• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

Recognized Consensus Standards: Medical Devices

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

New Search Back To Search Results
Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 056 Date of Entry 06/07/2021 
FR Recognition Number 19-40
Standard
IEC  60086-4 Edition 5.0 2019-04
Primary batteries - Part 4: Safety of lithium batteries [Including: Corrigendum 1 (2019) and Corrigendum 2 (2020)]
Scope/Abstract
This Part of IEC 60086 specifies tests and requirements for primary lithium batteries to ensure their safe operation under intended use and reasonably foreseeable misuse.

NOTE Primary lithium batteries that are standardized in IEC 60086-2 are expected to meet all applicable requirements herein. It is understood that consideration of this part of IEC 60086 might also be given to measuring and/or ensuring the safety of non-standardized primary lithium batteries. In either case, no claim or warranty is made that compliance or non-compliance with this standard will fulfil or not fulfil any of the user's particular purposes or needs.
Extent of Recognition
Complete standard
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.
Relevant FDA Guidance and/or Supportive Publications*
Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contacts
 Iacovos Kyprianou
  FDA/OC/CDRH/OSPTI/DAHRSSP/SSPI/
  301-796-2601
  iacovos.kyprianou@fda.hhs.gov
 Hamed Ghods
  FDA/OC/CDRH/OSEL/DBP/
  301-796-2656
  hamed.ghods@fda.hhs.gov
Standards Development Organization
IEC International Electrotechnical Commission http://www.iec.ch/
FDA Specialty Task Group (STG)
General II (ES/EMC)
*These are provided as examples and others may be applicable.
-
-