|
Class 2 Device Recall PercuNav Field Generator Stand |
|
Date Initiated by Firm |
April 01, 2011 |
Date Posted |
May 19, 2011 |
Recall Status1 |
Terminated 3 on March 15, 2012 |
Recall Number |
Z-2219-2011 |
Recall Event ID |
58429 |
510(K)Number |
K053610
|
Product Classification |
Computed Tomography X-Ray System - Product Code JAK
|
Product |
PercuNav Field Generator Stand is a Computed tomography, X-Ray System. Part Number 453561431911.
Designed to be used only with the PercuNav" image guided intervention system. The Field Generator Stand (with articulating Mounting Arm) is used to position the Field Generator near the area of interest. |
Code Information |
System Code Serial Numbers: 03B8Y3, 20071000005, 20071000006, 20071000007, 20071000008, 20071000009, 20071000010, 20081000011, 20081000013, 20091000019, 20091000020, 20091000021, 20091000022, 20091000023, 02RCHF, 02X3NK, 034K1R, 039VDK, 039W1H, 03B4F3, 03B59R, 03B8MW, 03B8N0, 03B8QR, 03B8XX, 03B8XY, 03B8XZ, 03B8Y0, 03B8Y1, 03B8Y4, 03B8Y5, B0091T, B009BJ, B009HV, B009VJ, B00DDJ, B00DQ2, B00DQX, B00DQZ, B00DRF, B00DRP, B00DT7, B00DXM, B00DY0, B00F1Q, B00F3H, B00F3M, B00HKW, B00HKZ, B00HV4, B00HV7, B00HW3, B00HZR, B00J05, B00P3X, B00P3Y, B00P59, B00P5P, B00PB4, B00PJJ, B00PPP, B00PQ7, and P002A20090500005. |
Recalling Firm/ Manufacturer |
Philips Medical Systems North America Co. Phillips 22100 Bothell Everett Hwy Bothell WA 98021-8431
|
For Additional Information Contact |
Philips Customer Service 425-487-7602
|
Manufacturer Reason for Recall |
The PercuNav Field Generator Stand could tip over during set up, during use, or during transport.
|
FDA Determined Cause 2 |
Device Design |
Action |
On 4/4/11, Philips Healthcare began sending out the URGENT-Medical Device Correction notification letter to their consignees.
Philips informed their customers of a potential tipping hazard of the PercuNav Field Generator Stand.
The notification letter also provides instructions to customers to ensure the articulating Mounting Arm is positioned so the Field Generator is not further than the specified distance from the back of the clamp on the stand's central pole during set up or during use.
Customers are instructed to apply a provided label on stand to warn of a potential for tipping due to excessive force. For transport, customers are instructed to remove the Field Generator from the Mounting Arm and place the Field Generator in the storage compartment on the system cart. The Mounting Arm should then be collapsed and folded into a 90 degree angle.
Customers are informed that a Philips field service representative will contact them to schedule installation of a service kit on the stand, free of charge, once service kits are available.
Customers can call Philips Customer Service at 1-866-767-7822 and reference to FCO 79500197 for any questions about this recall. |
Quantity in Commerce |
65 units total (33 units distributed to US and 32 units to foreign consignees) |
Distribution |
Worldwide Distribution -- USA, Australia, Austria, Belgium, China, France, Germany, Hong Kong, India, Korea, Netherlands, Singapore, Sweden, Switzerland, Taiwan, Thailand, and United Kingdom. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = JAK and Original Applicant = TRAXTAL TECHNOLOGIES INC.
|
|
|
|