• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR CP; IMMUNOASSAY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. ADVIA CENTAUR CP; IMMUNOASSAY ANALYZER Back to Search Results
Model Number ADVIA CENTAUR CP
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/13/2016
Event Type  malfunction  
Manufacturer Narrative
A siemens customer service engineer (cse) specialist was dispatched to the customer's site.The cse performed a total service call.The cse then recalibrated the reagent probe.The cause of the discordant, (b)(6) results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Event Description
Discordant, (b)(6), results were obtained on a patient sample on an advia centaur cp instrument.A (b)(6) result was reported out to the physician(s).The physician(s) requested a redraw.The new sample was tested on the same advia centaur cp instrument, resulting (b)(6).The customer then tested the original sample, in duplicate, resulting (b)(6).A corrected report was issued to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, (b)(6) results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVIA CENTAUR CP
Type of Device
IMMUNOASSAY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict ave.
tarrytown NY 10591
Manufacturer (Section G)
STRATEC BIOMEDICAL SWITZERLAND AG
registration #: 3008494306
neuwiesenstrasse 4
beringen 8222
SZ   8222
Manufacturer Contact
timothy rice
511 benedict ave
tarrytown, NY 10591
9145242406
MDR Report Key6159952
MDR Text Key62372068
Report Number2432235-2016-00751
Device Sequence Number0
Product Code LOL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 12/09/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberADVIA CENTAUR CP
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/14/2016
Initial Date FDA Received12/09/2016
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-