Class 2 Device Recall Signa Excite |
 |
Date Initiated by Firm |
November 09, 2010 |
Date Posted |
February 24, 2011 |
Recall Status1 |
Terminated 3 on July 03, 2012 |
Recall Number |
Z-1429-2011 |
Recall Event ID |
57084 |
510(K)Number |
K041476
|
Product Classification |
System, nuclear magnetic resonance imaging - Product Code LNH
|
Product |
6) Signa Excite (K041476 Signa Excite 1.5T and 3.0T) Signa Excite, 12x (K041476 Signa Excite 1.5T and 3.0T)
The GE Signa¿ HDe MR system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. The Signa HDe MR system is indicated for use as a diagnostic imaging device to produce axial sagittal, coronal and oblique images, spectra, dynamic images of the internal structures and organs of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. The images produced by the Signa¿ HDe system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
|
Code Information |
SERIAL 00000277692MR0 00000276632MR7 00000280465MR6 00000284315MR9 00000276732MR5 00000280892MR1 00000280464MR9 00000281796MR3 00000278623MR4 00000278196MR1 00000275595MR7 00000281935MR7 00000278626MR7 00000282909MR1 00000280461MR5 00000283074MR3 00000274891MR1 00000279922MR9 00000282548MR7 00000281267MR5 00000273517MR3 00000281020MR8 00000280954MR9 00000282105MR6 00000283998MR3 00000285302MR6 00000281346MR7 00000279502MR9 00000285599MR7 00000273536MR3 00000281185MR9 00000281312MR9 00000281456MR4 00000283663MR3 00000273487MR9 00000280009MR2 00000282253MR4 00000280276MR7 00000279925MR2 00000279887MR4 00000278472MR6 00000281445MR7 00000276633MR5 00000283316MR8 00000276728MR3 00000282695MR6 2284300 00000273537MR1 00000283452MR1 00000283468MR7 00000273265MR9 00000283345MR7 00000284360MR5 00000287579MR7 00000278245MR6 00000279503MR7 00000279078MR0 00000280435MR9 00000282830MR9 00000283077MR6 00000283653MR4 00000283437MR2 00000281681MR7 00000278622MR6 00000280620MR6 00000274711MR1 00000281022MR4 00000274713MR7 00000281314MR5 00000281273MR3 00000280274MR2 00000283665MR8 00000279915MR3 00000283365MR5 00000277599MR7 00000281265MR9 00000027862MR0 00000283530MR4 00000273541MR3 00000284786MR1 00000275937MR1 00000274937MR2 00000287317MR2 00000283624MR5 00000282788MR9 00000273516MR5 00000277763MR9 00000278726MR5 00000281227MR9 00000277965MR0 00000280094MR4 00000281347MR5 00000276330MR8 00000283377MR0 00000279076MR4 00000274717MR8 00002181458MR0 00000276735MR8 00000274931MR5 00000274170MR0 00000282463MR9 00000286406MR4 00000282524MR8 00000277788MR6 00000281937MR3 00000027888MR3 00000287864MR3 00000277959MR3 00000283151MR9 00000280278MR3 00000279627MR4 00000283473MR7 00000274930MR7 00000278619MR2 00000278246MR4 00000277700MR1 00000274889MR5 00000274933MR1 00000282849MR9 00000283471MR1 00000273514MR0 00000276737MR4 00000278561MR6 00000274939MR8 00000287086MR3 00000282812MR7 00000282492MR8 00000281457MR2 00000284903MR2 00000281526MR4 00000285664MR9 00000281776MR5 00000281703MR9 00000281188MR3 00000280374MR0 00000286309MR0 00000280462MR3 00000281442MR4 00000275597MR3 00000281829MR2 00000274709MR5 00000275474MR5 00000287063MR2 00000282792MR1 00000287617MR5 00000280069MR6 00000286391MR8 00000282107MR2 00000282106MR4 00000281644MR5 00000285951MR0 00000275936MR3 00000281678MR3 00000275938MR9 00000289205MR7 00000282857MR2 