|
Class 2 Device Recall Eclipse Treatment Planning System |
|
Date Initiated by Firm |
April 04, 2011 |
Date Posted |
May 19, 2011 |
Recall Status1 |
Terminated 3 on June 27, 2012 |
Recall Number |
Z-2220-2011 |
Recall Event ID |
58581 |
510(K)Number |
K073020
|
Product Classification |
System,planning,radiation therapy treatment - Product Code MUJ
|
Product |
Eclipse Treatment Planning System, Client software versions 8.0 through 8.6; Model Number: H48; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA
The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
|
Code Information |
LIST OF SERIAL NUMBERS: H480005 ,H480079 ,H480156 ,H480251 ,H480365 ,H480436 ,H480573 ,H480647 ,H480750 ,H480883 ,H480996 ,H480006 ,H480080 ,H480157 ,H480252 ,H480369 ,H480437 ,H480574 ,H480648 ,H480751 ,H480884 ,H480997 ,H480007 ,H480081 ,H480158 ,H480253 ,H480370 ,H480438 ,H480575 ,H480651 ,H480755 ,H480886 ,H480998 ,H480008 ,H480082 ,H480169 ,H480254 ,H480371 ,H480439 ,H480576 ,H480653 ,H480757 ,H480887 ,H480999 ,H480009 ,H480083 ,H480170 ,H480255 ,H480372 ,H480440 ,H480583 ,H480654 ,H480758 ,H480890 ,H481000 ,H480010 ,H480084 ,H480172 ,H480256 ,H480373 ,H480441 ,H480585 ,H480655 ,H480759 ,H480892 ,H481002 ,H480011 ,H480085 ,H480173 ,H480257 ,H480374 ,H480461 ,H480586 ,H480656 ,H480772 ,H480898 ,H481005 ,H480012 ,H480086 ,H480178 ,H480265 ,H480375 ,H480462 ,H480587 ,H480657 ,H480774 ,H480899 ,H481006 ,H480013 ,H480087 ,H480180 ,H480277 ,H480376 ,H480463 ,H480592 ,H480658 ,H480775 ,H480901 ,H481007 ,H480014 ,H480088 ,H480181 ,H480281 ,H480377 ,H480467 ,H480593 ,H480660 ,H480779 ,H480902 ,H481008 ,H480023 ,H480089 ,H480187 ,H480282 ,H480382 ,H480468 ,H480596 ,H480661 ,H480780 ,H480903 ,H481009 ,H480024 ,H480090 ,H480188 ,H480287 ,H480383 ,H480469 ,H480599 ,H480663 ,H480781 ,H480913 ,H481010 ,H480025 ,H480091 ,H480189 ,H480294 ,H480384 ,H480471 ,H480600 ,H480665 ,H480788 ,H480915 ,H481022 ,H480026 ,H480095 ,H480190 ,H480297 ,H480388 ,H480473 ,H480601 ,H480667 ,H480789 ,H480927 ,H481023 ,H480027 ,H480096 ,H480191 ,H480298 ,H480389 ,H480474 ,H480602 ,H480672 ,H480790 ,H480936 ,H481024 ,H480028 ,H480105 ,H480192 ,H480300 ,H480390 ,H480475 ,H480603 ,H480673 ,H480791 ,H480939 ,H481025 ,H480030 ,H480106 ,H480193 ,H480303 ,H480391 ,H480476 ,H480604 ,H480674 ,H480792 ,H480940 ,H481026 ,H480031 ,H480112 ,H480194 ,H480305 ,H480392 ,H480481 ,H480605 ,H480675 ,H480796 ,H480941 ,H481027 ,H480032 ,H480113 ,H480195 ,H480306 ,H480393 ,H480482 ,H480610 ,H480676 ,H480797 ,H480942 ,H481028 ,H480033 ,H480115 ,H480196 ,H480307 ,H480394 ,H480486 ,H480611 ,H480677 ,H480798 ,H480943 ,H481029 ,H480034 ,H480118 ,H480197 ,H480308 ,H480395 ,H480487 ,H480612 ,H480678 ,H480799 ,H480944 ,H481030 ,H480035 ,H480119 ,H480198 ,H480311 ,H480396 ,H480488 ,H480613 ,H480679 ,H480802 ,H480947 ,H481031 ,H480036 ,H480121 ,H480199 ,H480312 ,H480397 ,H480489 ,H480614 ,H480680 ,H480808 ,H480948 ,H481032 ,H480037 ,H480122 ,H480201 ,H480313 ,H480398 ,H480490 ,H480616 ,H480682 ,H480809 ,H480949 ,H481037 ,H480038 ,H480123 ,H480202 ,H480314 ,H480399 ,H480491 ,H480617 ,H480683 ,H480810 ,H480950 ,H481038 ,H480039 ,H480124 ,H480203 ,H480315 ,H480403 ,H480492 ,H480618 ,H480684 ,H480811 ,H480952 ,H481054 ,H480040 ,H480125 ,H480204 ,H480318 ,H480404 ,H480494 ,H480619 ,H480699 ,H480812 ,H480953 ,H481055 ,H480042 ,H480126 ,H480205 ,H480319 ,H480405 ,H480495 ,H480620 ,H480700 ,H480813 ,H480954 ,H481056 ,H480046 ,H480127 ,H480206 ,H480323 ,H480406 ,H480496 ,H480621 ,H480701 ,H480815 ,H480955 ,H481062 ,H480047 ,H480129 ,H480211 ,H480324 ,H480408 ,H480497 ,H480626 ,H480705 ,H480827 ,H480956 ,H481063 ,H480048 ,H480131 ,H480212 ,H480325 ,H480409 ,H480498 ,H480627 ,H480707 ,H480828 ,H480957 ,H481071 ,H480051 ,H480132 ,H480213 ,H480326 ,H480410 ,H480502 ,H480628 ,H480708 ,H480829 ,H480958 ,H481077 ,H480052 ,H480137 ,H480214 ,H480327 ,H480411 ,H480503 ,H480629 ,H480709 ,H480830 ,H480959 ,H481078 ,H480053 ,H480138 ,H480215 ,H480328 ,H480412 ,H480504 ,H480630 ,H480710 ,H480831 ,H480960 ,H481092 ,H480054 ,H480139 ,H480216 ,H480329 ,H480413 ,H480505 ,H480631 ,H480711 ,H480832 ,H480962 ,H481093 ,H480055 ,H480141 ,H480219 ,H480330 ,H480414 ,H480513 ,H480632 ,H480718 ,H480855 ,H480965 ,H481094 ,H480056 ,H480143 ,H480221 ,H480331 ,H480415 ,H480514 ,H480633 ,H480729 ,H480856 ,H480966 ,H481095 ,H480057 ,H480144 ,H480222 ,H480332 ,H480416 ,H480518 ,H480634 ,H480732 ,H480857 ,H480980 ,H481096 ,H480058 ,H480145 ,H480223 ,H480333 ,H480417 ,H480519 ,H480635 ,H480733 ,H480858 ,H480982 ,H481097 ,H480059 ,H480146 ,H480224 ,H480334 ,H480418 ,H480529 ,H480636 ,H480734 ,H480859 ,H480983 ,H481099 ,H480060 ,H480147 ,H480236 ,H480335 ,H480419 ,H480539 ,H480637 ,H480735 ,H480860 ,H480984 ,H481100 ,H480061 ,H480148 ,H480237 ,H480339 ,H480420 ,H480541 ,H480638 ,H480739 ,H480861 ,H480985 ,H481101 ,H480062 ,H480149 ,H480245 ,H480346 ,H480421 ,H480547 ,H480640 ,H480740 ,H480862 ,H480986 ,H481102 ,H480063 ,H480150 ,H480246 ,H480347 ,H480423 ,H480548 ,H480641 ,H480745 ,H480863 ,H480987 ,H481103 ,H480064 ,H480152 ,H480247 ,H480355 ,H480424 ,H480549 ,H480642 ,H480746 ,H480869 ,H480988 ,H481108 ,H480065 ,H480153 ,H480248 ,H480356 ,H480433 ,H480570 ,H480643 ,H480747 ,H480870 ,H480989 ,H481114 ,H480066 ,H480154 ,H480249 ,H480357 ,H480434 ,H480571 ,H480644 ,H480748 ,H480876 ,H480992 ,H481115 ,H480078 ,H480155 ,H480250 ,H480361 ,H480435 ,H480572 ,H480646 ,H480749 ,H480881 ,H480994 ,H481116 ,H481117 ,H481267 ,H481348 ,H481502 ,H481640 ,H481739 ,H481870 ,H481968 ,H482019 ,H482070 ,H482143 ,H481118 ,H481268 ,H481349 ,H481503 ,H481644 ,H481740 ,H481871 ,H481969 ,H482020 ,H482071 ,H482144 ,H481122 ,H481269 ,H481358 ,H481504 ,H481645 ,H481741 ,H481872 ,H481970 ,H482021 ,H482072 ,H482145 ,H481123 ,H481270 ,H481359 ,H481505 ,H481646 ,H481742 ,H481873 ,H481971 ,H482022 ,H482073 ,H482146 ,H481124 ,H481271 ,H481360 ,H481506 ,H481647 ,H481743 ,H481874 ,H481972 ,H482023 ,H482074 ,H482147 ,H481125 ,H481272 ,H481361 ,H481508 ,H481648 ,H481750 ,H481875 ,H481973 ,H482024 ,H482075 ,H482156 ,H481126 ,H481273 ,H481362 ,H481509 ,H481649 ,H481757 ,H481876 ,H481974 ,H482025 ,H482076 ,H482157 ,H481127 ,H481274 ,H481363 ,H481510 ,H481650 ,H481758 ,H481879 ,H481975 ,H482026 ,H482077 ,H482160 ,H481128 ,H481275 ,H481366 ,H481513 ,H481651 ,H481759 ,H481884 ,H481976 ,H482027 ,H482078 ,H482161 ,H481129 ,H481276 ,H481367 ,H481514 ,H481652 ,H481760 ,H481885 ,H481977 ,H482028 ,H482079 ,H482164 ,H481133 ,H481277 ,H481373 ,H481515 ,H481653 ,H481761 ,H481886 ,H481978 ,H482029 ,H482080 ,H482165 ,H481134 ,H481278 ,H481381 ,H481517 ,H481654 ,H481762 ,H481889 ,H481979 ,H482030 ,H482081 ,H482172 ,H481136 ,H481279 ,H481382 ,H481518 ,H481655 ,H481774 ,H481890 ,H481980 ,H482031 ,H482082 ,H482178 ,H481143 ,H481280 ,H481383 ,H481519 ,H481656 ,H481777 ,H481891 ,H481981 ,H482032 ,H482083 ,H482181 ,H481144 ,H481281 ,H481391 ,H481520 ,H481657 ,H481778 ,H481893 ,H481982 ,H482033 ,H482084 ,H482182 ,H481153 ,H481282 ,H481392 ,H481521 ,H481658 ,H481779 ,H481894 ,H481983 ,H482034 ,H482085 ,H482183 ,H481154 ,H481283 ,H481393 ,H481522 ,H481659 ,H481780 ,H481896 ,H481984 ,H482035 ,H482086 ,H482185 ,H481155 ,H481284 ,H481394 ,H481523 ,H481662 ,H481781 ,H481897 ,H481985 ,H482036 ,H482087 ,H482186 ,H481156 ,H481285 ,H481395 ,H481525 ,H481663 ,H481783 ,H481898 ,H481986 ,H482037 ,H482088 ,H482187 ,H481157 ,H481286 ,H481397 ,H481530 ,H481664 ,H481784 ,H481899 ,H481987 ,H482038 ,H482089 ,H482188 ,H481158 ,H481287 ,H481398 ,H481531 ,H481665 ,H481787 ,H481909 ,H481988 ,H482039 ,H482090 ,H482194 ,H481159 ,H481288 ,H481399 ,H481540 ,H481666 ,H481788 ,H481916 ,H481989 ,H482040 ,H482091 ,H482195 ,H481160 ,H481289 ,H481401 ,H481542 ,H481669 ,H481793 ,H481918 ,H481990 ,H482041 ,H482092 ,H482198 ,H481169 ,H481290 ,H481402 ,H481547 ,H481678 ,H481794 ,H481919 ,H481991 ,H482042 ,H482093 ,H482201 ,H481184 ,H481291 ,H481408 ,H481548 ,H481679 ,H481798 ,H481920 ,H481992 ,H482043 ,H482094 ,H482203 ,H481188 ,H481292 ,H481409 ,H481549 ,H481680 ,H481799 ,H481923 ,H481993 ,H482044 ,H482095 ,H482209 ,H481193 ,H481293 ,H481410 ,H481551 ,H481681 ,H481800 ,H481924 ,H481994 ,H482045 ,H482096 ,H482210 ,H481194 ,H481294 ,H481411 ,H481558 ,H481682 ,H481801 ,H481925 ,H481995 ,H482046 ,H482097 ,H482211 ,H481195 ,H481295 ,H481412 ,H481560 ,H481683 ,H481810 ,H481926 ,H481996 ,H482047 ,H482098 ,H482212 ,H481203 ,H481296 ,H481419 ,H481561 ,H481684 ,H481812 ,H481927 ,H481997 ,H482048 ,H482099 ,H482213 ,H481204 ,H481298 ,H481428 ,H481562 ,H481685 ,H481813 ,H481928 ,H481998 ,H482049 ,H482100 ,H482214 ,H481214 ,H481300 ,H481432 ,H481563 ,H481689 ,H481814 ,H481929 ,H481999 ,H482050 ,H482101 ,H482215 ,H481220 ,H481301 ,H481434 ,H481564 ,H481690 ,H481818 ,H481930 ,H482000 ,H482051 ,H482102 ,H482225 ,H481221 ,H481308 ,H481435 ,H481565 ,H481692 ,H481820 ,H481931 ,H482001 ,H482052 ,H482103 ,H482226 ,H481223 ,H481311 ,H481442 ,H481585 ,H481707 ,H481828 ,H481932 ,H482002 ,H482053 ,H482104 ,H482229 ,H481224 ,H481315 ,H481443 ,H481586 ,H481716 ,H481829 ,H481933 ,H482003 ,H482054 ,H482105 ,H482239 ,H481225 ,H481316 ,H481445 ,H481587 ,H481718 ,H481830 ,H481934 ,H482004 ,H482055 ,H482106 ,H482242 ,H481226 ,H481317 ,H481455 ,H481588 ,H481719 ,H481831 ,H481936 ,H482005 ,H482056 ,H482107 ,H482250 ,H481227 ,H481318 ,H481456 ,H481589 ,H481720 ,H481832 ,H481953 ,H482006 ,H482057 ,H482108 ,H482251 ,H481228 ,H481320 ,H481457 ,H481606 ,H481722 ,H481833 ,H481956 ,H482007 ,H482058 ,H482109 ,H482252 ,H481229 ,H481321 ,H481458 ,H481610 ,H481723 ,H481836 ,H481957 ,H482008 ,H482059 ,H482110 ,H482253 ,H481230 ,H481328 ,H481459 ,H481611 ,H481727 ,H481837 ,H481958 ,H482009 ,H482060 ,H482111 ,H482254 ,H481231 ,H481329 ,H481465 ,H481612 ,H481728 ,H481838 ,H481959 ,H482010 ,H482061 ,H482112 ,H482258 ,H481236 ,H481334 ,H481466 ,H481620 ,H481729 ,H481845 ,H481960 ,H482011 ,H482062 ,H482113 ,H482259 ,H481237 ,H481335 ,H481476 ,H481621 ,H481732 ,H481857 ,H481961 ,H482012 ,H482063 ,H482114 ,H482260 ,H481256 ,H481341 ,H481489 ,H481622 ,H481733 ,H481862 ,H481962 ,H482013 ,H482064 ,H482115 ,H482261 ,H481257 ,H481342 ,H481490 ,H481625 ,H481734 ,H481863 ,H481963 ,H482014 ,H482065 ,H482138 ,H482262 ,H481263 ,H481343 ,H481491 ,H481626 ,H481735 ,H481864 ,H481964 ,H482015 ,H482066 ,H482139 ,H482263 ,H481264 ,H481345 ,H481494 ,H481627 ,H481736 ,H481865 ,H481965 ,H482016 ,H482067 ,H482140 ,H482264 ,H481265 ,H481346 ,H481498 ,H481631 ,H481737 ,H481868 ,H481966 ,H482017 ,H482068 ,H482141 ,H482265 ,H481266 ,H481347 ,H481499 ,H481638 ,H481738 ,H481869 ,H481967 ,H482018 ,H482069 ,H482142 ,H482270 ,H482271 ,H482362 ,H482552 ,H482662 ,H482799 ,H482908 ,H483059 ,H483186 ,H483291 ,H483374 ,H483501 ,H482272 ,H482367 ,H482553 ,H482663 ,H482802 ,H482909 ,H483063 ,H483194 ,H483292 ,H483375 ,H483506 ,H482274 ,H482369 ,H482555 ,H482676 ,H482812 ,H482917 ,H483064 ,H483197 ,H483294 ,H483376 ,H483513 ,H482275 ,H482389 ,H482557 ,H482678 ,H482813 ,H482918 ,H483070 ,H483200 ,H483295 ,H483378 ,H483514 ,H482276 ,H482390 ,H482559 ,H482679 ,H482820 ,H482919 ,H483071 ,H483203 ,H483296 ,H483379 ,H483515 ,H482277 ,H482391 ,H482560 ,H482681 ,H482821 ,H482920 ,H483072 ,H483204 ,H483297 ,H483380 ,H483516 ,H482278 ,H482396 ,H482561 ,H482682 ,H482822 ,H482925 ,H483073 ,H483206 ,H483298 ,H483382 ,H483518 ,H482279 ,H482397 ,H482566 ,H482683 ,H482823 ,H482935 ,H483084 ,H483207 ,H483299 ,H483383 ,H483519 ,H482281 ,H482398 ,H482567 ,H482684 ,H482829 ,H482936 ,H483088 ,H483208 ,H483300 ,H483384 ,H483522 ,H482282 ,H482399 ,H482568 ,H482685 ,H482830 ,H482938 ,H483089 ,H483209 ,H483301 ,H483385 ,H483532 ,H482284 ,H482407 ,H482569 ,H482686 ,H482831 ,H482939 ,H483091 ,H483215 ,H483302 ,H483388 ,H483533 ,H482285 ,H482408 ,H482579 ,H482687 ,H482839 ,H482940 ,H483092 ,H483223 ,H483303 ,H483389 ,H483534 ,H482296 ,H482409 ,H482580 ,H482688 ,H482840 ,H482941 ,H483097 ,H483224 ,H483306 ,H483390 ,H483537 ,H482309 ,H482410 ,H482581 ,H482689 ,H482841 ,H482942 ,H483099 ,H483225 ,H483309 ,H483391 ,H483546 ,H482310 ,H482411 ,H482587 ,H482693 ,H482845 ,H482943 ,H483100 ,H483235 ,H483311 ,H483392 ,H483554 ,H482323 ,H482412 ,H482588 ,H482699 ,H482846 ,H482944 ,H483102 ,H483236 ,H483315 ,H483393 ,H483556 ,H482327 ,H482413 ,H482594 ,H482700 ,H482847 ,H482945 ,H483106 ,H483237 ,H483317 ,H483394 ,H483560 ,H482328 ,H482414 ,H482595 ,H482701 ,H482848 ,H482970 ,H483109 ,H483238 ,H483318 ,H483401 ,H483566 ,H482329 ,H482415 ,H482596 ,H482702 ,H482852 ,H482982 ,H483110 ,H483241 ,H483319 ,H483403 ,H483567 ,H482330 ,H482440 ,H482597 ,H482703 ,H482854 ,H482983 ,H483117 ,H483242 ,H483321 ,H483404 ,H483568 ,H482331 ,H482441 ,H482598 ,H482705 ,H482855 ,H482984 ,H483118 ,H483243 ,H483327 ,H483405 ,H483569 ,H482332 ,H482450 ,H482599 ,H482706 ,H482856 ,H482985 ,H483119 ,H483244 ,H483330 ,H483407 ,H483586 ,H482333 ,H482451 ,H482601 ,H482709 ,H482857 ,H482986 ,H483121 ,H483245 ,H483331 ,H483416 ,H483589 ,H482334 ,H482452 ,H482602 ,H482710 ,H482858 ,H482987 ,H483135 ,H483246 ,H483332 ,H483417 ,H483590 ,H482335 ,H482457 ,H482603 ,H482711 ,H482859 ,H482988 ,H483136 ,H483248 ,H483333 ,H483418 ,H483599 ,H482336 ,H482462 ,H482604 ,H482712 ,H482860 ,H482989 ,H483137 ,H483249 ,H483334 ,H483419 ,H483600 ,H482337 ,H482463 ,H482605 ,H482713 ,H482861 ,H482990 ,H483138 ,H483250 ,H483335 ,H483420 ,H483606 ,H482338 ,H482473 ,H482613 ,H482715 ,H482862 ,H482991 ,H483139 ,H483254 ,H483336 ,H483421 ,H483607 ,H482339 ,H482474 ,H482619 ,H482716 ,H482863 ,H482992 ,H483140 ,H483257 ,H483337 ,H483422 ,H483608 ,H482340 ,H482475 ,H482620 ,H482717 ,H482866 ,H483010 ,H483141 ,H483258 ,H483338 ,H483427 ,H483609 ,H482341 ,H482476 ,H482621 ,H482722 ,H482867 ,H483011 ,H483142 ,H483259 ,H483339 ,H483431 ,H483611 ,H482342 ,H482481 ,H482624 ,H482723 ,H482868 ,H483014 ,H483143 ,H483260 ,H483340 ,H483433 ,H483615 ,H482343 ,H482484 ,H482628 ,H482725 ,H482869 ,H483015 ,H483144 ,H483261 ,H483341 ,H483435 ,H483620 ,H482344 ,H482491 ,H482629 ,H482730 ,H482870 ,H483016 ,H483145 ,H483262 ,H483342 ,H483436 ,H483621 ,H482345 ,H482499 ,H482630 ,H482731 ,H482876 ,H483017 ,H483146 ,H483265 ,H483343 ,H483437 ,H483622 ,H482346 ,H482500 ,H482631 ,H482738 ,H482877 ,H483018 ,H483147 ,H483266 ,H483344 ,H483438 ,H483623 ,H482347 ,H482501 ,H482632 ,H482758 ,H482878 ,H483019 ,H483151 ,H483267 ,H483354 ,H483439 ,H483624 ,H482348 ,H482502 ,H482641 ,H482759 ,H482879 ,H483020 ,H483152 ,H483272 ,H483355 ,H483440 ,H483625 ,H482349 ,H482503 ,H482646 ,H482762 ,H482880 ,H483021 ,H483153 ,H483273 ,H483356 ,H483441 ,H483626 ,H482350 ,H482504 ,H482647 ,H482764 ,H482881 ,H483022 ,H483157 ,H483274 ,H483357 ,H483444 ,H483627 ,H482351 ,H482505 ,H482648 ,H482765 ,H482888 ,H483023 ,H483159 ,H483277 ,H483359 ,H483445 ,H483628 ,H482352 ,H482506 ,H482649 ,H482766 ,H482889 ,H483024 ,H483161 ,H483278 ,H483360 ,H483446 ,H483632 ,H482353 ,H482508 ,H482650 ,H482767 ,H482892 ,H483025 ,H483162 ,H483280 ,H483361 ,H483447 ,H483633 ,H482354 ,H482509 ,H482651 ,H482768 ,H482899 ,H483026 ,H483163 ,H483281 ,H483362 ,H483452 ,H483634 ,H482355 ,H482524 ,H482652 ,H482773 ,H482900 ,H483027 ,H483164 ,H483282 ,H483363 ,H483455 ,H483635 ,H482356 ,H482527 ,H482653 ,H482774 ,H482901 ,H483042 ,H483165 ,H483283 ,H483364 ,H483459 ,H483636 ,H482357 ,H482530 ,H482654 ,H482792 ,H482902 ,H483043 ,H483176 ,H483286 ,H483365 ,H483460 ,H483651 ,H482358 ,H482534 ,H482656 ,H482794 ,H482903 ,H483044 ,H483182 ,H483287 ,H483366 ,H483461 ,H483652 ,H482359 ,H482539 ,H482657 ,H482795 ,H482904 ,H483047 ,H483183 ,H483288 ,H483367 ,H483470 ,H483653 ,H482360 ,H482550 ,H482658 ,H482796 ,H482906 ,H483051 ,H483184 ,H483289 ,H483369 ,H483487 ,H483655 ,H482361 ,H482551 ,H482661 ,H482797 ,H482907 ,H483052 ,H483185 ,H483290 ,H483373 ,H483494 ,H483656 ,H483657 ,H483802 ,H483939 ,H484042 ,H484194 ,H484302 ,H484425 ,H484535 ,H484622 ,H484731 ,H484813 ,H483658 ,H483809 ,H483940 ,H484043 ,H484198 ,H484303 ,H484426 ,H484536 ,H484625 ,H484732 ,H484814 ,H483659 ,H483810 ,H483941 ,H484044 ,H484199 ,H484304 ,H484427 ,H484537 ,H484626 ,H484733 ,H484815 ,H483660 ,H483811 ,H483944 ,H484045 ,H484213 ,H484305 ,H484431 ,H484538 ,H484627 ,H484734 ,H484816 ,H483661 ,H483812 ,H483945 ,H484051 ,H484214 ,H484306 ,H484436 ,H484541 ,H484628 ,H484735 ,H484817 ,H483662 ,H483813 ,H483946 ,H484057 ,H484215 ,H484307 ,H484439 ,H484542 ,H484629 ,H484736 ,H484818 ,H483663 ,H483817 ,H483947 ,H484058 ,H484216 ,H484308 ,H484440 ,H484543 ,H484630 ,H484737 ,H484819 ,H483664 ,H483821 ,H483948 ,H484059 ,H484217 ,H484309 ,H484441 ,H484544 ,H484631 ,H484738 ,H484820 ,H483665 ,H483824 ,H483949 ,H484060 ,H484218 ,H484310 ,H484442 ,H484545 ,H484632 ,H484739 ,H484822 ,H483666 ,H483825 ,H483950 ,H484061 ,H484219 ,H484320 ,H484443 ,H484549 ,H484634 ,H484740 ,H484823 ,H483667 ,H483826 ,H483951 ,H484062 ,H484220 ,H484334 ,H484444 ,H484550 ,H484641 ,H484741 ,H484824 ,H483668 ,H483831 ,H483952 ,H484063 ,H484222 ,H484335 ,H484445 ,H484551 ,H484645 ,H484742 ,H484825 ,H483670 ,H483836 ,H483953 ,H484076 ,H484224 ,H484336 ,H484446 ,H484552 ,H484646 ,H484752 ,H484826 ,H483671 ,H483837 ,H483957 ,H484079 ,H484225 ,H484346 ,H484447 ,H484554 ,H484647 ,H484753 ,H484827 ,H483672 ,H483838 ,H483960 ,H484080 ,H484226 ,H484348 ,H484448 ,H484555 ,H484650 ,H484754 ,H484837 ,H483674 ,H483839 ,H483964 ,H484081 ,H484227 ,H484349 ,H484449 ,H484556 ,H484659 ,H484755 ,H484839 ,H483676 ,H483844 ,H483966 ,H484085 ,H484233 ,H484354 ,H484450 ,H484557 ,H484661 ,H484756 ,H484841 ,H483677 ,H483847 ,H483973 ,H484087 ,H484234 ,H484360 ,H484451 ,H484558 ,H484662 ,H484761 ,H484846 ,H483682 ,H483848 ,H483977 ,H484088 ,H484235 ,H484362 ,H484452 ,H484559 ,H484663 ,H484762 ,H484857 ,H483683 ,H483849 ,H483978 ,H484089 ,H484236 ,H484363 ,H484454 ,H484560 ,H484665 ,H484763 ,H484858 ,H483684 ,H483850 ,H483979 ,H484090 ,H484237 ,H484368 ,H484455 ,H484561 ,H484670 ,H484764 ,H484862 ,H483694 ,H483851 ,H483980 ,H484091 ,H484238 ,H484369 ,H484456 ,H484562 ,H484671 ,H484765 ,H484863 ,H483695 ,H483856 ,H483981 ,H484092 ,H484239 ,H484370 ,H484457 ,H484563 ,H484672 ,H484766 ,H484865 ,H483696 ,H483860 ,H483982 ,H484093 ,H484240 ,H484371 ,H484458 ,H484564 ,H484682 ,H484767 ,H484866 ,H483705 ,H483862 ,H483983 ,H484094 ,H484241 ,H484372 ,H484459 ,H484573 ,H484691 ,H484768 ,H484872 ,H483712 ,H483863 ,H483986 ,H484095 ,H484242 ,H484373 ,H484460 ,H484574 ,H484692 ,H484781 ,H484873 ,H483713 ,H483864 ,H483994 ,H484096 ,H484243 ,H484381 ,H484464 ,H484575 ,H484693 ,H484782 ,H484874 ,H483715 ,H483867 ,H483995 ,H484100 ,H484244 ,H484382 ,H484465 ,H484587 ,H484694 ,H484784 ,H484875 ,H483716 ,H483868 ,H483996 ,H484103 ,H484245 ,H484383 ,H484481 ,H484593 ,H484707 ,H484785 ,H484892 ,H483721 ,H483874 ,H483997 ,H484104 ,H484246 ,H484384 ,H484482 ,H484594 ,H484708 ,H484787 ,H484893 ,H483722 ,H483875 ,H483998 ,H484111 ,H484247 ,H484385 ,H484483 ,H484597 ,H484709 ,H484790 ,H484894 ,H483726 ,H483876 ,H483999 ,H484112 ,H484248 ,H484388 ,H484485 ,H484598 ,H484710 ,H484791 ,H484895 ,H483727 ,H483877 ,H484000 ,H484113 ,H484258 ,H484389 ,H484494 ,H484599 ,H484711 ,H484792 ,H484896 ,H483728 ,H483914 ,H484001 ,H484114 ,H484268 ,H484390 ,H484496 ,H484600 ,H484712 ,H484793 ,H484897 ,H483732 ,H483915 ,H484008 ,H484116 ,H484269 ,H484391 ,H484501 ,H484601 ,H484714 ,H484794 ,H484898 ,H483733 ,H483916 ,H484009 ,H484117 ,H484270 ,H484392 ,H484502 ,H484602 ,H484715 ,H484795 ,H484899 ,H483754 ,H483917 ,H484014 ,H484118 ,H484271 ,H484397 ,H484503 ,H484603 ,H484716 ,H484798 ,H484900 ,H483756 ,H483921 ,H484015 ,H484119 ,H484283 ,H484408 ,H484504 ,H484604 ,H484717 ,H484799 ,H484901 ,H483766 ,H483922 ,H484016 ,H484120 ,H484285 ,H484409 ,H484509 ,H484605 ,H484718 ,H484800 ,H484903 ,H483771 ,H483923 ,H484017 ,H484126 ,H484286 ,H484410 ,H484510 ,H484606 ,H484719 ,H484801 ,H484910 ,H483772 ,H483924 ,H484019 ,H484135 ,H484291 ,H484411 ,H484511 ,H484607 ,H484720 ,H484802 ,H484911 ,H483773 ,H483925 ,H484020 ,H484151 ,H484292 ,H484412 ,H484512 ,H484608 ,H484721 ,H484803 ,H484916 ,H483774 ,H483926 ,H484031 ,H484158 ,H484293 ,H484413 ,H484526 ,H484612 ,H484722 ,H484804 ,H484917 ,H483775 ,H483927 ,H484032 ,H484159 ,H484294 ,H484414 ,H484527 ,H484613 ,H484723 ,H484805 ,H484923 ,H483776 ,H483928 ,H484033 ,H484166 ,H484295 ,H484415 ,H484528 ,H484614 ,H484724 ,H484806 ,H484925 ,H483777 ,H483929 ,H484034 ,H484174 ,H484296 ,H484416 ,H484529 ,H484615 ,H484725 ,H484807 ,H484926 ,H483778 ,H483930 ,H484035 ,H484186 ,H484297 ,H484419 ,H484530 ,H484616 ,H484726 ,H484808 ,H484932 ,H483779 ,H483932 ,H484036 ,H484190 ,H484298 ,H484421 ,H484531 ,H484617 ,H484727 ,H484809 ,H484933 ,H483780 ,H483933 ,H484039 ,H484191 ,H484299 ,H484422 ,H484532 ,H484619 ,H484728 ,H484810 ,H484934 ,H483800 ,H483937 ,H484040 ,H484192 ,H484300 ,H484423 ,H484533 ,H484620 ,H484729 ,H484811 ,H484937 ,H483801 ,H483938 ,H484041 ,H484193 ,H484301 ,H484424 ,H484534 ,H484621 ,H484730 ,H484812 ,H484938 ,H484939 ,H485017 ,H485088 ,H485200 ,H485275 ,H485376 ,H485479 ,H485599 ,H485742 ,H485845 ,H485940 ,H484941 ,H485018 ,H485092 ,H485201 ,H485283 ,H485377 ,H485480 ,H485602 ,H485749 ,H485848 ,H485942 ,H484942 ,H485019 ,H485097 ,H485202 ,H485284 ,H485378 ,H485485 ,H485603 ,H485750 ,H485849 ,H485964 ,H484943 ,H485020 ,H485099 ,H485203 ,H485286 ,H485379 ,H485486 ,H485612 ,H485751 ,H485850 ,H485965 ,H484945 ,H485021 ,H485102 ,H485204 ,H485288 ,H485380 ,H485489 ,H485613 ,H485752 ,H485851 ,H485966 ,H484946 ,H485024 ,H485103 ,H485205 ,H485289 ,H485386 ,H485490 ,H485614 ,H485753 ,H485852 ,H485967 ,H484947 ,H485025 ,H485104 ,H485206 ,H485290 ,H485387 ,H485491 ,H485618 ,H485754 ,H485853 ,H485968 ,H484948 ,H485026 ,H485105 ,H485207 ,H485292 ,H485388 ,H485492 ,H485620 ,H485756 ,H485854 ,H485973 ,H484950 ,H485027 ,H485106 ,H485208 ,H485293 ,H485393 ,H485495 ,H485621 ,H485757 ,H485855 ,H485982 ,H484951 ,H485028 ,H485107 ,H485209 ,H485294 ,H485394 ,H485516 ,H485622 ,H485758 ,H485857 ,H485983 ,H484952 ,H485029 ,H485108 ,H485210 ,H485295 ,H485395 ,H485523 ,H485642 ,H485759 ,H485858 ,H485984 ,H484953 ,H485030 ,H485109 ,H485211 ,H485296 ,H485396 ,H485526 ,H485644 ,H485760 ,H485859 ,H485985 ,H484966 ,H485033 ,H485110 ,H485212 ,H485301 ,H485397 ,H485527 ,H485645 ,H485761 ,H485860 ,H485989 ,H484968 ,H485034 ,H485111 ,H485213 ,H485305 ,H485398 ,H485533 ,H485648 ,H485762 ,H485862 ,H485997 ,H484970 ,H485035 ,H485112 ,H485214 ,H485306 ,H485400 ,H485534 ,H485653 ,H485763 ,H485863 ,H485998 ,H484971 ,H485036 ,H485116 ,H485215 ,H485307 ,H485401 ,H485541 ,H485654 ,H485764 ,H485867 ,H485999 ,H484972 ,H485038 ,H485118 ,H485216 ,H485309 ,H485413 ,H485542 ,H485663 ,H485765 ,H485868 ,H486003 ,H484973 ,H485043 ,H485119 ,H485217 ,H485310 ,H485418 ,H485545 ,H485667 ,H485766 ,H485869 ,H486004 ,H484975 ,H485044 ,H485120 ,H485218 ,H485312 ,H485419 ,H485546 ,H485669 ,H485767 ,H485870 ,H486005 ,H484976 ,H485051 ,H485121 ,H485222 ,H485313 ,H485420 ,H485547 ,H485670 ,H485768 ,H485871 ,H486007 ,H484977 ,H485053 ,H485122 ,H485223 ,H485314 ,H485421 ,H485548 ,H485672 ,H485769 ,H485872 ,H486009 ,H484978 ,H485054 ,H485129 ,H485224 ,H485315 ,H485437 ,H485549 ,H485675 ,H485770 ,H485873 ,H486022 ,H484987 ,H485055 ,H485130 ,H485225 ,H485319 ,H485438 ,H485550 ,H485678 ,H485771 ,H485875 ,H486023 ,H484988 ,H485056 ,H485131 ,H485226 ,H485320 ,H485439 ,H485551 ,H485679 ,H485773 ,H485876 ,H486026 ,H484989 ,H485057 ,H485132 ,H485227 ,H485334 ,H485440 ,H485552 ,H485682 ,H485774 ,H485886 ,H486033 ,H484990 ,H485058 ,H485133 ,H485231 ,H485335 ,H485442 ,H485553 ,H485683 ,H485775 ,H485887 ,H486046 ,H484992 ,H485059 ,H485134 ,H485234 ,H485336 ,H485443 ,H485554 ,H485694 ,H485794 ,H485888 ,H486047 ,H484993 ,H485060 ,H485143 ,H485236 ,H485337 ,H485444 ,H485555 ,H485701 ,H485804 ,H485889 ,H486048 ,H484994 ,H485061 ,H485144 ,H485237 ,H485338 ,H485445 ,H485558 ,H485702 ,H485806 ,H485907 ,H486049 ,H484995 ,H485062 ,H485145 ,H485238 ,H485339 ,H485446 ,H485559 ,H485704 ,H485807 ,H485908 ,H486050 ,H484996 ,H485063 ,H485146 ,H485239 ,H485340 ,H485447 ,H485560 ,H485705 ,H485808 ,H485909 ,H486062 ,H484997 ,H485064 ,H485147 ,H485243 ,H485341 ,H485448 ,H485561 ,H485711 ,H485809 ,H485910 ,H486063 ,H484998 ,H485065 ,H485149 ,H485250 ,H485344 ,H485451 ,H485565 ,H485713 ,H485810 ,H485911 ,H486064 ,H484999 ,H485066 ,H485150 ,H485251 ,H485345 ,H485452 ,H485569 ,H485714 ,H485811 ,H485917 ,H486065 ,H485000 ,H485067 ,H485151 ,H485253 ,H485346 ,H485454 ,H485570 ,H485715 ,H485823 ,H485918 ,H486068 ,H485001 ,H485068 ,H485152 ,H485254 ,H485347 ,H485455 ,H485571 ,H485716 ,H485824 ,H485919 ,H486069 ,H485002 ,H485069 ,H485161 ,H485255 ,H485348 ,H485457 ,H485572 ,H485719 ,H485825 ,H485920 ,H486071 ,H485003 ,H485070 ,H485184 ,H485256 ,H485352 ,H485458 ,H485576 ,H485720 ,H485826 ,H485921 ,H486072 ,H485004 ,H485071 ,H485185 ,H485257 ,H485353 ,H485460 ,H485577 ,H485722 ,H485827 ,H485922 ,H486073 ,H485005 ,H485072 ,H485186 ,H485258 ,H485354 ,H485461 ,H485578 ,H485723 ,H485831 ,H485923 ,H486074 ,H485006 ,H485073 ,H485187 ,H485262 ,H485355 ,H485463 ,H485579 ,H485724 ,H485832 ,H485924 ,H486075 ,H485007 ,H485074 ,H485188 ,H485264 ,H485356 ,H485465 ,H485580 ,H485725 ,H485833 ,H485925 ,H486076 ,H485008 ,H485075 ,H485189 ,H485265 ,H485362 ,H485468 ,H485581 ,H485726 ,H485834 ,H485926 ,H486079 ,H485009 ,H485076 ,H485190 ,H485266 ,H485363 ,H485469 ,H485582 ,H485727 ,H485835 ,H485929 ,H486080 ,H485010 ,H485077 ,H485191 ,H485268 ,H485364 ,H485470 ,H485583 ,H485728 ,H485836 ,H485933 ,H486082 ,H485011 ,H485078 ,H485194 ,H485269 ,H485366 ,H485471 ,H485586 ,H485732 ,H485837 ,H485934 ,H486086 ,H485012 ,H485079 ,H485195 ,H485270 ,H485370 ,H485473 ,H485587 ,H485733 ,H485838 ,H485935 ,H486087 ,H485013 ,H485080 ,H485196 ,H485271 ,H485372 ,H485475 ,H485592 ,H485734 ,H485839 ,H485936 ,H486088 ,H485014 ,H485081 ,H485197 ,H485272 ,H485373 ,H485476 ,H485594 ,H485736 ,H485840 ,H485937 ,H486089 ,H485015 ,H485086 ,H485198 ,H485273 ,H485374 ,H485477 ,H485595 ,H485740 ,H485841 ,H485938 ,H486091 ,H485016 ,H485087 ,H485199 ,H485274 ,H485375 ,H485478 ,H485597 ,H485741 ,H485844 ,H485939 ,H486092 ,H486093 ,H486213 ,H486322 ,H486436 ,H486522 ,H486639 ,H486735 ,H486835 ,H486926 ,H486997 ,H487064 ,H486094 ,H486215 ,H486323 ,H486439 ,H486523 ,H486641 ,H486736 ,H486836 ,H486930 ,H486998 ,H487065 ,H486095 ,H486216 ,H486324 ,H486441 ,H486524 ,H486642 ,H486737 ,H486837 ,H486931 ,H486999 ,H487066 ,H486097 ,H486218 ,H486325 ,H486442 ,H486535 ,H486643 ,H486738 ,H486838 ,H486932 ,H487000 ,H487067 ,H486098 ,H486219 ,H486326 ,H486443 ,H486536 ,H486644 ,H486739 ,H486839 ,H486933 ,H487001 ,H487068 ,H486099 ,H486220 ,H486327 ,H486444 ,H486537 ,H486645 ,H486740 ,H486849 ,H486934 ,H487002 ,H487069 ,H486100 ,H486226 ,H486328 ,H486445 ,H486538 ,H486646 ,H486757 ,H486850 ,H486935 ,H487003 ,H487070 ,H486101 ,H486227 ,H486329 ,H486446 ,H486539 ,H486651 ,H486758 ,H486851 ,H486937 ,H487004 ,H487071 ,H486102 ,H486233 ,H486330 ,H486447 ,H486540 ,H486652 ,H486759 ,H486852 ,H486938 ,H487005 ,H487072 ,H486103 ,H486234 ,H486335 ,H486452 ,H486541 ,H486653 ,H486760 ,H486853 ,H486939 ,H487006 ,H487073 ,H486104 ,H486235 ,H486336 ,H486453 ,H486542 ,H486654 ,H486769 ,H486854 ,H486940 ,H487007 ,H487075 ,H486105 ,H486245 ,H486341 ,H486454 ,H486547 ,H486655 ,H486778 ,H486855 ,H486942 ,H487008 ,H487077 ,H486108 ,H486246 ,H486342 ,H486455 ,H486553 ,H486656 ,H486779 ,H486856 ,H486943 ,H487009 ,H487078 ,H486109 ,H486253 ,H486346 ,H486456 ,H486554 ,H486659 ,H486780 ,H486857 ,H486945 ,H487010 ,H487079 ,H486113 ,H486254 ,H486347 ,H486457 ,H486556 ,H486660 ,H486781 ,H486866 ,H486949 ,H487016 ,H487080 ,H486116 ,H486255 ,H486348 ,H486458 ,H486557 ,H486661 ,H486782 ,H486867 ,H486950 ,H487017 ,H487081 ,H486117 ,H486256 ,H486349 ,H486459 ,H486558 ,H486662 ,H486784 ,H486868 ,H486951 ,H487019 ,H487082 ,H486118 ,H486257 ,H486350 ,H486460 ,H486559 ,H486663 ,H486785 ,H486869 ,H486956 ,H487020 ,H487083 ,H486119 ,H486258 ,H486351 ,H486461 ,H486560 ,H486664 ,H486786 ,H486870 ,H486962 ,H487021 ,H487084 ,H486122 ,H486259 ,H486352 ,H486462 ,H486566 ,H486665 ,H486793 ,H486872 ,H486963 ,H487022 ,H487085 ,H486123 ,H486260 ,H486353 ,H486463 ,H486567 ,H486666 ,H486794 ,H486873 ,H486964 ,H487023 ,H487086 ,H486124 ,H486261 ,H486355 ,H486464 ,H486568 ,H486667 ,H486795 ,H486874 ,H486965 ,H487024 ,H487089 ,H486125 ,H486262 ,H486368 ,H486465 ,H486570 ,H486668 ,H486796 ,H486878 ,H486966 ,H487025 ,H487130 ,H486126 ,H486263 ,H486369 ,H486466 ,H486571 ,H486670 ,H486797 ,H486879 ,H486967 ,H487026 ,H487131 ,H486127 ,H486264 ,H486370 ,H486467 ,H486572 ,H486683 ,H486798 ,H486880 ,H486968 ,H487028 ,H487132 ,H486128 ,H486265 ,H486371 ,H486468 ,H486583 ,H486685 ,H486799 ,H486881 ,H486969 ,H487029 ,H487133 ,H486129 ,H486266 ,H486372 ,H486469 ,H486584 ,H486688 ,H486803 ,H486882 ,H486970 ,H487031 ,H487134 ,H486130 ,H486271 ,H486373 ,H486470 ,H486589 ,H486689 ,H486804 ,H486883 ,H486971 ,H487032 ,H487135 ,H486131 ,H486272 ,H486374 ,H486471 ,H486590 ,H486690 ,H486806 ,H486884 ,H486972 ,H487041 ,H487136 ,H486132 ,H486273 ,H486375 ,H486473 ,H486591 ,H486691 ,H486807 ,H486888 ,H486973 ,H487042 ,H487137 ,H486133 ,H486274 ,H486376 ,H486474 ,H486592 ,H486693 ,H486808 ,H486889 ,H486974 ,H487043 ,H487138 ,H486134 ,H486275 ,H486377 ,H486475 ,H486593 ,H486694 ,H486810 ,H486890 ,H486975 ,H487044 ,H487139 ,H486136 ,H486276 ,H486378 ,H486476 ,H486594 ,H486695 ,H486811 ,H486891 ,H486976 ,H487045 ,H487140 ,H486138 ,H486277 ,H486379 ,H486477 ,H486597 ,H486696 ,H486812 ,H486892 ,H486977 ,H487046 ,H487141 ,H486139 ,H486284 ,H486380 ,H486478 ,H486598 ,H486697 ,H486813 ,H486893 ,H486979 ,H487047 ,H487142 ,H486143 ,H486285 ,H486381 ,H486479 ,H486599 ,H486698 ,H486814 ,H486899 ,H486980 ,H487048 ,H487143 ,H486144 ,H486286 ,H486382 ,H486480 ,H486600 ,H486699 ,H486816 ,H486900 ,H486981 ,H487049 ,H487144 ,H486184 ,H486287 ,H486383 ,H486481 ,H486602 ,H486700 ,H486817 ,H486901 ,H486982 ,H487050 ,H487145 ,H486185 ,H486288 ,H486384 ,H486482 ,H486603 ,H486701 ,H486818 ,H486902 ,H486984 ,H487051 ,H487146 ,H486186 ,H486289 ,H486386 ,H486487 ,H486604 ,H486702 ,H486820 ,H486903 ,H486985 ,H487052 ,H487147 ,H486188 ,H486290 ,H486387 ,H486488 ,H486605 ,H486703 ,H486821 ,H486904 ,H486986 ,H487053 ,H487148 ,H486189 ,H486292 ,H486399 ,H486489 ,H486613 ,H486704 ,H486823 ,H486906 ,H486987 ,H487054 ,H487149 ,H486190 ,H486293 ,H486401 ,H486492 ,H486614 ,H486708 ,H486824 ,H486907 ,H486988 ,H487055 ,H487150 ,H486191 ,H486294 ,H486402 ,H486493 ,H486617 ,H486715 ,H486825 ,H486914 ,H486989 ,H487056 ,H487151 ,H486198 ,H486298 ,H486405 ,H486494 ,H486618 ,H486720 ,H486828 ,H486918 ,H486990 ,H487057 ,H487152 ,H486199 ,H486299 ,H486411 ,H486495 ,H486620 ,H486721 ,H486829 ,H486919 ,H486991 ,H487058 ,H487153 ,H486200 ,H486300 ,H486418 ,H486506 ,H486621 ,H486722 ,H486830 ,H486920 ,H486992 ,H487059 ,H487154 ,H486201 ,H486302 ,H486419 ,H486507 ,H486624 ,H486723 ,H486831 ,H486921 ,H486993 ,H487060 ,H487155 ,H486205 ,H486306 ,H486420 ,H486508 ,H486626 ,H486724 ,H486832 ,H486922 ,H486994 ,H487061 ,H487156 ,H486206 ,H486309 ,H486421 ,H486511 ,H486628 ,H486731 ,H486833 ,H486923 ,H486995 ,H487062 ,H487157 ,H486208 ,H486310 ,H486434 ,H486520 ,H486638 ,H486733 ,H486834 ,H486924 ,H486996 ,H487063 ,H487158 ,H487159 ,H487261 ,H487392 ,H487504 ,H487583 ,H487720 ,H487840 ,H487920 ,H488006 ,H488119 ,H488201 ,H487160 ,H487262 ,H487393 ,H487505 ,H487584 ,H487722 ,H487841 ,H487921 ,H488007 ,H488120 ,H488202 ,H487161 ,H487263 ,H487394 ,H487507 ,H487587 ,H487723 ,H487842 ,H487922 ,H488018 ,H488124 ,H488203 ,H487162 ,H487264 ,H487395 ,H487508 ,H487588 ,H487724 ,H487843 ,H487923 ,H488019 ,H488125 ,H488204 ,H487164 ,H487280 ,H487396 ,H487509 ,H487590 ,H487725 ,H487844 ,H487924 ,H488020 ,H488126 ,H488205 ,H487165 ,H487281 ,H487397 ,H487510 ,H487591 ,H487729 ,H487845 ,H487925 ,H488021 ,H488127 ,H488206 ,H487166 ,H487292 ,H487398 ,H487511 ,H487592 ,H487730 ,H487846 ,H487926 ,H488026 ,H488128 ,H488207 ,H487167 ,H487293 ,H487399 ,H487512 ,H487593 ,H487731 ,H487847 ,H487927 ,H488027 ,H488129 ,H488208 ,H487168 ,H487294 ,H487400 ,H487513 ,H487594 ,H487732 ,H487848 ,H487929 ,H488031 ,H488130 ,H488212 ,H487169 ,H487295 ,H487403 ,H487514 ,H487605 ,H487735 ,H487849 ,H487930 ,H488033 ,H488131 ,H488214 ,H487170 ,H487296 ,H487407 ,H487515 ,H487606 ,H487737 ,H487850 ,H487933 ,H488042 ,H488132 ,H488216 ,H487171 ,H487297 ,H487408 ,H487516 ,H487607 ,H487739 ,H487851 ,H487934 ,H488047 ,H488134 ,H488219 ,H487172 ,H487298 ,H487410 ,H487517 ,H487608 ,H487740 ,H487852 ,H487935 ,H488048 ,H488135 ,H488220 ,H487173 ,H487299 ,H487411 ,H487518 ,H487609 ,H487759 ,H487853 ,H487936 ,H488049 ,H488136 ,H488221 ,H487174 ,H487300 ,H487412 ,H487519 ,H487610 ,H487761 ,H487854 ,H487937 ,H488050 ,H488137 ,H488222 ,H487175 ,H487301 ,H487413 ,H487520 ,H487611 ,H487762 ,H487855 ,H487938 ,H488051 ,H488145 ,H488223 ,H487176 ,H487302 ,H487415 ,H487523 ,H487612 ,H487763 ,H487858 ,H487939 ,H488054 ,H488146 ,H488224 ,H487182 ,H487303 ,H487416 ,H487524 ,H487613 ,H487766 ,H487859 ,H487940 ,H488055 ,H488147 ,H488225 ,H487183 ,H487304 ,H487417 ,H487525 ,H487614 ,H487769 ,H487860 ,H487941 ,H488056 ,H488148 ,H488227 ,H487187 ,H487305 ,H487418 ,H487529 ,H487615 ,H487771 ,H487861 ,H487943 ,H488057 ,H488149 ,H488228 ,H487190 ,H487306 ,H487423 ,H487532 ,H487617 ,H487772 ,H487862 ,H487944 ,H488058 ,H488163 ,H488230 ,H487191 ,H487307 ,H487426 ,H487533 ,H487619 ,H487773 ,H487863 ,H487945 ,H488059 ,H488164 ,H488231 ,H487192 ,H487308 ,H487427 ,H487534 ,H487621 ,H487774 ,H487866 ,H487946 ,H488061 ,H488166 ,H488240 ,H487193 ,H487309 ,H487431 ,H487535 ,H487622 ,H487775 ,H487867 ,H487948 ,H488062 ,H488169 ,H488241 ,H487195 ,H487310 ,H487432 ,H487536 ,H487624 ,H487776 ,H487868 ,H487949 ,H488063 ,H488170 ,H488242 ,H487196 ,H487311 ,H487434 ,H487538 ,H487625 ,H487777 ,H487871 ,H487950 ,H488064 ,H488171 ,H488243 ,H487197 ,H487312 ,H487435 ,H487539 ,H487626 ,H487778 ,H487873 ,H487951 ,H488072 ,H488172 ,H488244 ,H487198 ,H487313 ,H487438 ,H487540 ,H487627 ,H487779 ,H487874 ,H487952 ,H488073 ,H488173 ,H488245 ,H487199 ,H487314 ,H487440 ,H487541 ,H487628 ,H487780 ,H487876 ,H487953 ,H488075 ,H488174 ,H488250 ,H487200 ,H487315 ,H487441 ,H487549 ,H487629 ,H487781 ,H487882 ,H487961 ,H488076 ,H488175 ,H488251 ,H487201 ,H487322 ,H487442 ,H487550 ,H487630 ,H487785 ,H487883 ,H487962 ,H488077 ,H488176 ,H488255 ,H487202 ,H487323 ,H487446 ,H487554 ,H487633 ,H487786 ,H487884 ,H487963 ,H488079 ,H488177 ,H488256 ,H487203 ,H487324 ,H487447 ,H487555 ,H487634 ,H487788 ,H487886 ,H487964 ,H488080 ,H488178 ,H488257 ,H487204 ,H487325 ,H487448 ,H487556 ,H487635 ,H487794 ,H487890 ,H487965 ,H488081 ,H488179 ,H488258 ,H487205 ,H487326 ,H487449 ,H487557 ,H487636 ,H487795 ,H487891 ,H487966 ,H488082 ,H488182 ,H488263 ,H487206 ,H487327 ,H487452 ,H487561 ,H487643 ,H487811 ,H487892 ,H487967 ,H488083 ,H488183 ,H488264 ,H487210 ,H487328 ,H487453 ,H487562 ,H487644 ,H487812 ,H487893 ,H487968 ,H488084 ,H488184 ,H488265 ,H487211 ,H487331 ,H487458 ,H487563 ,H487645 ,H487814 ,H487894 ,H487969 ,H488089 ,H488185 ,H488266 ,H487215 ,H487332 ,H487460 ,H487564 ,H487646 ,H487817 ,H487897 ,H487972 ,H488090 ,H488186 ,H488267 ,H487216 ,H487333 ,H487461 ,H487565 ,H487647 ,H487818 ,H487898 ,H487974 ,H488091 ,H488188 ,H488268 ,H487217 ,H487343 ,H487470 ,H487566 ,H487676 ,H487828 ,H487899 ,H487975 ,H488092 ,H488189 ,H488269 ,H487218 ,H487344 ,H487471 ,H487567 ,H487677 ,H487829 ,H487900 ,H487976 ,H488093 ,H488190 ,H488270 ,H487219 ,H487345 ,H487472 ,H487568 ,H487678 ,H487830 ,H487902 ,H487977 ,H488094 ,H488191 ,H488271 ,H487220 ,H487346 ,H487493 ,H487569 ,H487679 ,H487831 ,H487903 ,H487980 ,H488096 ,H488192 ,H488272 ,H487223 ,H487347 ,H487494 ,H487570 ,H487695 ,H487832 ,H487904 ,H487981 ,H488107 ,H488193 ,H488273 ,H487224 ,H487381 ,H487495 ,H487573 ,H487696 ,H487834 ,H487905 ,H487982 ,H488109 ,H488194 ,H488274 ,H487226 ,H487382 ,H487496 ,H487574 ,H487697 ,H487835 ,H487909 ,H487983 ,H488114 ,H488196 ,H488275 ,H487227 ,H487383 ,H487500 ,H487577 ,H487708 ,H487836 ,H487910 ,H487984 ,H488115 ,H488197 ,H488281 ,H487228 ,H487384 ,H487501 ,H487579 ,H487709 ,H487837 ,H487912 ,H487991 ,H488116 ,H488198 ,H488282 ,H487232 ,H487386 ,H487502 ,H487581 ,H487710 ,H487838 ,H487913 ,H487992 ,H488117 ,H488199 ,H488283 ,H487241 ,H487387 ,H487503 ,H487582 ,H487711 ,H487839 ,H487919 ,H487993 ,H488118 ,H488200 ,H488286 ,H488287 ,H488363 ,H488448 ,H488537 ,H488621 ,H488678 ,H488758 ,H488862 ,H488933 ,H489028 ,H489101 ,H488288 ,H488364 ,H488449 ,H488538 ,H488622 ,H488679 ,H488763 ,H488863 ,H488934 ,H489029 ,H489102 ,H488291 ,H488365 ,H488450 ,H488539 ,H488624 ,H488680 ,H488764 ,H488864 ,H488935 ,H489030 ,H489103 ,H488295 ,H488366 ,H488452 ,H488541 ,H488625 ,H488681 ,H488766 ,H488865 ,H488937 ,H489031 ,H489104 ,H488296 ,H488373 ,H488453 ,H488549 ,H488626 ,H488682 ,H488769 ,H488866 ,H488938 ,H489032 ,H489105 ,H488297 ,H488376 ,H488454 ,H488550 ,H488627 ,H488683 ,H488770 ,H488867 ,H488940 ,H489033 ,H489106 ,H488298 ,H488377 ,H488455 ,H488551 ,H488628 ,H488684 ,H488771 ,H488868 ,H488941 ,H489034 ,H489109 ,H488299 ,H488378 ,H488456 ,H488552 ,H488629 ,H488685 ,H488772 ,H488872 ,H488943 ,H489037 ,H489111 ,H488300 ,H488379 ,H488459 ,H488553 ,H488630 ,H488686 ,H488773 ,H488873 ,H488944 ,H489038 ,H489112 ,H488301 ,H488380 ,H488460 ,H488558 ,H488631 ,H488687 ,H488775 ,H488874 ,H488945 ,H489039 ,H489113 ,H488302 ,H488381 ,H488461 ,H488559 ,H488633 ,H488688 ,H488776 ,H488875 ,H488951 ,H489040 ,H489114 ,H488303 ,H488382 ,H488469 ,H488560 ,H488636 ,H488689 ,H488779 ,H488876 ,H488953 ,H489041 ,H489121 ,H488304 ,H488383 ,H488470 ,H488561 ,H488637 ,H488690 ,H488780 ,H488877 ,H488955 ,H489042 ,H489125 ,H488305 ,H488384 ,H488471 ,H488564 ,H488638 ,H488691 ,H488782 ,H488878 ,H488956 ,H489043 ,H489126 ,H488306 ,H488385 ,H488472 ,H488565 ,H488639 ,H488692 ,H488783 ,H488879 ,H488957 ,H489046 ,H489127 ,H488307 ,H488387 ,H488473 ,H488569 ,H488640 ,H488693 ,H488784 ,H488880 ,H488958 ,H489047 ,H489128 ,H488308 ,H488388 ,H488474 ,H488570 ,H488641 ,H488694 ,H488785 ,H488881 ,H488959 ,H489053 ,H489130 ,H488309 ,H488389 ,H488475 ,H488573 ,H488642 ,H488695 ,H488786 ,H488882 ,H488960 ,H489054 ,H489131 ,H488310 ,H488390 ,H488476 ,H488574 ,H488643 ,H488696 ,H488787 ,H488883 ,H488961 ,H489055 ,H489132 ,H488311 ,H488391 ,H488477 ,H488575 ,H488644 ,H488699 ,H488788 ,H488884 ,H488962 ,H489056 ,H489133 ,H488312 ,H488394 ,H488478 ,H488576 ,H488645 ,H488703 ,H488792 ,H488885 ,H488963 ,H489060 ,H489134 ,H488313 ,H488397 ,H488481 ,H488577 ,H488646 ,H488704 ,H488793 ,H488886 ,H488969 ,H489062 ,H489136 ,H488316 ,H488399 ,H488483 ,H488578 ,H488647 ,H488705 ,H488795 ,H488887 ,H488972 ,H489063 ,H489138 ,H488320 ,H488401 ,H488488 ,H488579 ,H488648 ,H488706 ,H488796 ,H488888 ,H488973 ,H489064 ,H489139 ,H488321 ,H488403 ,H488489 ,H488585 ,H488649 ,H488707 ,H488798 ,H488889 ,H488981 ,H489065 ,H489149 ,H488322 ,H488404 ,H488490 ,H488586 ,H488650 ,H488708 ,H488799 ,H488890 ,H488982 ,H489066 ,H489150 ,H488323 ,H488408 ,H488491 ,H488587 ,H488651 ,H488709 ,H488801 ,H488891 ,H488983 ,H489067 ,H489151 ,H488324 ,H488409 ,H488492 ,H488588 ,H488652 ,H488710 ,H488804 ,H488892 ,H488985 ,H489068 ,H489152 ,H488325 ,H488412 ,H488493 ,H488589 ,H488653 ,H488711 ,H488805 ,H488893 ,H488986 ,H489069 ,H489153 ,H488326 ,H488413 ,H488494 ,H488590 ,H488654 ,H488714 ,H488809 ,H488894 ,H488987 ,H489070 ,H489154 ,H488327 ,H488414 ,H488495 ,H488591 ,H488655 ,H488715 ,H488811 ,H488895 ,H488988 ,H489071 ,H489155 ,H488328 ,H488415 ,H488496 ,H488592 ,H488656 ,H488720 ,H488820 ,H488896 ,H488991 ,H489072 ,H489162 ,H488329 ,H488416 ,H488497 ,H488593 ,H488657 ,H488722 ,H488821 ,H488897 ,H488992 ,H489073 ,H489163 ,H488330 ,H488417 ,H488498 ,H488594 ,H488658 ,H488724 ,H488822 ,H488899 ,H488993 ,H489075 ,H489164 ,H488331 ,H488418 ,H488499 ,H488595 ,H488659 ,H488729 ,H488826 ,H488900 ,H488994 ,H489076 ,H489166 ,H488332 ,H488419 ,H488507 ,H488596 ,H488660 ,H488730 ,H488827 ,H488901 ,H488995 ,H489077 ,H489167 ,H488336 ,H488420 ,H488512 ,H488597 ,H488661 ,H488731 ,H488828 ,H488902 ,H489000 ,H489078 ,H489168 ,H488337 ,H488421 ,H488513 ,H488598 ,H488662 ,H488732 ,H488830 ,H488903 ,H489001 ,H489080 ,H489169 ,H488350 ,H488422 ,H488514 ,H488599 ,H488663 ,H488736 ,H488831 ,H488906 ,H489003 ,H489081 ,H489170 ,H488351 ,H488424 ,H488515 ,H488600 ,H488664 ,H488738 ,H488832 ,H488912 ,H489004 ,H489082 ,H489181 ,H488352 ,H488425 ,H488520 ,H488601 ,H488665 ,H488739 ,H488833 ,H488920 ,H489005 ,H489083 ,H489184 ,H488353 ,H488426 ,H488521 ,H488602 ,H488666 ,H488740 ,H488850 ,H488921 ,H489006 ,H489084 ,H489185 ,H488354 ,H488427 ,H488525 ,H488603 ,H488667 ,H488741 ,H488851 ,H488922 ,H489009 ,H489085 ,H489186 ,H488355 ,H488431 ,H488526 ,H488612 ,H488668 ,H488742 ,H488852 ,H488923 ,H489017 ,H489086 ,H489187 ,H488356 ,H488438 ,H488527 ,H488613 ,H488669 ,H488743 ,H488853 ,H488924 ,H489018 ,H489087 ,H489188 ,H488357 ,H488439 ,H488528 ,H488614 ,H488672 ,H488744 ,H488854 ,H488925 ,H489019 ,H489089 ,H489192 ,H488358 ,H488440 ,H488529 ,H488615 ,H488673 ,H488746 ,H488855 ,H488926 ,H489020 ,H489091 ,H489193 ,H488359 ,H488441 ,H488530 ,H488616 ,H488674 ,H488747 ,H488857 ,H488929 ,H489024 ,H489092 ,H489195 ,H488360 ,H488442 ,H488532 ,H488617 ,H488675 ,H488749 ,H488858 ,H488930 ,H489025 ,H489093 ,H489196 ,H488361 ,H488446 ,H488533 ,H488618 ,H488676 ,H488756 ,H488859 ,H488931 ,H489026 ,H489094 ,H489197 ,H488362 ,H488447 ,H488534 ,H488619 ,H488677 ,H488757 ,H488861 ,H488932 ,H489027 ,H489100 ,H489199 ,H489200 ,H489303 ,H489357 ,H489419 ,H489487 ,H489555 ,H489637 ,H489709 ,H489803 ,H489883 ,H48A000 ,H489201 ,H489304 ,H489358 ,H489420 ,H489488 ,H489556 ,H489638 ,H489710 ,H489804 ,H489884 ,H48A024 ,H489203 ,H489305 ,H489359 ,H489421 ,H489489 ,H489557 ,H489646 ,H489711 ,H489805 ,H489886 ,H48A025 ,H489204 ,H489306 ,H489360 ,H489422 ,H489490 ,H489558 ,H489647 ,H489712 ,H489806 ,H489890 ,H48A026 ,H489205 ,H489307 ,H489361 ,H489423 ,H489491 ,H489559 ,H489648 ,H489713 ,H489807 ,H489891 ,H48A027 ,H489206 ,H489308 ,H489362 ,H489424 ,H489492 ,H489560 ,H489657 ,H489714 ,H489808 ,H489898 ,H48A028 ,H489207 ,H489309 ,H489363 ,H489434 ,H489493 ,H489563 ,H489658 ,H489715 ,H489809 ,H489899 ,H48A032 ,H489208 ,H489310 ,H489364 ,H489435 ,H489494 ,H489565 ,H489659 ,H489716 ,H489810 ,H489911 ,H48A033 ,H489209 ,H489311 ,H489365 ,H489436 ,H489495 ,H489566 ,H489660 ,H489718 ,H489812 ,H489912 ,H48A034 ,H489210 ,H489312 ,H489366 ,H489437 ,H489496 ,H489568 ,H489661 ,H489719 ,H489813 ,H489913 ,H48A035 ,H489211 ,H489313 ,H489367 ,H489438 ,H489497 ,H489569 ,H489662 ,H489720 ,H489814 ,H489916 ,H48A036 ,H489212 ,H489314 ,H489368 ,H489439 ,H489498 ,H489570 ,H489664 ,H489724 ,H489815 ,H489917 ,H48A037 ,H489213 ,H489315 ,H489369 ,H489442 ,H489499 ,H489571 ,H489665 ,H489725 ,H489818 ,H489919 ,H48A039 ,H489214 ,H489316 ,H489370 ,H489443 ,H489506 ,H489573 ,H489666 ,H489729 ,H489819 ,H489920 ,H48A041 ,H489215 ,H489317 ,H489371 ,H489444 ,H489507 ,H489574 ,H489667 ,H489730 ,H489825 ,H489927 ,H48A042 ,H489216 ,H489318 ,H489376 ,H489445 ,H489508 ,H489575 ,H489668 ,H489731 ,H489826 ,H489933 ,H48A043 ,H489217 ,H489319 ,H489377 ,H489446 ,H489510 ,H489576 ,H489669 ,H489732 ,H489831 ,H489934 ,H48A047 ,H489219 ,H489320 ,H489378 ,H489447 ,H489511 ,H489579 ,H489670 ,H489733 ,H489834 ,H489935 ,H48A048 ,H489220 ,H489321 ,H489380 ,H489452 ,H489515 ,H489580 ,H489671 ,H489734 ,H489835 ,H489936 ,H48A049 ,H489221 ,H489322 ,H489381 ,H489453 ,H489516 ,H489581 ,H489672 ,H489735 ,H489836 ,H489937 ,H48A050 ,H489222 ,H489323 ,H489382 ,H489454 ,H489517 ,H489582 ,H489673 ,H489736 ,H489844 ,H489938 ,H48A051 ,H489228 ,H489324 ,H489383 ,H489455 ,H489518 ,H489583 ,H489674 ,H489737 ,H489845 ,H489939 ,H48A054 ,H489230 ,H489325 ,H489384 ,H489456 ,H489519 ,H489584 ,H489675 ,H489740 ,H489846 ,H489940 ,H48A055 ,H489231 ,H489326 ,H489385 ,H489457 ,H489520 ,H489585 ,H489676 ,H489741 ,H489847 ,H489941 ,H48A057 ,H489232 ,H489327 ,H489387 ,H489458 ,H489521 ,H489586 ,H489677 ,H489742 ,H489850 ,H489942 ,H48A058 ,H489243 ,H489328 ,H489389 ,H489459 ,H489522 ,H489588 ,H489678 ,H489744 ,H489851 ,H489948 ,H48A059 ,H489244 ,H489329 ,H489390 ,H489460 ,H489523 ,H489589 ,H489679 ,H489745 ,H489856 ,H489949 ,H48A060 ,H489245 ,H489330 ,H489391 ,H489461 ,H489524 ,H489590 ,H489681 ,H489746 ,H489857 ,H489952 ,H48A062 ,H489246 ,H489331 ,H489392 ,H489462 ,H489525 ,H489591 ,H489682 ,H489747 ,H489858 ,H489954 ,H48A063 ,H489251 ,H489332 ,H489393 ,H489463 ,H489526 ,H489592 ,H489683 ,H489754 ,H489859 ,H489956 ,H48A064 ,H489252 ,H489333 ,H489394 ,H489464 ,H489527 ,H489593 ,H489684 ,H489755 ,H489860 ,H489957 ,H48A065 ,H489253 ,H489334 ,H489395 ,H489465 ,H489530 ,H489594 ,H489685 ,H489756 ,H489861 ,H489958 ,H48A066 ,H489254 ,H489335 ,H489396 ,H489466 ,H489531 ,H489595 ,H489686 ,H489757 ,H489862 ,H489959 ,H48A068 ,H489255 ,H489336 ,H489397 ,H489467 ,H489532 ,H489598 ,H489688 ,H489758 ,H489864 ,H489961 ,H48A069 ,H489256 ,H489337 ,H489398 ,H489468 ,H489533 ,H489600 ,H489689 ,H489760 ,H489865 ,H489962 ,H48A070 ,H489260 ,H489338 ,H489399 ,H489469 ,H489534 ,H489601 ,H489691 ,H489767 ,H489866 ,H489963 ,H48A074 ,H489261 ,H489340 ,H489401 ,H489470 ,H489535 ,H489602 ,H489692 ,H489768 ,H489867 ,H489964 ,H48A075 ,H489262 ,H489341 ,H489402 ,H489471 ,H489536 ,H489603 ,H489694 ,H489780 ,H489868 ,H489965 ,H48A076 ,H489269 ,H489342 ,H489403 ,H489472 ,H489537 ,H489604 ,H489695 ,H489781 ,H489869 ,H489966 ,H48A077 ,H489270 ,H489343 ,H489404 ,H489473 ,H489538 ,H489605 ,H489696 ,H489782 ,H489870 ,H489970 ,H48A078 ,H489290 ,H489344 ,H489405 ,H489474 ,H489539 ,H489616 ,H489697 ,H489783 ,H489871 ,H489971 ,H48A079 ,H489291 ,H489346 ,H489406 ,H489475 ,H489540 ,H489617 ,H489698 ,H489784 ,H489872 ,H489972 ,H48A080 ,H489292 ,H489347 ,H489407 ,H489476 ,H489541 ,H489619 ,H489699 ,H489785 ,H489873 ,H489973 ,H48A081 ,H489293 ,H489348 ,H489408 ,H489478 ,H489542 ,H489620 ,H489700 ,H489788 ,H489874 ,H489974 ,H48A082 ,H489296 ,H489349 ,H489409 ,H489479 ,H489543 ,H489621 ,H489701 ,H489789 ,H489875 ,H489975 ,H48A106 ,H489297 ,H489350 ,H489410 ,H489480 ,H489544 ,H489629 ,H489702 ,H489790 ,H489876 ,H489976 ,H48A107 ,H489298 ,H489351 ,H489411 ,H489482 ,H489545 ,H489630 ,H489703 ,H489798 ,H489877 ,H489977 ,H48A108 ,H489299 ,H489352 ,H489412 ,H489483 ,H489546 ,H489633 ,H489704 ,H489799 ,H489878 ,H489981 ,H48A111 ,H489300 ,H489353 ,H489413 ,H489484 ,H489552 ,H489634 ,H489706 ,H489800 ,H489879 ,H489982 ,H48A112 ,H489301 ,H489354 ,H489414 ,H489485 ,H489553 ,H489635 ,H489707 ,H489801 ,H489880 ,H489990 ,H48A113 ,H489302 ,H489356 ,H489418 ,H489486 ,H489554 ,H489636 ,H489708 ,H489802 ,H489881 ,H489997 ,H48A114 ,H48A115 ,H48A358 ,H48A554 ,H48A888 ,H48B390 ,H48B791 ,H48A116 ,H48A360 ,H48A557 ,H48A896 ,H48B391 ,H48B804 ,H48A117 ,H48A362 ,H48A587 ,H48A912 ,H48B392 ,H48B805 ,H48A118 ,H48A363 ,H48A606 ,H48A913 ,H48B417 ,H48B806 ,H48A119 ,H48A364 ,H48A640 ,H48A914 ,H48B418 ,H48B807 ,H48A125 ,H48A365 ,H48A641 ,H48A915 ,H48B423 ,H48B808 ,H48A126 ,H48A366 ,H48A642 ,H48A916 ,H48B426 ,H48B809 ,H48A127 ,H48A367 ,H48A643 ,H48A917 ,H48B451 ,H48B810 ,H48A128 ,H48A374 ,H48A644 ,H48A960 ,H48B481 ,H48B817 ,H48A136 ,H48A375 ,H48A648 ,H48A961 ,H48B482 ,H48B820 ,H48A160 ,H48A376 ,H48A649 ,H48A962 ,H48B570 ,H48B843 ,H48A165 ,H48A377 ,H48A650 ,H48A972 ,H48B571 ,H48B900 ,H48A166 ,H48A378 ,H48A651 ,H48A988 ,H48B572 ,H48B937 ,H48A167 ,H48A379 ,H48A669 ,H48A991 ,H48B575 ,H48B938 ,H48A168 ,H48A380 ,H48A670 ,H48A992 ,H48B576 ,H48B939 ,H48A169 ,H48A381 ,H48A671 ,H48B007 ,H48B577 ,H48C018 ,H48A177 ,H48A382 ,H48A675 ,H48B008 ,H48B584 ,H48C022 ,H48A180 ,H48A383 ,H48A682 ,H48B009 ,H48B585 ,H48A181 ,H48A384 ,H48A685 ,H48B010 ,H48B619 ,H48A185 ,H48A385 ,H48A686 ,H48B011 ,H48B651 ,H48A194 ,H48A386 ,H48A687 ,H48B030 ,H48B664 ,H48A195 ,H48A387 ,H48A688 ,H48B031 ,H48B668 ,H48A204 ,H48A391 ,H48A689 ,H48B032 ,H48B669 ,H48A205 ,H48A392 ,H48A692 ,H48B033 ,H48B670 ,H48A209 ,H48A393 ,H48A693 ,H48B034 ,H48B671 ,H48A210 ,H48A406 ,H48A696 ,H48B043 ,H48B672 ,H48A213 ,H48A410 ,H48A697 ,H48B088 ,H48B673 ,H48A214 ,H48A411 ,H48A711 ,H48B091 ,H48B674 ,H48A215 ,H48A412 ,H48A760 ,H48B209 ,H48B675 ,H48A217 ,H48A413 ,H48A761 ,H48B210 ,H48B676 ,H48A231 ,H48A414 ,H48A775 ,H48B213 ,H48B677 ,H48A246 ,H48A422 ,H48A776 ,H48B214 ,H48B678 ,H48A253 ,H48A442 ,H48A777 ,H48B215 ,H48B679 ,H48A256 ,H48A443 ,H48A783 ,H48B218 ,H48B680 ,H48A257 ,H48A448 ,H48A784 ,H48B221 ,H48B681 ,H48A258 ,H48A454 ,H48A785 ,H48B222 ,H48B682 ,H48A267 ,H48A455 ,H48A794 ,H48B243 ,H48B683 ,H48A268 ,H48A456 ,H48A811 ,H48B244 ,H48B684 ,H48A269 ,H48A470 ,H48A815 ,H48B245 ,H48B685 ,H48A287 ,H48A471 ,H48A823 ,H48B246 ,H48B686 ,H48A288 ,H48A472 ,H48A824 ,H48B248 ,H48B687 ,H48A291 ,H48A480 ,H48A825 ,H48B249 ,H48B688 ,H48A292 ,H48A491 ,H48A837 ,H48B257 ,H48B689 ,H48A293 ,H48A492 ,H48A843 ,H48B258 ,H48B690 ,H48A294 ,H48A493 ,H48A844 ,H48B259 ,H48B762 ,H48A305 ,H48A494 ,H48A853 ,H48B260 ,H48B763 ,H48A306 ,H48A501 ,H48A863 ,H48B261 ,H48B764 ,H48A316 ,H48A532 ,H48A864 ,H48B288 ,H48B765 ,H48A322 ,H48A533 ,H48A865 ,H48B365 ,H48B788 ,H48A356 ,H48A535 ,H48A869 ,H48B376 ,H48B789 ,H48A357 ,H48A543 ,H48A870 ,H48B389 ,H48B790 ***12-20-12, AMENDED INFO ***: H488429, H489098, H485912, H485692 |
Recalling Firm/ Manufacturer |
Varian Medical Systems, Inc. Oncology Systems 911 Hansen Way Palo Alto CA 94304-1028
|
For Additional Information Contact |
Michael Pignataro 650-424-5731
|
Manufacturer Reason for Recall |
An anomaly in Eclipse Treatment Planning System, versions 8.0 to 8.6, whereby Actual SSD (source surface distance) value may not correspond to the Planned SSD value after Planning Approval.
|
FDA Determined Cause 2 |
Software design |
Action |
The firm, Varian Medical, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated April 4, 2011, to its customers.
***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.***
The letter describes the product, problem and actions to be taken. The customers were instructed to do the following:
1) Set the Planning Approved status in Eclipse (Eclipse displays both the Actual and Planned SSD values during plan approval) and verify both SSD values prior to approving the plan.
2)Verify the SSD values in the Eclipse Treatment Plan Report prior to approving the plan for treatment or proceeding with treatment delivery (this report lists both the Actual and Planned SSD values).
3) Prior to approving the plan in RT Chart, verify both the Actual and Planned SSD values in the Parameters Workspace by selecting Field Parameters>Geometry tab.
4) Prior to treatment, they should verify intended SSD values during independent MU calculation.
5)Re-familiarize themselves with the concepts of Actual SSD and Planned SSD by reviewing the listed information.
Please advise the appropriate personnel working in your radiotherapy department of the content of this letter.
Varian mitigated the product behavior in version 8.9.
If you require further clarification, please feel free to contact the Varian Oncology Help Desk ; phone: USA and Canada - 1888-827-4265; Europe: +41 41 749 8844; email: North America: support-americas@varian.com; Australia/New Zealand: support-anz@varian.com; Europe: support-emea@varian.com; South East Asia: seasia.apps.helpdesk@varian.com; China/Asia: china.apps.helpdesk@varian.com; Japan: Japan.Apps.Helpdesk@varian.com; Latin America: soporte.al@varian.com; Internet: Oncology Systems-www.myvarian.com; Varian Medical Systems -www.varian.com; or your local Varian Customer Support District or Regional Manager. |
Quantity in Commerce |
5288 units |
Distribution |
Worldwide distribution. |
Total Product Life Cycle |
TPLC Device Report
|
|
1 A record in this database is created when a firm initiates a correction or removal action. The record is updated if the FDA identifies a violation and classifies the action as a recall, and it is updated for a final time when the recall is terminated. Learn more about medical device recalls.
2 Per FDA policy, recall cause determinations are subject to modification up to the point of termination of the recall.
3 For details about termination of a recall see Code of Federal Regulations (CFR) Title 21 §7.55.
|
510(K) Database |
510(K)s with Product Code = MUJ and Original Applicant = VARIAN MEDICAL SYSTEMS, INC.
|
|
|
|