• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Recognition List Number: 034 FR Publication Date: 01/30/2014 

Part B: Supplementary Information

Recognition Number 9-68: ISO 23409 First edition 2011-02-15, male condoms - requirements and test methods for condoms made from synthetic materials. (ObGyn/Gastroenterology/Urology)

Date of Standard: 2011. 
Address of Standards Development Organization:
 International Organization for Standardization (ISO)*
  1, Rue de Varembe
  Case Postale 56
  CH 1211 Geneva 20, 0
  SWITZERLAND
CDRH Offices and Divisions associated with recognized standards:
(1)OFFICE OF DEVICE EVALUATION (ODE)
DIVISION OF REPRODUCTIVE, GASTRORENAL AND UROLOGY DEVICES (DRGUD)
(2)OFFICE OF SCIENCE AND ENGINEERING LABORATORIES (OSEL)
(3)OFFICE OF COMPLIANCE (OC)
Devices Affected:
Male Condoms made from Synthetic Materials
Processes Affected:
510(k), IDE, Design Control Input, QSR
Type of Standard:
 Vertical, International
Extent of Recognition:
Complete Standard, with the following comments and exceptions:

Annex J.3 Testing for holes, Electrical test is not recognized.
Related CFR Citations and Product Codes:
Regulation
Number
Device Name Device
Class
Product
Code
§884.5300 Condom, Synthetic Class 2 MOL
Relevant Guidance:
Testing Guidance for Male Condoms made from New Material (Non-Latex), DRAFT: June 29, 1995.

http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm080618.htm

Compliance Policy Guide and FDA Test Methods (Sub Chapter 345 Obstetrics and Gynecology (Sec. 345.100 Condoms; Defects - Criteria for Direct Reference Seizure (CPG 7124.21) (revised December 22, 2009)

Guidance for Industry and FDA Staff: Surveillance and Detention without Physical Examination of Condoms Document (issued on: July 11, 2008)

http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/ucm070731.htm
FDA Technical Contact:
 Sharon Andrews
  FDA/OMPT/CDRH/ODE
  10903 New Hampshire Avenue WO66, RMG110
  Silver Spring MD 20993
  301/796-6529
  email: sharon.andrews@fda.hhs.gov
* in the United States, copies of this standard can be obtained from:
  American National Standards Institute (ANSI)*
  25 West 43rd Street
  4th Floor
  New York, NY 10036
-
-