| Class 2 Device Recall UNICEL DXC SYNCHRON SYSTEMS | |
Date Initiated by Firm | November 16, 2010 |
Date Posted | April 13, 2011 |
Recall Status1 |
Terminated 3 on June 28, 2012 |
Recall Number | Z-1972-2011 |
Recall Event ID |
58022 |
510(K)Number | K042291 K060256 |
Product Classification |
UNICEL DXC SYNCHRON SYSTEMS - Product Code JJE
|
Product | UniCel DxC Clinical Systems with Serial Number less than 3412 Including: UniCel DxC 600 Part #: A10405,
UniCel DxC 600 PRO Part #: A11810,
UniCel DxC 600i Part #: A27318,
UniCel DxC 660i Part #: A64871,
UniCel DxC 680i Part #: A64903,
UniCel DxC 800 Part #: A11816,
UniCel DxC 800 PRO Part #: A11812,
UniCel DxC 860i Part #: A64935, and
UniCel DxC 880i Part #: A59102. |
Code Information |
UniCel DxC Clinical Systems with Serial Number less than 3412 (with serial numbers less than 3412, labeled behind the lower right side instrument door) UniCel DxC 600 Part #: A10405, UniCel DxC 600 PRO Part #: A11810, UniCel DxC 600i Part #: A27318, UniCel DxC 660i Part #: A64871, UniCel DxC 680i Part #: A64903, UniCel DxC 800 Part #:A11816, UniCel DxC 800 PRO Part #: A11812, UniCel DxC 860i Part #: A64935, and UniCel DxC 880i Part #: A59102 Subsequent Product Codes: CEK, CEO, CFJ, CGA, CGX, CJW, DCK, DLZ, JFL, JFP, JGJ, JGS, JIY, JJE, JXM, LFP, CFR, CDQ, CEM, CGN, CGZ, CHL, DCK, DFT, JFP, JGS, JHI, JIY, JJE, JLW, JMG, JXM, LCD & LCP. |
Recalling Firm/ Manufacturer |
Beckman Coulter Inc. 200 S Kraemer Blvd Brea CA 92821-6208
|
For Additional Information Contact | 714-993-5321 |
Manufacturer Reason for Recall | The hose clamps in the DxC (with serial numbers less than 3412, labeled behind the lower right side instrument door) hydropneumatic system are sharp and pose a possible risk of injury. An operator could be cut by the hose clamps, which could in turn result in increased exposure to any waste products (biohazards) that might be present on the assembly. |
FDA Determined Cause 2 | Pending |
Action | The recall communication was initiated with Beckman Coulter forwarding a Product Corrective Action Letter (PCA) dated November 16, 2010 to the affected customers. The recall notice explained the reason for recall and asked that customers take the following actions:
(1) Customers with a service contract will have hose clamp covers installed by their Field Service Engineer during the next scheduled Preventive Maintenance (PM). A kit can be ordered through their Customer Service Representative if consignees prefer not to wait for the next Preventive Maintenance visit and
(2) Customers without a service contract can obtain a kit by contacting their local Customer Service Representative.
Customers were also informed that Kit (PN A68986) containing hose clamp covers and installation instructions is available at no charge from their Customer Service Representatives. One kit is required for each DxC system.
The following telephone numbers were given for customers to obtain the kits: (800) 526-3821 (United States) and (800) 463-7828 (Canada).
Consignees were asked to share this information with their laboratory staff and retain this notification as part of their laboratory Quality System documentation. If any of the affected product was forwarded on to another laboratory they were asked to provide a copy of this letter to them.
Questions are to be directed to Beckman Coulter's Customer Support Center (Call Center) on the Beckman Coulter website at http://www.beckmancoulter.com/customersupport/support or call (800) 854-3633 in the United States and Canada. |
Quantity in Commerce | 1,535 units |
Distribution | Worldwide Distribution--United States and countries of Angola, Australia, Belgium, Bulgaria, China, Columbia, Czech Republic, Denmark, Eritrea, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Latvia, Lebanon, Lithuania, Mexico, Netherlands, Reunion, Serbia, South Korea, Spain, Slovakia, Swaziland, Sweden, Switzerland, Tunisia, Turkey, Taiwan, United Arab Emirates, and the United Kingdom. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = JJE
|
|
|
|