GA AAD06171522GA AAD06221761GA AAD06221780GA AAD06221783GA AAD06231785GA SCS06390095GA SCS06500654GA SCS06500659GA SCS07482033GA SCS08312920GA SCS08312921GA SCS08312922GA SCS08312938GA SCS08493414GA SCS08513453GA SCS08513454GA SCS09073623GA SCS09183734GA SCS09183736GA SCS09193750GA SCS09193751GA SCS09193756GA SCS09193757GA SCS09193758GA SCS09203759GA SCS09203761GA SCS09203764GA SCS09203765GA SCS09203766GA SCS09203769GA SCS09203770GA SCS09203771GA SCS09203772GA SCS09203774GA SCS09203777GA SCS09203779GA SCS09203780GA SCS09203783GA SCS09203785GA SCS09203788GA SCS09233789GA SCS09233790GA SCS09233792GA SCS09233794GA SCS09233795GA SCS09233796GA SCS09233797GA SCS09233798GA SCS09233802GA SCS09233803GA SCS09233804GA SCS09233805GA SCS09233806GA SCS09243807GA SCS09243815GA SCS09333958GA SCS09333964GA SCS09333967GA SCS09444231GA SCS09444234GA SCS09444238GA SCS09444239GA SCS09444250GA SCS09444251GA SCS09514363GA SCS10174625GA SCS10174629GA SCS08132558GA SCS08132562GA SCS07391644GA SCS07391651GA SCS08082361GA SCS08122493GA SCS08112432GA SCS10074433GA AAD06131475GA
AAD06131479GA SCS09233801GA SCS08092379GA SCS08162603GA SCS06420278GA SCS06420279GA SCS06420285GA SCS06420287GA AAD06261897GA SCS07441928GA SCS07441966GA SCS07482061GA SCS07482077GA SCS07482081GA SCS07482082GA SCS07482083GA SCS08252812GA SCS08262839GA SCS09514373GA SCS09514386GA SCS09514395GA SCS07391723GA SCS07391732GA SCS07391733GA SCS08132528GA AAD06031144GA AAD06071219GA AAD06071220GA AAD06071224GA AAD06111362GA AAD06111406GA AAD06121429GA AAD06121443GA AAD06121444GA AAD06171567GA AAD06171569GA AAD06171571GA AAD06171577GA AAD06171579GA AAD06171580GA AAD06171581GA AAD06171582GA AAD06171583GA AAD06171585GA AAD06181590GA AAD06181591GA AAD06181592GA AAD06181593GA AAD06181594GA AAD06181595GA AAD06181605GA AAD06181606GA AAD06181607GA AAD06181609GA AAD06181610GA AAD06181611GA AAD06181612GA AAD06181613GA AAD06181614GA AAD06181619GA AAD06191620GA AAD06191623GA AAD06191624GA AAD06191625GA AAD06191626GA AAD06191628GA AAD06191629GA AAD06201676GA AAD06201677GA AAD06201678GA AAD06201679GA SCS06430351GA SCS06440369GA SCS06480514GA SCS06480522GA SCS06480524GA SCS06480532GA SCS06480535GA SCS06480536GA SCS06480537GA SCS06480540GA SCS06480545GA SCS06480546GA SCS07080895GA SCS07080902GA SCS07090912GA SCS07171075GA SCS07391700GA SCS07391765GA SCS07391768GA SCS07391771GA SCS07462006GA SCS07522160GA SCS07522162GA SCS07522172GA SCS07522176GA SCS07522177GA SCS07522178GA AAD05210186GA AAD05210187GA AAD05210188GA AAD05220189GA AAD05220190GA AAD05220193GA AAD05430736GA AAD05430738GA AAD05430748GA AAD05430759GA AAD05430779GA AAD05450824GA AAD05450825GA AAD05450831GA AAD05450838GA AAD05450841GA AAD05450844GA AAD05460871GA AAD05521063GA AAD06171517GA AAD06171520GA AAD06332071GA AAD06332074GA AAD06342075GA AAD06342076GA AAD06342077GA AAD06342078GA AAD06342079GA AAD06342080GA AAD06342082GA AAD06342087GA AAD06342090GA AAD06352097GA AAD06352098GA AAD06352099GA AAD06352100GA AAD06352101GA AAD06352102GA AAD06352103GA AAD06352104GA AAD06352105GA AAD06352106GA AAD06352107GA AAD06352108GA AAD06352109GA AAD06352110GA AAD06352111GA AAD06352112GA AAD06352113GA SCS06390193GA SCS07090922GA SCS07110946GA SCS07441908GA SCS07441911GA SCS07441914GA SCS07441916GA SCS07441920GA SCS08242781GA SCS08242783GA SCS08242788GA SCS08242796GA SCS08393114GA SCS08393125GA SCS09173716GA AAD05340459GA AAD05340460GA AAD05340461GA AAD05340462GA AAD05340463GA AAD05340464GA AAD05340466GA AAD05340467GA AAD05340468GA AAD05340469GA AAD05340470GA AAD05340471GA AAD05350472GA AAD05350473GA AAD05350474GA AAD05350475GA AAD05350483GA AAD05350484GA AAD05350486GA AAD05350487GA AAD05350488GA AAD05350489GA AAD05350492GA AAD05350493GA AAD05350497GA AAD05350498GA AAD05350500GA AAD05350501GA AAD05350502GA AAD05360503GA AAD05360504GA AAD05360505GA AAD05360507GA AAD05360537GA AAD05360545GA AAD05380636GA AAD05430746GA AAD05450848GA SCS08202676GA SCS08202682GA SCS09023559GA SCS09023562GA SCS09143686GA SCS09364031GA SCS09364032GA AAD05330445GA AAD05330447GA SCS10174618GA SCS08413185GA SCS08413196GA SCS08423209GA SCS08423218GA SCS08503425GA SCS08503426GA SCS09093627GA SCS09093634GA SCS07391752GA SCS07391761GA SCS07391762GA SCS07391763GA SCS07391769GA SCS07391620GA SCS06500613GA SCS06500629GA SCS07030749GA SCS07030750GA SCS07030751GA SCS07030754GA SCS07030756GA SCS07030761GA SCS07030765GA SCS07040767GA SCS07040768GA SCS07040770GA SCS07040771GA SCS07040774GA SCS07040780GA SCS08373035GA SCS09073615GA SCS09153707GA SCS09454269GA SCS09454276GA SCS06430356GA SCS07171076GA SCS08473332GA SCS08473334GA SCS08473335GA SCS08493336GA SCS09193753GA AAD06041184GA SCS08393127GA SCS08393140GA SCS08132527GA SCS08132530GA AAD05240238GA SCS08042245GA AAD06121423GA AAD06121427GA SCS09153710GA SCS09153714GA AAD05240250GA AAD05250260GA AAD05250265GA SCS08252809GA SCS08252831GA SCS08292869GA SCS08292872GA SCS08363021GA SCS08363025GA SCS08363026GA SCS08373032GA SCS08252837GA SCS08302885GA SCS08302888GA SCS08162607GA SCS08162613GA SCS08172628GA SCS08212710GA SCS08232729GA SCS08242784GA SCS08242787GA SCS08252816GA SCS07171071GA SCS08252807GA SCS08302893GA SCS09283836GA SCS09293855GA SCS09404171GA SCS09404172GA SCS09404173GA SCS09404174GA SCS09404175GA SCS09414180GA SCS09414181GA SCS09414184GA SCS09414185GA SCS09414186GA SCS09414187GA SCS09454263GA SCS07371568GA SCS07371570GA SCS07371573GA SCS07371578GA SCS07371581GA SCS07371584GA SCS07371587GA SCS07371593GA SCS07371595GA SCS07371598GA SCS07381601GA SCS08433264GA SCS08493355GA SCS09023555GA SCS09023556GA SCS09023557GA SCS09023558GA SCS09023560GA SCS09023561GA SCS09143687GA SCS09143688GA SCS09364012GA SCS09364016GA