The customer stated that they received erroneous results for one patient sample tested for gluc3 glucose hk (gluc) on an cobas integra 400 plus (i400+).The erroneous results were not reported outside of the laboratory.The sample had a low volume.The sample initially resulted as 234.7 mg/dl.The sample was repeated twice, resulting as 286 mg/dl and 158.5 mg/dl respectively.The third repeat result was 162 mg/dl, which was considered to be correct and reported outside of the laboratory.No adverse events were alleged to have occurred with the patient.The gluc reagent lot number was 236290.The reagent expiration date was asked for, but not provided.The field service engineer checked the analyzer hardware and sample probe.These were ok.The gluc assay was successfully calibrated on (b)(6) 2017.Between 19-aug-2017 and 27-aug-2017, the level 1 quality control was within range, but there was a large variation in results.Between 27-aug-2017 and 29-aug-2017, the level 1 quality control was close to target and showed higher precision.The level 2 quality control was within range.
|