• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER ACCESS ACCUTNI+3 REAGENT; ANALYZER, CHEMISTRY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER ACCESS ACCUTNI+3 REAGENT; ANALYZER, CHEMISTRY Back to Search Results
Model Number A98143
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem Chest Pain (1776)
Event Date 03/19/2018
Event Type  Injury  
Event Description
Pt presented to the ed with epigastric and chest pain that has intensified over the last five days.Pt had troponin labs drawn, and samples were processed on beckman coulter dxi utilizing accutni+3 reagent.Troponin levels measured 2.67 ((b)(6) at 2305) that trended down to 2.59 ((b)(6) at 0450) and 2.49 ((b)(6) at 0555).During this time, point-of-care troponin levels remained within normal levels.Pt was taken to cath lab and did not have any discernable lesions per dr travis love in cardiology.Specimen collected at 0450 on (b)(6) was sent to outside lab and tested on a bitros analyzer.Result was <0.01.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCESS ACCUTNI+3 REAGENT
Type of Device
ANALYZER, CHEMISTRY
Manufacturer (Section D)
BECKMAN COULTER
brea CA
MDR Report Key7373250
MDR Text Key103691616
Report NumberMW5076159
Device Sequence Number0
Product Code MMI
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other
Type of Report Initial
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2018
Device Model NumberA98143
Device Lot Number724139
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received03/26/2018
Was Device Evaluated by Manufacturer? No Information
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age44 YR
Patient Weight95
-
-