A discordant, falsely high glucose result was obtained on one patient sample on a dimension vista 1500 instrument.The discordant result was reported to the physician(s), who questioned it.The customer tested a newly drawn sample on the original instrument, and recovered lower.The new sample was repeated two more times on the alternate instrument, and the result matched to the rerun result.The result obtained with the new draw was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant glucose result.
|