SCS09364024GA SCS09364025GA SCS09364026GA SCS09364028GA SCS09364029GA SCS09364030GA SCS09364035GA SCS09444232GA SCS09444233GA SCS09444235GA SCS09444236GA SCS10144477GA SCS10164593GA SCS06430314GA SCS06430315GA SCS06430316GA SCS06430317GA SCS06430319GA SCS06460480GA SCS06460487GA SCS06460491GA SCS07311436GA SCS07311438GA SCS07311440GA SCS07311441GA SCS07321452GA SCS07321456GA SCS07321457GA SCS07482036GA SCS07482038GA SCS07482045GA SCS07482046GA SCS07502130GA SCS07502131GA SCS07512147GA SCS07512152GA SCS07522156GA SCS07522166GA SCS07522168GA SCS07522170GA SCS07522179GA SCS08012185GA SCS08032221GA SCS08132542GA SCS08132544GA SCS08142568GA SCS08142570GA SCS08142573GA SCS08142574GA SCS08142575GA SCS08142594GA SCS08142598GA SCS08162601GA SCS08162611GA SCS08162612GA SCS08162615GA SCS08162617GA SCS08172625GA SCS08172643GA SCS08182652GA SCS08202666GA SCS08202667GA SCS08202668GA SCS08202672GA SCS08202673GA SCS08202675GA SCS08212690GA SCS08212691GA SCS08212706GA SCS08212707GA SCS08212708GA SCS08222717GA SCS08222718GA SCS08222720GA SCS08222721GA SCS08222722GA SCS08222723GA SCS08222725GA SCS08222726GA SCS08222727GA SCS08232737GA SCS08232738GA SCS08232739GA SCS08232740GA SCS08232743GA SCS08232746GA SCS08232747GA SCS08232748GA SCS08232749GA SCS08232750GA SCS08232751GA SCS08232752GA SCS08242766GA SCS08242767GA SCS08242769GA SCS08242770GA SCS08242771GA SCS08242772GA SCS08242773GA SCS08242778GA SCS08242779GA SCS08252798GA SCS08252813GA SCS08252818GA SCS08292863GA SCS08342960GA SCS08363004GA SCS08363005GA SCS08373044GA SCS08373045GA SCS08373046GA SCS08373048GA SCS08373049GA SCS08373050GA SCS08373051GA SCS08373053GA SCS08373054GA SCS08373055GA SCS08373056GA SCS08373057GA SCS08373059GA SCS08373060GA SCS08373062GA SCS08373063GA SCS08373064GA SCS08373065GA SCS08373066GA SCS08373067GA SCS08373068GA SCS08373069GA SCS08373070GA SCS08373077GA SCS08373078GA SCS09303878GA SCS09303879GA SCS09303880GA SCS09303881GA SCS09303890GA SCS09303891GA SCS09303899GA SCS09464291GA SCS09464301GA SCS09464302GA SCS07431861GA SCS07441910GA SCS07441915GA SCS07371554GA AAD05240222GA AAD05240223GA AAD05240247GA AAD05290346GA AAD05290347GA AAD05521041GA AAD05521043GA AAD05521045GA AAD05521047GA AAD05521048GA AAD05521049GA AAD05521052GA AAD05521057GA AAD06261918GA AAD06261925GA SCS06420288GA SCS06420289GA SCS07391716GA SCS08473280GA SCS08473283GA SCS08473284GA SCS08473288GA SCS08473289GA SCS09113636GA SCS07231351GA SCS07401776GA SCS08352968GA SCS06430293GA SCS06430301GA SCS06430304GA SCS07391744GA SCS07391766GA SCS08042250GA SCS08042251GA SCS08042253GA SCS08312926GA SCS08312929GA SCS08332948GA SCS08332952GA SCS08332953GA SCS08352969GA SCS08352974GA SCS08352979GA SCS09153692GA SCS09153695GA SCS09153696GA SCS07431869GA SCS07441965GA SCS07441967GA SCS07502100GA SCS07502101GA SCS07502102GA SCS07502104GA SCS07502107GA SCS07502108GA SCS07502109GA SCS07502111GA SCS07502112GA SCS07502114GA SCS07502116GA SCS07502118GA SCS07502120GA AAD05350485GA AAD05360509GA AAD05360525GA AAD05360528GA AAD05460857GA AAD05460860GA AAD05460868GA AAD05460869GA AAD05460872GA AAD05460876GA SCS07431845GA SCS07431847GA SCS07431849GA SCS07431851GA SCS07431865GA SCS07431866GA SCS07431867GA SCS07431871GA SCS07431874GA SCS07441923GA SCS07441924GA SCS07441929GA SCS07441932GA SCS07441963GA SCS07441972GA SCS07441975GA SCS07502105GA SCS07502106GA SCS07502110GA SCS07502113GA SCS07502122GA SCS08262842GA AAD05440815GA AAD05440821GA SCS08122502GA SCS08132531GA SCS08132532GA SCS08132534GA SCS08132535GA SCS08132540GA SCS08132551GA SCS08132553GA SCS08132564GA SCS08252801GA SCS08252817GA SCS07391657GA SCS07391658GA SCS07391662GA SCS07391663GA SCS07391666GA SCS07391678GA SCS07391679GA SCS07391682GA SCS07391685GA SCS07391690GA SCS07391759GA AAD05450823GA AAD05450846GA AAD06021113GA AAD06021115GA AAD06021116GA AAD06021124GA AAD06071217GA SCS08172639GA SCS08182654GA SCS09414192GA SCS09434214GA SCS09434215GA SCS10144482GA SCS10144491GA AAD05400725GA SCS09303865GA SCS09303868GA SCS09303873GA SCS09303874GA SCS09303875GA SCS09303876GA SCS09303882GA SCS09303884GA AAD06031141GA AAD06031145GA AAD06171514GA AAD06171515GA AAD06171516GA AAD06171524GA AAD06181608GA AAD06191647GA SCS06500655GA SCS06510664GA SCS06510668GA SCS06510680GA SCS07030743GA SCS07030748GA SCS08122497GA SCS08172635GA SCS08182653GA SCS08182655GA SCS09404161GA SCS07482056GA AAD05290344GA AAD05290345GA AAD05510979GA AAD05510981GA AAD06241800GA AAD06241803GA AAD06241805GA AAD06241810GA AAD06251825GA AAD06251829GA SCS08162609GA SCS08302896GA SCS08302897GA SCS09033569GA SCS09053600GA SCS09053601GA SCS09053613GA SCS09053614GA SCS09303888GA SCS07231327GA SCS07231339GA SCS07231342GA SCS07231344GA SCS08262850GA SCS08142582GA SCS08493353GA SCS08493356GA SCS08493379GA SCS09344004GA SCS09344006GA SCS09364014GA SCS09364021GA SCS08132546GA SCS07311407GA SCS08413174GA SCS08413177GA SCS07321446GA SCS07321451GA SCS07411803GA SCS07421816GA SCS07421822GA AAD06071235GA AAD06071236GA AAD06071237GA AAD06071246GA AAD06071247GA AAD06071248GA AAD06091261GA AAD06091276GA SCS07431864GA SCS07431875GA AAD06362121GA AAD06362130GA AAD06362131GA AAD06362132GA AAD06362133GA AAD06362134GA AAD06372147GA AAD06372148GA AAD06372149GA AAD06372150GA AAD06372152GA AAD06372154GA AAD06372155GA AAD06372156GA AAD06372157GA AAD06111370GA AAD06121458GA SCS07080873GA SCS07080881GA SCS07080889GA SCS07080893GA AAD05330442GA AAD06121447GA AAD06121448GA AAD06121451GA AAD06121452GA AAD06121454GA SCS06430325GA SCS08142589GA SCS08142591GA SCS08162604GA SCS08162605GA SCS08162608GA SCS08162614GA SCS08162616GA SCS08162618GA SCS08162619GA SCS08162622GA SCS08172627GA SCS08172629GA