• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Recognition List Number: 031 FR Publication Date: 08/05/2013 

Part B: Supplementary Information

Recognition Number 11-215: ASTM F897-02 (Reapproved 2013), standard test method for measuring fretting corrosion of osteosynthesis plates and screws. (Orthopedic )

Date of Standard: 2002. 
Address of Standards Development Organization:
 ASTM International (ASTM)
  100 Barr Harbor Drive
  West Conshohocken, PA 19428-2959
 
CDRH Office and Division associated with recognized standards:
OFFICE OF DEVICE EVALUATION (ODE)
Devices Affected:
Applies specifically to osteosynthesis plates and screws, but may also be used to study fretting corrosion of other devices where metal components are in contact.
Processes Affected:
510(K), PMA, IDE, PDP, DESIGN CONTROLS
Type of Standard:
 Test Methods, National, Horizontal
Extent of Recognition:
Complete standard
Related CFR Citations and Product Codes:
Regulation
Number
Device Name Device
Class
Product
Code
Method applies specifically to osteosynthesis plates and screws indicated by the following procodes. However, the method may be used to analyze susceptibility to fretting corrosion of material combinations in other orthopaedic device applications.
§888.3030 Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component Class 2 KTT
§888.3030 Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component, Metal Composite Class 2 LXT
§888.3030 Appliance, Fixation, Nail/Blade/Plate Combination, Single Component Class 2 KTW
§888.3030 Appliance, Nail/Blade/Plate Combination, Single Component Class 2 KWK
§888.3030 Condylar Plate Fixation Implant Class 2 JDP
§888.3030 Device, Fixation, Proximal Femoral, Implant Class 2 JDO
§888.3030 Plate, Fixation, Bone Class 2 HRS
§888.3040 Screw, Fixation, Bone Class 2 HWC
Relevant Guidance:
Guidance Document for Testing Non-Articulating, 'Mechanically Locked,' Modular Implant Components, May 1, 1995
FDA Technical Contact:
 Terry O. Woods
  FDA/OMPT/CDRH/OSEL
  10903 New Hampshire Avenue WO62 RM2116
  Silver Spring MD 20993
  301/796-2503
  email: terry.woods@fda.hhs.gov
-
-