|
Class 2 Device Recall Vital Signs Monitor, Model 93300 |
|
Date Initiated by Firm |
September 23, 2010 |
Date Posted |
August 15, 2011 |
Recall Status1 |
Terminated 3 on August 17, 2011 |
Recall Number |
Z-3024-2011 |
Recall Event ID |
56839 |
510(K)Number |
K093501
|
Product Classification |
Monitor, physiological, patient (with arrhythmia detection or alarms) - Product Code MHX
|
Product |
Spacelabs Medical elance Vital Signs Monitor, Model 93300.
The Spacelabs elance Vital Signs Monitor and elance Central Station is a prescription device intended for use by healthcare professionals in all areas of a healthcare facility. It is indicated for use in adult and pediatric populations. The device facilitates the monitoring of: ECG with arrhythmia detection; respiration; non-invasive blood pressures; invasive blood pressures; body temperature; functional arterial oxygen saturation; and entidal CO2. |
Code Information |
Model 93300. Serial numbers: 3300-000001, 3300-000002, 3300-000003, 3300-000005, 3300-000006, 3300-000007, 3300-000008, 3300-000009, 3300-000010, 3300-000011, 3300-000012, 3300-000013, 3300-000014, 3300-000015, 3300-000016, 3300-000017, 3300-000018, 3300-000019, 3300-000020, 3300-000021, 3300-000022, 3300-000023, 3300-000024, 3300-000026, 3300-000027, 3300-000028, 3300-000029, 3300-000030, 3300-000031, 3300-000032, 3300-000033, 3300-000034, 3300-000035, 3300-000036, 3300-000037, 3300-000038, 3300-000039, 3300-000040, 3300-000041, 3300-000042, 3300-000043, 3300-000044, 3300-000045, 3300-000046, 3300-000047, 3300-000049, 3300-000050, 3300-000052, 3300-000053, 3300-000054, 3300-000055, 3300-000056, 3300-000057, 3300-000058, 3300-000059, 3300-000060, 3300-000061, 3300-000062, 3300-000063, 3300-000064, 3300-000065, 3300-000066, 3300-000067, 3300-000068, 3300-000069, 3300-000070, 3300-000071, 3300-000072, 3300-000073, 3300-000074, 3300-000075, 3300-000076, 3300-000077, 3300-000078, 3300-000079, 3300-000080, 3300-000081, 3300-000082, 3300-000083, 3300-000084, 3300-000085, 3300-000086, 3300-000087, 3300-000088, 3300-000089, 3300-000090, 3300-000091, 3300-000092, 3300-000093, 3300-000094, 3300-000095, 3300-000096, 3300-000097, 3300-000098, 3300-000099, 3300-000100, 3300-000101, 3300-000102, 3300-000103, 3300-000104, 3300-000105, 3300-000106, 3300-000107, 3300-000108, 3300-000109, 3300-000110, 3300-000111, 3300-000112, 3300-000113, 3300-000114, 3300-000115, 3300-000116, 3300-000117, 3300-000118, 3300-000119, 3300-000120, 3300-000121, 3300-000122, 3300-000123, 3300-000124, 3300-000125, 3300-000126, 3300-000127, 3300-000128, 3300-000129, 3300-000130, 3300-000131, 3300-000132, 3300-000133, 3300-000134, 3300-000135, 3300-000136, 3300-000137, 3300-000138, 3300-000139, 3300-000140, 3300-000141, 3300-000142, 3300-000143, 3300-000144, 3300-000145, 3300-000146, 3300-000147, 3300-000148, 3300-000149, 3300-000150, 3300-000151, 3300-000152, 3300-000153, 3300-000154, 3300-000155, 3300-000156, 3300-000157, 3300-000158, 3300-000159, 3300-000160, 3300-000161, 3300-000162, 3300-000163, 3300-000164, 3300-000165, 3300-000166, 3300-000167, 3300-000168, 3300-000169, 3300-000170, 3300-000171, 3300-000172, 3300-000173, 3300-000174, 3300-000175, 3300-000176, 3300-000177, 3300-000178, 3300-000179, 3300-000180, 3300-000181, 3300-000182, 3300-000183, 3300-000184, 3300-000185, 3300-000186, 3300-000187, 3300-000188, 3300-000189, 3300-000190, 3300-000191, 3300-000192, 3300-000193, 3300-000194, 3300-000195, 3300-000196, 3300-000197, 3300-000198, 3300-000199, 3300-000200, 3300-000201, 3300-000202, 3300-000203, 3300-000204, 3300-000205, 3300-000206, 3300-000207, 3300-000208, 3300-000209, 3300-000210, 3300-000211, 3300-000212, 3300-000213, 3300-000214, 3300-000215, 3300-000216, 3300-000217, 3300-000218, 3300-000219, 3300-000220, 3300-000221, 3300-000222, 3300-000223, 3300-000224, 3300-000225, 3300-000226, 3300-000227, 3300-000228, 3300-000229, 3300-000230, 3300-000231, 3300-000232, 3300-000233, 3300-000234, 3300-000235, 3300-000236, 3300-000237, 3300-000238, 3300-000239, 3300-000240, 3300-000241, 3300-000242, 3300-000243, 3300-000244, 3300-000245, 3300-000246, 3300-000247, 3300-000248, 3300-000249, 3300-000250, 3300-000251, 3300-000252, 3300-000253, 3300-000254, 3300-000255, 3300-000256, 3300-000257, 3300-000258, 3300-000259, 3300-000260, 3300-000261, 3300-000262, 3300-000263, 3300-000264, 3300-000265, 3300-000266, 3300-000267, 3300-000268, 3300-000269, 3300-000270, 