SCS08172632GA SCS08202681GA SCS08202683GA SCS08212701GA SCS08212702GA SCS08212703GA SCS08212704GA SCS08413175GA SCS08413178GA SCS08413179GA SCS08413180GA SCS08413181GA SCS08413182GA SCS08413183GA SCS08413184GA SCS08413192GA SCS08413193GA SCS08443279GA AAD05430735GA AAD05430765GA AAD05521040GA AAD05521060GA AAD05521061GA AAD05521062GA AAD05521064GA AAD05521065GA AAD05521069GA AAD05521073GA AAD05521075GA AAD05521079GA AAD05521080GA AAD05521081GA AAD05521082GA AAD06111413GA AAD06171526GA AAD06171528GA AAD06171530GA AAD06171533GA AAD06171534GA AAD06171535GA SCS07060830GA SCS09113641GA SCS07512141GA SCS07522155GA AAD06041191GA SCS08132548GA SCS08132549GA SCS08493374GA SCS09203782GA SCS07381608GA SCS07381612GA SCS07381613GA SCS07381614GA SCS09283829GA SCS09283833GA SCS09293839GA SCS10164599GA SCS10164605GA SCS10164615GA AAD05370589GA AAD05370590GA AAD06111403GA AAD06111409GA AAD06111411GA AAD06111420GA SCS09203773GA SCS09203775GA SCS09203776GA SCS09233791GA SCS09243810GA SCS09243811GA SCS09243816GA SCS09293851GA SCS09293856GA SCS09293858GA SCS09323928GA SCS09323932GA
SCS09333953GA SCS06480501GA SCS09053603GA SCS09053606GA SCS09053607GA SCS09053608GA SCS09434225GA SCS09434226GA SCS07361539GA SCS07371548GA SCS07371555GA SCS07371560GA SCS08012198GA AAD06261932GA AAD06291956GA SCS06480520GA SCS07020726GA SCS07020732GA SCS07020735GA SCS07110950GA SCS07110954GA SCS07110969GA SCS07110983GA SCS07171100GA SCS07201185GA SCS07201187GA SCS07201188GA SCS07371591GA SCS07371596GA AAD06061214GA AAD05430739GA AAD05430763GA SCS07231325GA AAD05500960GA AAD05500963GA AAD05500964GA AAD05500965GA AAD05500966GA AAD05500967GA AAD05510978GA AAD05510980GA AAD05510983GA AAD06101294GA AAD06101295GA AAD06251818GA AAD06251819GA AAD06251820GA AAD06251823GA AAD06251824GA SCS07462011GA SCS09303897GA SCS09313911GA SCS09313912GA SCS09313913GA SCS09093631GA SCS09153697GA SCS09153698GA SCS09153699GA SCS09153703GA SCS06440405GA SCS07381604GA SCS07391641GA AAD05360514GA AAD05360515GA AAD05360516GA AAD05360517GA AAD05360518GA AAD05360526GA AAD05360527GA AAD06101318GA SCS09243809GA SCS09323924GA SCS09323927GA SCS09323940GA SCS09333960GA SCS09333965GA SCS09333976GA SCS09333980GA SCS09333983GA SCS09333984GA SCS09333985GA SCS09333986GA AAD05300388GA AAD05300389GA SCS10074430GA SCS08373030GA SCS08373031GA SCS08373082GA SCS08383092GA SCS08493384GA SCS08493385GA SCS08493386GA SCS09404169GA SCS09404170GA SCS09414178GA SCS09524402GA SCS09524405GA SCS07461985GA AAD05250271GA AAD06221763GA AAD06231797GA SCS07181159GA SCS08112443GA SCS08112449GA SCS08373058GA SCS08373079GA SCS09364027GA SCS08072324GA SCS08082341GA SCS08082343GA SCS08082358GA SCS08292875GA SCS08112441GA AAD05200173GA AAD05200176GA AAD05210184GA AAD05210185GA AAD05390661GA AAD05390662GA AAD05400727GA AAD06091275GA SCS07030760GA SCS08012195GA SCS09514364GA SCS09514368GA AAD05260299GA SCS06480542GA SCS07080872GA SCS07080874GA AAD05450836GA SCS08172644GA SCS08493373GA SCS09344003GA SCS09514362GA SCS09514372GA SCS10074425GA SCS07171108GA SCS08252805GA SCS08262843GA SCS09153691GA SCS09153693GA SCS07482089GA AAD05260306GA SCS08032223GA SCS08032227GA SCS08212697GA SCS08423250GA SCS08433261GA SCS09404160GA SCS09404162GA SCS07431862GA SCS07461978GA SCS07461997GA SCS10164598GA SCS10164603GA SCS10174621GA SCS10174630GA SCS10074432GA SCS10084448GA SCS10084451GA AAD05230197GA AAD05230198GA AAD05500954GA AAD05500955GA AAD05500957GA AAD05500958GA AAD06171527GA AAD06171537GA AAD06181615GA AAD06191621GA AAD06191632GA AAD06191635GA AAD06191648GA AAD06191650GA AAD06191653GA SCS06510676GA SCS06510678GA SCS07151038GA SCS07151042GA SCS07181126GA SCS07181140GA SCS07221310GA SCS07221312GA SCS07221321GA SCS07311416GA SCS07311434GA SCS07311435GA SCS07331490GA SCS07361522GA SCS07431876GA SCS07431877GA SCS07431878GA SCS07441941GA SCS07441944GA SCS07441958GA SCS07441973GA SCS07482049GA SCS07482052GA SCS07482053GA SCS07482065GA SCS07482067GA SCS07482068GA SCS07482069GA SCS07482071GA SCS07502125GA SCS07502127GA SCS07502128GA SCS07512143GA SCS07512151GA SCS08232736GA SCS08232756GA SCS08332943GA SCS08332944GA SCS08352967GA SCS08352976GA SCS08352983GA SCS08413188GA SCS08413189GA SCS08413197GA SCS08413198GA SCS08473281GA SCS08473282GA SCS08473285GA SCS08473286GA SCS08473287GA SCS08473291GA SCS08473298GA SCS08473300GA SCS08473303GA SCS08473304GA SCS08473305GA SCS08473309GA SCS08473311GA SCS08473314GA SCS08473317GA SCS08473319GA SCS08473322GA SCS08473330GA SCS08473333GA SCS09133670GA SCS09143671GA AAD06051199GA SCS08292865GA AAD05240239GA SCS07441927GA SCS07441955GA SCS07441956GA SCS07441957GA SCS08092386GA SCS08102391GA SCS08112428GA SCS08112431GA AAD06071251GA SCS07482041GA SCS07482088GA SCS07502103GA SCS07502115GA SCS07502117GA SCS07502119GA SCS07502121GA SCS07502123GA SCS07502124GA SCS07502126GA SCS08092387GA SCS08102396GA SCS09143683GA SCS06420280GA SCS06420290GA SCS08493383GA SCS08493388GA SCS08493394GA SCS08493397GA SCS08493401GA SCS08493419GA SCS08493420GA SCS08503428GA SCS08503430GA SCS08503431GA SCS09384101GA SCS09384107GA SCS09384108GA SCS09384109GA SCS09384110GA SCS09384111GA SCS09384112GA SCS09384114GA SCS09384117GA SCS09384118GA SCS09384120GA SCS09384123GA SCS09464288GA SCS09464304GA SCS09464309GA SCS08302884GA SCS08302895GA AAD06101312GA AAD06101313GA AAD06121439GA AAD06121440GA AAD06322037GA AAD06322038GA AAD06322039GA AAD06322040GA AAD06322042GA AAD06322044GA AAD06322050GA AAD06332051GA AAD06332052GA AAD06332055GA AAD06332059GA AAD06332060GA AAD06332061GA AAD06332064GA AAD06332066GA AAD06332070GA AAD06332072GA AAD06332073GA SCS07221302GA SCS07321442GA SCS07321450GA SCS08172638GA AAD05400718GA AAD05400719GA