WAVELIGHT GMBH WAVELIGHT FS200 FEMTOSECOND LASER; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Catalog Number 8065990941
Device Problems Suction Problem (2170); Suction Failure (4039)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/27/2018
Event Type  malfunction  
Manufacturer Narrative
Manufacturing record reviewed.No abnormalities that could have contributed to this event were found.Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.(b)(4).
 
Event Description
An optometrist reported during flap creation suction was lost at nine percent.The patient did not move.The procedure was completed the same day.
 
Manufacturer Narrative
A logfile review of the day of the treatment shows the three system tests for vacuum, energy and ablation were executed by the user without any problems.The reported treatment could be identified in the logfile.The treatment was aborted after a warning message displayed during canal cut.The message indicates that the user released laser pedals and pressed the right footswitch to turn off the vacuum.The user restarted the treatment and performed the treatment successfully on that day.No technical root cause is detectable.The root cause is user handling.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WAVELIGHT FS200 FEMTOSECOND LASER
Type of Device
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
WAVELIGHT GMBH
am wolfsmantel 5
erlangen 91058
GM  91058
MDR Report Key7938978
MDR Text Key123205673
Report Number3003288808-2018-01631
Device Sequence Number0
Product Code GEX
PMA/PMN Number
K101006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/14/2019
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065990941
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/17/2018
Initial Date FDA Received10/05/2018
Supplement Dates Manufacturer Received01/28/2019
Supplement Dates FDA Received02/14/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age26 YR
Patient Weight84
-
-