• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Recognition List Number: 031 FR Publication Date: 08/05/2013 

Part B: Supplementary Information

Recognition Number 11-253: ASTM F1800-12, standard test method for cyclic fatigue testing of metal tibial tray components of total knee joint replacements. (Orthopedic )

Date of Standard: 2012. 
Address of Standards Development Organization:
 ASTM International (ASTM)
  100 Barr Harbor Drive
  West Conshohocken, PA 19428-2959
 
CDRH Office and Division associated with recognized standards:
OFFICE OF DEVICE EVALUATION (ODE)
Devices Affected:
Metallic tibial trays used in total knee replacements (TKR)
Processes Affected:
510(K), HDE, IDE, PMA, PDP, Design Controls
Type of Standard:
 Test Methods, National, Vertical
Extent of Recognition:
Complete Standard
Related CFR Citations and Product Codes:
Regulation
Number
Device Name Device
Class
Product
Code
Unclassified Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Polymer/Metal/Polymer Class 3 LXY
Unclassified Prosthesis, Knee, Patello/Femorotibial, Unconstrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer Class 3 MBD
§888.3480 Metal Cemented Constrained Femorotibial Knee Prosthesis Class 3 KRN
§888.3510 Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer Class 2 KRO
§888.3520 Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented, Metal/Polymer Class 2 HSX
§888.3530 Prosthesis, Knee, Femorotibial, Semi-Constrained, Cemented, Metal/Polymer Class 2 HRY
§888.3550 Prosthesis, Knee, Patello/Femorotibial, Constrained, Cemented, Polymer/Metal/Metal Class 3 KRP
§888.3550 Prosthesis, Knee, Patello/Femorotibial, Constrained, Cemented, Polymer/Metal/Polymer Class 2 KRQ
§888.3560 Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uhmwpe, Pegged, Cemented, Polymer/Metal/Polymer Class 2 MBV
§888.3560 Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer Class 2 JWH
§888.3565 Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer Class 2 MBH
Relevant Guidance:
Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA, January 16, 2003
FDA Technical Contact:
-
-