AAD05400720GA AAD05400721GA SCS07391668GA SCS09173730GA AAD06342086GA AAD06352094GA SCS06500633GA SCS06500634GA SCS06510681GA SCS07391650GA SCS07391712GA SCS07391718GA SCS10174632GA SCS06430327GA AAD05430766GA AAD05430769GA AAD05430774GA AAD05440804GA AAD05440805GA AAD05440808GA AAD05440810GA AAD05440811GA AAD05440812GA AAD05510982GA AAD05510988GA SCS07070842GA SCS07070844GA SCS08473326GA SCS08473329GA SCS09043587GA SCS08413190GA SCS08413207GA AAD05250274GA SCS07291400GA SCS08122488GA AAD06141490GA AAD06211743GA SCS07080885GA SCS08423213GA SCS08423214GA SCS08493345GA SCS08493378GA SCS08493404GA SCS09323930GA SCS09323935GA SCS09323937GA SCS09394125GA SCS09394128GA SCS09394133GA SCS09333950GA SCS09333966GA SCS10144497GA SCS10144503GA SCS07401780GA AAD06362126GA SCS06440399GA SCS06440402GA SCS06440416GA SCS06460456GA SCS06460457GA SCS06460458GA SCS06460464GA SCS06460474GA SCS06460475GA SCS06460476GA SCS06460479GA SCS06460482GA SCS06460483GA SCS06460492GA SCS06460497GA SCS06460499GA SCS08423240GA SCS08423245GA AAD06111359GA SCS07141029GA SCS08393116GA SCS08393146GA SCS09303877GA SCS08513452GA SCS08513457GA SCS09283830GA SCS09283831GA SCS09283832GA SCS08112438GA SCS08112480GA SCS08122501GA SCS08122511GA SCS08122512GA SCS08132537GA SCS08132541GA SCS08142569GA SCS08423246GA SCS08423247GA SCS08433262GA SCS09384038GA SCS09384039GA SCS09384045GA AAD05260326GA SCS07411792GA SCS07221294GA SCS07221300GA SCS09093628GA SCS09093632GA AAD05500948GA SCS07211268GA SCS07221319GA SCS06390152GA SCS06390175GA SCS06390188GA SCS06390190GA SCS07391645GA SCS07391648GA SCS06460478GA SCS06460486GA SCS07050814GA SCS09464282GA SCS09464287GA SCS10164594GA SCS10164601GA AAD05430783GA SCS09183737GA SCS09183738GA SCS09203786GA SCS09384052GA AAD05200177GA AAD06231795GA SCS07391713GA SCS08022217GA SCS08052271GA SCS08212696GA SCS07411798GA SCS08092380GA SCS09043585GA SCS09043589GA SCS10144514GA SCS10144515GA SCS10144517GA SCS10144534GA SCS10144539GA SCS10144540GA SCS10214655GA SCS10214656GA SCS10224666GA SCS10224673GA SCS10224679GA SCS10224703GA SCS06390143GA SCS06390151GA SCS06500615GA SCS06500619GA SCS08202685GA AAD05330441GA AAD06181596GA AAD06181597GA AAD06181616GA AAD06181618GA AAD06191627GA AAD06191630GA SCS07391770GA SCS07431848GA SCS08373043GA SCS09203781GA SCS07492093GA SCS07492094GA SCS07141005GA SCS07151044GA SCS07161058GA SCS07161068GA SCS07161070GA SCS07311412GA SCS08232745GA SCS06480508GA SCS06480512GA SCS08473323GA SCS09293844GA SCS09384053GA SCS09384054GA AAD06021109GA SCS07311409GA SCS07311410GA SCS09524406GA SCS09524407GA SCS09524410GA AAD05380621GA AAD05380622GA AAD06261904GA AAD06261908GA AAD06261919GA AAD06261934GA AAD06281935GA AAD06281936GA AAD06281939GA AAD06281946GA AAD06291959GA AAD06291968GA AAD06291970GA AAD06291973GA AAD06291982GA AAD06291983GA SCS06430312GA SCS06430332GA SCS06460463GA SCS07020727GA SCS07020729GA SCS07050792GA SCS07050807GA SCS07050809GA SCS07050810GA SCS07050811GA SCS07110976GA SCS07110984GA SCS08072337GA SCS08092371GA SCS08092373GA SCS08493405GA SCS09434227GA SCS09434228GA AAD06031128GA AAD06031131GA SCS07431850GA SCS08383098GA SCS08383100GA AAD05380638GA AAD05380639GA SCS08232734GA SCS08232741GA SCS09444252GA SCS09444253GA SCS09444254GA SCS09454259GA SCS09454280GA AAD05460883GA SCS09293846GA AAD05390710GA AAD05460865GA AAD06151493GA SCS06430335GA SCS06500656GA SCS07080887GA SCS07161056GA SCS07181110GA SCS07181111GA SCS07181114GA SCS07181116GA SCS07181120GA SCS07181121GA SCS07181123GA SCS07181125GA SCS07181129GA SCS07181132GA SCS07181133GA SCS07181137GA SCS07181139GA SCS07201208GA SCS07201212GA SCS07201214GA SCS07201215GA SCS07201216GA SCS07201217GA SCS07201218GA SCS07201221GA SCS07201222GA SCS07201223GA SCS07201224GA SCS07201226GA SCS07201227GA SCS07201233GA SCS07201235GA SCS07201236GA SCS07201239GA SCS07201240GA SCS07201241GA SCS07201242GA SCS07201243GA SCS07201244GA SCS07201247GA SCS07201248GA SCS07211250GA SCS07211251GA SCS07211256GA SCS07211257GA SCS07211259GA SCS07211260GA SCS07211261GA SCS07211265GA SCS07211266GA SCS07211267GA SCS07211269GA SCS07211271GA SCS07211273GA SCS07211274GA SCS07211275GA SCS07231353GA SCS07231354GA SCS07231355GA SCS07231356GA SCS07231357GA SCS07291376GA SCS07291377GA SCS07291378GA SCS07291379GA SCS07291389GA SCS07371541GA SCS08302912GA SCS08312934GA SCS07391738GA SCS07391740GA SCS07391749GA SCS07431844GA AAD05240253GA AAD05240254GA AAD05240255GA AAD05250256GA AAD05250257GA AAD05250258GA AAD05250259GA AAD05250264GA SCS08503424GA SCS08503438GA SCS08503439GA SCS08503448GA SCS09333968GA SCS09394126GA SCS09394138GA SCS07331493GA SCS07331495GA SCS07331496GA SCS07331497GA SCS07331498GA SCS07331499GA SCS07331500GA SCS07331501GA SCS08082366GA SCS08082369GA SCS09113635GA SCS08302901GA SCS08302905GA SCS08332950GA SCS07050801GA SCS07050816GA SCS08423251GA SCS08423254GA SCS09384036GA SCS09384040GA SCS09384043GA SCS08413202GA SCS08423225GA SCS08423227GA SCS08433256GA SCS08433257GA SCS08433258GA SCS08352962GA SCS08352963GA SCS08352965GA SCS08352966GA SCS08352977GA SCS08352978GA SCS08352981GA SCS08352984GA SCS08352986GA SCS08352987GA SCS08352988GA SCS08362995GA SCS08363000GA SCS08363001GA SCS08363006GA SCS08363007GA SCS08332955GA SCS08342957GA SCS07211278GA AAD05250268GA AAD05250269GA AAD05250277GA SCS08102415GA SCS08102423GA SCS09394124GA SCS09394127GA SCS09394129GA SCS09394136GA SCS08302894GA SCS08433260GA SCS09173731GA SCS07431839GA SCS07431868GA SCS07441974GA SCS07461981GA SCS07461983GA SCS07461984GA SCS07492095GA SCS07492097GA SCS08012186GA SCS08022204GA SCS08022206GA SCS08062307GA SCS08082352GA SCS08082370GA SCS08102409GA SCS08112454GA SCS08302886GA SCS08312928GA