| | Class 2 Device Recall SmartPump Tourniquet Pump |  |
| Date Initiated by Firm | April 13, 2010 |
| Date Posted | January 18, 2011 |
| Recall Status1 |
Terminated 3 on May 16, 2011 |
| Recall Number | Z-0912-2011 |
| Recall Event ID |
56624 |
| 510(K)Number | K924273 |
| Product Classification |
Tourniquet, pneumatic - Product Code KCY
|
| Product | Stryker Instruments, SmartPump Tourniquet System, Dual Channel Tourniquet Pump, REF 5920-011-000, Rx only.
Intended to be used to temporarily occlude or decrease blood flow in a patient extremity during a surgical procedure of that extremity. |
| Code Information |
'0810000423, '0728303723, '0728303733, '0728303743, '0728303753, '0728303763, '0729800813, '0729800823, '0729800833, '0730909443, '0730909453, '0730909463, '0730909473, '0730909483, '0731005063, '0731005073, '0731005083, '0731005093, '0731005103, '0731705783, '0731705793, '0731705803, '0731705813, '0731705823, '0732403193, '0732404063, '0732404073, '0732404083, '0732404093, '0732404103, '0733401733, '0733401743, '0733401753, '0733401763, '0733401773, '0734404543, '0734404553, '0734404563, '0734404573, '0734404583, '0734404593, '0734404603, '0734404613, '0734404623, '0734404633, '0734517413, '0734517423, '0734517433, '0734517443, '0734517453, '0734517463, '0734517473, '0734517483, '0734517493, '0734517503, '0734800113, '0734800123, '0734800133, '0734800143, '0734800153, '0734800163, '0734800173, '0734800183, '0734800193, '0734800203, '0734800213, '0734800223, '0734800233, '0734800243, '0734800253, '0735102713, '0735102723, '0735102733, '0735102743, '0735102753, '0735102763, '0735102773, '0735102783, '0735102793, '0735102803, '0735212143, '0735212153, '0735212163, '0735212173, '0735212183, '0735405453, '0736100753, '0736100763, '0736100773, '0736100783, '0800201743, '0800201753, '0800201763, '0800306263, '0800306273, '0800306283, '0800306293, '0800306303, '0800306313, '0800306323, '0800306333, '0800306343, '0800306353, '0800402403, '0800402413, '0800402423, '0800402433, '0800402443, '0800402673, '0800402683, '0800402693, '0800402703, '0800402713, '0800903213, '0800903223, '0800903233, '0800903243, '0800903253, '0800903263, '0800903273, '0800903283, '0800903303, '0801003513, '0801003523, '0801003533, '0801003543, '0801003553, '0801003563, '0801003573, '0801003583, '0801003593, '0801003603, '0801103723, '0801103733, '0801103743, '0801103753, '0801401203, '0801401213, '0801401223, '0801401233, '0801401243, '0801401253, '0801401263, '0801401273, '0801401283, '0801401293, '0801408343, '0801408353, '0801408363, '0801408373, '0801408383, '0801408393, '0801408403, '0801408413, '0801408423, '0801408433, '0801507893, '0801507903, '0801507913, '0801507923, '0801507933, '0801606883, '0801606893, '0801606903, '0801606913, '0801606923, '0801606933, '0801606943, '0801606953, '0801606963, '0801606973, '0801700623, '0801700633, '0801700643, '0801700653, '0801700663, '0801700673, '0801700683, '0801700693, '0801700703, '0801700713, '0802203243, '0802203253, '0802203263, '0802203273, '0802406863, '0802406873, '0802406883, '0802500063, '0802500073, '0803117623, '0803117633, '0803117643, '0803117653, '0803117663, '0803117673, '0803117683, '0803117693, '0803117703, '0803117713, '0803214263, '0803214273, '0803214283, '0803214293, '0803214303, '0803214313, '0803214323, '0803214333, '0803214343, '0803214353, '0803502553, '0803502563, '0803502573, '0803502583, '0803502593, '0803700303, '0803700313, '0803700323, '0803700333, '0803700343, '0803700353, '0803700363, '0803801293, '0803801303, '0803801313, '0803801323, '0804216823, '0804216833, '0804216843, '0804216853, '0804401423, '0804401433, '0804401443, '0804401453, '0804401473, '0804401483, '0804401493, '0804401503, '0804401513, '0804401523, '0804401533, '0804401543, '0804401553, '0804401563, '0804601053, '0804601063, '0804601073, '0804601083, '0804601093, '0804601103, '0804601113, '0804601123, '0804601133, '0804601143, '0804900563, '0804900573, '0804900583, '0804900593, '0804900603, '0804900613, '0804900623, '0804900633, '0804900643, '0804900653, '0804900663, '0804900673, '0804900683, '0804900693, '0804900703, '0804904583, '0805203843, '0805203853, '0805203863, '0805203873, '0805203