• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 ASR ACETABULAR IMPLANT 54; ASR HIP SYSTEM : HIP METAL ACETABULAR CUPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 ASR ACETABULAR IMPLANT 54; ASR HIP SYSTEM : HIP METAL ACETABULAR CUPS Back to Search Results
Catalog Number 999804754
Device Problems Loss of or Failure to Bond (1068); Nonstandard Device (1420)
Patient Problems Tissue Damage (2104); Inadequate Osseointegration (2646); No Code Available (3191)
Event Date 02/25/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.No 510(k) number provided because this implant product code is sold internationally.It is sold in the us under a different product code.
 
Event Description
Revision of an asr cup.Cup was loose, but only minimal metallosis in soft tissues.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #
=
> (b)(4).Investigation summary
=
> the asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.Further investigation of this individual incident will not be undertaken.The investigation regarding the root cause(s) and/or corrective actions was conducted under mdd capa-001226.Ongoing post market surveillance is conducted per our procedures for this product.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASR ACETABULAR IMPLANT 54
Type of Device
ASR HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key8424361
MDR Text Key138908628
Report Number1818910-2019-87377
Device Sequence Number0
Product Code KWY
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Remedial Action Recall
Type of Report Initial,Followup
Report Date 02/25/2019
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number999804754
Device Lot Number1962995
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 02/25/2019
Initial Date FDA Received03/15/2019
Supplement Dates Manufacturer Received04/03/2019
Supplement Dates FDA Received04/23/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
-
-