SCS08373073GA SCS08383083GA SCS08383084GA SCS08383085GA SCS08383086GA SCS08383089GA SCS08413206GA SCS08413208GA SCS08423212GA SCS08423249GA SCS08493359GA SCS08503446GA SCS08503450GA SCS08513465GA SCS09384113GA SCS09384116GA SCS09384121GA SCS10154556GA SCS10234706GA SCS07482015GA AAD06181588GA SCS08493354GA SCS08493362GA SCS09093625GA SCS09093629GA SCS09093633GA SCS07171090GA SCS08292878GA SCS08493402GA SCS08493403GA SCS09414183GA SCS08352964GA SCS08363003GA SCS08393119GA SCS07391623GA SCS08352971GA SCS08352982GA SCS08362990GA SCS08363016GA SCS08373036GA SCS08373037GA SCS08373040GA SCS09053609GA SCS09053610GA SCS08172636GA SCS08102420GA SCS08112460GA SCS08122505GA SCS07421826GA SCS06390102GA SCS06390103GA SCS06390104GA SCS06390112GA SCS07421818GA SCS07421819GA SCS07421823GA SCS07421827GA SCS08292874GA SCS08292882GA SCS09414177GA SCS09414179GA SCS07221311GA SCS07221314GA SCS07321453GA SCS07391632GA SCS07411804GA SCS07411806GA SCS07411807GA SCS07411809GA SCS07411810GA SCS07411811GA SCS07411814GA SCS07411815GA SCS07421829GA SCS08172646GA SCS08182657GA SCS08202688GA
SCS08222715GA SCS08242765GA SCS08242792GA SCS08503421GA SCS08503436GA SCS09384041GA SCS09384047GA AAD06151496GA SCS09394130GA SCS09394132GA SCS09394137GA SCS09394139GA SCS09394140GA SCS09394144GA SCS09394146GA SCS06480525GA SCS10174640GA SCS10174649GA SCS10174652GA SCS08052273GA SCS08052279GA SCS08052283GA SCS08052289GA SCS08062297GA SCS08062298GA SCS09344005GA SCS09344007GA SCS09344008GA SCS09364013GA SCS09364017GA SCS09364019GA SCS09364020GA SCS09364022GA SCS09364023GA SCS09364033GA SCS09364034GA SCS09384037GA AAD06061208GA SCS09384115GA SCS09384119GA AAD05500945GA AAD05500947GA AAD05521066GA AAD05521067GA SCS08172630GA SCS09514389GA SCS10164606GA AAD06171565GA AAD06171570GA AAD06171572GA AAD06231798GA SCS09243808GA SCS09243812GA AAD06301990GA AAD06302000GA SCS09173729GA SCS09424206GA SCS06500643GA SCS07522161GA SCS07522169GA SCS08012188GA SCS08022216GA SCS08032229GA SCS08032238GA SCS08032239GA SCS08082351GA SCS08092377GA SCS08102389GA SCS08132543GA SCS10144480GA SCS10154553GA SCS10154554GA SCS10154557GA SCS08413166GA SCS08413167GA SCS08413176GA SCS08413187GA SCS08413191GA SCS08413199GA SCS09464290GA SCS09464297GA SCS09504350GA SCS09504351GA SCS09504352GA SCS09504353GA SCS09504355GA SCS09504356GA SCS09504358GA SCS09504359GA SCS09504360GA SCS09504361GA SCS10174619GA SCS09303872GA AAD06061215GA AAD05230194GA AAD05230199GA AAD05230200GA AAD05230201GA SCS07431836GA SCS07431838GA SCS07431870GA SCS07441976GA SCS07482048GA SCS07482070GA SCS07482072GA SCS07482073GA SCS07482074GA SCS07482075GA SCS07482076GA SCS07482079GA SCS07482090GA SCS07482091GA SCS07482092GA SCS07502133GA SCS07502135GA SCS07512142GA SCS07512146GA SCS07512149GA SCS08312937GA SCS08312941GA SCS08342959GA SCS08342961GA SCS08493418GA SCS07461982GA AAD05370575GA AAD05370576GA AAD05370579GA AAD05370580GA AAD05370581GA AAD05370582GA AAD05370583GA AAD05370584GA AAD05370591GA AAD05370592GA AAD05370593GA AAD05370594GA AAD05370595GA AAD05370596GA AAD06071250GA AAD06372161GA AAD06372164GA SCS06390114GA SCS06390115GA SCS06390117GA SCS07030762GA SCS07030763GA SCS07070848GA SCS07070850GA SCS07070852GA SCS07070853GA SCS07070854GA SCS07070855GA SCS07070858GA SCS07070864GA SCS07080876GA SCS07080880GA SCS07110963GA SCS07110967GA SCS07361517GA SCS09293842GA SCS09293860GA SCS08493389GA SCS09313910GA AAD06261877GA AAD06261878GA AAD06261881GA AAD06261886GA AAD06261896GA AAD06291974GA AAD06291975GA AAD06291976GA AAD06291977GA AAD06291978GA SCS07060828GA SCS07060831GA SCS09043581GA SCS09043583GA SCS09043584GA SCS08042260GA SCS07391714GA AAD06171554GA SCS07070843GA SCS07070847GA SCS07431863GA SCS08112473GA SCS08142592GA AAD06091263GA AAD06091269GA AAD06091274GA AAD06091278GA SCS09384097GA SCS09514396GA SCS09514397GA SCS07401775GA SCS06490570GA SCS07482037GA SCS07482043GA AAD05380615GA AAD06091277GA AAD06091280GA SCS06390171GA SCS06390173GA SCS08252820GA SCS08312925GA SCS08312927GA SCS08423211GA SCS08423215GA SCS08423216GA SCS08423219GA SCS08072336GA SCS08162600GA SCS08172640GA SCS06520708GA SCS07010722GA SCS09514375GA SCS09514381GA SCS09514384GA SCS07381610GA SCS07381611GA SCS07391617GA SCS07391618GA SCS07391621GA SCS07391622GA SCS07391624GA SCS07391630GA SCS07391634GA SCS07391636GA SCS09454257GA SCS09454272GA SCS09454273GA SCS09454274GA SCS09454275GA AAD06031151GA AAD06031152GA SCS06440390GA SCS06460500GA SCS06480507GA SCS06480534GA SCS06480557GA SCS06480563GA SCS06480566GA SCS07020733GA SCS07030741GA SCS07291401GA SCS08493375GA SCS09133654GA SCS09133660GA SCS09153690GA SCS09153694GA SCS09293849GA SCS09384102GA SCS09384105GA SCS08262849GA SCS08292879GA SCS08292880GA SCS08292881GA SCS06440382GA SCS06460473GA SCS07391747GA SCS08292877GA SCS08302887GA SCS08493415GA SCS07391695GA SCS08383101GA SCS09474340GA AAD05320427GA AAD05320428GA AAD05521042GA AAD06111372GA AAD06111377GA SCS07371577GA SCS07371580GA SCS07371586GA SCS07371588GA SCS07371589GA SCS07371594GA SCS08212711GA SCS08383103GA SCS08383104GA SCS08383106GA SCS08383108GA SCS08383109GA SCS08383111GA SCS08393117GA SCS08393118GA SCS08413168GA SCS08142567GA AAD06101317GA AAD06101321GA AAD06101323GA SCS08332945GA SCS08352970GA SCS08352973GA SCS07291380GA SCS07291382GA SCS07291386GA SCS07371571GA SCS07371575GA SCS07371579GA SCS07371590GA SCS07431832GA SCS07431833GA SCS07431834GA SCS07431835GA SCS07431837GA SCS07431856GA SCS07441925GA SCS07441933GA SCS07461988GA SCS07461989GA SCS07492096GA SCS08042256GA SCS08042257GA SCS08042258GA SCS08042259GA SCS08052265GA SCS08052267GA SCS08052268GA SCS08052269GA SCS08052275GA SCS08062299GA SCS08102398GA SCS08252800GA