3300-000271, 3300-000272, 3300-000273, 3300-000274, 3300-000275, 3300-000276, 3300-000277, 3300-000278, 3300-000279, 3300-000280, 3300-000281, 3300-000282, 3300-000283, 3300-000284, 3300-000285, 3300-000286, 3300-000287, 3300-000288, 3300-000289, 3300-000290, 3300-000291, 3300-000292, 3300-000293, 3300-000294, 3300-000295, 3300-000296, 3300-000297, 3300-000298, 3300-000299, 3300-000300, 3300-000301, 3300-000302, 3300-000303, 3300-000304, 3300-000305, 3300-000306, 3300-000307, 3300-000308, 3300-000309, 3300-000310, 3300-000311, 3300-000312, 3300-000313, 3300-000314, 3300-000315, 3300-000316, 3300-000317, 3300-000318, 3300-000319, 3300-000320, 3300-000321, 3300-000322, 3300-000323, 3300-000324, 3300-000325, 3300-000326, 3300-000327, 3300-000328, 3300-000329, 3300-000330, 3300-000331, 3300-000332, 3300-000333, 3300-000334, 3300-000335, 3300-000336, 3300-000337, 3300-000338, 3300-000339, 3300-000340, 3300-000341, 3300-000342, 3300-000343, 3300-000344, 3300-000345, 3300-000346, 3300-000347, 3300-000348, 3300-000349, 3300-000350, 3300-000351, 3300-000352, 3300-000353, 3300-000354, 3300-000355, 3300-000356, 3300-000357, 3300-000358, 3300-000359, 3300-000360, 3300-000361, 3300-000362, 3300-000363, 3300-000364, 3300-000365, 3300-000366, 3300-000367, 3300-000368, 3300-000369, 3300-000370, 3300-000371, 3300-000372, 3300-000373, 3300-000374, 3300-000375, 3300-000376, 3300-000377, 3300-000378, 3300-000379, 3300-000380, 3300-000381, 3300-000382, 3300-000383, 3300-000384, 3300-000385, 3300-000386, 3300-000387, 3300-000388, 3300-000389, 3300-000390, 3300-000391, 3300-000392, 3300-000393, 3300-000394, 3300-000395, 3300-000396, 3300-000397, 3300-000398, 3300-000399, 3300-000400, 3300-000401, 3300-000402, 3300-000403, 3300-000404, 3300-000405, 3300-000406, 3300-000407, 3300-000408, 3300-000409, 3300-000410, 3300-000411, 3300-000412, 3300-000413, 3300-000414, 3300-000415, 3300-000416, 3300-000417, 3300-000418, 3300-000419, 3300-000420, 3300-000421, 3300-000422, 3300-000423, 3300-000424, 3300-000425, 3300-000426, 3300-000427, 3300-000428, 3300-000429, 3300-000430, 3300-000431, 3300-000432, 3300-000433, 3300-000434, 3300-000435, 3300-000436, 3300-000437, 3300-000438, 3300-000439, 3300-000440, 3300-000441, 3300-000442, 3300-000443, 3300-000444, 3300-000445, 3300-000446, 3300-000447, 3300-000448, 3300-000449, 3300-000450, 3300-000451, 3300-000452, 3300-000453, 3300-000454, 3300-000455, 3300-000456, 3300-000457, 3300-000458, 3300-000459, 3300-000460, 3300-000461, 3300-000462, 3300-000463, 3300-000464, 3300-000465, 3300-000466, 3300-000467, 3300-000468, 3300-000469, 3300-000470, 3300-000471, 3300-000472, 3300-000473, 3300-000474, 3300-000475, 3300-000476, 3300-000477, 3300-000478, 3300-000479, 3300-000480, 3300-000481, 3300-000482, 3300-000483, 3300-000484, 3300-000485, 3300-000486, 3300-000487, 3300-000488, 3300-000489, 3300-000490, 3300-000491, 3300-000492, 3300-000493, 3300-000494, 3300-000495, 3300-000496, 3300-000497, 3300-000498, 3300-000499, 3300-000500, 3300-000501, 3300-000502, 3300-000503, 3300-000504, 3300-000505, 3300-000506, 3300-000507, 3300-000508, 3300-000509, 3300-000510, 3300-000511, 3300-000512, 3300-000513, 3300-000514, 3300-000515, 3300-000516, 3300-000517, 3300-000518, 3300-000519, 3300-000520, 3300-000521, 3300-000522, 3300-000523, 3300-000524, 3300-000525, 3300-000526, 3300-000527, 3300-000528, 3300-000529, 3300-000530, 3300-000531, 3300-000532, 3300-000533, 3300-000534, 3300-000535, 3300-000536, 3300-000537, 3300-000538, 3300-000539, 3300-000540, 3300-000541, 3300-000542, 3300-000543, 3300-000544, 3300-000545, 3300-000546, 3300-000547, 3300-000548, 3300-000549, 3300-000550, 3300-000551, 3300-000552, 3300-000553, 3300-000554, 3300-000555, 3300-000556, 3300-000557, 3300-000558, 3300-000559, 3300-000560, 3300-000561, 3300-000562, 3300-000563, 3300-000564, 3300-000565, 3300-000566, 3300-000567, 3300-000568, 3300-000569, 3300-000570, 3300-000571, 3300-000572, 3300-000573, 3300-000574, 3300-000577, 3300-000578, 3300-000579, 3300-000580, 3300-000581, 3300-000582, 3300-000583, 3300-000584, 3300-000585, 3300-000586, 3300-000587, 3300-000588, 3300-000589, 3300-000590, 3300-000591, 3300-000592, 3300-000593, 3300-000594, 3300-000595, 3300-000596, 3300-000597, 3300-000598, 3300-000599, 3300-000600, 3300-000601, 3300-000602, 3300-000603, 3300-000604, 3300-000605, 3300-000607, 3300-000608, 3300-000609, 3300-000610, 3300-000611, 3300-000612, 3300-000613, 3300-000614, 