883, '0805302503, '0805302513, '0805302523, '0805302533, '0805302543, '0805302553, '0805302563, '0805302573, '0805302583, '0805302593, '0805702383, '0805702393, '0805702403, '0805702413, '0805702423, '0805803833, '0805803843, '0806002443, '0806002453, '0806002463, '0806002473, '0806002483, '0806002503, '0806002513, '0806002533, '0807006223, '0807006233, '0807006243, '0807006253, '0807006263, '0807719493, '0807903073, '0807903083, '0807903093, '0807903103, '0807903113, '0807903123, '0807903133, '0807903143, '0807903153, '0807903163, '0807903173, '0807903183, '0807903193, '0807903203, '0807903213, '0807903223, '0807903233, '0807903243, '0807903253, '0807903263, '0807903273, '0807903283, '0807903293, '0807903303, '0807903313, '0807903323, '0807903333, '0807903343, '0807903353, '0807903363, '0807903373, '0807903383, '0807903393, '0807903403, '0807903413, '0807903423, '0807903433, '0807903443, '0807903453, '0807903473, '0807903483, '0807903493, '0807903503, '0807903513, '0807903523, '0807903533, '0807903543, '0807903553, '0807903563, '0808000063, '0808504403, '0808712153, '0808712163, '0808712173, '0808712193, '0808712203, '0808712213, '0808712223, '0808712233, '0808712243, '0808712253, '0808712263, '0808712273, '0808712283, '0808712293, '0808712303, '0808712313, '0808712323, '0808800903, '0808800913, '0808800923, '0808800933, '0808800943, '0808800953, '0808800963, '0808800973, '0808800983, '0808800993, '0809304473, '0809304523, '0809304533, '0809304543, '0809304553, '0809304563, '0809513623, '0809513633, '0809513643, '0809513653, '0809513663, '0809513673, '0809513683, '0809513693, '0809513703, '0809513713, '0809829903, '0809829913, '0809829923, '0810000403, '0810000413, '0810000433, '0810000443, '0810700373, '0810700383, '0810700393, '0810700403, '0810700413, '0812102393, '0812102403, '0813700613, '0813700623, '0813700633, '0813700643, '0813700653, '0813700663, '0813700673, '0813700683, '0813700693, '0813700703, '0814006963, '0814006973, '0814006983, '0814006993, '0814007003, '0814101303, '0814101313, '0814101323, '0814101333, '0814101343, '0814400653, '0814400663, '0814400673, '0814400683, '0814400693, '0814900263, '0814900273, '0814900283, '0814900293, '0814900303, '0815001143, '0815001153, '0815001163, '0815001173, '0815001183, '0815120653, '0815120663, '0815120673, '0815120683, '0815120693, '0815120703, '0815120713, '0815120723, '0815120733, '0815120743, '0815502593, '0815502603, '0815502613, '0815502623, '0815502633, '0815601333, '0815601343, '0815601353, '0815601363, '0815601373, '0815704723, '0815704733, '0815704743, '0815704753, '0815704763, '0815802823, '0815802833, '0815802843, '0815802853, '0815802863, '0817710333, '0817710343, '0817710353, '0817710363, '0817710373, '0817710383, '0817710393, '0817710403, '0817710413, '0817710423, '0817800403, '0817800413, '0817800423, '0817800433, '0817800443, '0817800453, '0817800463, '0817800473, '0817800483, '0817800493, '0817900013, '0817900023, '0817900033, '0818200973, '0818200983, '0818200993, '0818201003, '0818201013, '0818201023, '0818201033, '0818201043, '0818201053, '0818201063, '0818903363, '0818903373, '0818903383, '0818903393, '0818903403, '0819002463, '0819002473, '0819002483, '0819002493, '0819002503, '0819100893, '0819100903, '0819100913, '0819100923, '0819100933, '0819605163, '0819605173, '0819605183, '0819605193, '0819605203, '0819700613, '0819700623, '0819700633, '0819700643, '0819700653, '0819800013, '0819800023, '0819800033, '0819800043, '0819800053, '0819802653, '0819802663, '0819802673, '0820013433, '0820013443, '0820013453, '0820013463, '0820013473, '0820702473, '0820702483, '0820702493, '0820702503, '0820702513, '0820702523, '0820702533, '0820702543, '0820702553, '0820702563, '0821103513, '0821103523, '0821103533, '0821103543, '0821103553, '0821103563, '0821103573, '0821103583, '0821103593, '0821103603, '0821201743, '0821201753, '0821201763, '0821201773, '0821201783, '0821201793, '0821201803, '0821201813, '0821201823, '0821201833, '0821903223, '0821903233, '0821903243, '0821903253, '0821903263, '0821903273, '0821903283, '0821903293, '0821903303, '0821903313, '0822408763, '0822408773, '0822408783, '0822408793, '0822408803, '0822408813, '0822408823