SCS08252808GA SCS08252811GA SCS08252835GA SCS08262847GA SCS08262848GA SCS08332951GA SCS08332954GA SCS08503441GA SCS09404165GA SCS09404166GA SCS09404167GA SCS09404168GA AAD06051198GA SCS06430300GA SCS06430302GA SCS06430303GA SCS06430307GA SCS07030755GA SCS07030757GA SCS07462007GA SCS08082346GA SCS08082349GA SCS08082355GA SCS08092372GA SCS08092376GA SCS08122492GA SCS08493377GA SCS10074429GA SCS10074436GA SCS06420277GA SCS06420286GA SCS07391619GA SCS07391626GA SCS07391640GA SCS08252802GA SCS07391665GA SCS07391674GA SCS07391676GA SCS07391684GA SCS07391688GA SCS07482024GA SCS07482027GA AAD06031140GA SCS08362992GA SCS08363014GA SCS08312933GA SCS08503445GA SCS08513455GA SCS08513459GA SCS09263817GA SCS09293847GA SCS09333961GA SCS10164608GA SCS10164609GA SCS10164612GA SCS10164616GA SCS10174643GA SCS10174644GA SCS10174646GA SCS10174650GA SCS06430341GA SCS06440367GA SCS06440373GA SCS07070863GA SCS07311415GA SCS07311425GA SCS07311432GA SCS08423253GA SCS09143681GA SCS09143682GA SCS09143684GA SCS07231343GA SCS07231345GA SCS07231346GA SCS07231349GA SCS08132522GA AAD06261913GA SCS08222719GA SCS08242763GA SCS08292851GA SCS08292857GA SCS08292858GA SCS08292859GA SCS08292861GA SCS08302883GA SCS08302898GA SCS08302899GA SCS08312913GA SCS08312915GA SCS07311430GA SCS07311431GA SCS07090921GA SCS07090923GA SCS08112436GA SCS08112439GA SCS10084445GA SCS10084447GA SCS08473316GA AAD05521092GA AAD05350499GA AAD05380647GA AAD05380648GA AAD05380649GA AAD05380650GA AAD05380651GA AAD05380652GA AAD05390653GA AAD05390654GA AAD05521094GA SCS09384088GA SCS09384094GA SCS08082357GA SCS10144468GA SCS10144508GA SCS10154571GA SCS09444249GA AAD05310397GA SCS08072318GA SCS08072331GA SCS08072332GA SCS08082348GA SCS08082350GA SCS08102394GA SCS08232742GA SCS09424203GA SCS09424204GA SCS10164596GA SCS08262844GA SCS08302889GA SCS08302891GA SCS08302892GA SCS07120988GA SCS07120989GA SCS07161061GA SCS07161063GA SCS08122504GA SCS09133656GA AAD06251822GA SCS08493367GA SCS08473328GA SCS08493371GA SCS08493381GA SCS08493391GA SCS08493396GA SCS06430334GA SCS06440364GA SCS06440370GA SCS06440411GA SCS07461999GA SCS07462004GA SCS07462005GA SCS07462012GA SCS07482019GA SCS07482023GA SCS07482025GA SCS07482030GA SCS08042244GA SCS09303889GA AAD05240202GA AAD05240203GA AAD05240204GA SCS07080867GA SCS08473315GA SCS08473318GA SCS09133646GA SCS09133649GA SCS09133652GA AAD05430784GA AAD05430789GA AAD05430790GA AAD05430791GA AAD05430792GA AAD05430793GA AAD05430795GA AAD05430799GA AAD05430800GA AAD05430801GA AAD05430802GA AAD05440807GA AAD05440809GA AAD05440813GA AAD05440818GA AAD05440819GA SCS06390105GA SCS06390106GA SCS06390107GA SCS06390108GA SCS10074427GA SCS08112451GA SCS09384086GA SCS08493376GA SCS09133664GA SCS09133666GA SCS09424212GA SCS09404159GA SCS09424213GA AAD05450832GA AAD05450833GA SCS07060819GA SCS07060839GA SCS08433268GA SCS09153712GA SCS10164607GA SCS09193749GA SCS09303864GA SCS09303869GA SCS09303871GA AAD05250284GA AAD05430758GA AAD05430794GA AAD05430796GA AAD06191638GA SCS06430344GA SCS07411791GA SCS08423229GA SCS06390147GA SCS06390156GA SCS06390160GA SCS06390166GA SCS09414188GA SCS09414189GA SCS09414190GA SCS09414191GA SCS09414193GA SCS09414194GA SCS09414195GA SCS09414196GA SCS09414197GA SCS09414198GA SCS09414199GA SCS09424200GA SCS09414182GA SCS08423221GA SCS08423224GA SCS08042263GA SCS08242780GA SCS07231326GA SCS08082365GA SCS08032228GA AAD06131465GA SCS07461980GA SCS07461990GA SCS07461991GA SCS07522157GA SCS07522158GA SCS07522164GA SCS07522171GA SCS07522175GA SCS07522181GA SCS07522182GA SCS08012194GA SCS08042261GA SCS08493347GA SCS08493351GA SCS10164600GA SCS08423243GA SCS08423244GA SCS08493370GA SCS08423232GA SCS08443275GA SCS08363022GA SCS08383093GA SCS08383094GA SCS08383096GA SCS08383099GA SCS10144455GA SCS10144467GA AAD05310421GA AAD05310422GA AAD05310423GA AAD05320426GA AAD05480903GA AAD06241801GA SCS07201228GA SCS07201230GA SCS07231332GA SCS07231336GA SCS07231348GA SCS07522159GA SCS07522167GA SCS07522180GA SCS07522183GA SCS07522184GA SCS08012187GA SCS08012190GA SCS08012191GA SCS08022201GA SCS08022202GA SCS08182651GA SCS08202686GA SCS07441962GA SCS07482084GA SCS07482087GA SCS09434217GA SCS09434218GA SCS09434219GA SCS10214653GA SCS10214654GA SCS06430349GA SCS08362991GA SCS08362996GA SCS09033577GA SCS09033578GA SCS09043592GA SCS09043594GA AAD06171523GA SCS08312930GA SCS09053602GA SCS09053605GA SCS09323941GA AAD05430778GA AAD05500939GA AAD05500952GA AAD06312017GA AAD06312019GA AAD06312020GA AAD06312024GA AAD06312031GA AAD06312035GA SCS06500660GA SCS08413186GA SCS08413195GA SCS08423210GA SCS09404176GA SCS10144463GA SCS10144522GA SCS10144527GA SCS10144529GA SCS10144530GA SCS10144536GA SCS06440386GA SCS06460466GA SCS06460472GA SCS06460488GA SCS06460489GA SCS08032222GA SCS08032226GA SCS08032230GA SCS08032231GA SCS08032232GA SCS08032233GA SCS08042246GA SCS08042247GA SCS08062296GA SCS07151036GA SCS07151043GA SCS07201189GA SCS07201195GA SCS07201220GA SCS07361535GA SCS07371574GA SCS07391721GA SCS07411799GA SCS07421820GA SCS07441940GA SCS07441952GA SCS08062305GA SCS08112462GA SCS08122494GA SCS08292855GA SCS08292860GA SCS08292866GA SCS08292868GA SCS08302902GA SCS08302903GA SCS08302904GA SCS08302906GA SCS08312916GA SCS08312918GA AAD06241804GA SCS06430361GA SCS06430363GA SCS08393112GA SCS08393121GA SCS08393155GA SCS09053599GA SCS09073616GA SCS09173717GA SCS09173728GA SCS09303894GA SCS09303895GA SCS09303898GA SCS09303901GA SCS09303902GA SCS09303903GA SCS09303904GA SCS09333957GA SCS09424205GA SCS09424208GA SCS09424209GA SCS09424210GA SCS09424211GA SCS09434229GA SCS09524412GA SCS10144473GA SCS10224676GA SCS10224677GA SCS10224680GA SCS10224681GA SCS10224683GA