3300-000615, 3300-000616, 3300-000617, 3300-000618, 3300-000619, 3300-000620.3300-000621, 3300-000622, 3300-000623, 3300-000624, 3300-000625, 3300-000626, 3300-000627, 3300-000628, 3300-000629, 3300-000630, 3300-000631, 3300-000632, 3300-000633, 3300-000634, 3300-000635, 3300-000636, 3300-000637, 3300-000638, 3300-000639, 3300-000640, 3300-000641, 3300-000642, 3300-000643, 3300-000644, 3300-000645, 3300-000646, 3300-000647, 3300-000648, 3300-000649, 3300-000650, 3300-000651, 3300-000652, 3300-000653, 3300-000654, 3300-000655, 3300-000656, 3300-000657, 3300-000658, 3300-000659, 3300-000660, 3300-000661, 3300-000662, 3300-000663, 3300-000664, 3300-000665, 3300-000666, 3300-000667, 3300-000668, 3300-000669, 3300-000670, 3300-000671, 3300-000672, 3300-000673, 3300-000674, 3300-000675, 3300-000676, 3300-000677, 3300-000678, 3300-000679, 3300-000680, 3300-000681, 3300-000682, 3300-000683, 3300-000684, 3300-000685, 3300-000686, 3300-000687, 3300-000688, 3300-000689, 3300-000690, 3300-000691, 3300-000692, 3300-000693, 3300-000694, 3300-000695, 3300-000696, 3300-000697, 3300-000698, 3300-000699, 3300-000700, 3300-000701, 3300-000702, 3300-000703, 3300-000704, 3300-000705, 3300-000706, 3300-000707, 3300-000708, 3300-000709, 3300-000710, 3300-000711, 3300-000712, 3300-000713, 3300-000714, 3300-000715, 3300-000716, 3300-000717, 3300-000718, 3300-000719, 3300-000720, 3300-000721, 3300-000722, 3300-000723, 3300-000724, 3300-000725, 3300-000726, 3300-000727, 3300-000728, 3300-000729, 3300-000730, 3300-000731, 3300-000732, 3300-000733, 3300-000734, 3300-000735, 3300-000736, 3300-000737, 3300-000738, 3300-000739, 3300-000740, 3300-000741, 3300-000742, 3300-000743, 3300-000744, 3300-000745, 3300-000746, 3300-000747, 3300-000748, 3300-000749, 3300-000750, 3300-000751, 3300-000752, 3300-000753, 3300-000754, 3300-000755, 3300-000756, 3300-000757, 3300-000758, 3300-000759, 3300-000760, 3300-000761, 3300-000762, 3300-000763, 3300-000764, 3300-000765, 3300-000766, 3300-000767, 3300-000768, 3300-000769, 3300-000770, 3300-000771, 3300-000772, 3300-000773, 3300-000774, 3300-000775, 3300-000776, 3300-000777, 3300-000778, 3300-000779, 3300-000780, 3300-000781, 3300-000782, 3300-000783, 3300-000784, 3300-000785, 3300-000786, 3300-000787, 3300-000788, 3300-000789, 3300-000790, 3300-000791, 3300-000792, 3300-000793, 3300-000794, 3300-000795, 3300-000796, 3300-000797, 3300-000798, 3300-000799, 3300-000800, 3300-000801, 3300-000802, 3300-000803, 3300-000804, 3300-000805, 3300-000806, 3300-000807, 3300-000808, 3300-000809, 3300-000810, 3300-000811, 3300-000812, 3300-000813, 3300-000814, 3300-000815, 3300-000816, 3300-000817, 3300-000818, 3300-000819, 3300-000820, 3300-000821, 3300-000822, 3300-000823, 3300-000824, 3300-000825, 3300-000826, 3300-000827, 3300-000828, 3300-000829, 3300-000830, 3300-000831, 3300-000832, 3300-000833, 3300-000834, 3300-000835, 3300-000836, 3300-000837, 3300-000838, 3300-000839, 3300-000840, 3300-000841, 3300-000842, 3300-000843, 3300-000844, 3300-000845, 3300-000846, 3300-000847, 3300-000848, 3300-000849, 3300-000850, 3300-000851, 3300-000852, 3300-000853, 3300-000854, 3300-000855, 3300-000856, 3300-000857, 3300-000858, 3300-000859, 3300-000860, 3300-000861, 3300-000862, 3300-000863, 3300-000864, 3300-000865, 3300-000866, 3300-000867, 3300-000868, 3300-000869, 3300-000870, 3300-000871, 3300-000872, 3300-000873, 3300-000874, 3300-000875, 3300-000876, 3300-000877, 3300-000878, 3300-000879, 3300-000880, 3300-000881, 3300-000882, 3300-000883, 3300-000884, 3300-000885, 3300-000886, 3300-000887, 3300-000888, 3300-000889, 3300-000890, 3300-000891, 3300-000892, 3300-000893, 3300-000894, 3300-000895, 3300-000896, 3300-000897, 3300-000898, 3300-000899, 3300-000900, 3300-000901, 3300-000902, 3300-000903, 3300-000904, 3300-000905, 3300-000906, 3300-000907, 3300-000908, 3300-000909, 3300-000910, 3300-000911, 3300-000912, 3300-000913, 3300-000914, 3300-000915, 3300-000916, 3300-000917, 3300-000918, 3300-000919, 3300-000920, 3300-000921, 3300-000922, 3300-000923, 3300-000924, 3300-000925, 3300-000926, 3300-000927, 3300-000928, 3300-000929, 3300-000930, 3300-000931, 3300-000932, 3300-000933, 3300-000934, 3300-000935, 3300-000936, 3300-000937, 3300-000938, 3300-000939, 3300-000940, 3300-000941, 3300-000942, 3300-000943, 3300-000944, 3300-000945, 3300-000946, 3300-000947, 3300-000948, 3300-000949, 3300-000950, 3300-000951, 3300-000952, 3300-000953, 3300-000954, 