Class 2 Device Recall Fresenius 2008T Hemodialysis Machine

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 


New Search Back to Search Results
  Class 2 Device Recall Fresenius 2008T Hemodialysis Machine see related information
Date Initiated by Firm January 19, 2011
Date Posted March 04, 2011
Recall Status1 Terminated 3 on July 08, 2013
Recall Number Z-1526-2011
Recall Event ID 57851
510(K)Number K0933902  
Product Fresenius 2008T Hemodialysis Machine
Part Number 190573, 2008T Hemodialysis System, OLCIDIASAFE PLUS
Part Number 190713, 2008T Hemodialysis System with CDX
Part Number 190766, 2008T Hemodialysis System with biBag

The Fresenius 2008T is indicated for acute and chronic dialysis therapy.
Code Information Units distributed between Sept 1, 2010 and January 5, 2011. Serial Numbers: 0T0S100031 0T0S100035 0T0S100036 0T0S100037 0T0S100039 0T0S100040 0T0S100041 0T0S100042 0T0S100046 0T0S100048 0T0S100050 0T0S100052 0T0S100058 0T0S100059 0T0S100060 0T0S100061 0T0S100062 0T0S100063 0T0S100064 0T0S100065 0T0S100067 0T0S100068 0T0S100070 0T0S100071 0T0S100072 0T0S100080 0T0S100084 0T0S100085 0T0S100086 0T0S100087 0T0S100088 0T0S100089 0T0S100090 0T0S100093 0T0S100094 0T0S100095 0T0S100096 0T0S100097 0T0S100098 0T0S100099 0T0S100100 0T0S100101 0T0S100102 0T0S100103 0T0S100105 0T0S100107 0T0S100109 0T0S100110 0T0S100111 0T0S100113 0T0S100114 0T0S100115 0T0S100116 0T0S100117 0T0S100118 0T0S100119 0T0S100121 0T0S100122 0T0S100123 0T0S100124 0T0S100125 0T0S100126 0T0S100127 0T0S100129 0T0S100130 0T0S100131 0T0S100132 0T0S100133 0T0S100135 0T0S100136 0T0S100137 0T0S100138 0T0S100139 0T0S100140 0T0S100141 0T0S100142 0T0S100143 0T0S100144 0T0S100145 0T0S100146 0T0S100147 0T0S100148 0T0S100149 0T0S100150 0T0S100151 0T0S100152 0T0S100153 0T0S100154 0T0S100155 0T0S100156 0T0S100157 0T0S100158 0T0S100159 0T0S100160 0T0S100161 0T0S100162 0T0S100163 0T0S100164 0T0S100165 0T0S100166 0T0S100167 0T0S100168 0T0S100170 0T0S100171 0T0S100172 0T0S100173 0T0S100174 0T0S100175 0T0S100176 0T0S100177 0T0S100179 0T0S100180 0T0S100181 0T0S100182 0T0S100183 0T0S100184 0T0S100185 0T0S100186 0T0S100187 0T0S100188 0T0S100189 0T0S100190 0T0S100191 0T0S100192 0T0S100193 0T0S100194 0T0S100195 0T0S100196 0T0S100197 0T0S100198 0T0S100199 0T0S100200 0T0S100202 0T0S100203 0T0S100204 0T0S100205 0T0S100206 0T0S100207 0T0S100208 0T0S100210 0T0S100211 0T0S100212 0T0S100213 0T0S100214 0T0S100215 0T0S100216 0T0S100217 0T0S100218 0T0S100220 0T0S100221 0T0S100222 0T0S100223 0T0S100224 0T0S100225 0T0S100226 0T0S100227 0T0S100228 0T0S100229 0T0S100230 0T0S100231 0T0S100232 0T0S100233 0T0S100234 