00000283506MR4 00000280767MR5 00000286526MR9 00000280095MR1 00000282961MR2 00000281266MR7 00000286425MR4 00000273008MR3 00000281840MR9 00000283435MR6 00000280434MR2 00000279080MR6 00000280526MR5 00000282080MR1 00000286282MR9 00000283317MR6 00000279892MR4 00000286476MR7 00000280903MR6 00000283357MR2 00000283385MR3 00000285384MR4 00000287396MR6 00000274892MR9 00000283696MR3 00000285023MR8 00000286267MR0 00000283474MR5 00000282911MR7 00000284214MR4 00000275475MR2 00000286149MR0 00000275736MR7 00000283697MR1 00000277600MR3 00000289162MR0 00000286250MR6 00000280619MR8 00000281056MR2 00000280713MR9 00000280759MR2 00000281313MR7 00000275473MR7 00000281865MR6 00000281857MR3 00000278307MR4 00000282079MR3 00000283150MR1 00000285301MR8 00000283755MR7 00000282464MR7 00000282958MR8 00000282083MR5 00000273009MR1 00000282029MR8 00000274710MR3 00000274150MR2 00000275935MR5 00000276332MR4 00000282084MR3 00000283300MR2 00000285914MR8 00000285300MR0 00000286060MR9 00000286058MR3 00000286061MR7 00000278463MR5 00000277602MR9 00000284236MR7 00000285020MR4 00000276276MR3 00000276273MR0 00000288212MR4 00000273518MR1 00000286146MR6 00000279077MR2 00000274708MR7 00000278305MR8 00000273267MR5 00000272999MR4 00000274169MR2 00000275735MR9 00000275599MR9 00000275737MR5 00000274173MR4 00000279404MR8 00000273489MR5 00000277961MR9 00000284464MR5 00000287399MR0 00000278469MR2 00000279407MR1 00000283358MR0 00000283576MR7 00000276631MR9 00000284296MR1 00000283792MR0 00000293589MR8 00000282730MR1 00000275738MR3 00000284337MR3 00000281858MR1 00000278613MR5 00000283039MR6 00000277863MR7 00000281866MR4 00000278558MR2 00000280273MR4 00000290775MR6 00000284297MR9 00000286145MR8 00000280004MR3 00000286281MR1 00000284279MR7 00000278621MR8 00000278748MR9 00000284642MR6 00000274936MR4 00000278468MR4 00000282682MR4 00000280769MR1 00000281344MR2 00000278304MR1 00000284355MR5 00000278195MR3 00000274171MR8 00000280432MR6 00000279630MR8 00000293588MR0 00000274890MR3 00000287850MR2 00000273006MR7 00000273540MR5 00000282959MR6 00000282686MR5 00000281704MR7 00000276731MR7 00000280768MR3 00000283470MR3 00000286392MR6 00000287810MR6 00000274166MR8 00000289004MR4 00000287879MR1 00000288901MR2 00000283386MR1 00000282501MR6 00000288667MR9 00000293587MR2 00000285172MR3 00000282791MR3 00000283526MR2 00000282790MR5 00000288244MR7 00000286559MR0 00000285019MR6 00000278749MR7 00000277601MR1 00000280898MR8 00000280951MR5 00000285380MR2 00000283761MR5 00000286800MR8 00000289798MR1 00000288811MR3 00000282088MR4 00000284787MR9 00000285222MR6 00000276635MR0 00000280005MR0 00000283378MR8 00000286538MR4 00000280621MR4 00000285303MR4 00000282831MR7 00000283174MR1 00000286897MR4 00000284644MR2 00000275598MR1 00000280319MR5 00000288589MR5 00000281936MR5 00000284457MR9 00000282858MR0 00000285219MR2 00000283662MR5 00000281317MR8 00000273535MR5 00000287590MR4 00000278566MR5 00000280002MR7 00000283076MR8 00000279423MR8 00000281603MR1 00000278243MR1 00000274168MR4 00000275596MR5 00000273488MR7 00000273533MR0 00000274172MR6 00000277864MR5 00000274938MR0 00000282525MR5 00000284717MR6 00000284596MR4 00000285026MR1 00000279924MR5 00000279890MR8 00000281257MR6 00000279869MR2 00000288196MR9 00000284643MR4 00000286778MR6 00000287619MR1 00000288204MR1 00000288774MR3 00000283893MR6 