3300-000955, 3300-000956, 3300-000957, 3300-000958, 3300-000959, 3300-000960, 3300-000961, 3300-000962, 3300-000963, 3300-000964, 3300-000965, 3300-000967, 3300-000968, 3300-000969, 3300-000970, 3300-000971, 3300-000972, 3300-000973, 3300-000974, 3300-000975, 3300-000976, 3300-000977, 3300-000978, 3300-000979, 3300-000980, 3300-000981, 3300-000982, 3300-000983, 3300-000984, 3300-000985, 3300-000986, 3300-000987, 3300-000988, 3300-000989, 3300-000990, 3300-000991, 3300-000992, 3300-000993, 3300-000994, 3300-000996, 3300-000997, 3300-000998, 3300-000999, 3300-001000, 3300-001001, 3300-001002, 3300-001003, 3300-001004, 3300-001005, 3300-001006, 3300-001107, 3300-001108, 3300-001109, 3300-001111, 3300-001112, 3300-001113, 3300-001114, 3300-001115, 3300-001116, 3300-001117, 3300-001118, 3300-001119, 3300-001120, 3300-001121, 3300-001122, 3300-001123, 3300-001124, 3300-001125, 3300-001126, 3300-001127, 3300-001128, 3300-001129, 3300-001130, 3300-001131, 3300-001132, 3300-001133, 3300-001134, 3300-001135, 3300-001136, 3300-001137, 3300-001138, 3300-001139, 3300-001140, 3300-001141, 3300-001142, 3300-001143, 3300-001144, 3300-001145, 3300-001146, 3300-001147, 3300-001148, 3300-001149, 3300-001150, 3300-001151, 3300-001152, 3300-001153, 3300-001154, 3300-001155, 3300-001156, 3300-001157, 3300-001158, 3300-001159, 3300-001160, 3300-001161, 3300-001162, 3300-001163, 3300-001164, 3300-001165, 3300-001166, 3300-001167, 3300-001168, 3300-001169, 3300-001170, 3300-001171, 3300-001172, 3300-001173, 3300-001174, 3300-001175, 3300-001176, 3300-001177, 3300-001178, 3300-001179, 3300-001180, 3300-001181, 3300-001182, 3300-001183, 3300-001184, 3300-001185, 3300-001186, 3300-001187, 3300-001188, 3300-001189, 3300-001190, 3300-001191, 3300-001192, 3300-001193, 3300-001194, 3300-001195, 3300-001196, 3300-001197, 3300-001198, 3300-001199, 3300-001200, 3300-001201, 3300-001202, 3300-001203, 3300-001204, 3300-001205, 3300-001206, 3300-001207, 3300-001208, 3300-001209, 3300-001210, 3300-001211, 3300-001212, 3300-001213, 3300-001214, 3300-001215, 3300-001216, 3300-001217, 3300-001218, 3300-001219, 3300-001220, 3300-001221, 3300-001222, 3300-001223, 3300-001224, 3300-001225, 3300-001226, 3300-001227, 3300-001228, 3300-001229, 3300-001230, 3300-001231, 3300-001232, 3300-001233, 3300-001234, 3300-001235, 3300-001236, 3300-001237, 3300-001238, 3300-001239, 3300-001240, 3300-001241, 3300-001242, 3300-001243, 3300-001244, 3300-001245, 3300-001246, 3300-001247, 3300-001248, 3300-001249, 3300-001250, 3300-001251, 3300-001252, 3300-001253, 3300-001254, 3300-001255, 3300-001256, 3300-001257, 3300-001258, 3300-001259, 3300-001260, 3300-001261, 3300-001262, 3300-001263, 3300-001264, 3300-001265, 3300-001266, 3300-001267, 3300-001268, 3300-001269, 3300-001270, 3300-001271, 3300-001272, 3300-001273, 3300-001274, 3300-001275, 3300-001276, 3300-001277, 3300-001278, 3300-001279, 3300-001280, 3300-001281, 3300-001282, 3300-001283, 3300-001284, 3300-001285, 3300-001286, 3300-001287, 3300-001288, 3300-001289, 3300-001290, 3300-001291, 3300-001292, 3300-001293, 3300-001294, 3300-001295, 3300-001296, 3300-001297, 3300-001298, 3300-001299, 3300-001300, 3300-001301, 3300-001302, 3300-001303, 3300-001304, 3300-001305, 3300-001306, 3300-001307, 3300-001308, 3300-001309, 3300-001310, 3300-001311, 3300-001312, 3300-001313, 3300-001314, 3300-001315, 3300-001316, 3300-001317, 3300-001318, 3300-001319, 3300-001320, 3300-001321, 3300-001322, 3300-001323, 3300-001324, 3300-001325, 3300-001326, 3300-001327, 3300-001328, 3300-001329, 3300-001330, 3300-001331, 3300-001332, 3300-001333, 3300-001334, 3300-001335, 3300-001336, 3300-001337, 3300-001338, 3300-001339, 3300-001340, 3300-001341, 3300-001342, 3300-001343, 3300-001344, 3300-001345, 3300-001346, 3300-001347, 3300-001348, 3300-001349, 3300-001350, 3300-001351, 3300-001352, 3300-001353, 3300-001354, 3300-001355, 3300-001356, 3300-001357, 3300-001358, 3300-001359, 3300-001360, 3300-001361, 3300-001362, 3300-001363, 3300-001364, 3300-001365, 3300-001366, 3300-001367, 3300-001368, 3300-001369, 3300-001370, 3300-001371, 3300-001372, 3300-001373, 3300-001374, 3300-001375, 3300-001376, 3300-001377, 3300-001378, 3300-001379, 3300-001380, 3300-001381, 3300-001382, 3300-001383, 3300-001384, 3300-001386, 3300-001387, 3300-001388, 3300-001389, 3300-001390, 3300-001391, 3300-001392, 3300-001393, 3300-001394, 3300-001395, 3300-001396, 3300-001397, 3300-001398, 