A siemens customer service engineer (cse) and application specialist (as) were dispatched to the customer site.The cse verified the probe positions, probe priming, and functioning of the reaction cuvette washer (wud) and dilution cuvette washer (dwud).The cse then cleaned the wud and dwud with hypochlorite solution.The cse also checked the functioning of valves, system bottles and filters (cuvette conditioner, cuvette wash, water, and sodium chloride), and vertical pumps.The as checked the method setting.A siemens customer application specialist reviewed the instrument data and determined that in all cases, lipase was measured in the rotary reaction vessel cuvette prior to processing of crea_2.The cause of the discordant, falsely elevated crea_2 results on multiple patient samples is unknown.The instrument is performing within manufacturing specifications.No further evaluation of device is required.
|
Discordant falsely elevated creatinine (crea_2) results were obtained on multiple patient samples on an advia 2400 instrument.The discordant results were not reported to the physician(s).The samples were repeated on the same instrument, resulting lower.The repeat results were reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated crea_2 results.
|
The initial mdr 2432235-2017-00171 was filed on march 9, 2017.Additional information (04/19/2017): a siemens headquarters support center specialist reviewed the service report and suspected that there may have been contamination due to carryover from lipase in the rotary reaction vessel cuvette.After the siemens customer service engineer performed instrument maintenance and decontamination procedure, the issue resolved.
|