Model Number N/A
Device Problem Temperature Problem (3022)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/04/2018
Event Type  malfunction  
Manufacturer Narrative
The production device history record (dhr) for this iabp unit is not required to be reviewed per (b)(4) since the device manufacture date is greater than one year from the event date.Full event site name: (b)(6) medical center.A getinge service territory manager (stm) evaluated this unit and was unable to reproduce the reported issue.The attached logs show nothing abnormal with the unit and the fans are operational.The unit calibrated and passed all functional and safety tests performed and was returned to the customer and cleared for clinical use.
 
Event Description
It was reported that while in use on a patient the cardiosave intra-aortic balloon pump (iabp) showed a message that the users interpreted as "overheating" related.Even thought the nurse indicated that it was very unclear what they saw in the screen.It was also reported that the batteries were hot.There was no patient harm and no adverse event was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer (Section G)
DATASCOPE MAHWAH
1300 macarthur blvd.
mahwah NJ 07430
Manufacturer Contact
1300 macarthur blvd.
mahwah, NJ 07430
MDR Report Key7726201
MDR Text Key115419619
Report Number2249723-2018-01265
Device Sequence Number0
Product Code DSP
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2018
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age37
Patient Weight124
-
-