00000280373MR2 00000282922MR4 00000288254MR6 00000266145MR2 00000274693MR1 00000277191MR3 00000276153MR4 00000280627MR1 00000269437MR0 00000265177MR6 00000265870MR6 00000280678MR4 00000264407MR8 00000274691MR5 00000265173MR5 00000272348MR4 00000273196MR6 00000270975MR6 00000273228MR7 00000272622MR2 00000270569MR7 00000262068MR0 00000265184MR2 00000269044MR4 00000275961MR1 00000267324MR2 00000274998MR4 00000280674MR3 00000268471MR0 00000275015MR6 00000265183MR4 00000262258MR7 00000272339MR3 00000270244MR7 00000279483MR2 00000270383MR3 00000270250MR4 00000262459MR1 00000269326MR5 00000261062MR4 00000265718MR7 00000276417MR3 00000262669MR5 00000267653MR4 00000271250MR3 00000275355MR6 00000276411MR6 00000274997MR6 00000266463MR9 00000263834MR4 00000266344MR1 00000264402MR9 00000272629MR7 00000264744MR4 00002700253MR8 00000269038MR6 00000274072MR8 00000269438MR8 00000270459MR1 00000280056MR3 00000268288MR8 00000267111MR3 00000269343MR0 00000238864MR3 00000267657MR5 00000267113MR9 00000263977MR1 00000280085MR2 00000274835MR8 00000278722MR4 00000261879MR1 00000267468MR7 00000264093MR6 00000269377MR8 00000260691MR1 00000273459MR8 00000272621MR4 00000263880MR7 00000264406MR0 00000280735MR2 00000260481MR7 00000277185MR5 00000266753MR3 00000272430MR0 00000277489MR1 00000267804MR3 00000266461MR3 00000278772MR9 00000269384MR4 00000280385MR6 00000266071MR0 00000268286MR2 00000267978MR5 00000267112MR1 00000266068MR6 00000269324MR0 00000261677MR9 00000280089MR4 00000260451MR0 00000260698MR6 00000269040MR2 00000272890MR5 00000261867MR6 00000260650MR7 00000256434MR2 00000262464MR1 00000262075MR5 00000271663MR7 00000265061MR2 00000271083MR8 00000266754MR1 00000265592MR6 00000267805MR0 00000264403MR7 00000263266MR9 00000273672MR6 00000271234MR7 00000269456MR0 00000275959MR5 00000265914MR2 00000273365MR7 00000272341MR9 00000272343MR5 00000275846MR4 00000278669MR7 00000270388MR2 00002686687MR1 00000262299MR1 00000263879MR9 00000260699MR4 00000274078MR5 00000267208MR7 00000266677MR4 00000280057MR1 00000266828MR3 00000275089MR1 00000276249MR0 00000266469MR6 00000273200MR6 00000260649MR9 00000280698MR2 00000267974MR4 00000272529MR9 00000270627MR3 00000267040MR4 00000273621MR3 00000267211MR1 00000264408MR6 00000264090MR2 00000272283MR3 00000269342MR2 00000273341MR8 00000269382MR8 00000264091MR0 00000272105MR8 00000273368MR1 00000272962MR2 00000266823MR4 00000278768MR7 00000278076MR5 00000276486MR8 00000266822MR6 00000279798MR3 00000278875MR0 00000279755MR3 00000267180MR8 00000269460MR2 00000270386MR6 00000269340MR6 00000276414MR0 00000273230MR3 00000262458MR3 00000262800MR6 00000269380MR2 00000269574MR0 00000269457MR8 00000261068MR1 00000267811MR8 00000262665MR3 00000264617MR2 00000266668MR3 00000236924MR7 00000228285MR3 00000270625MR7 00000268478MR5 00000269042MR8 00000277575MR7 00000278688MR7 00000265182MR6 00000277494MR1 00000272344MR3 00000273367MR3 00000262668MR7 00000264401MR1 00000266747MR5 00000271246MR1 00000267464MR6 00000261877MR5 00000267181MR6 00000268290MR4 00000270971MR5 00000267469MR5 00000277862MR9 00000270243MR9 00000269378MR6 00000272627MR1 00000275583MR3 00000260626MR7 00000270968MR1 00000270461MR7 00000270462MR5 00000280702MR2 347400 00000270011MR0 00000262196MR9 00000270629MR9 00000267210MR3 