0T0S100235 0T0S100236 0T0S100237 0T0S100238 0T0S100239 0T0S100240 0T0S100241 0T0S100242 0T0S100243 0T0S100244 0T0S100245 0T0S100246 0T0S100247 0T0S100248 0T0S100249 0T0S100250 0T0S100251 0T0S100252 0T0S100253 0T0S100254 0T0S100255 0T0S100256 0T0S100257 0T0S100258 0T0S100259 0T0S100260 0T0S100261 0T0S100262 0T0S100263 0T0S100264 0T0S100265 0T0S100266 0T0S100267 0T0S100268 0T0S100270 0T0S100271 0T0S100272 0T0S100274 0T0S100275 0T0S100276 0T0S100277 0T0S100280 0T0S100281 0T0S100282 0T0S100284 0T0S100285 0T0S100286 0T0S100288 0T0S100289 0T0S100291 0T0S100293 0T0S100294 0T0S100295 0T0S100296 0T0S100297 0T0S100298 0T0S100299 0T0S100300 0T0S100301 0T0S100302 0T0S100303 0T0S100304 0T0S100305 0T0S100306 0T0S100307 0T0S100308 0T0S100309 0T0S100310 0T0S100311 0T0S100312 0T0S100313 0T0S100314 0T0S100315 0T0S100316 0T0S100317 0T0S100318 0T0S100319 0T0S100320 0T0S100321 0T0S100322 0T0S100323 0T0S100324 0T0S100325 0T0S100326 0T0S100327 0T0S100328 0T0S100329 0T0S100330 0T0S100331 0T0S100332 0T0S100333 0T0S100334 0T0S100335 0T0S100336 0T0S100337 0T0S100338 0T0S100339 0T0S100340 0T0S100341 0T0S100342 0T0S100343 0T0S100344 0T0S100345 0T0S100346 0T0S100347 0T0S100348 0T0S100349 0T0S100350 0T0S100351 0T0S100352 0T0S100353 0T0S100354 0T0S100355 0T0S100356 0T0S100357 0T0S100358 0T0S100359 0T0S100360 0T0S100361 0T0S100362 0T0S100363 0T0S100364 0T0S100365 0T0S100366 0T0S100367 0T0S100368 0T0S100369 0T0S100370 0T0S100371 0T0S100372 0T0S100373 0T0S100374 0T0S100375 0T0S100376 0T0S100377 0T0S100378 0T0S100379 0T0S100380 0T0S100381 0T0S100382 0T0S100383 0T0S100384 0T0S100385 0T0S100386 0T0S100387 0T0S100388 0T0S100389 0T0S100390 0T0S100391 0T0S100392 0T0S100393 0T0S100394 0T0S100395 0T0S100396 0T0S100397 0T0S100398 0T0S100399 0T0S100400 0T0S100401 0T0S100402 0T0S100403 0T0S100404 0T0S100405 0T0S100406 0T0S100407 0T0S100408 0T0S100409 0T0S100410 0T0S100411 0T0S100412 0T0S100413 0T0S100414 0T0S100415 0T0S100416 0T0S100417 0T0S100418 0T0S100419 0T0S100420 0T0S100421 0T0S100422 0T0S100423 0T0S100424 0T0S100425 0T0S100426 0T0S100427 0T0S100428 0T0S100429 0T0S100430 0T0S100431 0T0S100432 0T0S100433 0T0S100434 0T0S100435 0T0S100436 0T0S100437 0T0S100438 0T0S100439 0T0S100440 0T0S100441 0T0S100442 0T0S100443 0T0S100444 0T0S100445 0T0S100446 0T0S100447 0T0S100448 0T0S100449 0T0S100450 0T0S100451 0T0S100452 0T0S100453 0T0S100454 0T0S100455 0T0S100456 0T0S100457 0T0S100458 0T0S100459 0T0S100460 0T0S100461 0T0S100462 0T0S100463 0T0S100464 0T0S100465 0T0S100466 0T0S100467 0T0S100468 0T0S100469 0T0S100470 0T0S100471 0T0S100472 0T0S100473 0T0S100474 0T0S100475 0T0S100477 0T0S100478 0T0S100479 0T0S100480 