, '0822408833, '0822408843, '0822408853, '0822603643, '0822603653, '0822603663, '0822603673, '0822603683, '0824203083, '0824203093, '0824203103, '0824203113, '0824203123, '0824203133, '0824203143, '0824203153, '0824203163, '0824203173, '0824901813, '0824901823, '0824901833, '0824901853, '0824901863, '0824901873, '0824901883, '0824901893, '0824901903, '0825400133, '0825400143, '0825400153, '0825500013, '0825500023, '0825500033, '0825500043, '0825500063, '0825500743, '0825500753, '0826000113, '0826000123, '0826000133, '0826000143, '0826000153, '0826000163, '0826000173, '0826000183, '0826000193, '0826000203, '0826000213, '0826000223, '0826000233, '0826000243, '0826000253, '0826000263, '0826000283, '0826000293, '0826000303, '0827016833, '0827016843, '0827016853, '0827016863, '0827016873, '0827016883, '0827016893, '0827016903, '0827016913, '0827016923, '0827016933, '0827016943, '0827016953, '0827016963, '0827016973, '0827401243, '0828401043, '0828401053, '0828401063, '0828401073, '0828401083, '0828401093, '0828401103, '0828401113, '0828401123, '0828401133, '0829014313, '0829014323, '0829014333, '0829014343, '0829014353, '0829014363, '0829014383, '0829014393, '0829014403, '0829700293, '0830101113, '0830101123, '0830101133, '0830101143, '0830101153, '0830101163, '0830101173, '0830101183, '0830101193, '0830101203, '0830101213, '0830101223, '0830101233, '0830101243, '0830101253, '0830101263, '0830101273, '0830101283, '0830101293, '0830101303, '0830203803, '0830203813, '0830203823, '0830203833, '0830203843, '0830203853, '0830203863, '0830203873, '0830203883, '0830203893, '0830502663, '0830502673, '0830502683, '0830502693, '0830502703, '0830502713, '0830502723, '0830502733, '0830502743, '0830502753, '0830503853, '0830902323, '0830902333, '0830902343, '0830902353, '0830902363, '0831000813, '0831000823, '0831503073, '0831503083, '0831503093, '0831503103, '0831503113, '0831801593, '0831801603, '0831801613, '0831801623, '0831801633, '0831801643, '0831801653, '0831801663, '0831801673, '0831801683, '0831801693, '0831801703, '0831801713, '0831801723, '0831801733, '0833904503, '0833904513, '0833904523, '0833904533, '0833904543, '0834604373, '0834604383, '0834604393, '0834604403, '0834604413, '0834703153, '0834703163, '0834703173, '0834703183, '0834703193, '0834703203, '0834703213, '0834703223, '0834703233, '0834703243, '0836400013, '0836400023, '0836400033, '0836400043, '0836400053, '0836400063, '0836400073, '0836400083, '0836400093, '0836400103, '0900700013, '0900700023, '0900700033, '0900700043, '0900700053, '0900700063, '0900700073, '0901200013, '0901200023, '0901200033, '0901200043, '0901200053, '0901200063, '0901200073, '0901304683, '0901304693, '0901304703, '0901304713, '0901504623, '0901504633, '0901504643, '0901504653, '0901504663, '0901504673, '0901504683, '0901504693, '0901504703, '0901504713, '0901504723, '0901504733, '0901504743, '0901504753, '0901504763, '0901617253, '0901617263, '0901617273, '0901617283, '0901617303, '0901617313, '0901617323, '0901617333, '0901617343, '0902000413, '0902000423, '0902000433, '0902000443, '0902000453, '0902109223, '0902109233, '0902109243, '0902109253, '0902109263, '0902109273, '0902109283, '0902109293, '0902109313, '0902109323, '0902109333, '0902109343, '0902109353, '0902109373, '0902109383, '0902109393, '0902109403, '0902109413, '0902305053, '0902305063, '0902305073, '0902305083, '0903402273, '0903402283, '0903402293, '0903402303, '0903402313, '0903402323, '0903402333, '0903402343, '0903402353, '0903402363, '0903402373, '0903402383, '0903402393, '0903402403, '0903402413, '0903402423, '0903402433, '0903402443, '0903402453, '0903402463, '0903703223, '0903703233, '0903703253, '0903703273, '0903703283, '0903703293, '0903703303, '0903703313, '0904027583, '0904027593, '0904027603, '0904027613, '0904027623, '0904027633, '0904027643, '0904027653, '0904027663, '0904027673, '0904210053, '0904210063, '0904210073, '0904210083, '0904210093, '0904210103, '0904210113, '0904210123, '0904210133, '0904210143, '0904302883, '0904302893, '0904302903, '0904302913, '0904302923, '0904302933, '0904302953, '0904302973, '0904401493, '0904401503, '0904401513, '0904401523, '0904401533, '0904401543, '0904401553, '0904401563, '0904401573, '0904401583, '0904727563, '0904727573, '0904727583, '0904727593, '0904727603, '0905001623, '0905001633, '0905001643, '0905001653, '0905001663, '0905001673, '0905001683, '0905001693, '0905001703, '0905001713, '0905103513, '0905103523, '0905103533, '0905103543, '0905103553, '0905103563, '0905103573, '0905103583, '0905103593, '0905103603, '0905802993, '0905803003, '0905803013, '0905803033, '0906106533, '0906106543, '0906106553, '0906106563, '0906106573, '0906106583, '0906106593, '0906106603, '0906106613, '0906106623, '0906106633, '0906106643, '0906106653, '0906106663, '0906106673, '0906106683, '0906106693, '0906106703, '0906106713, '0906106723, '0906106733, '0906106743, '0906106753, '0906106763, '0906106773, '0906106783, '0906106793, '0906106803, '0906106813, '0906106823, '0907003853, '0907003863, '0907003873, '0907003883, '0907003893, '0907003903, '0907003913, '0907003923, '0907003933, '0907003943, '0907003953, '0907003963, '0907003973, '0907003983, '0907003993, '0907004003, '0907004013, '0907004023, '0907004033, '0907004043, '0908405603, '0908405613, '0908405623, '0908405643, '0908405653, '0908405663, '0908405673, '0908405683, '0908405693, '0908405713, '0908405723, '0908405733, '0908405743, '0909105733, '0909105743, '0909105753, '0909105763, '0909105773, '0909622563, '0909622573, '0909622583, '0909622593, '0909622603, '0909622613, '0909622623, '0909622633, '0909622643, '0909622653, '0909622663, '0909622673, '0909622683, '0909622693, '0909622703, '0909622713, '0909622733, '0909622743, '0909622753, '0909622773, '0909622783, '0909622793, '0909622803, '0909622813, '0909622823, '0909622833, '0909622843, '0909622853, '0911207073, '0911207083, '0911207093, '0911207103, '0911207113, '0911207123, '0911207133, '0911207143, '0911207153, '0911207163, '0911207173, '0911207183, '0911207193, '0911207203, '0911207223, '0911207233, '0911207253, '0911207263, '0911207273, '0911207283, '0911207293, '0911207303, '0911207313, '0912422223, '0912422233, '0912422243, '0912422253, '0912422273, '0912422283, '0912422293, '0912422303, '0912422313, '0912422323, '0912422333, '0912422343, '0912422353, '0912422363, '0912708883, '0912708893, '0912708903, '0912708913, '0912708923, '0912708933, '0912708943, '0912708953, '0912708963, '0912708973, '0912708983, '0912708993, '0912709003, '0912709013, '0912709023, '0912709033, '0912709043, '0912709073, '0913900393, '0913900403, '0913900413, '0913900423, '0913900433, '0913900473, '0913900493, '0913900503, '0913900513, '0913900523, '0913900533, '0914727153, '0914727163, '0914727173, '0914727183, '0914727193, '0915200313, '0915200323, '0915200333, '0915200343, '0915200353, '0915301483, '0915301503, '0915301513, '0915301523, '0915303563, '0915303573, '0915303583, '0915303593, '0915303603, '0915303613, '0915303623, '0915303633, '0915303643, '0915303653, '0915501523, '0915501533, '0915501543, '0915501553, '0915501563, '0915501573, '0915501583, '0915501593, '0915501603, '0915502073, '0915502083, '0915502093, '0915502103, '0915502113, '0915900313, '0915900323, '0915900333, '0915900343, '0915900353, '0916001743, '0916001753, '0916001763, '0916001773, '0916001783, '0916101233, '0916101253, '0916101263, '0916101273, '0916101683, '0916101703, '0916101713, '0916101723, '0916900163, '0917300013, '0917300023, '0917300033, '0917300043, '0918102663, '0918102673, '0918102683, '0918102693, '0918102703, '0918700173, '0918700183, '0918700193, '0918700203, '0918700213, '0918700223, '0918700233, '0918700243, '0918700253, '0918700263, '0918700273, '0918700283, '0918700293, '0918700303, '0918700313, '0918700323, '0919402513, '0919402523, '0919402533, '0919402543, '0919402553, '0919402563, '0919402573, '0919402583, '0919402593, '0919402603, '0919601043, '0919601053, '0919601063, '0919601073, '0919601083, '0920100213, '0920100223, '0920100233, '0920100243, '0920100263, '0920100273, '0920100283, '0920100293, '0920100303, '0920100313, '0920100323, '0920100333, '0920100343, '0920100353, '0920100363, '0920100373, '0920100383, '0920100393, '0920100403, '0921101203, '0922409693, '0922409703, '0922409713, '0923603233, '0923603243, '0923603263, '0923603283, '0923603293, '0923603303, '0923603313, '0923603323, '0