00000273221MR2 00000276591MR9 00000276594MR9 891011 00000265181MR8 00000263971MR4 00000268688MR9 00000026572MR1 00000265596MR7 00000269577MR3 00000266827MR5 00000271118MR2 00000276592MR3 00000261205MR9 00000262666MR1 00000266067MR8 00000273680MR9 00000268367MR0 00002675655MR9 00000262673MR7 00000269458MR6 00000268365MR4 00000271251MR1 00000260480MR9 00000272531MR5 00000225375MR5 00000270392MR4 00000265917MR5 00000261201MR8 00000277496MR6 00000263974MR8 00000270247MR0 00000273528MR0 00000268363MR9 00000263968MR0 00000272342MR7 00000279480MR8 00000277192MR1 00000272103MR3 00000262067MR2 00000277923MR9 00000269323MR2 00000270008MR6 00000275670MR8 00000262667MR9 00000266673MR3 00000263875MR7 00000275669MR0 00000273458MR0 00000272892MR1 00000267178MR2 00000270460MR9 00000260009MR6 00000279147MR3 00000267321MR8 00000261066MR5 00000266672MR5 00000271124MR0 00000262066MR4 00000263969MR8 00000277955MR1 00000272286MR6 2333 00000272072MR0 00000273460MR6 00000272428MR4 00000269047MR7 00000256367MR4 00000279607MR6 00000272891MR3 00000261069MR9 00000264087MR6 00000277188MR9 00000267322MR6 00000263874MR0 00000260697MR8 00000274839MR0 00000267467MR9 00000272620MR6 00000265007MR5 00000268691MR3 00000272895MR4 00000272101MR7 00000277193MR9 00000267472MR9 00000266755MR8 00000267470MR3 00000278871MR9 00000280088MR6 00000262930MR1 00000264619MR8 00000266449MR8 00000265919MR1 00000266756MR6 00000267114MR7 00000266674MR1 00000267213MR7 00000268419MR9 00000268282MR1 00000269381MR0 00000269462MR8 00000269385MR1 00000269579MR9 00000274838MR2 00000273462MR2 00000276589MR9 00000273675MR9 00000276415MR7 00000262804MR8 00000269043MR6 00000272963MR0 00000266450MR6 00000262297MR5 00000263871MR6 00000263972MR2 00000267648MR4 00000279450MR1 00000280301MR3 00000275086MR7 00000272628MR9 00000275344MR0 00000261870MR0 00000272347MR6 00000261197MR8 00000264404MR5 00000269045MR1 00000272104MR1 00000274079MR3 00000264095MR1 00000265588MR4 00000279753MR8 00000271121MR6 00000267651MR8 00000268119MR5 00000273679MR1 00000270007MR8 00000262457MR5 00000279799MR1 00000278131MR8 00000261676MR1 00000266466MR2 00000272624MR8 00000271249MR5 00000272970MR5 00000268364MR7 00000267654MR2 00000264405MR2 00000271660MR3 00000270390MR8 00000259956MR1 UNK51078490003 00000265724MR5 00000267328MR3 00000262300MR7 00000275844MR9 00000278667MR1 00000266757MR4 00000279083MR0 00000280676MR8 00000276248MR2 00000271120MR8 00000274830MR9 00000262456MR7 00000266746MR7 00000273525MR6 00000275093MR3 00000273681MR7 00000261868MR4 00000277195MR4 00000275580MR9 00000265058MR8 00000261199MR4 00000275014MR9 00000266669MR1 00000266826MR7 00000269581MR5 00000268685MR5 00000264620MR6 00000274697MR2 00000275699MR7 00000271239MR6 00000266821MR8 00000265719MR5 UNK51078490008 00000267810MR0 00000272099MR3 00000268476MR9 00000279800MR7 00000280700MR6 00000280734MR5 00000267036MR2 00000278210MR0 00000275965MR2 00000268418MR1 00000272436MR7 00000275642MR7 00000260026MR0 00000277921MR3 00000268291MR2 00000277582MR3 00000262792MR5 00000270626MR5 00000268474MR4 00000272427MR6 00000269046MR9 00000266670MR9 00000271658MR7 00000273227MR9 00000274071MR0 00000276494MR2 00000273457MR2 00000275017MR2 00000268284MR7 00000275845MR6 00000276590MR7 00000260692MR9 00000261202MR6 00000263341MR0 