The investigation has determined that lower than expected vitros glu results were obtained when testing non-vitros quality control samples on a vitros 5,1 fs chemistry system.In addition, unexpected results were also observed across multiple assays on the same vitros 5,1 fs system.The assignable cause of the event is likely instrument related.The customer performed maintenance actions to the vitros 5,1 fs system.Following these actions the customer observed several condition codes and the quality control results for multiple assays were not within the customer¿s established baseline range.An ortho field engineer (fe) was sent to the site to mitigate the condition codes.The fe performed service actions including replacement of the immuno rate wash cam, interrupter optical switch, spring torsion drive pulley, bearing pads, and timing belt were completed to resolve the issue.The quality control results post service were acceptable for all assays including vitros glu.There was no evidence to suggest the vitros reagent malfunctioned.The historical glu qc data shows the reagent was performing as expected.
|
A customer obtained lower than expected vitros glu results from non-vitros quality control samples on a vitros 5,1 fs chemistry system.Non-vitros multichem (lot 15706161) level 1 vitros glu result <1.1 mmol/l versus the expected result 2.59 mmol/l.Non-vitros multichem (lot 15706161) level 2 vitros glu result <1.1 mmol/l versus the expected result 7.23 mmol/l a biased result of the direction and magnitude observed may lead to inappropriate physician action if undetected.The lower than expected vitros glu results were obtained when testing non-vitros quality control samples.Although there were no reports of affected patient sample results, it cannot be confirmed that patient sample results were not affected or would not be affected if the event were to recur undetected.There was no allegation of patient harm as a result of the event.This report corresponds to ortho clinical diagnostics inc.Complaint number (b)(4).
|