923603333, '0923603343, '0923603353, '0923603363, '0923603383, '0923603393, '0923603403, '0923603423, '0923603433, '0923603443, '0923603453, '0923603463, '0923603473, '0923800483, '0923800493, '0923800503, '0923800513, '0923800523, '0923800533, '0923800543, '0923800553, '0923800563, '0923800573, '0923800583, '0923800593, '0923800603, '0923800613, '0923800623, '0923800633, '0923800643, '0923800653, '0923800663, '0923800673, '0923800683, '0923800693, '0923800713, '0923800723, '0929300013, '0929902843, '0929902853, '0929902863, '0929902873, '0929902893, '0929902903, '0929902913, '0929902923, '0929902933, '0929902943, '0929902953, '0929902963, '0929902973, '0929902983, '0929902993, '0929903003, '0929903013, '0929903023, '0929903033, '0929903043, '0929903053, '0929903063, '0929903073, '0929903083, '0929903093, '0929903103, '0929903113, '0929903123, '0929903133, '0929903143, '0929903153, '0929903163, '0929903183, '0929903193, '0929903203, '0929903213, '0929903223, '0929903233, '0929903243, '0929903253, '0929903263, '0929903273, '0929903283, '0929903293, '0929903303, '0929903313, '0929903323, '0929903333, '0930203273, '0930203283, '0930203293, '0930203303, '0930203313, '0930203323, '0930203333, '0930203343, '0930203353, '0930203363, '0930203373, '0930203383, '0930203393, '0930203403, '0930203413, '0930203423, '0930203433, '0930203443, '0930203453, '0930203463, '0930203473, '0930203483, '0930203493, '0930203503, '0930203513, '0930203523, '0930203533, '0930203543, '0930203553, '0930203563, '0930203573, '0930203583, '0930203593, '0930203603, '0930203613, '0930203623, '0930203633, '0930203643, '0930203653, '0930203663, '0930203673, '0930203683, '0930203693, '0930203703, '0930203713, '0930203723, '0930203733, '0930203743, '0930203753, '0930203763, '0932004403, '0932004413, '0932004423, '0932004433, '0932004443, '0932004453, '0932004463, '0932004473, '0932004483, '0932004493, '0932004503, '0932004513, '0932004523, '0932004533, '0932004543, '0932004553, '0932004563, '0932004573, '0932004583, '0932004593, '0932004603, '0932004613, '0932004623, '0932004633, '0932004643, '0932004653, '0932004663, '0932004673, '0932004683, '0932004693, '0932004703, '0932004713, '0932004723, '0932004733, '0932004743, '0933400713, '0933400723, '0933400733, '0933400743, '0933700613, '0933700623, '0933700633, '0933700643, '0933700653, '0933700663, '0933700673, '0933700683, '0933700693, '0933700703, '0933810913, '0933810923, '0933810933, '0933810943, '0933810953, '0934100013, '0934100023, '0934100033, '0934100043, '0934100053, '0934200013, '0934200023, '0934200033, '0934200043, '0934200053, '0934307083, '0934307093, '0934307103, '0934307113, '0934307123, '0934307533, '0934307543, '0934307553, '0934307563, '0934307573, '0934500213, '0934500223, '0934500233, '0934500243, '0934500263, '0934500273, '0934500283, '0934500293, '0934500303, '0934500313, '0934500323, '0934500333, '0934500343, '0934500353, '0934900313, '0934900323, '0934900333, '0935201013, '0935201023, '0935201033, '0935201043, '0935201053, '0935600413, '0935600423, '0935600433, '0935600443, '0935600453, '0935600463, '0935600473, '0935600483, '0935600493, '0935600503, '0935702193, '0935702203, '0935702213, '0935702223, '0935702233, '0935709383, '0935709393, '0935709403, '0935709413, '0935709423, '0935800013, '0935800023, '0935800033, '0935800043, '0935800053, '0935800093, '0935800103, '0935800113, '0935800123, '0935800133, '0936300013, '0936300023, '0936300033, '0936300043, '0936300053, '0936301613, '0936301623, '0936301633, '0936301643, '0936301653, '1000400013, '1000400023, '1000400033, '1000400043, '1000400053, '1000500763, '1000500773, '1000500783, '1000500793, '1000500803, '1000500813, '1000500823, '1000500833, '1000500843, '1000500853, '1000511813, '1000511823, '1000511833, '1000511843, '1000511853, '1001200013, '1001200023, '1001200033, '1001200043, '1001200053, '1001300013, '1001300023, '1001300033, '1001300043, '1001300053, '1001400013, '1001400023, '1001400033, '1001500013, '1001500023, '1001500033, '1001500043, '1001800333, '1001800343, '1001800353, '1001800363, '1001800373, '1001800383, '1001800393, '1001800403, '1001800413, '1001800423, '1001800433, '1001800443, '1001800483, '1001