00000267806MR8 00000270464MR1 00000270391MR6 00000273362MR4 00000273522MR3 00000275847MR2 00000234805MR0 00000276591MR5 00000273366MR5 00000279460MR0 00000272533MR1 00000268009MR8 00000262671MR1 00000273344MR2 00000278687MR9 00000277574MR0 00000270468MR2 1112 00000269344MR8 00000262257MR9 00000273465MR5 00000270466MR6 00000282503MR2 00000262798MR2 00000266464MR7 00000272528MR1 00000269435MR4 00000270387MR4 00000268518MR8 00000268477MR7 00000268515MR4 00000269575MR7 00000272968MR9 00000270026MR8 00000272969MR7 00000272070MR4 00000268692MR1 00000271662MR9 00000268417MR3 00000270467MR4 00000272536MR4 00000263338MR6 00000263878MR1 00000263873MR2 00000261880MR9 00000265918MR3 00000262077MR1 00000265921MR7 00000266202MR1 00000266676MR6 00000268366MR2 00000263265MR1 00000268010MR6 00000267812MR6 00000268289MR6 00000268693MR9 00000268008MR0 00000266824MR2 00000267471MR1 00000268472MR8 00000268420MR7 00000272071MR2 00000269461MR0 00000271661MR1 00000271664MR5 00000272625MR5 00000269341MR4 00000268686MR3 00000266462MR1 00000272073MR8 00000272074MR6 00000267465MR3 00000272100MR9 00000272889MR7 00000272898MR8 00000270565MR5 00000273360MR8 00000271665MR2 00000275582MR5 00000273225MR3 00000272965MR5 00000275087MR5 00000275964MR5 00000276492MR6 00000266146MR0 00000266148MR6 00000276272MR2 00000266745MR9 00000269463MR6 00000279459MR2 00000270022MR7 00000272337MR7 00000272289MR0 00000273464MR8 00000266064MR5 00000275358MR0 00000270624MR0 00000279148MR1 00000278955MR0 00000277924MR7 00000279451MR9 00000260693MR7 00000262078MR9 00000262197MR7 00000261060MR8 00000262461MR7 00000263819MR5 00000266675MR8 00000271659MR5 00000268117MR9 00000272075MR3 00000266204MR7 00000272287MR4 00000268510MR5 00000273622MR1 00000276138MR5 00000264096MR9 00000266743MR4 00000268115MR3 00000263872MR4 00000273226MR1 00000264094MR4 00000265065MR3 00000272079MR5 00000279452MR7 00000260452MR8 00000266070MR2 00000267320MR0 00000273461MR4 00000277952MR8 00000268114MR6 00000279454MR3 00000270628MR1 00000275668MR2 00000273530MR6 00000261204MR2 00000275667MR4 00000273677MR5 00000272433MR4 00000272435MR9 00000265178MR4 00000274832MR5 00000270025MR0 00000264745MR1 00000265059MR6 00000276585MR7 00000270970MR7 00000273343MR4 00000260627MR5 00000262460MR9 00000264616MR4 00000270245MR4 00000262198MR5 00000262932MR7 00000268763MR0 00000277579MR9 00000266465MR4 00000270633MR1 00000266448MR0 00000262805MR5 00000274690MR7 00000267466MR1 00000265180MR0 00000265922MR5 00000274837MR4 00000265006MR7 00000263340MR2 00000261063MR2 00000262799MR0 00000268473MR6 00000269039MR4 00000265920MR9 00000265868MR0 00000262065MR6 00000263976MR3 00000260696MR0 00000267327MR5 00000272340MR1 00000263339MR4 00000267206MR1 00000275645MR0 00000275088MR3 00000268509MR7 00000263876MR5 00000265060MR4 00000264618MR0 00000260695MR2 00000275584MR1 00000266144MR5 00000259842MR3 00000265869MR8 00000266467MR0 00000266201MR3 00000265172MR7 00000267182MR4 00000276250MR8 00000272078MR7 00000261675MR3 00000272535MR6 00000275095MR8 00000269339MR8 00000272346MR8 00000263263MR6 00000263264MR4 00000267326MR7 00000271122MR4 00000276418MR1 00000270384MR1 00000272336MR9 00000278842MR0 00000261200MR0 00000261065MR7 00000262931MR9 00000262801MR4 00000264741MR0 00000265179MR2 00000267973MR6 