0T0S100481 0T0S100483 0T0S100484 0T0S100485 0T0S100486 0T0S100487 0T0S100488 0T0S100489 0T0S100490 0T0S100491 0T0S100492 0T0S100493 0T0S100494 0T0S100495 0T0S100496 0T0S100497 0T0S100498 0T0S100499 0T0S100500 0T0S100501 0T0S100502 0T0S100503 0T0S100504 0T0S100505 0T0S100506 0T0S100507 0T0S100508 0T0S100509 0T0S100510 0T0S100511 0T0S100512 0T0S100513 0T0S100514 0T0S100515 0T0S100516 0T0S100517 0T0S100518 0T0S100519 0T0S100520 0T0S100521 0T0S100522 0T0S100523 0T0S100524 0T0S100525 0T0S100526 0T0S100527 0T0S100528 0T0S100529 0T0S100530 0T0S100531 0T0S100532 0T0S100533 0T0S100534 0T0S100535 0T0S100536 0T0S100537 0T0S100538 0T0S100539 0T0S100540 0T0S100541 0T0S100542 0T0S100543 0T0S100544 0T0S100545 0T0S100546 0T0S100547 0T0S100548 0T0S100549 0T0S100550 0T0S100551 0T0S100552 0T0S100553 0T0S100554 0T0S100555 0T0S100556 0T0S100557 0T0S100558 0T0S100559 0T0S100560 0T0S100561 0T0S100562 0T0S100563 0T0S100564 0T0S100565 0T0S100566 0T0S100567 0T0S100568 0T0S100569 0T0S100570 0T0S100571 0T0S100572 0T0S100573 0T0S100574 0T0S100575 0T0S100576 0T0S100577 0T0S100578 0T0S100579 0T0S100580 0T0S100581 0T0S100582 0T0S100583 0T0S100584 0T0S100585 0T0S100586 0T0S100587 0T0S100588 0T0S100589 0T0S100590 0T0S100591 0T0S100592 0T0S100593 0T0S100594 0T0S100595 0T0S100596 0T0S100597 0T0S100598 0T0S100599 0T0S100600 0T0S100601 0T0S100602 0T0S100603 0T0S100604 0T0S100605 0T0S100606 0T0S100607 0T0S100608 0T0S100609 0T0S100610 0T0S100611 0T0S100612 0T0S100613 0T0S100614 0T0S100615 0T0S100616 0T0S100617 0T0S100618 0T0S100619 0T0S100620 0T0S100621 0T0S100622 0T0S100623 0T0S100624 0T0S100625 0T0S100626 0T0S100627 0T0S100628 0T0S100629 0T0S100630 0T0S100631 0T0S100632 0T0S100633 0T0S100634 0T0S100635 0T0S100636 0T0S100637 0T0S100638 0T0S100639 0T0S100640 0T0S100641 0T0S100642 0T0S100643 0T0S100644 0T0S100645 0T0S100646 0T0S100647 0T0S100648 0T0S100649 0T0S100650 0T0S100651 0T0S100652 0T0S100653 0T0S100654 0T0S100655 0T0S100656 0T0S100657 0T0S100658 0T0S100659 0T0S100660 0T0S100661 0T0S100662 0T0S100663 0T0S100664 0T0S100665 0T0S100666 0T0S100667 0T0S100668 0T0S100669 0T0S100670 0T0S100671 0T0S100672 0T0S100673 0T0S100674 0T0S100675 0T0S100676 0T0S100677 0T0S100678 0T0S100679 0T0S100680 0T0S100681 0T0S100682 0T0S100683 0T0S100684 0T0S100685 0T0S100686 0T0S100687 0T0S100688 0T0S100689 0T0S100690 0T0S100691 0T0S100692 0T0S100693 0T0S100694 0T0S100695 0T0S100696 0T0S100697 0T0S100698 0T0S100699 0T0S100700 0T0S100701 0T0S100702 0T0S100703 0T0S100704 0T0S100705 0T0S100706 0T0S100707 0T0S100708 0T0S100709 0T0S100710 0T0S100711 0T0S100712 0T0S100713 0T0S100714 0T0S100715 0T0S100716 0T0S100717 