| Class 2 Device Recall Toshiba AQUILION ONE | |
Date Initiated by Firm | September 07, 2010 |
Date Posted | February 15, 2011 |
Recall Status1 |
Terminated 3 on April 20, 2012 |
Recall Number | Z-1257-2011 |
Recall Event ID |
57499 |
510(K)Number | K083282 |
Product Classification |
X-Ray CT Scanner - Product Code JAK
|
Product | Toshiba TSX-301A AQUILION ONE (MPDCT0297EA) - Whole Body X-Ray CT Scanner |
Code Information |
Affected Serial No.: 2DA0872038, 2DA0962116, 2DA0862034, 2DA09Z2161, 2DA07X2008, 2DA1012165, 2DA08X2072, 2DA09Y2151, 2DA0842023, 2DA07X2005, 2DA1012168, 2DA08X2067, 2DA0882053, 2DA0862036, 2DA08Z2092, 2DA0882051, 2DA09Z2162, 2DA1072211, 2DA09X2143, 2DA08Y2077,2DA09X2139, 2DA08X2070, 2DA0982133 2133, 2DA0842026, 2DA0882052, 2DA0972123, 2DA08Y2080, 2DA0882048, 2DA08X2063, 2DA09Y2148, 2DA1012170, 2DA08X2071, 2DA1012169, 2DA0952115, 2DA0982131, 2DA07Y2010, 2DA08X2061, 2DA1012167, 2DA0882049, 1DA0962006, 1DA09Y2014, 1DA0992011, 1DA0972007, 1DA0972008, 2DA1012166, 2DA09Z2158, 2DA1072212, 2DA1012174, 2DA0792004, 2DA1062208, 2DA0792002, 1DA1062015, 1DB1072012 |
Recalling Firm/ Manufacturer |
Toshiba American Medical Systems Inc 2441 Michelle Dr P.O. Box 2068 Tustin CA 92781-2068
|
For Additional Information Contact | 714-730-5000 |
Manufacturer Reason for Recall | Incorrect Dose Calculation. The software that performs the pre-scan dose calculation uses an incorrect parameter for an exam Plan using a tube voltage of 80kV or 100kV. This results in a pre-scan expected dose that is approximately 10% lower than what the displayed value should be when the scanner executes the exam Plan. |
FDA Determined Cause 2 | Software design |
Action | Toshiba America Medical Systems (TAMS) issued notification to consignees 9/7/2010, via USPS return receipt mail. The letter contains a return reply form that is to be faxed to TAMS as proof of notification. |
Quantity in Commerce | 44 units distributed between 09/2007 - 07/2010 |
Distribution | Nationwide distribution. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JAK
|
|
|
|