00000267977MR7 00000271126MR5 00000266147MR8 00000267037MR0 00000279159MR8 00000280302MR1 00000262803MR0 00000277194MR7 00000279751MR2 00000270974MR9 00000272893MR9 00000262664MR6 00000273229MR5 00000274996MR8 00000260625MR9 00000265590MR0 00000273676MR7 00000277190MR5 00000272894MR7 00000277583MR1 00000279802MR3 00000264089MR4 00000274688MR1 00000277487MR5 00000265593MR4 00000273223MR8 00000275091MR7 00000277189MR7 00000279752MR0 00000279458MR4 00000280299MR9 00000267809MR2 00000261198MR6 00000267650MR0 00000261206MR7 00000272432MR6 00000266072MR8 00000279482MR4 00000274074MR4 00000270563MR0 00000238361MR3 00000270572MR1 00000270631MR5 00000264747MR7 00000266345MR8 00000272102MR5 00000269572MR4 00000264742MR8 00000261679MR5 00000264088MR6 00000265866MR4 00000269379MR4 00000272335MR1 00000276493MR4 00000260628MR3 00000264740MR2 00000267323MR4 00000265594MR2 00000268421MR5 00000275094MR1 00000267807MR6 00000272532MR3 00000029079MR2 00000266713MR7 00000290984MR4 00000029464MR7 00000280313MR8 00000265966MR4 00000269583MR1 00000280724MR6 00000271052MR3 00000263314MR7 00000265801MR1 00000266043MR9 00000268371MR2 00000275097MR4 000A0000012MR2 00000290961MR2 00000275469MR5 00000241749MR1 00000266129MR6 00000280723MR8 00000280046MR4 00000264693MR3 00000269393MR5 00000291209MR5 00000270266MR0 00000269792MR8 00000295454MR3 00000291666MR6 00000289771MR8 00000286629MR1 00000286299MR3 00000280098MR5 00000270607MR5 00000281225MR3 00000263935MR9 1 00000227408MRI 00000264599MR2 00000227571MR7 00000279874MR2 00000270550MR7 00000270321MR3 00000271141MR4 00000266011MR6 00000267359MR8 00000228289MR3 00000296210MR8 00000266838MR2 00000265687MR4 0 00000290528MR9 00000263660MR3 00000292604MR6 00000266009MR0 00000266161MR9 0000266161MR9A 00000026584MR6 00000265968MR8 00000266214MR6 00000281077MR8 00000271753MR6 00000295451MR9 151 00000275023MR0 00000293024MR6 00000269359MR6 00000266740MR0 00000271656MR1 00000269061MR8 00000265939MR9 00000268672MR3 00000029027MR9 00000029187MR8 00000280946MR5 00000263603MR3 00000237968MR3 00000289882MR3 00000296420MR3 00000275920MR7 00000283249MR1 104 00000271269MR3 00000283321MR8 00000271776MR7 00000266605MR5 00000271268MR5 00000266837MR4 00000275062MR8 00000266457MR1 00000000BYV7UN 00000268745MR7 00000268049MR4 00000296638MR0 00000283464MR6 00000291910MR8 00000271055MR6 00000290712MR9 00000266764MR0 00000267646MR8 00000269483MR4 00000270381MR7 00000294646MR5 00000293706MR8 00000291852MR2 00000269392MR7 00000296098MR7 00000267728MR4 00000266162MR7 00000269584MR9 00000278364MR5 00000271049MR9 00000276436MR3 00000271780MR9 00000026397MR9 00000268501MR4 00000271139MR8 00000268359MR7 00000295558MR1 00000266442MR3 00000027974MR5 00000268306MR8 00000272973MR9 00000276370MR4 00000266653MR5 00000265579MR3 00000291254MR1 00000295221MR6 00000268453MR8 00000265960MR5 00000251871MR0 00000274180MR9 00000285956MR9 |
Recalling Firm/ Manufacturer |
GE Healthcare, LLC 3000 N Grandview Blvd Waukesha WI 53188-1615
|
For Additional Information Contact |
262-513-4122
|
Manufacturer Reason for Recall |
GE Healthcare has become aware of a potential issue associated with the gradient cables being inadvertently swapped while servicing the gradients on some GE Healthcare MR Systems that may impact patient safety.