0T0S100718 0T0S100719 0T0S100720 0T0S100721 0T0S100722 0T0S100723 0T0S100724 0T0S100725 0T0S100726 0T0S100727 0T0S100728 0T0S100729 0T0S100730 0T0S100731 0T0S100732 0T0S100733 0T0S100734 0T0S100735 0T0S100736 0T0S100737 0T0S100738 0T0S100739 0T0S100740 0T0S100741 0T0S100742 0T0S100743 0T0S100744 0T0S100745 0T0S100746 0T0S100747 0T0S100748 0T0S100749 0T0S100750 0T0S100751 0T0S100752 0T0S100753 0T0S100754 0T0S100755 0T0S100756 0T0S100757 0T0S100758 0T0S100759 0T0S100760 0T0S100761 0T0S100762 0T0S100763 0T0S100764 0T0S100765 0T0S100766 0T0S100767 0T0S100768 0T0S100770 0T0S100771 0T0S100772 0T0S100773 0T0S100774 0T0S100775 0T0S100776 0T0S100777 0T0S100778 0T0S100779 0T0S100780 0T0S100781 0T0S100782 0T0S100783 0T0S100784 0T0S100785 0T0S100786 0T0S100787 0T0S100788 0T0S100789 0T0S100790 0T0S100791 0T0S100792 0T0S100793 0T0S100794 0T0S100795 0T0S100796 0T0S100797 0T0S100798 0T0S100799 0T0S100800 0T0S100801 0T0S100802 0T0S100803 0T0S100804 0T0S100805 0T0S100806 0T0S100807 0T0S100808 0T0S100810 0T0S100811 0T0S100812 0T0S100813 0T0S100814 0T0S100815 0T0S100816 0T0S100817 0T0S100818 0T0S100819 0T0S100820 0T0S100821 0T0S100822 0T0S100823 0T0S100824 0T0S100825 0T0S100826 0T0S100827 0T0S100828 0T0S100829 0T0S100830 0T0S100831 0T0S100832 0T0S100833 0T0S100834 0T0S100835 0T0S100836 0T0S100837 0T0S100838 0T0S100839 0T0S100840 0T0S100841 0T0S100842 0T0S100843 0T0S100844 0T0S100845 0T0S100846 0T0S100847 0T0S100848 0T0S100849 0T0S100851 0T0S100852 0T0S100853 0T0S100854 0T0S100855 0T0S100856 0T0S100857 0T0S100858 0T0S100859 0T0S100860 0T0S100861 0T0S100862 0T0S100863 0T0S100864 0T0S100865 0T0S100866 0T0S100867 0T0S100868 0T0S100869 0T0S100870 0T0S100871 0T0S100872 0T0S100873 0T0S100874 0T0S100875 0T0S100876 0T0S100877 0T0S100878 0T0S100879 0T0S100880 0T0S100881 0T0S100882 0T0S100883 0T0S100884 0T0S100885 0T0S100886 0T0S100887 0T0S100888 0T0S100889 0T0S100890 0T0S100891 0T0S100892 0T0S100893 0T0S100894 0T0S100895 0T0S100896 0T0S100897 0T0S100898 0T0S100899 0T0S100900 0T0S100901 0T0S100902 0T0S100903 0T0S100904 0T0S100905 0T0S100906 0T0S100907 0T0S100908 0T0S100909 0T0S100910 0T0S100911 0T0S100912 0T0S100913 0T0S100914 0T0S100915 0T0S100916 0T0S100917 0T0S100918 0T0S100919 0T0S100920 0T0S100921 0T0S100922 0T0S100923 0T0S100924 0T0S100925 0T0S100926 0T0S100927 0T0S100928 0T0S100929 0T0S100930 0T0S100931 0T0S100932 0T0S100933 0T0S100934 0T0S100935 0T0S100936 0T0S100937 0T0S100938 0T0S100939 0T0S100940 0T0S100941 0T0S100942 0T0S100943 0T0S100944 0T0S100945 0T0S100946 0T0S100947 0T0S100948 0T0S100949 0T0S100950 0T0S100951 0T0S100952 0T0S100953 0T0S100954 0T0S100955 0T0S100956 0T0S100957 