Recognition List Number: 034 FR Publication Date: 01/30/2014 

Part B: Supplementary Information

Recognition Number 16-185: ANSI / RESNA WC-2:2009, american national standard for wheelchairs - volume 2, additional requirements for wheelchairs (including scooters) with electrical systems section 21: requirements and test methods for electromagnetic compatibility of electrically powered wheelchairs and motorized scooters. (Physical Medicine)

Date of Standard: 2009. 
Addresses of Standards Development Organizations:
 American National Standards Institute (ANSI)
  25 West 43rd Street
  4th Floor
  New York, NY 10036
 
 Rehabilitation Engineering and Assistive Technology Society of North America (RESNA)
  1700 N. Moore Street
  Suite 1540
  Arlington, VA 22209-1903
 
CDRH Office and Division associated with recognized standards:
OFFICE OF DEVICE EVALUATION (ODE)
Devices Affected:
Powered wheelchairs and motorized scooters
Processes Affected:
510(k), PMA, IDE, PDP, HDE, Design Controls
Type of Standard:
 Vertical, National
Extent of Recognition:
Complete standard with the following exceptions:

Clause 9.2.1 is not recognized
Related CFR Citations and Product Codes:
Regulation
Number
Device Name Device
Class
Product
Code
§890.3800 Vehicle, Motorized 3-Wheeled Class 2 INI
§890.3860 Wheelchair, Powered Class 2 ITI
§890.3880 Wheelchair, Special Grade Class 2 IQC
§890.3890 Wheelchair, Stair Climbing Class 2 IMK
§890.3900 Wheelchair, Standup Class 2 IPL
§890.3930 Elevator, Wheelchair, Portable Class 2 ING
Relevant Guidance:
"Guidance Document for the Preparation Of Premarket Notification [510(k)] Applications for Mechanical and Powered Wheelchairs And Motorized Three Wheeled Vehicles" July 26, 1995 (reformatted 12/18/97).

"Letter To Powered Wheelchair/Scooter Or Accessory/Component Manufacturers, May 26, 1994"

Additional general information on medical device EMC can be found in guidance documents on EMC in the following device areas: Respiratory and Anesthesia Devices, Ventilators, Apnea Monitors, Powered Wheelchairs and Motorized Scooters, Cardiac Pacemakers, and Implantable Defibrillators. These and any other guidance documents with additional EMC requirements should be utilized.
FDA Technical Contact:
 John Marszalek
  OMPT/CDRH/OC/DICO/FEB
  10903 New Hampshire Avenue WO66 RM3661
  Silver Spring MD 20993
  301/796-7067
  email: john.marszalek@fda.hhs.gov
-
-