ENCORE MEDICAL L.P. RSP SHOULDER; RSP STANDARD HUMERAL SOCKET INSERT, 32MM, HXE-PLUS Back to Search Results
Model Number 509-00-032
Device Problems Device Dislodged or Dislocated (2923); Appropriate Term/Code Not Available (3191)
Patient Problem Joint Dislocation (2374)
Event Date 09/26/2018
Event Type  Injury  
Event Description
Second revision surgery - due to the patient having dislocated again.
 
Manufacturer Narrative
The reason for this revision surgery was due to a dislocation.The previous revision surgery and the revision surgery detailed in this investigation occurred 18 days apart.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr), shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to a dislocation.There were no findings during this investigation that indicate that the reported device was defective.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP SHOULDER
Type of Device
RSP STANDARD HUMERAL SOCKET INSERT, 32MM, HXE-PLUS
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key7951671
MDR Text Key123234043
Report Number1644408-2018-00941
Device Sequence Number0
Product Code PHX
PMA/PMN Number
K051075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 12/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Device Operator Health Professional
Device Expiration Date07/19/2023
Device Model Number509-00-032
Device Catalogue Number509-00-032
Device Lot Number377P1395
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 09/26/2018
Initial Date FDA Received10/10/2018
Supplement Dates Manufacturer Received12/18/2018
Supplement Dates FDA Received12/19/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number0
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age73 YR
-
-