2/2/2016 integra investigation completed.Manufacture date unknown.Method: failure analysis, device history evaluation.Results: failure analysis - scissors were returned in used condition, not showing any unusual markings.The returned scissors showing excessive wear, blackish/dark brown staining, repair markings and a broken tip.While visually inspecting the scissors, it is noticed that there is dark brown/black staining.This is the result from detergents used to clean the instrument.The black acid staining can be caused by low ph (less than six) during autoclaving.It is also noticed that the tip is broken.It appears the damage to the tip may have been due to the corrosion weakening the instrument.This type of damage is typically the result of improper processing.The complaint is confirmed; damage/ worn.Device history evaluation - dhr review was completed with all history available.Nonconforming product report / nonconforming material report history: none.Variance authorization / deviation history: none.Engineering change order/manufacturing change order history: there is no applicable engineering change order/manufacturing change order history.Corrective action preventive action history: none.Health hazard evaluation history: none.Conclusion: the root cause has not been identified as a workmanship or material deficiency.
|