Date Initiated by Firm | March 10, 2011 |
Date Posted | June 24, 2011 |
Recall Status1 |
Terminated 3 on September 16, 2011 |
Recall Number | Z-2662-2011 |
Recall Event ID |
58186 |
510(K)Number | K981313 |
Product Classification |
Accelerator, linear, medical - Product Code IYE
|
Product | Elekta Impac Software Sequencer Verify & Record System, for MOSAIQ
The intended use of Sequencer (generically referred to as a verification system) is to assist in the process of process of patient treatment on a radiotherapy treatment machine. It gives the capability to notify users of actions that need to take place prior to treatment, to display reference images for set up purposes and to auto setup the machine to predefined settings. |
Code Information |
MOSAIQ 1.30, 1.60, 2.00, 2.10, and 2.20; including all service packs. |
Recalling Firm/ Manufacturer |
Impac Medical Systems Inc 100 Mathilda Pl Fl 5th Sunnyvale CA 94086-6076
|
For Additional Information Contact | Christopher Ivicevich 408-830-8023 |
Manufacturer Reason for Recall | There's a problem in the database conversion program that is used when upgrading from the Multi-ACCS to MOSAIQ or from one version of MOSAIQ to another. If the gantry start angle is incorrect prior to pressing beam-on, the patient will be mistreated. |
FDA Determined Cause 2 | Software design (manufacturing process) |
Action | The firm, Elekta, sent a "USER NOTICE" to its customers with an attached confirmation of receipt form. The notice describes the product, problem and actions to be taken. The customers were instructed to follow the Interim mitigation instructions noted in the notice -verify the Arc start angle of the Simple ARC with MLC treatment field definition prior to delivery and if the customer identifes a field with an errant gantry start angle, to cease using that plan immediately and consult with the original plan for that field and make the correction to the gantry start angle. In addition, the customers were instructed to complete and return the Confirmation of Receipt via fax to: North America and Asia at 702-992-5002 or Europe and Africa +44 (0)1293 654401, Attn: Elekta OIS Su[pport and/or scan and email to: support@impact.com, europe.support@impact.com or by mail; and distribute this notice to any all users of IMPAC software at their organization who are potentially affected by this issue.
For questions or requests for upgrade, email support@impac.com (North America/Asia) or europe.support@impac.com (Europe) or call USA 1-800-488-4672. |
Quantity in Commerce | 840 |
Distribution | Worldwide distribution: USA (nationwide) and country of Canada. |
Total Product Life Cycle | TPLC Device Report |
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls. 2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall. 3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database | 510(K)s with Product Code = IYE
|