The customer contacted the siemens customer care center (ccc).A ccc specialist reviewed the instrument data and found no reagent or instrument issues associated with this event.Quality control (qc) was within range on the day the event occurred.The ccc had performed diagnostic troubleshooting, which passed.The customer ran precision, resulting within specifications.A siemens customer service engineer (cse) was dispatched to the customer site.After analyzing the instrument, the cse cleaned the aliquot probe, reinstalled and aligned it.A siemens headquarters support center (hsc) specialist has reviewed the instrument data which indicated there was no issue related to a reagent non-conformance or instrument malfunction.Hsc has concluded that this issue is an isolated event specific to sample handling and or sample integrity.The instrument is performing according to specifications.No further evaluation of the device is required.
|
A discordant, falsely low glucose (glu) result was obtained on a patient sample on a dimension vista 1500 instrument.The discordant result was reported to a nurse(s) who questioned it.The patient was redrawn and the test was repeated via a handheld device, resulting higher.The original sample was then repeated twice, on the same dimension vista instrument, resulting higher than the initial result and matching the patient's clinical picture.The first of the replicate results was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely low glu result.
|