SCS08252836GA SCS09153711GA SCS09333973GA SCS09333974GA SCS09333981GA SCS08292862GA SCS08342958GA SCS08503433GA SCS09073618GA SCS09073620GA SCS09193754GA SCS09193755GA SCS09323938GA SCS09333975GA SCS09434220GA SCS09434221GA SCS09434222GA SCS09434223GA SCS09434224GA SCS09444243GA SCS09444244GA SCS09444245GA SCS09444246GA SCS09444247GA SCS09454262GA SCS09474312GA SCS09524408GA SCS09524409GA SCS09524411GA SCS10144494GA SCS10144543GA SCS10154566GA AAD06312032GA SCS08503444GA SCS08503449GA SCS09293841GA SCS09303862GA SCS09303863GA SCS09303866GA SCS09303867GA SCS09303870GA SCS09323916GA SCS09323917GA SCS09323919GA SCS09323920GA SCS09323933GA SCS09323936GA SCS09323939GA SCS09323942GA SCS09333943GA SCS09333944GA SCS09333945GA SCS09333946GA SCS09333947GA SCS09333948GA SCS09333949GA SCS09333951GA SCS09333952GA SCS09333954GA SCS09333955GA SCS09333959GA SCS09333962GA SCS09333969GA SCS09333970GA SCS09333972GA SCS09344011GA SCS09384044GA SCS09384046GA SCS09384048GA SCS10144518GA SCS10144523GA SCS10144524GA SCS10144528GA SCS10144532GA SCS10144535GA SCS10144537GA SCS10144538GA
SCS10144542GA SCS10144521GA AAD05240225GA AAD05240229GA AAD05240231GA SCS06500622GA SCS07110956GA SCS07391735GA SCS07401774GA SCS07411790GA SCS07421824GA SCS08052285GA SCS08052286GA SCS08052287GA SCS08062309GA SCS08062310GA SCS08062313GA SCS08072322GA SCS08072327GA SCS08072333GA SCS08072335GA AAD05500944GA SCS07191170GA SCS07231341GA SCS08503440GA SCS09073617GA SCS09153706GA SCS09454264GA SCS10144470GA AAD05370561GA AAD05370562GA SCS08252838GA SCS07040788GA SCS07421821GA SCS08042248GA SCS08042249GA SCS08052266GA SCS08052276GA SCS08062294GA SCS08112433GA SCS08112440GA SCS08112446GA SCS08112448GA SCS08112452GA SCS08112453GA SCS08112458GA SCS08112481GA SCS08122496GA SCS08122519GA SCS08122520GA SCS08132526GA SCS08132545GA SCS08132547GA SCS08132550GA SCS08132560GA SCS08132563GA SCS08142579GA SCS08202671GA SCS08262846GA SCS08423231GA SCS09133653GA SCS09173723GA SCS09173725GA SCS09173727GA SCS10074438GA SCS10074440GA SCS10144483GA SCS10144493GA SCS08132523GA SCS07411796GA SCS08212714GA SCS08413203GA SCS10174647GA AAD06332058GA SCS07040789GA SCS07191168GA SCS07191175GA SCS07391680GA SCS07411782GA SCS07431840GA SCS07431846GA SCS08032220GA SCS08122518GA SCS08373076GA SCS09313905GA SCS09313908GA SCS10084441GA AAD06171538GA AAD06171541GA SCS09444237GA SCS09444240GA SCS09444241GA SCS09444242GA SCS09454277GA SCS09514366GA SCS09514371GA SCS10074434GA SCS10084442GA SCS10084444GA SCS10084446GA SCS10144512GA SCS10154562GA SCS10164610GA SCS10174620GA SCS10174626GA SCS10174641GA SCS10174642GA SCS10224695GA SCS09033565GA SCS09033566GA SCS09033567GA SCS09033568GA SCS09033570GA SCS09033572GA SCS09033573GA SCS09033575GA SCS09033576GA SCS09033579GA SCS09043580GA SCS09043586GA SCS09043593GA SCS09053595GA SCS09053596GA SCS09053598GA AAD05240237GA SCS06380052GA SCS06380065GA SCS06440426GA SCS06440444GA SCS06460462GA SCS07231335GA SCS07371583GA SCS07381607GA SCS07381609GA SCS07391710GA SCS07391720GA SCS07391730GA SCS07391731GA SCS07391767GA SCS07401773GA SCS08503422GA SCS08513460GA SCS09333977GA SCS09333978GA SCS09384042GA SCS08393130GA SCS07421830GA SCS08373047GA SCS08172631GA SCS08172634GA AAD06111367GA AAD06121457GA SCS07411797GA SCS07411801GA SCS07411812GA SCS07411813GA SCS07421825GA SCS07421828GA SCS08132529GA SCS08142593GA SCS08142596GA SCS08242786GA SCS08252799GA SCS09524417GA SCS09183740GA SCS08102421GA SCS07381605GA SCS07381606GA SCS08082356GA SCS08212712GA SCS08112475GA SCS08252806GA SCS09113642GA SCS06440448GA SCS07161066GA SCS07291390GA SCS08102426GA SCS07461994GA AAD05310407GA AAD05510994GA SCS06440396GA SCS06440413GA SCS07171097GA SCS07391633GA SCS07391734GA SCS07391736GA AAD05480911GA SCS08022213GA SCS08022215GA SCS08082354GA SCS08423230GA SCS08423238GA SCS08122515GA SCS09474326GA SCS09474330GA AAD06131473GA AAD05330444GA AAD05360533GA AAD06261880GA AAD06261892GA SCS07231323GA SCS07231328GA SCS07231329GA SCS07231330GA SCS07231333GA SCS07231334GA SCS07231337GA SCS07231347GA SCS07231350GA SCS07231352GA SCS07311419GA SCS07311428GA SCS07371582GA SCS08032240GA SCS08032242GA SCS08112435GA SCS09113639GA SCS07482018GA SCS10144454GA SCS10144459GA SCS10144460GA SCS10144461GA SCS10144465GA SCS10144466GA SCS10144472GA SCS10144475GA SCS10144481GA SCS10144488GA SCS10144495GA SCS10144498GA SCS10144500GA SCS10144531GA SCS10174628GA SCS10174631GA SCS07361531GA SCS07361532GA SCS07361533GA SCS07361536GA SCS07371550GA SCS07371551GA SCS07371552GA SCS07371553GA SCS07391711GA SCS07391725GA SCS07391728GA SCS07161053GA SCS07161054GA SCS07161055GA SCS08413201GA SCS08413205GA SCS08433255GA SCS08433259GA SCS07020734GA AAD05030136GA AAD05030137GP AAD05030146GP AAD05040150GP AAD05040151GP AAD05040153GP AAD05040155GP SCS07040776GA SCS07060836GA SCS07060837GA SCS07090918GA SCS07391642GA SCS07391646GA SCS07391649GA SCS07401778GA SCS09464307GA SCS09474336GA SCS10144516GA SCS06430337GA SCS06430339GA SCS06430348GA SCS07391638GA SCS08503437GA SCS08132552GA SCS09424207GA SCS09474343GA SCS09494345GA SCS09494347GA SCS08433263GA AAD06191643GA AAD06191658GA SCS06390185GA SCS06390189GA SCS08052272GA SCS08112442GA SCS08112465GA SCS08112468GA SCS08112478GA SCS08112483GA SCS08383097GA SCS07482034GA SCS07482080GA SCS08042252GA SCS08042254GA SCS08363019GA SCS08363020GA SCS09474318GA SCS09474337GA SCS08202674GA SCS07231324GA SCS07231331GA SCS07231340GA SCS08393122GA SCS08423242GA SCS09113637GA SCS09113638GA SCS09143685GA SCS09283838GA SCS09333998GA SCS09333999GA SCS09334000GA SCS09334001GA SCS09334002GA SCS09364015GA SCS09384051GA SCS09384087GA SCS09384090GA SCS09384091GA SCS09434216GA SCS09474310GA