800493, '1001800503, '1002901573, '1002901583, '1002901603, '1002901613, '1002901623, '1002901643, '1002901653, '1002901663, '1002901673, '1002901683, '1002901693, '1002901703, '1002901713, '1003200933, '1003200943, '1003200953, '1003200963, '1003200973, '1003200983, '1003200993, '1003201003, '1003201013, '1003201023, '1003201033, '1003201043, '1003201053, '1003201063, '1003201073, '1003201093, '1003201103, '1003201123, '1003504953, '1003504963, '1003504973, '1003504983, '1003504993, '1003505003, '1003505013, '1003505023, '1003505033, '1003505053, '1003505063, '1003505073, '1003505093, '1003505103, '1003505173, '1003900493, '1003900523, '1003900533, '1003900543, '1003900583, '1003900593, '1003900623, '1003900633, '1003900653, '1005010013, '1005300023, '1005300033, '1005300053, '1006100033, '1006100043, '1006100053, '1006300373, '1006300393, '12359, '2115943, '215318, '215333, '215343, '215345, '215349, '215431, '215432, '215433, '215438, '215439, '215441, '215442, '215443, '215444, '215445, '215446, '215452, '215453, '215455, '215456, '215457, '215459, '215462, '215463, '215465, '215466, '215467, '215471, '215472, '215473, '215474, '215475, '215476, '215480, '215481, '215482, '215483, '215487, '215488, '215489, '215490, '215491, '215494, '215496, '215497, '215499, '215500, '215502, '215504, '215508, '215509, '215510, '215511, '215512, '215513, '215514, '215515, '215516, '215517, '215520, '215524, '215525, '215527, '215528, '215529, '215531, '215532, '215533, '215536, '215537, '215538, '215539, '215540, '215542, '215543, '215544, '215545, '215547, '215549, '215550, '2155551, '215558, '215560, '215561, '215562, '215564, '215566, '215567, '215568, '215570, '215571, '215572, '215573, '215576, '215582, '215583, '215587, '215589, '215590, '215591, '215592, '215595, '215598, '215599, '215600, '215601, '215603, '215606, '215607, '215609, '215612, '215614, '215616, '215620, '215621, '215623, '215624, '215625, '215626, '215627, '215629, '215631, '215633, '215634, '215636, '215639, '215640, '215643, '215645, '215646, '215647, '215648, '215649, '215652, '215653, '215654, '215655, '215656, '215658, '215659, '215660, '215661, '215663, '215664, '215665, '215666, '215668, '215671, '215672, '215674, '215675, '215676, '215677, '215678, '215680, '215684, '215685, '215687, '215689, '215691, '215692, '215694, '215696, '215697, '215698, '215699, '215700, '215701, '215702, '215703, '215704, '215709, '215710, '215711, '215712, '215714, '215718, '215719, '215720, '215721, '215722, '215723, '215726, '215728, '215729, '215730, '215731, '215732, '215733, '215735, '215736, '215738, '215739, '215742, '215744, '215745, '215746, '215747, '215749, '215750, '215753, '215755, '215756, '215757, '215758, '215760, '215761, '215762, '215763, '215766, '215767, '215768, '215769, '215770, '215771, '215774, '215776, '215777, '215778, '215779, '215780, '215781, '215782, '215783, '215785, '215786, '215787, '215788, '215789, '215790, '215791, '215792, '215793, '215794, '215795, '215796, '215798, '215799, '215800, '215801, '215803, '215804, '215805, '215806, '215807, '215808, '215809, '215810, '215811, '215812, '215813, '215814, '215815, '215816, '215819, '215820, '215821, '215823, '215824, '215825, '215826, '215827, '215829, '215830, '215831, '215832, '215833, '215834, '215835, '215836, '215837, '215838, '215839, '215840, '215842, '215846, '215847, '215850, '215851, '215853, '215854, '215855, '215856, '215857, '215858, '215859, '215860, '215861, '215864, '215865, '215866, '215867, '215868, '215869, '215870, '215871, '215872, '215873, '215874, '215875, '215877, '215878, '215879, '215880, '215881, '215882, '215883, '215884, '215885, '215886, '215887, '215888, '215889, '215890, '215891, '215892, '215893, '215896, '215897, '215899, '215900, '215904, '215905, '215907, '215908, '215910, '215911, '215912, '215913, '215914, '215915, '215916, '215917, '215918, '215919, '215920, '215921, '215922, '215923, '215924, '215925, '215927, '215928, '215929, '215930, '215931, '215932, '215933, '215934, '215939, '215940, '215941, '215944, '215945, '215946, '215947, '215949, '215951, '215953, '215954, '215955, '215956, '215957, '215959, '215960, '215961, '215962, '215963, '215964