MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR; IMMUNOASSAY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR; IMMUNOASSAY ANALYZER Back to Search Results
Model Number ADVIA CENTAUR
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/15/2015
Event Type  malfunction  
Event Description
A discordant, falsely low cancer antigen 19-9 (ca 19-9) result was obtained on one patient sample on an advia centaur instrument.The discordant result was reported to the physician(s), who questioned it.The sample was diluted with a 1:10 dilution factor and was tested on the same instrument, resulting higher.The corrected result was reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely low ca 19-9 result.
 
Manufacturer Narrative
A siemens customer service engineer (cse) specialist was dispatched to the customer site.After evaluating the instrument, the cse performed a total service call but could not find an instrument malfunction.The quality controls were within the acceptable ranges.The cause of the discordant, falsely low ca 19-9 result on one patient sample is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Manufacturer Narrative
The initial mdr 2432235-2015-00319 was filed on july 8, 2015.A siemens customer service engineer (cse) specialist was dispatched to the customer site.After evaluating the instrument, the cse performed a total service call but could not find an instrument malfunction.The quality controls were within the acceptable ranges.The cause of the discordant, falsely low ca 27.29 result on one patient sample is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.Corrected information (10/18/2016): the initial mdr states the method name as cancer antigen 19-9.The correct method name is cancer antigen 27.29.
 
Event Description
A discordant, falsely low cancer antigen 27.29 (ca 27.29) result was obtained on one patient sample on an advia centaur instrument.The discordant result was reported to the physician(s), who questioned it.The sample was diluted with a 1:10 dilution factor and was tested on the same instrument, resulting higher.The corrected result was reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely low ca 27.29 result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR
Type of Device
IMMUNOASSAY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict ave
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS
manufacturing limited
registration number: 8020888
chapel lane, swords, co., dublin
EI  
Manufacturer Contact
cassandra kocsis
511 benedict avenue
tarrytown, NY 10591
9145242687
MDR Report Key4899350
MDR Text Key6829184
Report Number2432235-2015-00319
Device Sequence Number0
Product Code NIG
Reporter Country CodeUS
PMA/PMN Number
K971418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,health professional,use
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 10/18/2016
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Date FDA Received07/08/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA CENTAUR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/18/2016
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/25/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-