• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. GLENOSPHERE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. GLENOSPHERE Back to Search Results
Catalog Number 9999
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problems Irritation (1941); No Code Available (3191)
Event Date 06/21/2017
Event Type  Injury  
Manufacturer Narrative
This event report was received through clinical data collection activities.Pending evaluation.
 
Event Description
Index surgery: (b)(6) 2015.Revision due to suprascapular nerve irritation from screw.The case report form indicates this event is possibly related to devices and definitely related to procedure.
 
Manufacturer Narrative
Engineering evaluation noted that the revision reported was likely the result of nerve irritation due to the positioning and/or length of one of the compression screws used to secure the glenoid plate.
 
Manufacturer Narrative
Engineering evaluation noted that the revision reported was likely the result of nerve irritation due to the positioning and/or length of one of the compression screws used to secure the glenoid plate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLENOSPHERE
Type of Device
GLENOSPHERE
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66th court
gainesville FL 32653
MDR Report Key8085801
MDR Text Key127612017
Report Number1038671-2018-00887
Device Sequence Number0
Product Code KWT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup,Followup
Report Date 02/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received11/19/2018
Is this an Adverse Event Report? Yes
Device Operator Health Professional
Device Catalogue Number9999
Was Device Available for Evaluation? No
Date Manufacturer Received10/29/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age65 YR
Patient Weight68
-
-