3300-001399, 3300-001400, 3300-001401, 3300-001402, 3300-001403, 3300-001404, 3300-001405, 3300-001406, 3300-001407, 3300-001408, 3300-001409, 3300-001412, 3300-001413, 3300-001414, 3300-001415, 3300-001416, 3300-001417, 3300-001418, 3300-001419, 3300-001420, 3300-001421, 3300-001422, 3300-001423, 3300-001424, 3300-001425, 3300-001426, 3300-001427, 3300-001428, 3300-001429, 3300-001430, 3300-001431, 3300-001432, 3300-001433, 3300-001434, 3300-001435, 3300-001436, 3300-001437, 3300-001438, 3300-001439, 3300-001440, 3300-001444, 3300-001445, 3300-001446, 3300-001447, 3300-001448, 3300-001449, 3300-001450, 3300-001451, 3300-001452, 3300-001453, 3300-001454, 3300-001455, 3300-001456, 3300-001457, 3300-001458, 3300-001459, 3300-001460, 3300-001461, 3300-001462, 3300-001463, 3300-001464, 3300-001465, 3300-001466, 3300-001467, 3300-001468, 3300-001469, 3300-001470, 3300-001471, 3300-001472, 3300-001473, 3300-001474, 3300-001475, 3300-001476, 3300-001477, 3300-001478, 3300-001479, 3300-001480, 3300-001481, 3300-001482, 3300-001483, 3300-001484, 3300-001485, 3300-001486, 3300-001487, 3300-001488, 3300-001489, 3300-001490, 3300-001491, 3300-001492, 3300-001493, 3300-001494, 3300-001495, 3300-001496, 3300-001497, 3300-001499, 3300-001500, 3300-001501, 3300-001502, 3300-001503, 3300-001504, 3300-001505, 3300-001506, 3300-001507, 3300-001508, 3300-001509, 3300-001510, 3300-001511, 3300-001512, 3300-001513, 3300-001514, 3300-001515, 3300-001516, 3300-001517, 3300-001518, 3300-001519, 3300-001520, 3300-001521, 3300-001522, 3300-001523, 3300-001524, 3300-001525, 3300-001526, 3300-001527, 3300-001528, 3300-001529, 3300-001530, 3300-001531, 3300-001532, 3300-001533, 3300-001534, 3300-001535, 3300-001536, 3300-001537, 3300-001538, 3300-001539, 3300-001540, 3300-001541, 3300-001542, 3300-001543, 3300-001544, 3300-001545, 3300-001546, 3300-001547, 3300-001548, 3300-001551, 3300-001552, 3300-001553, 3300-001554, 3300-001555, 3300-001556, 3300-001557, 3300-001558, 3300-001559, 3300-001560, 3300-001561, 3300-001562, 3300-001563, 3300-001564, 3300-001565, 3300-001566, 3300-001567, 3300-001568, 3300-001569, 3300-001570, 3300-001571, 3300-001572, 3300-001573, 3300-001574, 3300-001575, 3300-001576, 3300-001577, 3300-001578, 3300-001579, 3300-001580, 3300-001581, 3300-001582, 3300-001583, 3300-001584, 3300-001585, 3300-001586, 3300-001587, 3300-001588, 3300-001589, 3300-001590, 3300-001591, 3300-001592, 3300-001593, 3300-001594, 3300-001595, 3300-001596, 3300-001597, 3300-001598, 3300-001599, 3300-001600, 3300-001601, 3300-001602, 3300-001603, 3300-001604, 3300-001605, 3300-001606, 3300-001607, 3300-001608, 3300-001609, 3300-001610, 3300-001611, 3300-001612, 3300-001613, 3300-001614, 3300-001615, 3300-001616, 3300-001617, 3300-001618, 3300-001619, 3300-001620, 3300-001621, 3300-001622, 3300-001623, 3300-001624, 3300-001625, 3300-001626, 3300-001627, 3300-001628, 3300-001629, 3300-001630, 3300-001631, 3300-001632, 3300-001633, 3300-001634, 3300-001635, 3300-001636, 3300-001637, 3300-001638, 3300-001639, 3300-001640, 3300-001641, 3300-001642, 3300-001643, 3300-001644, 3300-001645, 3300-001646, 3300-001647, 3300-001648, 3300-001649, 3300-001650, 3300-001651, 3300-001652, 3300-001653, 3300-001654, 3300-001655, 3300-001656, 3300-001657, 3300-001658, 3300-001659, 3300-001660, 3300-001661, 3300-001662, 3300-001663, 3300-001664, 3300-001665, 3300-001666, 3300-001667, 3300-001668, 3300-001669, 3300-001670, 3300-001671, 3300-001672, 3300-001673, 3300-001674, 3300-001675, 3300-001676, 3300-001677, 3300-001678, 3300-001679, 3300-001680, 3300-001681, 3300-001682, 3300-001683, 3300-001684, 3300-001685, 3300-001686, 3300-001687, 3300-001688, 3300-001689, 3300-001690, 3300-001691, 3300-001692, 3300-001693, 3300-001694, 3300-001695, 3300-001696, 3300-001697, 3300-001698, 3300-001699, 3300-001700, 3300-001702, 3300-001703, 3300-001704, 3300-001705, 3300-001706, 3300-001707, 3300-001708, 3300-001709, 3300-001710, 3300-001711, 3300-001712, 3300-001713, 3300-001714, 3300-001715, 3300-001716, 3300-001717, 3300-001718, 3300-001719, 3300-001720, 3300-001721, 3300-001722, 3300-001723, 3300-001724, 3300-001725, 3300-001726, 3300-001727, 3300-001728, 3300-001729, 3300-001730, 3300-001731, 3300-001732, 3300-001733, 3300-001734, 3300-001735, 3300-001736, 3300-001737, 3300-001738, 3300-001739, 3300-001740, 3300-001741, 3300-001742, 3300-001743, 3300-001744, 3300-001745, 3300-001746, 3300-001747, 