If the gradient cables are inadvertently swapped while servicing the gradients, and the Service Engineer
does not perform the geometry check as required in the service procedure, images m
|
FDA Determined Cause 2 |
Nonconforming Material/Component |
Action |
On November 9, 2010, the firm, GE Healthcare, sent an "Urgent Medical Device Correction" letter dated November 5, 2010 to all Consignees/Customers. The letter described the product, problem and action to be taken by the firm. GE instructed the customers to contact their local GEHC support Engineer, if they have any questions about the image orientation, or the procedures for geometry checks on thier system. Additionally, a GE Healthcare representative will install a device, which will reduce the potiential for the gradient cables to be swapped.
Contact your local GEHC Support Engineer if you have any questions concerning this notification or phone: 262-548-2731. |
Quantity in Commerce |
1219 |
Distribution |
Worldwide distribution: USA including DC and PR; and countries including: VIET NAM, VENEZUELA, URUGUAY, UNITED KINGDOM, UNITED ARAB EMIRATES, UKRAINE, TURKEY, TUNISIA, FORMER YUGOSLAV REPUBLIC OF MACEDONIA, THAILAND, TAIWAN, SYRIAN ARAB REPUBLIC, SWITZERLAND, SWEDEN, SPAIN, SOUTH AFRICA, SLOVENIA, SLOVAKIA, SINGAPORE, SAUDI ARABIA, RUSSIAN FEDERATION, ROMANIA, REUNION, REPUBLIC OF KOREA, PORTUGAL POLAND, PHILIPPINES, PERU,PARAGUAY, PANAMA, PAKISTAN, NORWAY, NEW ZEALAND, NETHERLANDS, MOROCCO, MEXICO, MAURITIUS, MARTINIQUE,MALTA, MALAYSIA, LUXEMBOURG,LITHUANIA, LIBYAN ARAB JAMAHIRIYA, LEBANON, LATVIA, KUWAIT, KENYA, KAZAKHSTAN, JORDAN, JAPAN, JAMAICA, ITALY, ISRAEL, ISLAMIC REPUBLIC OF IRAN, IRELAND, INDONESIA, INDIA, ICELAND, HUNGARY, HONG KONG, HONDURAS, GUYANA, GUATEMALA, GUADELOUPE, GREECE, GERMANY, GEORGIA, FRENCH POLYNESIA, FRANCE, FINLAND, EL SALVADOR, EGYPT, ECUADOR, DOMINICAN REPUBLIC, DENMARK, CECH REPUBLIC, CYPRUS, CROATIA, COLOMBIA, CHINA, CHILE, CANDA, BULGARIA, BRAZIL, BOSNIA AND HERZEGOVINA, BOLIVIA, BELGIUM, BELARUS, AZERBAIJAN, AUSTRIA, AUSTRALIA, ARGENTINA, ALGERIA. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = LNH and Original Applicant = GE MEDICAL SYSTEMS, LLC
|
|
|
|