Model Number DIMENSION VISTA 500 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/22/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The customer contacted a siemens customer care center (ccc) and stated that their quality controls (qc) were within acceptable range on the day the event occurred.The customer determined the issue was with the original sample from the patient and the customer was satisfied with the instrument performance.The instrument is performing according to specifications.No further evaluation of this device is required.
|
|
Event Description
|
A discordant, falsely elevated creatinine (cre2) result was obtained on a patient sample on a dimension vista 500 instrument.The discordant result was reported to a nurse practitioner(s) who questioned it.The patient was redrawn and the sample was tested on the same dimension vista instrument, resulting lower and matching the patient's historical results.The original sample was then spun and repeated on two dimension vista instruments, matching the original high result.The redraw result was reported to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely low cre2 result.
|
|
Manufacturer Narrative
|
The initial mdr 2517506-2017-00380 was filed on april 12, 2017.Additional information (05/22/2017): a siemens headquarters support center (hsc) specialist evaluated the data related to the event and could not determine the cause of the discordant, ecre2 result.Corrections (05/26/2017): reworded to reflect falsely elevated ecre2, sample id were added, additional text was added, correct serial number and udi number were added and correct instrument manufacturing date was added.
|
|
Event Description
|
There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated cre2 result.
|
|
Search Alerts/Recalls
|