A beckman coulter (bec) field service engineer (fse) was at the customer's site and evaluated the instruments performance.The fse inspected the instrument and replaced pipettor one (1) piston, guide set and pressure monitor to address the event log messages displayed on the unicel dxi 800 access immunoassay system.Although the customer reported that the non-reproducible results were obtained from pipettor one, analysis of the patient results confirmed that the non-reproducible results were obtained utilizing all four reagent pipettors.As the non-reproducible results were obtained from all four (4) pipettors, the fse rebuilt pipettors two, three and four.The fse also rebuilt the sample pump to complete the repair.The fse successfully verified the performance of the unicel dxi 800 access immunoassay system after hardware parts replacement.In conclusion, the cause of this event is a hardware malfunction.Although repair of several hardware components resolved the issue, no one component can be implicated as the sole contributor to this event.
|