0T0S100958 0T0S100959 0T0S100960 0T0S100961 0T0S100962 0T0S100963 0T0S100964 0T0S100965 0T0S100966 0T0S100967 0T0S100968 0T0S100969 0T0S100970 0T0S100971 0T0S100972 0T0S100973 0T0S100974 0T0S100975 0T0S100976 0T0S100977 0T0S100978 0T0S100979 0T0S100980 0T0S100981 0T0S100982 0T0S100983 0T0S100984 0T0S100985 0T0S100986 0T0S100987 0T0S100988 0T0S100989 0T0S100990 0T0S100991 0T0S100992 0T0S100993 0T0S100994 0T0S100995 0T0S100996 0T0S100997 0T0S100998 0T0S100999 0T0S101000 0T0S101001 0T0S101002 0T0S101003 0T0S101004 0T0S101005 0T0S101006 0T0S101007 0T0S101008 0T0S101009 0T0S101010 0T0S101011 0T0S101012 0T0S101013 0T0S101014 0T0S101015 0T0S101016 0T0S101017 0T0S101018 0T0S101019 0T0S101020 0T0S101021 0T0S101022 0T0S101023 0T0S101024 0T0S101025 0T0S101026 0T0S101027 0T0S101028 0T0S101029 0T0S101030 0T0S101031 0T0S101032 0T0S101033 0T0S101034 0T0S101035 0T0S101036 0T0S101037 0T0S101038 0T0S101039 0T0S101040 0T0S101041 0T0S101042 0T0S101043 0T0S101044 0T0S101045 0T0S101046 0T0S101047 0T0S101048 0T0S101049 0T0S101050 0T0S101051 0T0S101052 0T0S101053 0T0S101054 0T0S101055 0T0S101056 0T0S101057 0T0S101058 0T0S101059 0T0S101060 0T0S101061 0T0S101062 0T0S101063 0T0S101064 0T0S101065 0T0S101066 0T0S101067 0T0S101068 0T0S101069 0T0S101070 0T0S101071 0T0S101072 0T0S101073 0T0S101074 0T0S101075 0T0S101076 0T0S101077 0T0S101078 0T0S101079 0T0S101080 0T0S101081 0T0S101082 0T0S101083 0T0S101084 0T0S101085 0T0S101086 0T0S101087 0T0S101088 0T0S101089 0T0S101090 0T0S101091 0T0S101092 0T0S101093 0T0S101094 0T0S101095 0T0S101096 0T0S101097 0T0S101098 0T0S101099 0T0S101100 0T0S101101 0T0S101102 0T0S101103 0T0S101104 0T0S101105 0T0S101106 0T0S101107 0T0S101108 0T0S101109 0T0S101110 0T0S101111 0T0S101112 0T0S101113 0T0S101114 0T0S101115 0T0S101116 0T0S101117 0T0S101118 0T0S101119 0T0S101120 0T0S101121 0T0S101122 0T0S101123 0T0S101124 0T0S101125 0T0S101126 0T0S101127 0T0S101128 0T0S101129 0T0S101130 0T0S101131 0T0S101132 0T0S101133 0T0S101134 0T0S101135 0T0S101136 0T0S101137 0T0S101138 0T0S101139 0T0S101140 0T0S101141 0T0S101142 0T0S101143 0T0S101144 0T0S101145 0T0S101146 0T0S101147 0T0S101148 0T0S101149 0T0S101150 0T0S101151 0T0S101152 0T0S101153 0T0S101154 0T0S101155 0T0S101156 0T0S101157 0T0S101158 0T0S101159 0T0S101160 0T0S101161 0T0S101162 0T0S101163 0T0S101164 0T0S101165 0T0S101166 0T0S101167 0T0S101168 0T0S101169 0T0S101170 0T0S101171 0T0S101172 0T0S101173 0T0S101174 0T0S101175 0T0S101176 0T0S101177 0T0S101178 0T0S101179 0T0S101180 0T0S101181 0T0S101182 0T0S101183 0T0S101184 0T0S101185 0T0S101186 0T0S101187 0T0S101188 0T0S101189 0T0S101190 0T0S101191 0T0S101192 0T0S101193 