Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 050 Date of Entry 09/17/2018 
FR Recognition Number 8-479
Standard
ISO  13781 Second edition 2017-07
Implants for surgery - Homopolymers, copolymers and blends on poly(lactide) - In vitro degradation testing
Scope/Abstract
This document describes methods for the determination of chemical and mechanical changes in poly(lactide)-based homopolymers, copolymers and/or blends induced under in vitro degradation testing conditions. This document covers polymers based on L-lactide, D-lactide, and/or D, L-lactide monomeric units.
The purpose of this document is to compare and/or evaluate materials or processing conditions. This document also describes the fundamental physical and mechanical evaluations needed for an in vitro degradation characterization of an absorbable poly(lactide) or other hydrolysable material or device.
This document is applicable to poly(lactide)-based homopolymers, copolymers and/or blends in bulk or processed forms and used for the manufacture of surgical implants, including finished products (packaged and sterilized implants).
The test methods specified in this document are also intended to determine the in vitro degradation rate and related changes in material properties of polylactide-based copolymers and/or blends with various other comonomers, such as glycolid, trimethylene, carbonate and/or ε-caprolactone. Unless otherwise validated for a specific device, these in vitro methods cannot be used to definitively predict device behaviour under in vivo conditions.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies
Relevant FDA Guidance and/or Supportive Publications*
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 John Goode
  FDA/OC/CDRH/OPEQ/OHTVI/DHTVIA/
  301-796-6407
  john.goode@fda.hhs.gov
Standards Development Organization
ISO International Organization for Standardization https://www.iso.org/
FDA Specialty Task Group (STG)
Materials
*These are provided as examples and others may be applicable.
-
-