• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS 6000 C (501) MODULE; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS 6000 C (501) MODULE; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number C501
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/12/2017
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).Unique identifier (udi)#: (b)(4).
 
Event Description
The customer stated that they received an erroneous result for one patient sample tested for gluc3 glucose hk (glu) on a cobas 6000 c (501) module - c501.The customer also mentioned that they had issues with low quality controls for the albumin assay.The initial glu as 0.17 mmol/l.The sample was repeated, resulting as 4.77 mmol/l.It was asked, but it is not known if the erroneous result was reported outside of the laboratory.No adverse events were alleged.The glu reagent lot number and expiration date were requested but not provided.The gear pump pressure of the analyzer was low.
 
Manufacturer Narrative
The field service engineer exchanged and adjusted the gear pump head.He also exchanged the sample probe.Component code - device subassembly = gear pump head.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 6000 C (501) MODULE
Type of Device
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
HITACHI HIGH TECH CORP.
882 ichige hitachinaka
na
ibaraki 312-8 504
JA   312-8504
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key6757611
MDR Text Key81525443
Report Number1823260-2017-01578
Device Sequence Number0
Product Code CFR
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K060373
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2017
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberC501
Device Catalogue Number04745914001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-