• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC DIMENSION VISTA 1500; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC DIMENSION VISTA 1500; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number DIMENSION VISTA 1500
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/25/2017
Event Type  malfunction  
Manufacturer Narrative
The customer contacted a siemens customer care center (ccc).The quality controls (qc) were within range on the day of the event.The ccc specialist reviewed the instrument data, which indicated a sample mix issue.A siemens customer service engineer (cse) was dispatched to the customer site.The cse replaced the sample probe 1 (s1) mixer.The cse aligned the arm and ran mix test, which passed.The cse ran an alignment test, which passed.The cse ran a quick check, which passed.The customer ran calibrations and qc, which passed.The cause of falsely depressed glu results on patient samples is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Event Description
Discordant, falsely depressed glucose (glu) results were obtained on two patient samples on a dimension vista 1500 instrument.The discordant results were not reported to the physician(s).The samples were repeated on the alternate dimension vista instruments, resulting higher.The corrected results were reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely depressed glu results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIMENSION VISTA 1500
Type of Device
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC
500 gbc drive
po box 6101
newark DE 19702
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC
registration number: 1226181
101 silvermine road
brookfield CT 06804
Manufacturer Contact
cassandra kocsis
511 benedict ave
tarrytown, NY 10591
9145242687
MDR Report Key6343968
MDR Text Key67927324
Report Number2517506-2017-00185
Device Sequence Number0
Product Code CFR
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051087
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/20/2017
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDIMENSION VISTA 1500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/28/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-