Discordant, false negative total human chorionic gonadotropin (thcg) results were obtained upon initial and repeat run on one patient sample on an advia centaur xp instrument.The patient initially tested positive using an hcg urine test.The discordant result was reported to the physician(s).The patient was re-tested with hcg urine test, resulting positive.A new sample was obtained from the patient and was tested on the same instrument with the dilution factor as 1:200, resulting positive.The patient was also tested in an alternate laboratory on an unknown methodology, resulting positive.The corrected result from the advia centaur xp instrument was not reported, while it is unknown if the result from the alternate laboratory was reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, false negative thcg results.
|
The initial mdr 2432235-2015-00360 was filed on august 14, 2015.Additional information (09/08/2015): a siemens headquarters support center (hsc) specialist reviewed the sample data.The sample data suggested that the initial serum sample was compromised.The customer did not provide the sample for further testing.The customer did not obtain any additional discordant results.The cause of the discordant, false negative thcg results on one patient sample is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|