0T0S101194 0T0S101195 0T0S101196 0T0S101197 0T0S101198 0T0S101199 0T0S101200 0T0S101201 0T0S101202 0T0S101203 0T0S101204 0T0S101205 0T0S101206 0T0S101207 0T0S101208 0T0S101209 0T0S101210 0T0S101211 0T0S101212 0T0S101213 0T0S101214 0T0S101215 0T0S101217 0T0S101218 0T0S101219 0T0S101220 0T0S101221 0T0S101222 0T0S101223 0T0S101224 0T0S101225 0T0S101226 0T0S101227 0T0S101228 0T0S101229 0T0S101230 0T0S101231 0T0S101232 0T0S101233 0T0S101234 0T0S101235 0T0S101236 0T0S101237 0T0S101238 0T0S101239 0T0S101240 0T0S101241 0T0S101242 0T0S101243 0T0S101244 0T0S101245 0T0S101246 0T0S101247 0T0S101248 0T0S101249 0T0S101250 0T0S101251 0T0S101252 0T0S101253 0T0S101254 0T0S101255 0T0S101256 0T0S101257 0T0S101258 0T0S101259 0T0S101260 0T0S101261 0T0S101262 0T0S101263 0T0S101264 0T0S101265 0T0S101266 0T0S101267 0T0S101268 0T0S101269 0T0S101270 0T0S101271 0T0S101272 0T0S101273 0T0S101274 0T0S101275 0T0S101276 0T0S101277 0T0S101278 0T0S101279 0T0S101280 0T0S101281 0T0S101282 0T0S101283 0T0S101284 0T0S101286 0T0S101287 0T0S101288 0T0S101289 0T0S101290 0T0S101292 0T0S101293 0T0S101294 0T0S101295 0T0S101296 0T0S101297 0T0S101298 0T0S101299 0T0S101300 0T0S101301 0T0S101302 0T0S101303 0T0S101304 0T0S101305 0T0S101306 0T0S101307 0T0S101308 0T0S101309 0T0S101310 0T0S101311 0T0S101312 0T0S101313 0T0S101314 0T0S101315 0T0S101316 0T0S101317 0T0S101318 0T0S101319 0T0S101320 0T0S101321 0T0S101322 0T0S101323 0T0S101324 0T0S101325 0T0S101326 0T0S101327 0T0S101328 0T0S101329 0T0S101330 0T0S101331 0T0S101332 0T0S101333 0T0S101334 0T0S101335 0T0S101336 0T0S101337 0T0S101338 0T0S101339 0T0S101340 0T0S101341 0T0S101342 0T0S101343 0T0S101344 0T0S101345 0T0S101347 0T0S101348 0T0S101349 0T0S101350 0T0S101351 0T0S101352 0T0S101353 0T0S101354 9T0S100012   
Recalling Firm/
Manufacturer
Fresenius Medical Care Holdings, Inc.
920 Winter St
Waltham MA 02451-1521
For Additional Information Contact SAME
800-662-1237
Manufacturer Reason
for Recall
Cursor on the dialysis screen may disappear and not respond to user input
FDA Determined
Cause 2
Device Design
Action Fresenius Medical Care contacted customers by mailing a Customer Notification letter "Urgent correction FMNCA Model 2008T Hemodialysis Machine" via Certified-Mail, Return Receipt requested on January 19, 2011. The Technical Services department began contacting consignees by telephone on January 13, 2011, to schedule an appointment for installing the electronic board on the 2008T machines at the customers' facilities. For questions call Technical services 1-800-227-2572.
Quantity in Commerce 1268 units
Distribution Worldwide Distribution - USA and Canada