3300-001748, 3300-001749, 3300-001750, 3300-001751, 3300-001752, 3300-001753, 3300-001754, 3300-001755, 3300-001756, 3300-001757, 3300-001758, 3300-001759, 3300-001760, 3300-001761, 3300-001762, 3300-001763, 3300-001764, 3300-001765, 3300-001766, 3300-001767, 3300-001768, 3300-001769, 3300-001770, 3300-001771, 3300-001772, 3300-001773, 3300-001774, 3300-001775, 3300-001776, 3300-001777, 3300-001778, 3300-001779, 3300-001780, 3300-001781, 3300-001782, 3300-001783, 3300-001784, 3300-001785, 3300-001786, 3300-001787, 3300-001788, 3300-001789, 3300-001790, 3300-001791, 3300-001792, 3300-001793, 3300-001794, 3300-001795, 3300-001796, 3300-001797, 3300-001798, 3300-001799, 3300-001800, 3300-001801, 3300-001802, 3300-001803, 3300-001804, 3300-001805, 3300-001806, 3300-001807, 3300-001808, 3300-001809, 3300-001810, 3300-001811, 3300-001812, 3300-001813, 3300-001814, 3300-001815, 3300-001816, 3300-001817, 3300-001818, 3300-001819, 3300-001820, 3300-001821, 3300-001822, 3300-001823, 3300-001824, 3300-001825, 3300-001826, 3300-001827, 3300-001828, 3300-001829, 3300-001830, 3300-001831, 3300-001832, 3300-001833, 3300-001834, 3300-001835, 3300-001836, 3300-001837, 3300-001838, 3300-001839, 3300-001840, 3300-001841, 3300-001842, 3300-001843, 3300-001844, 3300-001845, 3300-001846, 3300-001847, 3300-001848, 3300-001849, 3300-001850, 3300-001851, 3300-001852, 3300-001853, 3300-001854, 3300-001855, 3300-001856, 3300-001857, 3300-001858, 3300-001859, 3300-001860, 3300-001861, 3300-001862, 3300-001863, 3300-001864, 3300-001865, 3300-001866, 3300-001867, 3300-001868, 3300-001869, 3300-001870, 3300-001871, 3300-001872, 3300-001873, 3300-001874, 3300-001875, 3300-001876, 3300-001877, 3300-001878, 3300-001879, 3300-001880, 3300-001881, 3300-001882, 3300-001883, 3300-001884, 3300-001885, 3300-001886, 3300-001887, 3300-001888, 3300-001889, 3300-001890, 3300-001891, 3300-001892, 3300-001893, 3300-001894, 3300-001895, 3300-001896, 3300-001897, 3300-001898, 3300-001899, 3300-001900, 3300-001901, 3300-001902, 3300-001903, 3300-001904, 3300-001905, 3300-001906, 3300-001907, 3300-001908, 3300-001909, 3300-001910, 3300-001911, 3300-001912, 3300-001913, 3300-001914, 3300-001915, 3300-001916, 3300-001917, 3300-001918, 3300-001919, 3300-001920, 3300-001921, 3300-001922, 3300-001923, 3300-001924, 3300-001925, 3300-001926, 3300-001927, 3300-001928, 3300-001929, 3300-001930, 3300-001931, 3300-001932, 3300-001933, 3300-001934, 3300-001935, 3300-001936, 3300-001937, 3300-001938, 3300-001939, 3300-001940, 3300-001941, 3300-001942, 3300-001943, 3300-001944, 3300-001945, 3300-001946, 3300-001947, 3300-001948, 3300-001949, 3300-001950, 3300-001951, 3300-001951, 3300-001952, 3300-001953, 3300-001954, 3300-001955, 3300-001956, 3300-001957, 3300-001958, 3300-001959, 3300-001960, 3300-001961, 3300-001962, 3300-001963, 3300-001964, 3300-001965, 3300-001966, 3300-001967, 3300-001968, 3300-001969, 3300-001970, 3300-001971, 3300-001972, 3300-001973, 3300-001974, 3300-001975, 3300-001976, 3300-001977, 3300-001978, 3300-001979, 3300-001980, 3300-001981, 3300-001982, 3300-001983, 3300-001984, 3300-001985, 3300-001986, 3300-001987, 3300-001988, 3300-001989, 3300-001990, 3300-001991, 3300-001992, 3300-001993, 3300-001994, 3300-001995, 3300-001996, 3300-001997, 3300-001998, 3300-001999, 3300-002000, 3300-002001, 3300-002002, 3300-002003, 3300-002004, 3300-002005, 3300-002006, 3300-002007, 3300-002008, 3300-002009, 3300-002010, 3300-002011, 3300-002012, 3300-002013, 3300-002014, 3300-002015, 3300-002016, 3300-002017, 3300-002018, 3300-002019, 3300-002020, 3300-002021, 3300-002022, 3300-002023, 3300-002024, 3300-002025, 3300-002026, 3300-002027, 3300-002028, 3300-002029, 3300-002030, 3300-002031, 3300-002032, 3300-002033, 3300-002034, 3300-002035, 3300-002036, 3300-002037, 3300-002038, 3300-002039, 3300-002040, 3300-002041, 3300-002042, 3300-002043, 3300-002044, 3300-002045, 3300-002046, 3300-002047, 3300-002048, 3300-002049, 3300-002050, 3300-002051, 3300-002052, 3300-002053, 3300-002054, 3300-002055, 3300-002056, 3300-002057, 3300-002058, 3300-002059, 3300-002060, 3300-002062, 3300-002063, 3300-002064, 3300-002065, 3300-002066, 3300-002067, 3300-002068, 3300-002069, 3300-002070, 3300-002071, 3300-002072, 3300-002076, 3300-002077, 3300-002078, 3300-002079, 3300-002080, 3300-002081, 3300-002082, 3300-002083, 3300-002084, 3300-002085, 3300-002086, 3300-002087, 3300-002088, 3300-002089, 3300-002090, 