Part B: Supplementary Information Sheet (SIS) |
FR Recognition List Number
|
052
|
Date of Entry 07/15/2019
|
FR Recognition Number
|
8-516
|
Standard | |
ISO 13779-4 Second edition 2018-12 Implants for surgery - Hydroxyapatite - Part 4: Determination of coating adhesion strength |
|
Scope/AbstractThis document specifies a test method for measurement of the adhesion strength of hydroxyapatite coatings intended for use on metallic-substrate components of surgical implants.
NOTE Requirements for the competence of testing laboratories can be found in ISO/IEC 17025.
|
|
Extent of Recognition
|
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies. |
|
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number |
Device Name |
Device Class |
Product Code |
§888.3045 |
Filler, Bone Void, Osteoinduction (W/O Human Growth Factor)
|
Class 2
|
MBP
|
§888.3045 |
Filler, Bone Void, Calcium Compound
|
Class 2
|
MQV
|
§888.3045 |
Calcium Salt Bone Void Filler, Drillable, Non-Screw Augmentation
|
Class 2
|
OIS
|
§888.3353 |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
|
Class 2
|
LZO
|
§888.3353 |
Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous Cemented, Osteophilic Finish
|
Class 2
|
MAY
|
§888.3353 |
Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
|
Class 2
|
MEH
|
§888.3353 |
Hip, Semi-Constrained, Cemented, Metal/Ceramic/Polymer + Additive, Porous Uncemented
|
Class 2
|
OQI
|
§888.3358 |
Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
|
Class 2
|
LPH
|
§888.3358 |
Prosthesis, Hip, Semi-Constrained, Uncemented, Metal/Polymer, Porous
|
Class 2
|
MBL
|
§888.3358 |
Hip Prosthesis, Semi-Constrained, Cemented, Metal/Polymer, + Additive, Porous, Uncemented
|
Class 2
|
OQG
|
§888.3535 |
Prosthesis, Knee, Femorotibial, Unicompartmental/Unicondylar, Uncemented, Porous-Coated, Metal/Polymer
|
Class 2
|
NJD
|
§888.3565 |
Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
|
Class 2
|
MBH
|
§888.3670 |
Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer, Uncemented
|
Class 2
|
MBF
|
|
Relevant FDA Guidance and/or Supportive Publications*
Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement, Issued April 1994
510(k) Information Needed for Hydroxyapatite Coated Orthopedic Implants, Issued March 1995 (revised 2/20/97)
Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments, Issued May 2004
Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device, Issued June 2003
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018. |
|
FDA Technical Contact
|
Standards Development Organization
|
FDA Specialty Task Group (STG)
|
*These are provided as examples and others may be applicable. |
|
|