The customer stated that they received erroneous high results for a total of five patient samples tested for the elecsys ft4 ii assay (ft4) on a cobas 8000 e 602 module (e602).The customer stated that some erroneous values were reported outside of the laboratory and amended.The customer could not specify which values were reported outside of the laboratory.The first sample initially resulted as 28.87 pmol/l and repeated as 18.73 pmol/l.The second sample initially resulted as 72.45 pmol/l and repeated as 58.91 pmol/l.The third sample initially resulted as 31.8 pmol/l and repeated as 13.5 pmol/l.The fourth sample initially resulted as 89.20 pmol/l and repeated as 39.3 pmol/l.On (b)(6) 2017, field service engineer cleaned a wash station waste line and replaced the reagent probe since it was intermittently sticking in the upper position.He replaced seals and pinch tubing.He checked for materials in the incubator and ran performance testing.He noted that precision looked good in the performance test, but some values for another parameter were poor.He ran quality controls.After the service visit, the customer stated that they had erroneous results for an additional 3 patient samples.The customer provided data for one of these additional samples (sample 5).On (b)(6) 2017, this fifth sample initially resulted as 28.57 pmol/l and repeated as 14.98 pmol/l.No adverse events were alleged to have occurred with the patients.The ft4 reagent lot number was 24682500, with an expiration date of 30-jun-2018.
|