3300-002091, 3300-002092, 3300-002093, 3300-002094, 3300-002095, 3300-002096, 3300-002097, 3300-002098, 3300-002099, 3300-002100, 3300-002101, 3300-002102, 3300-002103, 3300-002104, 3300-002105, 3300-002106, 3300-002107, 3300-002108, 3300-002109, 3300-002110, 3300-002111, 3300-002112, 3300-002113, 3300-002114, 3300-002115, 3300-002116, 3300-002117, 3300-002118, 3300-002119, 3300-002120, 3300-002121, 3300-002122, 3300-002123, 3300-002124, 3300-002125, 3300-002126, 3300-002127, 3300-002128, 3300-002129, 3300-002130, 3300-002131, 3300-002132, 3300-002133, 3300-002134, 3300-002135, 3300-002136, 3300-002137, 3300-002138, 3300-002139, 3300-002140, 3300-002141, 3300-002142, 3300-002143, 3300-002144, 3300-002145, 3300-002146, 3300-002147, 3300-002148, 3300-002149, 3300-002150, 3300-002151, 3300-002152, 3300-002153, 3300-002154, 3300-002155, 3300-002156, 3300-002157, 3300-002158, 3300-002159, 3300-002160, 3300-002161, 3300-002162, 3300-002163, 3300-002164, 3300-002165, 3300-002166, 3300-002167, 3300-002168, 3300-002169, 3300-002170, 3300-002171, 3300-002172, 3300-002173, 3300-002174, 3300-002175, 3300-002176, 3300-002177, 3300-002178, 3300-002179, 3300-002180, 3300-002181, 3300-002182, 3300-002183, 3300-002184, 3300-002185, 3300-002186, 3300-002187, 3300-002188, 3300-002190, 3300-002191, 3300-002192, 3300-002193, 3300-002194, 3300-002195, 3300-002196, 3300-002197, 3300-002198, 3300-002199, 3300-002200, 3300-002201, 3300-002202, 3300-002203, 3300-002204, 3300-002205, 3300-002206, 3300-002207, 3300-002208, 3300-002209, 3300-002210, 3300-002211, 3300-002212, 3300-002213, 3300-002214, 3300-002215, 3300-002216, 3300-002217, 3300-002218, 3300-002219, 3300-002220, 3300-002221, 3300-002222, 3300-002223, 3300-002224, 3300-002225, 3300-002226, 3300-002227, 3300-002228, 3300-002229, 3300-002230, 3300-002231, 3300-002232, 3300-002233, 3300-002234, 3300-002235, 3300-002236, 3300-002237, 3300-002238, 3300-002239, 3300-002240, 3300-002241, 3300-002242, 3300-002243, 3300-002244, 3300-002245, 3300-002246, 3300-002247, 3300-002248, 3300-002249, 3300-002250, 3300-002251, 3300-002252, 3300-002253, 3300-002254, 3300-002256, 3300-002257, 3300-002258, 3300-002259, 3300-002260, 3300-002261, 3300-002262, 3300-002263, 3300-002264, 3300-002265, 3300-002266, 3300-002267, 3300-002268, 3300-002269, 3300-002270, 3300-002271, 3300-002280, 3300-002281, 3300-002282, 3300-002283, 3300-002290, 3300-002291, 3300-002292, 3300-002293, 3300-002294, 3300-002295, 3300-002296, 3300-002297, 3300-002298, 3300-002299, 3300-002300, 3300-002301, 3300-002302, 3300-002303, 3300-002304, 3300-002305, 3300-002306, 3300-002307, 3300-002308, 3300-002309, 3300-002310, 3300-002311, 3300-002319, 3300-002320, 3300-002335, 3300-002336, 3300-002341, 3300-002344, 3300-002345, 3300-002346, 3300-002347, 3300-002348, 3300-002349, 3300-002350, 3300-002352, 3300-002353, 3300-002354, 3300-002355, 3300-002356, 3300-002357, 3300-002358, 3300-002359, 3300-002367, 3300-002368, 3300-002369, 3300-002370, 3300-002371, 3300-002372, 3300-002385, 3300-002386, 3300-002387, 3300-002388, 3300-002389, 3300-001549, 3300-001550, 3300-100004, 3300-100008, 3300-100011, 3300-100012, 3300-100015, 3300-100017, 3300-100018, 3300-100019, 3300-100022, 3300-100023, 3300-100024, 3300-100025, 3300-100026, 3300-100027, 3300-100028, 3300-100029, 3300-100030, 3300-100034, 3300-100037, 3300-100039, 3300-100041, 3300-100042, 3300-100043, 3300-100045, 3300-100048, 3300-100051, 3300-100052, 3300-100055, 3300-100056, 3300-100057, 3300-100058, 3300-100059, 3300-100060, 3300-100061, 3300-100066, 3300-100069, 3300-100071, 3300-100072, 3300-100078, 3300-100079, 3300-100082, 3300-100083, 3300-100086, 3300-100087, 3300-100088, 3300-100089, 3300-100091, 3300-100093, 3300-100094, 3300-100095, 3300-100101, 3300-100102, 3300-100103, 3300-100106, 3300-100119, 3300-100126, 3300-100134, 3300-100135, 3300-100136, 3300-100137, 3300-100138, 3300-100139, 3300-100140, 3300-100141, 3300-100142, 3300-100143, 3300-100144, 3300-100145, 3300-100146, 3300-100147, 3300-100148, 3300-100149, 3300-100150, 3300-100151, 3300-100152, 3300-100153, 3300-100154, 3300-100155, 3300-100156, 3300-100157, 3300-100158, 3300-100159, 3300-100160, 3300-100161, 3300-100162, 3300-100163, 3300-100164, 3300-100165, 3300-100166, 3300-100167, 3300-100168, 3300-100169, 3300-100170, 3300-100171, 3300-100174, 3300-100175, 3300-100176, 3300-100178, 3300-100179, 3300-100180, 3300-100181, 3300-100182, 3300-100183, 