The customer contacted a siemens customer care center (ccc) and stated that their quality controls were within acceptable range.The customer also stated that they did not get any flag on the patient sample, however, they had obtained abnormal assay flag on hemolysis, icterus and lipemia interference.A siemens headquarters support center (hsc) specialist reviewed the instrument.The hsc specialist recommended that the customer replace the source lamp, as two filter reading were above the recommended ranges.The filters are not likely related the hcg issue.The hsc specialist concluded that the issue may have been due to a specimen integrity issue affecting the sampling of the original result, the tube or cup being removed after a level sense and before sampling to process another test, or processing of the wrong sample.The cause of the discordant, falsely low hcg result on one patient sample is unknown.The instrument is performing within specifications.No further evaluation of device is required.
|
A discordant, falsely low human chorionic gonadotropin (hcg) result was obtained on one patient sample on a dimension exl with lm instrument.The discordant result was reported to the physician(s), who questioned it.The sample was repeated on the same instrument, resulting higher.The corrected result was reported to the physician(s).There are no reports of patient intervention or adverse health consequence due to the discordant, falsely low hcg result.
|