Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 047 Date of Entry 08/21/2017 
FR Recognition Number 6-400
Standard
ASTM  F3212-16(2023)
Standard Test Method for Coring Testing of Huber Needles
Scope/Abstract
1.1 This test method covers the qualitative measurement of Huber-type needles' potential to remove septum material during implantable port access (1)2.
1.2 This test method does not address other issues that may include, but are not limited to, force measurement during the perforation/withdrawal, septum integrity, and any safety issues.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§880.5570 Needle, Hypodermic, Single Lumen Class 2 FMI
Relevant FDA Guidance and/or Supportive Publications*
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Oleg Vesnovsky
  FDA/OC/CDRH/OSEL/DAM/
  301-796-2527
  oleg.vesnovsky@fda.hhs.gov
Standards Development Organization
ASTM ASTM International http://www.astm.org/
FDA Specialty Task Group (STG)
General Plastic Surgery/General Hospital