, '215965, '215966, '215968, '215969, '215970, '215972, '215975, '215979, '215981, '215983, '215996, '216000, '216008, '216014, '216024, '216026, '216028, '216029, '216035, '216037, '216039, '216042, '216044, '216049, '216058, '216062, '216064, '216066, '216071, '216074, '216085, '216092, '216095, '216104, '216107, '216111, '216119, '216120, '216122, '216127, '216128, '251753, '276514, '276515, '276516, '276517, '276519, '276520, '276522, '276523, '276525, '276527, '276528, '276529, '276531, '276532, '276535, '276536, '276538, '276539, '276541, '276543, '276548, '276549, '276550, '276551, '276553, '276554, '276555, '276556, '276558, '276560, '276563, '276564, '276565, '276566, '276567, '276568, '276569, '276570, '276571, '276572, '276573, '276574, '276575, '276576, '276577, '276578, '276579, '276580, '276583, '276586, '276588, '276589, '276591, '276592, '276595, '276596, '276597, '276599, '276601, '276603, '276604, '276607, '276608, '276609, '276611, '276615, '276616, '276617, '276618, '276623, '276624, '276625, '276627, '276628, '276629, '276630, '276631, '276632, '276633, '276634, '276637, '276639, '276640, '276641, '276642, '276643, '276644, '276645, '276646, '276647, '276648, '276649, '276651, '276655, '276657, '276658, '276665, '276666, '276667, '276668, '276670, '276671, '276673, '276675, '276678, '276679, '276680, '276681, '276682, '276683, '276684, '276686, '276689, '276691, '276692, '276693, '276697, '276698, '276699, '276700, '276701, '276702, '276703, '276704, '276705, '276706, '276707, '276708, '276709, '276710, '276711, '276713, '276714, '276715, '276716, '276718, '276719, '276721, '276722, '276723, '276724, '276725, '276726, '276733, '276739, '276740, '276741, '276751, '276752, '276754, '276755, '276756, '276757, '276760, '276762, '276765, '276766, '276767, '276768, '276769, '276770, '276772, '276779, '276780, '276781, '276782, '276783, '276787, '276790, '276791, '276792, '276793, '276796, '276797, '276799, '276802, '276804, '276807, '276815, '276816, '276817, '276818, '276820, '276821, '276824, '276827, '276829, '276830, '277000, '277001, '277002, '277003, '277004, '277005, '277006, '277007, '277008, '277009, '277010, '277011, '277012, '277013, '277014, '277015, '277016, '277017, '277018, '277019, '277020, '277021, '277022, '277023, '277024, '277025, '277026, '277027, '277028, '277029, '277030, '277031, '277032, '277033, '277034, '277035, '277036, '277037, '277038, '277039, '277040, '277041, '277042, '277043, '277044, '277045, '277046, '277047, '277048, '277049, '277050, '277051, '277052, '277053, '277054, '277055, '277056, '277057, '277058, '277059, '277060, '277061, '277062, '277063, '277064, '277065, '277066, '277067, '277068, '277069, '277070, '277071, '277072, '277073, '277074, '277075, '277076, '277077, '277078, '277079, '277080, '277081, '277082, '277083, '277084, '277085, '277086, '277087, '277088, '277089, '277090, '277091, '277092, '277093, '277094, '277095, '277096, '277097, '277098, '277099, '277100, '277101, '277102, '277103, '277104, '277105, '277106, '277107, '277108, '277109, '277110, '277111, '277112, '277113, '277114, '277115, '277116, '277117, '277118, '277119, '277120, '277121, '277122, '277123, '277124, '277125, '277126, '277127, '277128, '277129, '277130, '277131, '277132, '277133, '277134, '277135, '277136, '277137, '277138, '277139, '277140, '277141, '277142, '277143, '277144, '277145, '277146, '277147, '277148, '277149, '277150, '277151, '277152, '277153, '277154, '277155, '277156.'277157, '277158, '277160, '277161, '277162, '277163, '277164, '277165, '277166, '277167, '277168, '277169, '277170, '277171, '277172, '277173, '277174, '277175, '277176, '277177, '277178, '277179, '277180, '277181, '277182, '277183, '277184, '277185, '277186, '277187, '277188, '277189, '277190, '277191, '277192, '277193, '277194, '277195, '277196, '277197, '277198, '277199, '277200, '277201, '277202, '277203, '277204, '277205, '277206, '277207, '277208, '277209, '277210, '277211, '277212, '277213, '277214, '277215, '277216, '277217, '277218, '277219, '277220, '277221, '277222, '277223, '277224, '277225, '277226, '277227, '277228, '277229, '277230, '277231, '277232, '277233, '277234, '277235, '