The customer received questionable high gluc3 glucose hk results for an unknown number of patient samples.One example was provided as an initial result of 150 mg/dl and a repeat result of 107 mg/dl.Erroneous results were reported outside the laboratory.There was no allegation of an adverse event.The reagent lot number was 193436.The expiration date was requested but was not provided.A specific root cause could not be determined.Review of the provided calibration data did not indicate a reagent issue but due to multiple errors found, an issue with the calibrator handling or the analyzer performance was suspected.Imprecision and sporadic outliers in the qc data indicated an issue with the gear pump, the wash efficiency, or the sample/reagent probe pipetting was possible.
|