Part B: Supplementary Information Sheet (SIS)
FR Recognition List Number 060 Date of Entry 05/29/2023 
FR Recognition Number 6-487
Standard
ANSI AAMI  PB70:2022
Liquid barrier performance and classification of protective apparel and drapes intended for use in health care facilities
Scope/Abstract
Establishes minimum barrier performance requirements, a classification system, and associated labeling requirements for protective apparel, surgical drapes, and drape accessories intended for use in health care facilities.
Extent of Recognition
Partial recognition. The following part(s) of the standard is (are) not recognized:
3 Definitions, Procedure gown
4.1 Labeling
Annex B (Informative) Examples of barrier performance classification of surgical gowns, surgical gowns-E, isolation gowns, other protective gowns, surgical drapes, hoods, togas, and other protective apparel
Rationale for Recognition
This standard is relevant to medical devices and is recognized on its scientific and technical merit and/or because it supports existing regulatory policies.

This standard is recognized in part because:
3 Definitions "Procedure gown" is in conflict with existing regulations, 21 CFR 878.4040 and the guidance listed below (Ref (2)). FDA considers procedure gowns that claim moderate to high barrier penetration to be "surgical gowns" within the meaning of 21 CFR 878.4040.
4.1 Labeling is in conflict with 21 CFR 810.5(a) and final FDA Guidance (Ref (2)), which states that the level of liquid barrier protection must be clearly identified in labeling.
Annex B (informative), Examples of barrier performance classification of surgical gowns, surgical gowns-E, isolation gowns, other protective gowns, surgical drapes, hoods, togas, and other protective apparel", Figure B.2, "Example of a surgical gown-E intended for surgical applications" is in conflict with the published FDA Medical Gown website (Ref (3)), Surgical Gowns, Figure 1,"Critical Zones for Surgical Gowns."
Transition Period
FDA recognition of ANSI, AAMI PB70:2012 [Rec# 6-296] will be superseded by recognition of ANSI, AAMI PB70:2022 [Rec# 6-487]. FDA will accept declarations of conformity, in support of premarket submissions, to [Rec# 6-296] until July 6, 2025. After this transition period, declarations of conformity to [Rec# 6-296] will not be accepted.
Public Law, CFR Citation(s) and Procode(s)*
Regulation Number Device Name Device Class Product Code
§878.4040 Gown, Surgical Class 2 FYA
§878.4040 Gown, Isolation, Surgical Class 2 FYC
§878.4040 Non-Surgical Isolation Gown Class 1 OEA
§878.4040 Gown, Non-Sterile, Non-Isolation, Intended To Provide Moderate Or High Barrier Protection Class 2 QPC
§878.4040 Medical Gowns With Chemotherapy Labeling Claims - Tested For Use With Chemotherapy Drugs Class 2 QSO
§878.4370 Pad, Kelly Class 2 FNW
§878.4370 Drape, Patient, Ophthalmic Class 2 HMT
§878.4370 Drape, Microscope, Ophthalmic Class 2 HMW
§878.4370 Ring (Wound Protector), Drape Retention, Internal Class 2 KGW
§878.4370 Drape, Surgical Class 2 KKX
§878.4370 General Surgery Tray Class 2 LRO
§878.4370 Drape, Surgical, Ent Class 2 ERY
§878.4370 Drape, Pure Latex Sheet, With Self-Retaining Finger Cot Class 2 EYX
§878.4370 Drape, Urological, Disposable Class 2 EYY
§878.4380 Drape, Adhesive, Aerosol Class 1 KGT
Relevant FDA Guidance and/or Supportive Publications*
(1) Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes, issued August 1993.

(2) Premarket Notification Requirements Concerning Gowns Intended for Use in Health Care Settings, Guidance for Industry and Food and Drug Administration Staff, issued December 2015.

(3) Medical Gowns. https://www.fda.gov/medical-devices/personal-protective-equipment-infection-control/medical-gowns.

Appropriate Use of Voluntary Consensus Standards in Premarket Submissions for Medical Devices - Guidance for Industry and Food and Drug Administration Staff, issued September 2018.
FDA Technical Contact
 Yongqing Chen
  FDA/OC/CDRH/OPEQ/OHTIV/DHTIVB
  240-402-9433
  Yongqing.Chen@fda.hhs.gov
Standards Development Organizations
ANSI American National Standards Institute https://www.ansi.org/
AAMI Association for the Advancement of Medical Instrumentation http://www.aami.org
FDA Specialty Task Group (STG)
General Plastic Surgery/General Hospital
*These are provided as examples and others may be applicable.
-
-