• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDELA LLC PUMP IN STYLE ADVANCED - ON-THE-GO TOTE; PUMP, BREAST, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDELA LLC PUMP IN STYLE ADVANCED - ON-THE-GO TOTE; PUMP, BREAST, POWERED Back to Search Results
Model Number 57063
Device Problems Crack (1135); Failure to Power Up (1476); Electrical Power Problem (2925)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/01/2018
Event Type  malfunction  
Manufacturer Narrative
The customer was sent a replacement power supply and return of her original power supply was requested for testing/evaluation.A visual inspection of the power supply indicated that the power supply housing was breached and the prongs were broken/damaged.This issue with a damaged rev m power supply for the pump in style device was addressed in investigation (b)(4).The investigation found that they were being damaged during shipment from the manufacturer to medela.This damage was causing the plastic housing to fail prematurely when subjected to normal use and foreseeable misuse.The packaging used by the manufacturer to ship the power supply to medela was not robust enough to handle all of the potential shipping, handling, and abuse conditions that could arise from logistics of the consolidation process.As a result of the investigation, the shipping and consolidation process was modified to reduce the handling and potential for double stacking of the skids.The shipping packaging strength was also increased to further protect the power supply during shipping.Since then, as a part of routine continuous improvement activities, a rev p power supply has been distributed to market, manufactured under a revised design and by a different manufacturer.
 
Event Description
On (b)(6) 2018, the customer alleged to medela llc that her pump in style pump would not power on with the power supply, but would with the battery pack.The customer additionally alleged that the power supply housing was cracked and being held together by tape.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE ADVANCED - ON-THE-GO TOTE
Type of Device
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA LLC
1101 corporate drive
mchenry IL 60014
Manufacturer (Section G)
MEDELA LLC
1101 corporate drive
mchenry IL 60014
Manufacturer Contact
jan kloiber
1101 corporate drive
mchenry, IL 60014
8004358316
MDR Report Key8051848
MDR Text Key128789035
Report Number1419937-2018-00247
Device Sequence Number0
Product Code HGX
Reporter Country CodeUS
PMA/PMN Number
K031614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/08/2018
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number57063
Device Catalogue Number57063
Device Lot Number9207010 REV M-3212
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2018
Date Manufacturer Received10/15/2018
Was Device Evaluated by Manufacturer? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-