The customer contacted the siemens customer care center (ccc).The customer stated that the quality controls (qc) were within range on the day of the event.As per ccc specialist's recommendations, the customer reran qc, which were within range and also, ran precision testing on random patient samples, which passed.However, the sample in question gave imprecise results.A siemens headquarters support center (hsc) reviewed the information related to the event and found no sign of an instrument or method related issue on the date of occurrence.The cause of the discordant, falsely depressed glu result on one patient sample is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|
A discordant, falsely depressed glucose (glu) result was obtained on one patient sample on a dimension exl 200 instrument.The discordant result was not reported to the physician(s).The sample was repeated on the same instrument, resulting higher.The customer informed the nurse that the initial result was incorrect.The customer ran a precision study on the patient sample and obtained imprecise results.The customer obtained a new sample from the patient.The redrawn sample was tested, which resulted higher than previously obtained results.There are no reports of patient intervention or adverse health consequences due to the discordant, falsely depressed glu result.
|