277236, '277237, '277238, '277239, '277240, '277241, '277242, '277243, '277244, '277245, '277246, '277247, '277248, '277249, '277250, '277251, '277252, '277253, '277254, '277255, '277256, '277257, '277258, '277259, '277260, '277261, '277263, '277264, '277265, '277266, '277267, '277268, '277269, '277270, '277271, '277272, '277273, '277274, '277275, '277276, '277277, '277278, '277279, '277280, '277281, '277282, '277283, '277284, '277285, '277286, '277288, '277289, '277290, '277291, '277292, '277293, '277294, '277295, '277296, '277297, '277298, '277299, '277300, '277301, '277302, '277303, '277304, '277305, '277306, '277308, '277309, '277310, '277311, '277313, '277314, '277315, '277316, '277317, '277318, '277319, '277320, '277321, '277322, '277323, '277324, '277325, '277326, '277327, '277328, '277329, '277330, '277331, '277332, '277333, '277334, '277335, '277336, '277337, '277338, '277339, '277340, '277341, '277342, '277343, '277344, '277345, '277346, '277347, '277348, '277349, '277350, '277351, '277352, '277353, '277354, '277355, '277356, '277357, '277359, '277360, '277362, '277363, '277364, '277365, '277366, '277367, '277368, '277369, '277370, '277371, '277372, '277373, '277374, '277385, '277386, '277387, '277388, '277389, '277390, '277391, '277392, '277393, '277394, '277395, '277396, '277397, '277398, '277399, '277400, '277401, '277402, '277403, '277404, '277405, '277406, '277407, '277408, '277409, '277410, '277411, '277412, '277413, '277414, '277415, '277416, '277417, '277418, '277419, '277420, '277421, '277422, '277423, '277424, '277425, '277426, '277427, '277428, '277429, '277430, '277431, '277432, '277433, '277434, '277435, '277436, '277437, '277438, '277439, '277440, '277441, '277442, '277443, '277444, '277445, '277446, '277447, '277448, '277449, '277450, '277451, '277452, '277453, '277454, '277455, '277456, '277457, '277458 and '277459. |
| FEI Number |
1811755
|
Recalling Firm/ Manufacturer |
Stryker Instruments Div. of Stryker Corporation
4100 East Milham Ave.
Kalamazoo
MI
49001
|
| For Additional Information Contact | 800-800-4236 Ext. 3808 |
Manufacturer Reason for Recall | This recall is related to the Electi-cord recall. There is a risk when excessive force is applied to the plugs, a fracture can develop on the prongs inside the molded section of the plug. Over time the fracture may lead to arcing inside of the molded sction of the plug and a build up of heat which can cause charring, electric shock or fire. To this date there have been zero complaint for this fa |
FDA Determined Cause 2 | Component design/selection |
| Action | Stryker Instruments sent an Urgent - Medical Device Recall Notification letter dated April 13, 2010, to all affected customers. The letter identified the product, the problem,and the action to be taken by the customer. Stryker notified customers again on May 24, 2010, and again on September 14, 2010, via email, fax, or telephone.
Customers were instructed to locate the units and visually inspect each affected cord. Remove the affected cord and dispose of it using their healthcare facilities disposal procedure. If the cord did not show signs of damage, the unit could remain in service until a replacement cord was available.
Customers were to return the enclosed business reply form to confirm receipt of the notification letter and identify how many affected units were currently in their inventory. Stryker instruments would send out replacement cords.
If the customer loaned or sold any of the units listed in the notification, the customer was to forward a copy of the notice to the new users and advise Stryker of the new user's name and their new location.
If products were disposed of and were no longer in use, the customer was to advise Stryker of their obsolescence by providing them with their serial numbers.
For questions regarding this recall call 800-800-4236, ext 3808 or ext 3584 Monday -Friday 8am-5pm (EST). |
| Distribution | Worldwide Distribution - USA (nationwide) and the countries of Portugal, Canada, China, Taiwan, France, Germany, Greece, Spain, India, Italy, Korea, Switzerland, Singapore, and England. |
| Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 �7.55.
|