3300-100184, 3300-100185, 3300-100186, 3300-100187, 3300-100188, 3300-100196, 3300-100197, 3300-100198, 3300-100199, 3300-100200, 3300-100201, 3300-100202, 3300-100203, 3300-100204, 3300-100205, 3300-100206, 3300-100207, 3300-100208, 3300-100209, 3300-100210, 3300-100211, 3300-100212, 3300-100213, 3300-100214, 3300-100215, 3300-100216, 3300-100217, 3300-100218, 3300-100219, 3300-100220, 3300-100221, 3300-100222, 3300-100223, 3300-100224, 3300-100225, 3300-100226, 3300-100227, 3300-100228, 3300-100229, 3300-100230, 3300-100231, 3300-100232, 3300-100233, 3300-100234, 3300-100235, 3300-100236, 3300-100237, 3300-100238, 3300-100239, 3300-100240, 3300-100241, 3300-100242, 3300-100243, 3300-100244, 3300-100245, 3300-100246, 3300-100247, 3300-100248, 3300-100249, 3300-100250, 3300-100251, 3300-100252, 3300-100253, 3300-100254, 3300-100255, 3300-100256, 3300-100260, 3300-100261, 3300-100262, 3300-100264, 3300-100266, 3300-100267, 3300-100268, 3300-100269, 3300-100270, 3300-100273, 3300-100274, 3300-100275, 3300-100276, 3300-100277, 3300-100278, 3300-100279, 3300-100280, 3300-100281, 3300-100282, 3300-100283, 3300-100284, 3300-100285, 3300-100286, 3300-100287, 3300-100288, 3300-100289, 3300-100290, 3300-100291, 3300-100292, 3300-100293, 3300-100294, 3300-100295, 3300-100296, 3300-100297, 3300-100298, 3300-100299, 3300-100300, 3300-100301, 3300-100303, 3300-100304, 3300-100305, 3300-100306, 3300-100307, 3300-100308, 3300-100309, 3300-100310, 3300-100311, 3300-100312, 3300-100313, 3300-100314, 3300-100315, 3300-100316, 3300-100317, 3300-100318, 3300-100319, 3300-100320, 3300-100321, 3300-100322, 3300-100323, 3300-100324, 3300-100325, 3300-100326, 3300-100327, 3300-100328, 3300-100330, 3300-100336, 3300-100337, 3300-100341, 3300-100363, 3300-100364, 3300-100365, 3300-100366, 3300-100368, 3300-100369, 93300-100006, 93300-100007, 93300-100009, 93300-100010, 93300-100013, 93300-100014, 93300-100016, 93300-100020, 93300-100036, 93300-100038, 93300-100040, 93300-100046, 93300-100047, 93300-100049, 93300-100050, 93300-100053, 93300-100054, 93300-100062, 93300-100063, 93300-100064, 93300-100067, 93300-100068, 93300-100070, 93300-100073, 93300-100074, 93300-100075, 93300-100076, 93300-100077, 93300-100080, 93300-100081, 93300-100084, 93300-100085, 93300-100090, 93300-100092, 93300-100096, 93300-100097, 93300-100100, 93300-100104, 93300-100105, 93300-100106, 93300-100107, 93300-100108, 93300-100109, 93300-100110, 93300-100111, 93300-100112, 93300-100113, 93300-100114, 93300-100115, 93300-100116, 93300-100117, 93300-100118, 93300-100120, 93300-100121, 93300-100122, 93300-100123, 93300-100124, 93300-100125, 93300-100126, 93300-100127, 93300-100128, 93300-100129, 93300-100130, 93300-100131, 93300-100132, 93300-100133, 93300-100258, 93300-100259, 93300-100263, AND 93300-100265. |
Recalling Firm/ Manufacturer |
Spacelabs Healthcare, Llc 5150 220th Ave Se Issaquah WA 98029-6834
|
For Additional Information Contact |
Al Van Houdt 425-657-7200 Ext. 5970
|
Manufacturer Reason for Recall |
The monitor has a 3 second delay in the waveforms and respiration alarm; the delay is not disclosed in the IFU.
|
FDA Determined Cause 2 |
Error in labeling |
Action |
Spacelabs Healthcare sent an "URGENT-MEDICAL DEVICE CORRECTION" letter dated September 24, 2010 to all affected customers. The letter describes the product, problem, and the action to be taken.
The letter includes the revised Operator's Manual as a replacement for customers current manual.
For Technical Support call Spacelabs Medical at 1-800-522-7025 and select 2. |
Quantity in Commerce |
2517 |
Distribution |
Worldwide Distribution-USA (Puerto Rico) and countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BOLIVIA, CHILE, CHINA, COLOMBIA, CYPRUS, CZECH REPUBLIC, DOMINICAN REPUBLIC, ECUADOR, EGYPT, EL SALVADOR, FRANCE, GERMANY, GREECE, GUATEMALA, HONDURAS, HONG KONG, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, KENYA, KOREA, REPUBLIC OF, KUWAIT, LEBANON, MACEDONIA, MALAYSIA, NEPAL, NETHERLANDS, NICARAGUA, OMAN, PALESTINE, PANAMA, PARAGUAY, PERU, PUERTO RICO, PHILIPPINES, QATAR, ROMANIA, SAUDI ARABIA, SINGAPORE, SOUTH AFRICA, SPAIN, SRI LANKA, SWITZERLAND, SYRIAN ARAB REPUBLIC, THAILAND, TRINIDAD AND TOBAGO, TUNISIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, VENEZUELA, and VIET NAM. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MHX and Original Applicant = SPACELABS MEDICAL INC.
|
|
|
|