The customer contacted the siemens customer care center (ccc).The ccc specialist instructed the customer to clean the aliquot probe and drain, the integrated multi-sensor technology (imt) probe and perform alignments.The customer did not perform troubleshooting as indicated by ccc.The customer instead, ran precision on patient samples, which resulted satisfactory.The customer determined the cause of the falsely low cre2 result was associated with sample integrity.The instrument is performing according to specifications.No further evaluation of the device is required.
|
A discordant, falsely low urine creatinine (cre2) result was obtained on one patient sample on a dimension vista 500 instrument.The discordant result was not reported to the physician(s).The sample was repeated twice on the same dimension vista instrument, resulting higher than the initial result.The second repeat result was obtained at the end of the reagent flex well.The first repeat result was reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely low urine creatinine (cre2) result.
|