The customer received questionable high crea plus creatinine plus results for an unknown number of patient samples.The samples were repeated on an analytical p module analyzer.Of the data provided for 21 patient samples, only the result for 6 patient samples were discrepant.Refer to the attachment to the medwatch for all patient data.The results were not reported outside of the laboratory.There was no adverse event.The reagent lot number and expiration date were requested but were not provided.The field service representative could not determine a cause.He checked the sample probe and all the probes alignments, performed an air purge, rinse, and prime.He ran qc and linearity testing with all results within specifications.
|