SCS09474311GA SCS09474331GA SCS09474332GA SCS09474333GA SCS09474334GA SCS09474335GA SCS09474339GA SCS09474342GA SCS09474344GA SCS09494346GA SCS09494348GA SCS09494349GA SCS10144504GA SCS10144509GA SCS10144519GA SCS10144533GA SCS10164595GA AAD05040156GP SCS07391673GA AAD05460888GA SCS07371559GA SCS09454255GA SCS09474315GA AAD05380624GA AAD05380635GA AAD05380637GA AAD05460853GA AAD05460856GA AAD05460874GA AAD05460877GA AAD05460882GA AAD05460884GA AAD05480896GA AAD06322041GA AAD06322047GA SCS07040787GA SCS07391739GA SCS07391742GA SCS07391748GA SCS08242782GA SCS08423239GA SCS08433269GA SCS08433273GA SCS09043582GA SCS09113643GA SCS09113644GA SCS09133648GA SCS08232754GA SCS08232755GA SCS09474323GA SCS07421831GA SCS08503429GA AAD05360553GA SCS08493350GA SCS08493358GA SCS08493361GA SCS08513464GA SCS09333971GA SCS09333979GA SCS09333982GA SCS09133667GA SCS09133668GA SCS09133669GA SCS06480541GA SCS06480548GA SCS06480564GA SCS08232732GA SCS08242774GA SCS08242775GA SCS08242776GA SCS08252797GA SCS07151047GA SCS07151049GA SCS07171094GA SCS07171098GA SCS07191166GA SCS07191167GA SCS07191169GA SCS07191171GA SCS07191177GA SCS07191178GA SCS08072338GA SCS08332949GA SCS08413171GA SCS08232731GA SCS08242757GA SCS08242759GA SCS08242760GA AAD06031134GA SCS08122503GA SCS08122508GA SCS08122509GA SCS08122514GA SCS08122516GA SCS10144458GA SCS06460468GA SCS06480506GA SCS08423222GA SCS07080883GA SCS07080897GA SCS07080903GA SCS07371565GA SCS07371566GA SCS07371597GA SCS09474321GA SCS09474324GA SCS09474329GA SCS09464295GA SCS09474341GA SCS07171093GA SCS07171095GA SCS07171103GA SCS07171105GA SCS08022209GA SCS08022212GA SCS08022214GA AAD05350491GA SCS08122499GA SCS08122500GA SCS08192664GA SCS08202665GA SCS08012189GA SCS08012192GA SCS08022200GA SCS08022203GA SCS08022207GA SCS08022208GA SCS08022210GA SCS08022211GA SCS08102414GA SCS08102424GA SCS08142576GA SCS08142578GA AAD06131474GA SCS08383105GA SCS08383107GA SCS09153702GA SCS09153704GA SCS09193742GA SCS09193743GA SCS09193744GA SCS09193745GA SCS09193746GA SCS09193747GA SCS09193748GA SCS09193752GA SCS09203762GA SCS09203763GA SCS09203767GA SCS09203768GA SCS08212705GA SCS08212709GA SCS08222716GA SCS08312935GA SCS08312940GA SCS08342956GA SCS08252814GA SCS08393131GA SCS08393133GA SCS08393139GA SCS08393143GA SCS08393147GA SCS08393148GA SCS08393160GA SCS08393162GA SCS08413169GA SCS08413172GA SCS10144490GA SCS10144492GA SCS10144507GA SCS10154550GA AAD06352093GA AAD06352095GA SCS06500631GA SCS07030739GA SCS07030740GA SCS07030742GA SCS07030745GA SCS07030747GA SCS07030752GA SCS07030764GA SCS09454258GA SCS09454260GA SCS08082359GA SCS08092374GA SCS08092375GA SCS08092384GA SCS08092385GA SCS08102407GA SCS08102411GA SCS09143676GA SCS09143677GA AAD06362139GA SCS07090924GA SCS07441931GA SCS08072319GA SCS08072340GA SCS09133659GA SCS09524418GA SCS09524419GA AAD05330449GA SCS07060817GA SCS07070865GA SCS07311418GA SCS07311422GA SCS07361505GA SCS08062301GA SCS08062302GA SCS08503442GA SCS08513451GA SCS08513456GA SCS08513458GA SCS08513461GA SCS08513462GA SCS09384049GA SCS09384050GA SCS09434230GA SCS09464286GA SCS09464308GA SCS10144545GA SCS10154569GA SCS10154572GA SCS06430353GA SCS06440412GA SCS07151037GA SCS07221313GA SCS07221320GA SCS07311427GA SCS08032224GA AAD05470893GA AAD05500949GA AAD05500950GA AAD05500951GA AAD05500961GA AAD06081257GA AAD06091267GA AAD06091271GA AAD06091279GA AAD06171518GA AAD06171519GA AAD06171525GA AAD06191649GA SCS06440376GA SCS07080888GA SCS07090909GA SCS07090917GA SCS07201211GA SCS07291383GA SCS07411783GA SCS07411784GA SCS07462003GA SCS08052264GA SCS08052270GA SCS08072329GA SCS08092378GA SCS08112461GA SCS08142587GA SCS08292873GA SCS08493366GA SCS09333956GA SCS09504354GA SCS09504357GA SCS10154564GA SCS10154575GA SCS10164588GA SCS10164591GA SCS10164611GA |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
262-513-4122
|
Manufacturer Reason for Recall |
Transport Pro Monitor stops communication with the CARESCAPE Patient Data Module (PDM) after 414 days of continuous run time. This time will equate to different amounts of real time depending on how much the units is actually in service per day. Transport Pro contains an internal timer that is used to control the software and remind users to perform preventive maintenance. When this internal timer
|
FDA Determined Cause 2 |
Software design |
Action |
GE Healthcare issued an "Urgent Medical Device Correction" letter dated September 17, 2010 to consignees. The letter was addressed to Hospital Administrator, Head of Biomedical Engineering and Nursing Manager. The letter described the product, Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. Service representatives will updates all of the affected Transport Pros with PDM that were distributed.
Customers may contact GE at 262-548-2731 about this correction. |
Quantity in Commerce |
3256 |
Distribution |
Worldwide Distribution: Domestic in the states of AL, AZ, AR, CA,CT, CO, DC, FL, GA, ID, IL, IN, IA, KY, LA, MD, MA, ME, MI, MN, MS, MO, NE, NV, NJ, NM, NY, NC, OH, OK, PA, SC, TN, TX, VA, VT, WA, WI, and WY, and Internation in the countries of BRAZIL, BERMUDA, BELGIUM, AUSTRIA, AUSTRALIA, CANADA, CHINA, COLOMBIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FINLAND, FRANCE, GERMANY GREECE, INDONESIA, IRELAND, ITALY, JAPAN, JORDAN, KUWAIT, LEBANON, LITHUANIA, MALAYSIA, MAURITIUS, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, REPUBLIC OF KOREA, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD 7 TOBAGO, and the UNITED KINGDOM. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MHX and Original Applicant = GE MEDICAL SYSTEMS INFORMATION TECHNOLOGIES
|
|
|
|